    Chapter 3 - System Setup 3 - 37
    3-7-3-2 External Peripherals (Optional) The external printers are connected via Ethernet (TCP/IP network).
    Color Printer, HP OfficeJet PRO K550
    Installation instructions are available in the  Printer Driver Installation Manual for EchoPAC PC 
    and Vivid 7 , Direction Number FC294837.
    Color Laser Printer, HP  Color LaserJet 3600n
    Please use the installation instructions  from HP, delivered with the printer. 
    Software driver installation instructions are available in the  Printer Driver Installation Manual for 
    EchoPAC PC and Vivid 7 , Direction Number FC294837.
    Color Printer, HP Inkjet 1200DTN (Replaces HP Inkjet 1100DTN)
    Installation instructions are available in the HP Inkjet 1200 Installation Manual for EchoPAC PC 
    and Vivid 7 , Direction Number FC194654.
    Color Printer, HP Inkjet 1100DTN (Obsolete from factory)
    Installation instructions  are available in the HP Inkjet 1100 Installation Manual for EchoPAC PC 
    and Vivid 7 , Direction Number EP194599.
    Color Printer, HP Deskjet 6127 (Obsolete from factory)
    Installation instructions are available in the  HP Deskjet 6127 Installation Manual for EchoPAC 
    PC and Vivid 7 , Direction Number EP194516.
    Color Printer, HP Deskjet 990 Cxi Network Printer (Obsolete)
    (This printer replaces the EPSO N 980N printer, see below.)
    Installation instructions are available in the  HP Deskjet 990 Cxi Network Printer Installation 
    Manual , Direction Number EP194422.
    The printer has been delivered in two versions: - EP100422, 110 VAC.
    - EP100423, 220 VAC.
    Color Inkjet Printer, EPSON 980N. (Obsolete)
    Installation instructions are available in the  EchoPAC PC Service Manual, 
    Direction Number EP091298. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 38 Section 3-7 - .  Configuration
    3-7-3-2      External Pe ripherals (Optional) (cont’d)
    Lexmark Color LaserWriter Network Printer.
    These versions are available:
    - 100 VAC: * Lexmark c762 Color Printer
    * Lexmark c752 Color Printer (Going Obsolete)
    * Lexmark c750 Colo r Printer (Obsolete)
    - 110 VAC: * Lexmark c762 Color Printer
    * Lexmark c752 Color Printer (Going Obsolete)
    * Lexmark c750 Colo r Printer (Obsolete)
    - 220 VAC: * Lexmark c762 Color Printer
    * Lexmark c752 Color Printer (Going Obsolete)
    * Lexmark c750 Colo r Printer (Obsolete)
    - Installation of the  Lexmark c762n Color Printer  and Lexmark c752n Color Printer  is 
    described in  Lexmark c752n/c762n Color Printer Installation Manual , Direction Number 
    FC194475.
    - Installation of the  Lexmark c750n Color Printer  is described in Lexmark c750 Color Printer 
    Installation Manual , Direction Number FC194252.
    3-7-4 Available Probes
    See Section 9-26 - Probes  on page 9-80.
    3-7-5 Video Specification
    See 5-8-9 - Video Specifications  on page 5-122. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 39
    3-7-6 Software Options Configuration
    3-7-6-1 Software Option Installation A Password (Software Option String) enables a softw are option or a combination of software options. 
    This password is specific for each Vivid 7.
    3-7-6-2 Install Software Option 1.) Press CONFIG
     (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    2.) Select  ADMIN
     (lower part of window), 
    3.) Select the  SYSTEM ADMIN
     tab.
      
    4.) Select NEW to open the New Key dialog where you type the Software Option Key (Alphanumeric 
    String).
        
    Figure 3-29   System Admin screen
    Figure 3-30   Type Software Option Key (Alphanumeric string)
    SYSTEM ADMIN TAB NEW
    ADMIN
    Field for Software Option Key (Alphanumeric String) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 40 Section 3-7 - .  Configuration
    3-7-6-2      Install Software Option (cont’d) 5.) Type the Password (Software Option Key (Alphanumeric string)).
    6.) Select OK to save the new setting.
    7.) Close the Config window.
    CAUTIONIncorrect password entry will resu lt in loss of system options. 
    If password is incorrect, please contact your local GE Service representative or the Online 
    Center. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 41
    Section 3-8
    Connectivity Overview
    NOTE: If you are new or unfamiliar to co nnectivity on Vivid 7, you should read Section 5-15 
    Connectivity Theory on page 5-148  before you continue with the next descriptions and 
    procedures.
