GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 63 8-9-11 Record TCP/IP Setti ngs - Software Version v2.x 1.) Select Connectivity . 2.) Select the Tcpip tab. 3.) Record all settings in Table 8-16 Record settings from Tcpip screen (v2.x) on page 8-63 . Figure 8-57 TCP/IP Example Settings Table 8-16 Record settings from Tcpip screen (v2.x) PARAMETERVALUE Computer Name VIVID7-00 _ _ _ _ AE Title VIVID7-00 _ _ _ _ IP Settings IP Address_ _ _ . _ _ _ . _ _ _ Subnet mask_ _ _ . _ _ _ . _ _ _ Default Gateway_ _ _ . _ _ _ . _ _ _ Remote Archive Setup Remote Archive IP-Addr _ _ _ . _ _ _ . _ _ _ Remote Archive Name
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 64 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-12 Record Default Dataflow - Software Version v2.x 1.) In the Connectivity window, select Dataflow. 2.) Verify that Default Dataflow is selected. 3.) Record Default Dataflow Name in table Table 8-17 on page 8-64. Figure 8-58 Select Dataflow Table 8-17 Record Default Dataflow Name DEFAULT DATAFLOW NAME
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 65 8-9-13 Record Standard Printer IP and Port Settings - Software Version v2.x 1.) In Connectivity window, select Services. 2.) In the Services list, Scroll down until Standard Print appears. 3.) Select Standard Print . 4.) Select Open. 5.) Select Printer to open the Printer scroll down menu. 6.) Select Properties from the Printer scroll down menu.Figure 8-59 Choosing Standard Print Figure 8-60 Properties CONNECTIVITY STANDARD PRINT OPEN SERVICES SCROLL DOWN MENU PRINTER PROPERTIES
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 66 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-13 Record Standard Printer IP and Port Settings - Software Version v2.x (cont’d) 7.) Select Ports. 8.) Select Standard TCP\IP Port . 9.) Select Configure Port .Figure 8-61 Configure Port Figure 8-62 Configure Port (Port Settings dialog) STANDARD TCP\IP PORT CONFIGURE PORT PORTS
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 67 8-9-13 Record Standard Printer IP and Port Settings - Software Version v2.x (cont’d) 10.) Select and record Printer Name or IP Address in Table 8-18 on page 8-67 . 11.) Select Raw. 12.) Select and record Port Number in Table 8-18 on page 8-67 . 13.) Select Cancel. 14.) Close rest of the Windows by selecting Ok. Table 8-18 Printer Name or IP Address and Port Number PRINTER NAME OR IP ADDRESSPORT NUMBER Figure 8-63 Configure Port PORT NUMBER OK PRINTER NAME OR IP ADDRESS
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 68 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-13-1 Record Service Settings - Software Version v2.x 1.) Select Service. . 2.) Record Video Setting Format and VCR Type in Table 8-19 Video Settings and VCR Type on page 8-68 . Figure 8-64 Service Settings Table 8-19 Video Settings and VCR Type Video Setting FormatVCR Type
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 69 8-9-13-2 Information on the About tab - Software Version v2.x 1.) Select the About tab. 2.) Record Application software version, System software version and Processor model (BEP) in Table 8-20 Record Software and hardware versions on page 8-69 . Figure 8-65 Information available on the About tab Table 8-20 Record Software and hardware versions DESCRIPTIONRECORD VERSION APPLICATION SW SYSTEM SW PROCESSOR (BEP)
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 70 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-14 Software Loading - Software Version v2.x 8-9-14-1 Overview This software loading procedure covers: • Installation of BT’02 Software (application software version v2.x) on Vivid 7 units. Refer to Table 8-20 on page 8-69 where you recorded the unit’s software versions. The software for Vivid 7 is divided on two CDs, one CD with the System Software and one with the Application Software. Table 8-21 Vivid 7 Models Versus Minimum Software Requirement GE VINGMED PART NUMBER DESCRIPTION APPLICATION SOFTWARE VERSION FB000030 VIVID 7 (BT’01) V2.3.1 (or higher) FC000060 VIVID 7 (BT’01) V2.3.1 (or higher) FC000180 VIVID 7 PRO (BT’02) V2.3.1 (or higher) FC000190VIVID 7 PRO (BT’02)V2.3.1 (or higher) FC000200VIVID 7 (BT’02)V2.3.1 (or higher) FC000210VIVID 7 (BT’02)V2.3.1 (or higher)
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 71 8-9-14-2 Vivid 7 System Software Loading - Software Version v2.x 1.) Insert the applicable Vivid 7 System Software CD into the CD drive, see Table 8-20 Record Software and hardware versions on page 8-69 . 2.) Power down the System, see 4-2-3 Power Shut Down on page 4-7 . 3.) Wait until the LCD’s on the Operator Panel are s witched off and the ON/Standby switch has turned amber. 4.) Power ON the System, see 4-2-2 Power ON/ Boot UP on page 4-3 . The scanner will now boot from the CD you inserted. Figure 8-66 BIOS loading example
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 72 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-14-2 Vivid 7 Syst em Software Loading - Software Version v2.x (cont’d) 5.) When requested, enter the missing part of the Co mputer Name (the unit’s serial number) as you recorded it in Table 8-16 Record settings from Tcpip screen (v2.x) on page 8-63 . Example: VIVID7-001890 Always enter two zeros before you enter the serial number. This means that when you have a serial number which is 1890 you enter 001890. When the serial number is 1123 you enter 001123. 6.) In this step, you can choose if you want to do a co mplete Vivid 7 installation or if you only want to update or reinstall the software on the C: partition (C: drive). Usually it should be OK to select “B” to update the C-drive only . This will keep Presets etc. a.) Press B to update/re-install the softw are on the C: partition. (Only data on the C: partition will be erased. System configurations et c. will be kept.) Continue with step 7. b.) Use ONLY if a “B” update cannot be used: Press A to do a complete Vivid 7 software installation. (All data will be erased.) c.) Press C to Exit. Figure 8-67 Serial Number entry example NOTICEIf you select “ A” in the next step, ALL existing software and data on the hard disk will be erased. If backup has not been performed as described earlier in this manual, all data like Patient Database, System Configuration and User Configur ations (Customer Presets), will be lost. Figure 8-68 Info entry page Enter the Serial Number here Press A, B or C to choose the wanted option.