Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Vivid 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 858
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 2 - Site Preparations 2 - 1
    Chapter 2
    Site Preparations
    Section 2-1
    Overview
    2-1-1 Purpose of Chapter 2
    This chapter provides the information required to  plan and prepare for the installation of a Vivid 7. 
    Included are descriptions of the fa cility and electrical needs to be met by the  purchaser of the unit. 
    2-1-2 Contents in Chapter 2
    Table 2-1    Contents in Chapter 2
    SectionDescriptionPage Number
    2-1 Overview 2-1
    2-2 General Console Requirements 2-2
    2-3 Facility Needs 2-10 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    2 - 2 Section 2-2 - General Console Requirements
    Section 2-2
    General Console Requirements
    2-2-1 Console Environmental Requirements
    2-2-1-1 If the unit is very cold or hot
     
    2-2-1-2 Environmental specifications
    2-2-1-3 Cooling The cooling requirement for the Vivid 7 is 3500 BTU/hr . This figure does not include cooling needed for 
    lights, people, or other equipment in the room. Each  person in the room places an additional 300 BTU/
    hr. demand on the cooling system.
    2-2-1-4 Lighting Bright light is needed for system  installation, updates and repairs. However, operator and patient 
    comfort may be optimized if the room light is su bdued and indirect. Therefore a combination lighting 
    system (dim/bright) is recommended. Keep in mind t hat lighting controls and dimmers can be a source 
    of EMI which could degrade image quality. These co ntrols should be selected to minimize possible 
    interference.
    CAUTIONIf the unit is very cold or hot, do not turn on it s power until it has had a chance to acclimate to 
    its operating environment.
    Table 2-2    Vivid 7 Acclimate Time
    DON’T STORE AT  THESE 
    TEMPERATURES KEEP STORAGE TEMPERATURE WITHIN THESE LIMITS
    oC
    -40-35-30-25-20-15-10-5 0 5 1015202530354045505560
    oF-40 -31 -22 -13 -4 5 14 23 32 41 50 59 68 77 86 95 104 113 122 131 140
    Hrs 201816141210864200000002468
    Table 2-3    Environmental Specifications for Vivid 7 Scanners
    Operating temperatureStorage temperatureHumidityHeat dissipation
    10 to 35 oC (50 to 95 oF) -20 to 60 oC (-4 to 140 oF) < 90% rH non-condensing 3500 BTU pr hour 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 2 - Site Preparations 2 - 3
    2-2-2 Electrical Requirements
    2-2-2-1 General Requirements NOTE: GE Medical Systems requires a dedicated power a nd ground for the proper operation of its Ultrasound 
    equipment. This dedicated power shall originate at  the last distribution panel before the system. 
    Sites with a mains power system with defined Neutral and Live: 
    The dedicated line shall consist of  one phase, a neutral (not shared with any other circuit), and a full 
    size ground wire from the distribution panel to the Ultrasound outlet. 
    Sites with a mains power system without a defined Neutral: 
    The dedicated line shall consist of on e phase (two lines), not shared with any other circuit, and a full 
    size ground wire from the distribution panel to the Ultrasound outlet. 
    Please note that image artifacts ca n occur, if at any time within the facility, the ground from the main 
    facilitys incoming power source to th e Ultrasound unit is only a conduit. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    2 - 4 Section 2-2 - General Console Requirements
    2-2-2-2 Electrical Requirements for Vivid 7  Dimension, Vivid 7 and Vivid 7 PRO
    Electrical Specifications for Vivid 7 and Vivid 7 PR O. Monitor and on board peripherals are included.
    Table 2-4    Electrical Specifications for Vivid 7   sheet 1 of 3
    GE VINGMED 
    PART NUMBER
    DESCRIPTIONVOLTAGETOLERANCESCURRENTFREQUENCY
    FD000010 VIVID 7 DIMENSION CONSOLE, 
    LCD MONITOR 230 VAC ±10% 5 A 50-60 Hz
    FD000020  VIVID 7 DIMENSION CONSOLE, 
    LCD MONITOR 100-120 VAC ±10% 10 A 50-60 Hz
    FD000120  VIVID 7 PRO CONSOLE, 
    LCD MONITOR 230 VAC ±10% 5 A 50-60 Hz
    FD000130VIVID 7 PRO CONSOLE, 
    LCD MONITOR100-120 VAC±10%10 A50-60 Hz
    FD000140 VIVID 7 DIM.  W/  MULTI-DIM. & 
    VOLUME, 
    LCD MONITOR230 VAC±10%5 A50-60 Hz
    FD000150 VIVID 7 DIM.  W/  MULTI-DIM. & 
    VOLUME, 
    LCD MONITOR100-120 VAC±10%10 A50-60 Hz
    FD000160  VIVID 7 DIM.  W/  MULTI-DIM. 