    3-8-1 Physical Connection
    3-8-1-1 Stand-alone Vivid 7
    No network connection needed.
    3-8-1-2 “Sneaker Net” Environment No network connection needed.
    3-8-1-3 Connection from Vivid 7 to an EchoPAC PC Workstation
    Direct Cable Connection from Vivid 7 to an  EchoPAC PC Workstation via a Crossover Cable
    You will only need a Crossover  Cable for network use to connect the two units this way.
    1.) Connect one end of the crossed network cabl e to the network connector on the Vivid 7.
    2.) Connect the other end to the network connector to  the EchoPAC PC Workstation, see chapter 3 in 
    the EchoPAC PC Workstation Service Manual.
    Connection via a Peer-to-Peer Network
    You will need a network hub and one network ca ble for each unit connected to the hub.
    See  Ethernet Switch / Hub on page 3-133 .
    Connection via Hospital Network
    You will need one network cable to co nnect the Vivid 7 to a wall outlet on the hospital’s network. 
    3-8-1-4 Connection between a Vivid 7 and a DICOM Server on a Network You will need one network cable.
    1.) Connect one end of the cable to  the Ethernet connector on Vivid 7
    2.) Connect the other end of the cable to the wall outlet
    If a Peer-to-Peer Network is connected to the hospi tal’s network, you may connect the Vivid 7 to Peer-
    to-Peer Network. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 42 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    Section 3-9
    Connectivity Setup - Software  v7.x (BT’08), v6.x, v5.x and v4.x 
    NOTE: If connected to a stand-alone network (P eer-to-Peer network with a Vivid 7 scanner, an 
    EchoPAC PC workstation and an optional netw ork printer), you should use default delivery 
    settings. 
    3-9-1 Introduction
    To be able to use the network functions when con nected to a hospital network, the scanner must have 
    a proper network address. 
    • Before you can set up the scanner, you need to collect some information. 
    •The  Pre-quote Worksheet for DICOM Network Information on page 2-14  can be used for gathering 
    this information. 
    • Typical source for this information is the network administrator.
    NOTE: The remote use of Common Se rvice Platform is not compatible with use of DHCP. If DHCP is 
    turned on, remote dia gnostics will be disabled.
    3-9-2 Contents in this Section
    •3-9-3 Compatibility on page 3-43
    • 3-9-4 Select TCP/IP Set-up Screen on page 3-44
    • 3-9-5 Changing the Computer Name, AE Title and/or Port Number (Port No.) on page 3-45
    • 3-9-6 Set the Scanner’s Network Information -  S/W v4.2/v5.2/v6.1/v7.x and later on page 3-48
    • 3-9-7 Set the Scanner’s Network Information  - S/W v4.0/v4.1/v5.0/v5.1/v6.0 on page 3-50
    • 3-9-8 Advanced Settings on page 3-51
    • 3-9-9 DHCP Configuration on page 3-52
    • 3-9-10 Set the Remote Archive’s Network Information on page 3-55
    • 3-9-11 Save the New Settings on page 3-55
    • 3-9-14 Set Up Connection to a DICOM Server in a Network on page 3-58
    • 3-9-15 Export Configuration on page 3-70
    • 3-9-16 Create a New Dataflow on page 3-75
    • 3-9-17 Query/Retrieve (Q/R) Setup for So ftware v7.x/v6.x/v5.x/v4.x on page 3-78
    • 3-9-18 Mapping of Parameters from Vivid 7 to DICOM on page 3-85 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 43
    3-9-3 Compatibility
    3-9-3-1 Vivid 7 with Software Version v7.x Vivid 7 with software version v7.x can communicate with:
    • EchoPAC PC with software  version v7.x or newer.
    • Image Vault 4.3 or newer.
    3-9-3-2 Vivid 7 with Software Version v6.x Vivid 7 with software version v6.x can communicate with:
    • EchoPAC PC with software version v6.x (use the  same, or newer software version as used on the 
    Vivid 7)
    • Image Vault 4.1 or newer
    • EchoServer 7
    3-9-3-3 Vivid 7 with Software Version v5.x Vivid 7 with software version v5.x can communicate with:
    • EchoPAC PC with software version v5.x (use the  same, or newer software version as used on the 
    Vivid 7)
    • Image Vault 4 or newer
    • EchoServer 7
    3-9-3-4 Vivid 7 with Software Version v4.x Vivid 7 with Software Versio n v4.x can communicate with:
    • EchoPAC PC with software version v4.x (use the  same, or newer software version as used on the 
    Vivid 7)
    • EchoPAC PC with software version v5.x
    • Image Vault 4 or newer
    • EchoServer 7 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 44 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-4 Select TCP/ IP Set-up Screen
    1.) Press Config (F2 ) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    2.) If not already selected, select  Connectivity from the bottom row of “buttons” on the screen.
