GE Vivid 7 Service Manual
Have a look at the manual GE Vivid 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 7 3-3-4 Receiving the Vivid 7 3-3-4-1 Examine All Packages Examine all packages closely at time of de livery, as described in the procedure below. Table 3-5 Examine All Packages STEPTASKILLUSTRATIONS 1 Is damage apparent? • If yes; continue with the instructions in subsection 3-3-4-2 - Damage in Transportation . • If no; continue with the next step. 2 Is the Shock Indicator red colored inside the middle of the indicator? • If yes: The Shock Indicator has been activated. Continue with the instru ctions in subsection 3-3-4-2 - Damage in Transportation . • If no: Continue with step 2. 3 Is the Tilt Indicator red colored inside the middle of the indicator? • If yes: The Tilt Indicator has been activated. Continue with the instru ctions in subsection 3-3-4-2 - Damage in Transportation before you continue with the next step . • If no: Continue with the instructions in 3-3-5 - Unpacking Vivid 7 . 4The two upper hinges on the Front Side and the Rear Side of the wooden transportation bow have been sealed with red plastic seals, marked GE Vi ngmed Ultrasound and a serial number. Verify that the four red plastic seals are intact at arrival. If seals are broken: If broken, it indicates that the container may have been opened after it left the manufacturer. Continue with the instructions in subsection 3-3-4-2 - Damage in Transportation . Continue with the instructions in 3-3-5 - Unpacking Vivid 7. RED COLOR RED COLOR
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 8 Section 3-3 - Receiving and Unpacking the Equipment 3-3-4-2 Damage in Transportation Follow this procedure if damage is apparent, or if the Tilt & Drop Indicators show failure: Table 3-6 Damage in Transportation STEPTASK 1 Write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving agent. 2 Report the damage to the carrier. • Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the content s and containers held for inspection by the carrier. • A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period. 3 Report the damage on the Post Delivery Checklist. Specify if the tilt & drop indicato rs show failure in the “Packing” field on the Post Delivery Checklist.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 9 3-3-4-3 Vivid 7 Transportation Box Label The Vivid 7 Transportation Box Label is located at the front of the transportation box. Figure 3-3 Vivid 7 Transportation Box Label HANDLE WITH CARE KEEP DRY, PROTECT FROM MOISTURE TOP. UPRIGHT TRANSPORTATION & STORAGE RECYCLING. RECYCLABLE WOOD KEEP TRANSPORTATION TEMPERATURE BETWEEN -20°C and +60°C RELATIVE HUMIDITY BETWEEN 30 and 95% FRAGILE, HANDLE WITH CARE
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 10 Section 3-3 - Receiving and Unpacking the Equipment 3-3-5 Unpacking Vivid 7 CAUTIONTwo people are needed to unpack the unit because of its weight. Attempts to move the unit considerable distances or on an incline by one person could result in injury or damage or both. Two people are required whenever a part weighing 16 kg (35 lbs) or more must be lifted or pushed up or down a slope. Table 3-7 Unpacking Vivid 7 StepTask 1. Open the four hinges on each door and remove the doors. One of the doors are used as ramp out off and into the transportation box. On the first version of the Transportation Box, only the front door is meant to be used as a ramp. It has bevel cut in one end. 2. Place the front door as a ramp against the rear edge of the pallet. 3. Carefully remove the accessory box, and any other item s, including the wooden shelf above the scanners keyboard. - Vivid 7 with LCD monitor: Part #10, support for the LCD monitor, can’t be removed at this time. Remove the rest of the packing material from the transportation box. - Vivid 7 with CRT monitor: Remove all the filling material, from the transportation box. Front door bevel cut. This end of the door should be used as the lower end of the ramp. On the first version transportation box, which is no longer manufactured by GE, a 10 cm board (not furnished) is placed as a support under the upper end of the ramp.On the second version transportation box, the ramp is placed directly on the labank ends. “Labank”
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 11 4. • Press once on the brake pedal to release the brakes. • Keep direction lock activated. The direction lock keeps the fr ont wheels from swiveling and blocking the system inside the narrow transportation box. Table 3-7 Unpacking Vivid 7 (cont’d) StepTask Transport notice FC314187-03 NOTIC E M ovin g V iv id 7/L ogiq 9 i n a nd o ut o f t h e t r a nsp orta tio n b ox Vivid 7/Logiq 9 has brakes on all wheels, but direction lock only on the front wheels. The wheels position when moving the system into its transportation box are therefore vital. If the wheels are swiveling when the system is inside the transportation box, it may jam the system inside the box. The system must be moved backwards both into and out of the transportation box. M ovin g t h e s yste m i n to t h e t r a nsp orta tio n b ox 1. Align the front wheel under the front end of the system as shown in Figure 1. 2. Push the direction lock pedal to lock the front wheels in this direction. Check that they are locked. 3. Push the Top Console to its lowest and most backward locked position. 4. Pull the system into the box with the rear end first. M ovin g t h e s yste m o ut o f t h e t r a nsp orta tio n b ox 1. Unlock brakes but keep direction lock activated. The direction lock keeps the front wheels in position, and secures the direction stability when the system is rolled out of the transportation box. 2. Pull the system out of the box with rear end first. Figure 1: Brakes and direction locks Figure 2: Console and Wheels position Front Wheels Rear Wheels Direction LockBrakeBrakes INOUT Transportation box C AU TIO N Have two people available to unpack the system. Attempts to move the system considerable distances or on an incline by one person could result in injury or damage or both.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 12 Section 3-3 - Receiving and Unpacking the Equipment 5.Carefully move the instrument out of the Transportation Box, down the ramp, with rear end first. 6.Vivid 7 with LCD: Remove part number 10 (LCD monitor enforcement). 7. Remove the clear plastic (wrapped around the scanner) from the unit. 8. Place all the filling material inside the Transportation Box, close it and store it for possible use in the future. Table 3-7 Unpacking Vivid 7 (cont’d) StepTask
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 13 Section 3-4 Packing Materials - Re cycling Information The packing materials for Vivid 7 are recyclable: • The Transportation Box is made of spruce or similar material. (“PHYTOSANITARY CERTIFICATE” incl uded in all shipments to The Peoples Republic of China.) • Lever lockings (hinges) are made of zinc plated steel. • The inner reinforcements are made of Ethafoam (Polyethylene foam). • The plastic foil is made of LDPE (Low Density Polyethylene).