    IMAGING, 
    LCD MONITOR 230 VAC ±10% 5 A 50-60 Hz
    FD000170  VIVID 7 DIM.  W/  MULTI-DIM. 
    IMAGING, 
    LCD MONITOR  100-120 VAC ±10% 10 A 50-60 Hz
    FD000180  VIVID 7 DIM.  W/  VOLUME 
    ULTRASOUND, 
    LCD MONITOR 230 VAC ±10% 5 A 50-60 Hz
    FD000190 VIVID 7 DIM.  W/  VOLUME 
    ULTRASOUND, 
    LCD MONITOR100-120 VAC±10%10 A50-60 Hz
    FC000890VIVID 7 DIMENSION 
    MULTIDIMENSIONAL & Volume 
    (BT’06) 230 VAC230 VAC±10%5 A50-60 Hz
    FC000900VIVID 7 DIMENSION 
    MULTIDIMENSIONAL & Volume 
    (BT’06) 120 VAC110-120 VAC±10%10 A50-60 Hz
    FC000910 VIVID 7 DIMENSION (BT’06) 230 VAC 230 VAC ±10% 5 A 50-60 Hz
    FC000920 VIVID 7 DIMENSION (BT’06) 
    100-120 VAC 110-120 VAC ±10% 10 A 50-60 Hz
    FC000930 VIVID 7 PRO (BT’06) 230 VAC 230 VAC ±10% 5 A 50-60 Hz
    FC000940VIVID 7 PRO (BT’06) 100-120 VAC110-120 VAC±10%10 A50-60 Hz
    FC000950VIVID 7 DIMENSION 
    MULTIDIMENSIONAL IMAGING 
    (BT’06) 230 VAC230 VAC±10%5 A50-60 Hz 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 2 - Site Preparations 2 - 5
    FC000960VIVID 7 DIMENSION 
    MULTIDIMENSIONAL IMAGING 
    (BT’06) 120 VAC110-120 VAC±10%10 A50-60 Hz
    FC000970 VIVID 7 DIMENSION VOLUME 
    ULTRASOUND (BT’06) 230 VAC 230 VAC ±10% 5 A 50-60 Hz
    FC000980 VIVID 7 DIMENSION VOLUME 
    ULTRASOUND (BT’06) 120 VAC 110-120 VAC ±10% 10 A 50-60 Hz
    FC000760 VIVID 7 DIMENSION, 
    MULTIDIMENSIONAL RENDERING 
    AND VOLUME 230 VAC ±10% 5 A 50-60 Hz
    FC000770VIVID 7 DIMENSION, 
    MULTIDIMENSIONAL RENDERING 
    AND VOLUME110-120 VAC±10%10 A50-60 Hz
    FC000780VIVID 7 DIMENSION230 VAC±10%5 A50-60 Hz
    FC000790VIVID 7 DIMENSION100-120 VAC±10%10 A50-60 Hz
    FC000800 VIVID 7 PRO (BT’05), 230 VAC ±10% 5 A 50-60 Hz
    FC000810 VIVID 7 PRO (BT’05) 100-120 VAC ±10% 10 A 50-60 Hz
    FC000820 VIVID 7 DIMENSION MULTI-
    DIMENSIONAL IMAGING (BT’05) 230 VAC ±10% 5 A 50-60 Hz
    FC000830VIVID 7 DIMENSION MULTI-
    DIMENSIONAL IMAGING (BT’05)100-120 VAC±10%10 A50-60 Hz
    FC000840VIVID 7 DIMENSION w/ VOLUME 
    ULTRASOUND (BT’05)230 VAC±10%5 A50-60 Hz
    FC000850VIVID 7 DIMENSION w/ VOLUME 
    ULTRASOUND (BT’05)100-120 VAC±10%10 A50-60 Hz
    FC000699 Vivid 7 (BT ’04) 100-120 VAC ±10% 10 A 50-60 Hz
    FC000660 Vivid 7 PRO (BT ’04) 100-120 VAC ±10% 10 A 50-60 Hz
    FC000650 Vivid 7 PRO (BT ’04) 230 VAC ±10% 5 A 50-60 Hz
    FC000640Vivid 7 Dimension (BT ’04)100-120 VAC±10%10 A50-60 Hz
    FC000630Vivid 7 Dimension (BT ’04)230 VAC±10%5 A50-60 Hz
    FC000620Vivid 7 Dimension with 4D (BT’04)100-120 VAC±10%10 A50-60 Hz
    FC000610 Vivid 7 Dimension with 4D (BT’04) 230 VAC ±10% 5 A 50-60 Hz
    FC000440 Vivid 7 PRO with RFI 100-120 VAC ±10% 10 A 50-60 Hz
    FC000430 Vivid 7 PRO with RFI 230 VAC ±10% 5 A 50-60 Hz
    FC000420Vivid 7 with RFI100-120 VAC±10%10 A50-60 Hz
    FC000410Vivid 7 with RFI230 VAC±10%5 A50-60 Hz
    FC000340Vivid 7 PRO (BT ’03)100-120 VAC±10%10 A50-60 Hz
    FC000330 Vivid 7 PRO (BT ’03) 230 VAC ±10% 5 A 50-60 Hz
    FC000320 Vivid 7 (BT ’03) 100-120 VAC ±10% 10 A 50-60 Hz
    FC000310 Vivid 7 (BT ’03) 230 VAC ±10% 5 A 50-60 Hz
    Table 2-4    Electrical Specifications
     for Vivid 7 (cont’d) sheet 2 of 3
    GE VINGMED 
    PART NUMBER
    DESCRIPTIONVOLTAGETOLERANCESCURRENTFREQUENCY 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    2 - 6 Section 2-2 - General Console Requirements
    2-2-2-3 Inrush Current Inrush current is not a factor to consider due to  the inrush current limiting properties of the power 
    supplies. 