    3.) Select the  TCP/IP TAB (it is named  Tcpip). The resulting screen gives  you an overview of many of 
    the network settings for Vivid 7. 
    .
    NOTICEIf possible, do not change  the Computer Name . If you change the Computer Name, you will loose 
    the access to all images in the Local Archive (on t he internal hard disk drive). If you must change the 
    Computer Name, please refer to  3-9-5 Changing the Computer Name, AE Title and/or Port Number 
    (Port No.) on page 3-45 .
    Figure 3-31   TCP/IP Overview screen for Vivid 7
    COMPUTER NAME / AE TITLE: 
    FOR Vivid 7, THIS NAME IS ON THE FORM: 
    VIVIDX-00NNNN , WHERE “00NNNN” IS A 
    NUMBER (NNNN IS THE SCANNER’S SERIAL 
    NUMBER). 
    PORT NO:
    DEFAULT PORT NUMBER: 104
    IP SETTINGS:
    IP SETTINGS AREA FOR THE Vivid 7 
    SCANNER. 
    DEFAULT SETUP:
    IP-ADDRESS: 10.0.0.3
    SUBNET MASK: 255.255.255.0
    GATEWAY: 0.0.0.0
    REMOTE ARCHIVE SETUP AREA:
    IP SETTINGS AREA FOR A REMOTE 
    ARCHIVE.
    EXAMPLE:  ECHOPAC PC (ECHOPAC7) .
    DEFAULT SETUP:
    REMOTE ARCHIVE IP-ADDR:  10.0.0.4
    REMOTE ARCHIVE NAME (THE NAME OF 
    THE PC OR SERVER WITH THE REMOTE 
    ARCHIVE):  ECHOPAC7-000001
    DETAILED DICOM LOG:
    SELECT TO TURN  DETAILED DICOM LOG  ON.
    (IT WILL BE TURNED OFF AFTER A RESTART)
    TURN ON TO TROUBLESHOOT DICOM 
    ISSUES.
    ADVANCED SETTINGS:
    USE  ADVANCED SETTINGS  IF YOU NEED 
    TO CHANGE Vivid 7’s IP SETTINGS OR 
    TURN DHCP ON OR OFF. SAVE SETTINGS
    SELECT 
    SAVE SETTINGS  TO 
    ARCHIVE ANY CHANGES YOU HAVE 
    DONE TO THE TCP/IP SETTINGS
    SELECT MODIFY TO CHANGE 
    THE COMPUTER NAME 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 45
    3-9-5 Changing the Computer  Name, AE Title and/or Port Number (Port No.)
    NOTE: To change the  AE Title and/or Port No. , please refer to:  3-9-5-3 Change AE Title and/or Port 
    Number (Port No.) on page 3-47 .
    3-9-5-1 Before you change the Computer Name Move all images to remote media before you change the Computer Name. See:  4-2-15 Space 
    Management on page 4-20 , or refer to the User Manual, for a detailed description.
    3-9-5-2 Change the Computer Name
    .
    1.) When in the  Tcpip screen, select  Modify. This will bring up a Warning,  see the illustration below.
    2.) Select  Ok to continue or  Cancel if you don’t want to change the  Computer Name.
    If you selected  Ok, the  Computer Name  will be editable.
    NOTICEIf possible, do not change  the Computer Name . If you change the Computer Name, you will loose 
    the access to all images in the Local Archive (on t he internal hard disk drive). If you must change the 
    Computer Name, please refer to  3-9-5 Changing the Computer Name, AE Title and/or Port Number 
    (Port No.) on page 3-45 .
    Figure 3-32   Select ModifyFigure 3-33   Warning
    MODIFY 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 46 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-5-2      Change the Computer Name (cont’d) 3.) Edit the  Computer Name .
    4.) Select  Save settings  to store your changes. This  will bring up a new Warning screen.
    5.) Select  Ok to save your changes or  Cancel to return without saving any changes.
    6.) Reboot Vivid 7 to activate the settings  or continue with other Tcpip set-up tasks.
    Figure 3-34   Edit Computer Name
    Figure 3-35   Warning
    EDIT THE 
    COMPUTER NAME 
    						