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 14 Section 3-5 - Preparing for Setup Section 3-5 Preparing for Setup 3-5-1 Physical Inspection • Verify that the system arrive d intact (visual inspection). If the system has been damaged, please refer to “Damage in Transportation“ on page viii in the beginning of this manual. • Verify that all items are present (see the Pa cking List). Report any missing parts on the Post Delivery Checklist . 3-5-2 EMI Protection This unit has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system from image artifacts caused by this interference. For this reason, it is imperative that all covers and hardware are installed and secured before the unit is put into operation. See 2-2-3 EMI Limitations on page 2-7 for more information about EMI protection. Section 3-6 Completing the Setup 3-6-1 System Specifications 3-6-1-1 System Requirements Verification • Verify that the site meet s the requirements listed in Chapter 2. • Verify that the s pecifications below don’t conf lict with any on-site conditions. 3-6-1-2 Physical Dimensions The physical dimensions of the Vivid 7 unit are summarized in Table 3-8. 3-6-1-3 Mass (Weight) with Monitor and Peripherals 3-6-1-4 Maximum Audio Noise Less than 70dB(A) according to the standard DIN 45635 - 19 - 01 - KL2. Table 3-8 Physical Dimensions of Vivid 7 with monitor and peripherals HeightWidthDepthUnit 137.5 - 157.5 64 90 cm 54.1 - 62 25.2 35,4 Inches Table 3-9 Weight of Vivid 7 with monitor and peripherals ModelWeight [kg]Weight [lbs] Vivid 7 200 410
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 15 3-6-2 Electrical Specifications 3-6-2-1 Verification of the System’s Voltage Settings Verify that the mains voltage specif ied for the unit is available on-site. The Voltage setting for the unit is found on a label ne ar the Mains Power Circuit Breaker at the rear of the system, see Figure 3-4 on page 3-15 . See also Table 3-10 Electrical Specifications for Vivid 7 on page 3-16 . WARNINGWARNINGCONNECTING A Vivid 7 UNIT TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY DESTROY THE UNIT. Figure 3-4 Mains Circuit Breaker and Power Rating Label Power Rating Label
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 16 Section 3-5 - Preparing for Setup 3-6-2-2 Electrical Specifications for Vivid 7 Table 3-10 Electrical Specifications for Vivid 7 sheet 1 of 3 GE VINGMED PART NUMBER DESCRIPTIONVOLTAGETOLERANCESCURRENTFREQUENCY FD000010 VIVID 7 DIMENSION CONSOLE, LCD MONITOR 230 VAC ±10% 5 A 50-60 Hz FD000020 VIVID 7 DIMENSION CONSOLE, LCD MONITOR 100-120 VAC ±10% 10 A 50-60 Hz FD000120 VIVID 7 PRO CONSOLE, LCD MONITOR 230 VAC ±10% 5 A 50-60 Hz FD000130VIVID 7 PRO CONSOLE, LCD MONITOR100-120 VAC±10%10 A50-60 Hz FD000140 VIVID 7 DIM. W/ MULTI-DIM. & VOLUME, LCD MONITOR230 VAC±10%5 A50-60 Hz FD000150 VIVID 7 DIM. W/ MULTI-DIM. & VOLUME, LCD MONITOR100-120 VAC±10%10 A50-60 Hz FD000160 VIVID 7 DIM. W/ MULTI-DIM. IMAGING, LCD MONITOR 230 VAC ±10% 5 A 50-60 Hz FD000170 VIVID 7 DIM. W/ MULTI-DIM. IMAGING, LCD MONITOR 100-120 VAC ±10% 10 A 50-60 Hz FD000180 VIVID 7 DIM. W/ VOLUME ULTRASOUND, LCD MONITOR 230 VAC ±10% 5 A 50-60 Hz FD000190 VIVID 7 DIM. W/ VOLUME ULTRASOUND, LCD MONITOR100-120 VAC±10%10 A50-60 Hz FC000890VIVID 7 DIMENSION MULTIDIMENSIONAL & Volume (BT’06) 230 VAC230 VAC±10%5 A50-60 Hz FC000900VIVID 7 DIMENSION MULTIDIMENSIONAL & Volume (BT’06) 120 VAC110-120 VAC±10%10 A50-60 Hz FC000910 VIVID 7 DIMENSION (BT’06) 230 VAC 230 VAC ±10% 5 A 50-60 Hz FC000920 VIVID 7 DIMENSION (BT’06) 100-120 VAC 110-120 VAC ±10% 10 A 50-60 Hz FC000930 VIVID 7 PRO (BT’06) 230 VAC 230 VAC ±10% 5 A 50-60 Hz FC000940VIVID 7 PRO (BT’06) 100-120 VAC110-120 VAC±10%10 A50-60 Hz FC000950VIVID 7 DIMENSION MULTIDIMENSIONAL IMAGING (BT’06) 230 VAC230 VAC±10%5 A50-60 Hz FC000960VIVID 7 DIMENSION MULTIDIMENSIONAL IMAGING (BT’06) 120 VAC110-120 VAC±10%10 A50-60 Hz