    Maximum inrush current at startup:
    • 230 VAC: 12 A
    • 120 VAC: 7 A
    2-2-2-4 Site Circuit Breaker
    It is recommended that the br anch circuit breaker for the unit be readily accessible.
    2-2-2-5 Site Power Outlets A dedicated AC power outlet must be within reach of  the unit without extension cords. Other outlets 
    adequate for the external peripherals, medical and test  equipment needed to support this unit must also 
    be present within 1 m (3.2 ft.) of th e unit. Electrical installation must meet all current local, state, and 
    national electrical codes.
    2-2-2-6 Unit Power Plug If the unit arrives without a power plug, or with the  wrong plug, you must contact your GE dealer or the 
    installation engineer must s upply what is locally required.
    2-2-2-7 Power Stability Requirements Voltage drop-out
    Max 10 ms.
    Power Transients
    (All applications)
    Less than 25% of nominal peak volt age for less than 1 millisecond for any type of transient, including 
    line frequency, synchronous, asynchronous, or aperiodic transients.
    FC000210Vivid 7 (BT ’02)100-120 VAC±10%10 A50-60 Hz
    FC000200Vivid 7 (BT ’02)230 VAC±10%5 A50-60 Hz
    FC000190Vivid 7 PRO (BT ’02)100-120 VAC±10%10 A50-60 Hz
    FC000180 Vivid 7 PRO (BT ’02) 230 VAC ±10% 5 A 50-60 Hz
    FC000060 Vivid 7 (BT ’01) 100-120 VAC ±10% 10 A 50-60 Hz
    FC000030 Vivid 7 (BT ’01) 230 VAC ±10% 5 A 50-60 Hz
    CAUTIONPOWER OUTAGE MAY OCCUR. The 
    Vivid 7 requires a dedicated single branch circuit. To avoid 
    circuit overload and possible loss of critical  care equipment, make sure you DO NOT have any 
    other equipment operating on the same circuit.
    Table 2-4    Electrical Specifications
     for Vivid 7 (cont’d) sheet 3 of 3
    GE VINGMED 
    PART NUMBER
    DESCRIPTIONVOLTAGETOLERANCESCURRENTFREQUENCY 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 2 - Site Preparations 2 - 7
    2-2-3 EMI Limitations
    Ultrasound machines are susceptible to Electromagne tic Interference (EMI) from radio frequencies, 
    magnetic fields, and transients in the air or wiring.  They also generate EMI. The Vivid 7 complies with 
    limits as stated on the EMC label. Ho wever there is no guarantee that interference will not occur in a 
    particular installation.
    Possible EMI sources should be identi fied before the unit is installed. 
    Electrical and electronic equipment may produce EMI  unintentionally as the result of a defect. These 
    sources include: 
    • medical lasers, 
    • scanners, 
    • cauterizing guns, 
    • computers, 
    •monitors, 
    • fans, 
    • gel warmers, 
    • microwave ovens,
    • light dimmers 
    • portable phones. 
    The presence of a broadcast station or broadcast van may also cause interference.
    See  Table 2-5 on page 2-8  for EMI Prevention tips. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    2 - 8 Section 2-2 - General Console Requirements
    2-2-3 EMI Limitations (cont’d).
