GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 43 8-8-7 Record SW Opti on Keys - Software v3.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12. 2.) From the Admin tab, select SYSTEM ADMIN . 3.) Record SW Options Keys from SW Options Key window. Figure 8-37 System Admin WARNINGWARNINGMake sure that the SW Option Keys (passwords) have been recorded correctly. If the key is incorrect, you will not be able to log on after the SW installation has been completed. The password is case sensitive. Hyphens must also be recorded. There may be more than one password. Table 8-10 SW Options Keys SW OPTIONS KEYS
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 44 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-8 Record TCP/IP Settings - Software Version v3.x 1.) Select Connectivity . 2.) Select the Tcpip tab. 3.) Record all settings in Table 8-11 Record settings from Tcpip screen (v3.x) on page 8-44 . Figure 8-38 TCP/IP Example Settings Table 8-11 Record settings from Tcpip screen (v3.x) PARAMETERVALUE Computer Name VIVID7-00 _ _ _ _ AE Title VIVID7-00 _ _ _ _ IP Settings IP Address_ _ _ . _ _ _ . _ _ _ Subnet mask_ _ _ . _ _ _ . _ _ _ Default Gateway_ _ _ . _ _ _ . _ _ _ Remote Archive Setup Remote Archive IP-Addr _ _ _ . _ _ _ . _ _ _ Remote Archive Name
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 45 8-8-9 Record Service Settings - Software Version v3.x 1.) Select Service. . 2.) Record Video Setting Format and VCR Type in Table 8-12 on page 8-45. Figure 8-39 Service Settings Table 8-12 Video Setting and VCR type VIDEO SETTING FORMATVCR TYPE ___ PAL ___NTSC
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 46 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-10 Information on the About tab - Software Version v3.x 1.) Select the About tab. 2.) Record software and hardvare versions in Table 8-13 on page 8-46. Figure 8-40 Information available on the About tab Table 8-13 Record Software and Hardware versions (v3.x) DESCRIPTIONRECORD VERSION APPLICATION SW SYSTEM SW Processor (BEP)
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 47 8-8-11 Software Loading - Software Version v3.x 8-8-11-1 Overview This software loading procedure covers: • Installation of BT’03 software (sof tware version v3.x) on Vivid 7 units. Refer to Table 8-13 on page 8-46 where you recorded the unit’s software versions. The software for Vivid 7 is divided on two CDs, on e CD with the System Software and one with the Application Software. 8-8-11-2 Vivid 7 System SW Loading - Software Version v3.x 1.) Insert the applicable Vivid 7 System Software CD into the CD drive. 2.) Power down the System, see 4-2-3 Power Shut Down on page 4-7 . 3.) Wait until the LCD’s on the Operator Panel are s witched off and the ON/Standby switch has turned yellow. 4.) Power ON the System, see 4-2-3 Power Shut Down on page 4-7 . The scanner will now boot from the CD you inserted. Figure 8-41 BIOS loading example
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 48 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-11-2 Vivid 7 S ystem SW Loading - Software Version v3.x (cont’d) 5.) When requested, enter the missing part of the Co mputer Name (the unit’s serial number) as you recorded it in Table 8-6 Record settings from Tcpip screen - software version v7.x/v6.x/v5.x/v4.x on page 8-21 . Example: VIVID7-001890 Always enter two zeros before you enter the serial number. This means that when you have a serial number which is 1890 you enter 001890. When the serial number is 1123 you enter 001123. 6.) In this step, you can choose if yo u want to do a complete Vivid 7 installation and deleta all data or if you only want to update or reinstall the software on the C: partition (C: drive). Usually it should be OK to select “ B” to update the C-dr ive only. This will keep Patient Archive and Presets intact. a.) Select B to update/re-install th e software on the C: partition. (Only data on the C partition will be erased. This will keep Pati ent Archive and Presets intact. b.) Use “ A” ONLY if a “ B” update cannot be used: Select A to do a complete Vivid 7 software installation. (All data will be erased.) c.) Press C if you want to Exit. Figure 8-42 Serial Number entry example NOTICEIf you select “ A” in the next step, ALL existi ng software and data will be erased. If backup has not been performed as described earlier in this manual; All data like Patient Database, System Configuration and User Configurations (custo mer presets) will be lost. Figure 8-43 Info entry page Enter the Serial Number here Press A, B or C to choose the wanted option.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 49 8-8-11-2 Vivid 7 S ystem SW Loading - Software Version v3.x (cont’d) 7.) The Software loader program starts and the below Time Bar Info Screen appears. 8.) Wait for the SW Installation to be completed. (Typic al installation time: 5 minutes). Status bar on the screen indicates process. 9.) You will now be prompted to remove the first CD and restart th e unit when the SW Installation is completed. Remove the CD, but don’t restart the System this time . The next step is to load the application software. Figure 8-44 Software Loading Underway (Picture distorted here due to photo technical limitations, it should look OK on screen) WARNINGWARNINGDo not interrupt the SW loading at any time.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 50 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-11-3 Application SW Installation - Software Version v3.x 1.) Insert the applicable CD labelled Vivid7 & EchoPAC PC Application SW , see Table 8-9 CDs with BT’03 Software, SW. v3.x on page 8-39 . 2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the scanner. (This is called Forced Shut Down). 3.) Wait until the LCD’s on the Oper ator Panel is switched OFF and the ON/Standby Switch on the Operator Panel has turned yellow. 4.) Power ON the System, see 4-2-2 Power ON/ Boot UP on page 4-3 . 5.) A dialogue box, called Start Ap plication, will appear, select Install SW. 6.) A dialogue box, called Star tloader, will appear, select OK to continue. 7.) A dialogue box will appear, al so named Startloader stating “New System SW installation will start now”, select OK to continue. 8.) A Command prompt window will ap pear, hit any key on the alphanumeric keyboard to continue. 9.) A message in the Command prompt window may prompt you that copying application software takes three minutes. This may ta ke as much as five minutes. 10.) Several dialogue boxe s will appear on the screen during the SW load. 11.) Remove the CD from the CD drive when prompted. 12.) Press any key to continue. 13.) The Scanner will now pow er down and reboot. WARNINGWARNINGDo not interrupt the SW loading at any time.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 51 8-8-12 Verifications - Software Version v3.x NOTE: Software Option Strings are software versio n dependant. If a new software version has been installed, please follow installa tion instructions for that softwar e. Usually, new software option strings are needed. 8-8-12-1 Verify SW Option Key Se ttings - Software Version v3.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12. 2.) Select Admin (lower part of window). 3.) Select the SYSTEM ADMIN tab. 4.) Verify that the SW Options Keys (alphanumeric strings) in the SW Options Key window are the same as recorded in 8-8-7 Record SW Option Keys - Software v3.x on page 8-43. 5.) Select New to open the New Key dialog where you can type any missing Software Option Key(s) (alphanumeric string(s)). Figure 8-45 System Admin screen Figure 8-46 Type Software Option Key System Admin Tab New Admin Field for Software Option Key (Serial String)
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 52 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-12-2 Verify Video and VCR Se ttings - Software Version v3.x 1.) Select Service (lower, right corner) . 6.) From Video Settings pull down menu, verify Format as recorded in Table 8-12 on page 8-45 . Correct the selection if needed. 7.) From the VCR pull down menu, verify VCR type as recorded in Table 8-12 on page 8-45 . Correct the selection if needed. Figure 8-47 Video Settings