    Table 2-5    EMI Prevention/abatement
    EMI RuleDetails
    Be aware of RF sources Keep the unit at least 5 meters or 15 feet away from other EMI sources. Special shielding may 
    be required to eliminate interference problem
    s caused by high frequency, high powered radio 
    or video broadcast signals.
    Ground the unit Poor grounding is the most likely reason a unit
     will have noisy images. Check grounding of the 
    power cord and power outlet.
    Replace all screws, RF  gaskets, covers, cores After you finish repairing or updating the syste
    m, replace all covers and tighten all screws. Any 
    cable with an external connection requires a ma gnet wrap at each end. Install the Card Rack 
    cover over the Card Rack. Loose or missing cove rs or RF gaskets allow radio frequencies to 
    interfere with the ultrasound signals.
    Replace broken RF  gasketsIf more than 20% or a pair of the fingers on an RF  gasket are broken, replace the gasket. Do 
    not turn on the unit until any  loose metallic part is removed.
    Do not place labels where 
    RF gaskets touch metalNever place a label where RF gaskets meet  the unit. Otherwise, the gap created will permit 
    RF leakage. Or, if a label has been found in such a position, move the label.
    Use GE specified 
    harnesses and peripheralsThe interconnect cables are grounded and require  ferrite beads and other shielding. Also, 
    cable length, material, and routing are all im portant; do not change from what is specified.
    Take care with cellular  phones Cellular phones may transmit a 5 V/m sig
    nal; that could cause image artifacts. 
    Properly dress peripheral  cables Do not allow cables to lie across the top of th
    e Card Rack or hang out of the peripheral bays. 
    Loop the excess length for peripheral cables in side the peripheral bays. Attach the monitor 
    cables to the frame. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 2 - Site Preparations 2 - 9
    2-2-4 Probes Environmental Requirements
    2-2-5 Time and Manpower Requirements
    Site preparation takes time. Begin Pre-installation c hecks as soon as possible, if possible, six weeks 
    before delivery, to allow enough time to make any changes.
     
    Table 2-6    Operation and Storage Temperatures for Probes.
    ElectronicPA M P T E
    Operation: 10 to 40 
    oC (50 to 104 oF) 5 to 42.7 oC (41 to 108,9 oF)
    Storage: -20 to 50 
    oC (-4 to 122 oF) -20 to 60 oC (-4 to 140 oF)
    Temperatures in degrees Celsius (oC) conversion to degrees F: (oF) = (oC * 9/5) + 32
    CAUTIONSystems and PAMPTE probes are designed for storage temperatures of -20 oC to + 60 oC 
    (-4 
    oF to +140 oF). 
    Electronic probes are designed for storage temperatures of -20 
    oC to + 50 oC (-4 oF to +122 oF).
    When exposed to large temperature variations, the product should be kept at room temperature 
    the needed time to stabilize its temperature before use. 
    Refer to  Table 2-2 Vivid 7 Acclimate Time on page 2-2  to determine the needed settlement 
    time.
    CAUTIONThe Vivid 7 weighs at least 190 kg (419 lbs.) when ready to use, depending on installed monitors 
    and peripherals. Care must be used when moving it or replacing its parts. Failure to follow the 
    precautions listed could result in injury, uncontrolled motion and costly damage.
    ALWAYS:
    Be sure the pathway is clear. Use slow, ca reful motions.Use two people when moving on 
    inclines or lifting more than 16 kg (35 lbs). 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    2 - 10 Section 2-3 - Facility Needs
    Section 2-3
    Facility Needs
    2-3-1 Purchaser Responsibilities
    The work and materials nee ded to prepare the site is the responsibility of the purchaser. Delay, 
    confusion, and waste of manpower can be avoided by  completing pre installation work before delivery. 
    Purchaser responsibility includes:
    • Procuring the materials required.
    • Completing the preparations before delivery of the ultrasound system.
    • Paying the costs for any alterations and modification s not specifically provided in the sales contract.
    NOTE: All electrical installations th at are preliminary to the positioning of the equipment at the site 
    prepared for the equipment must be performed  by licensed electrical contractors. Other 
    connections between pieces of electrical equi pment, calibrations, and testing must also be 
    performed by qualified personnel. The products  involved (and the accompanying electrical 
    installations) are highly sophi sticated and special engineering competence is required. All 
    electrical work on these products must comply  with the requirements of applicable electrical 
    codes. The purchaser of GE  equipment must only utilize q ualified personnel to perform 
    electrical servicing on the equipment. 
    The desire to use a non–listed or customer provided product or to place an approved product further 
    from the system than the interfac e kit allows, presents challenges to the installation team. To avoid 
    delays during installation, such variances should be made known to the individuals or group performing 
    the installation at the earliest possible  date (preferably prior to the purchase).
    The ultrasound suite must be clean prior to delivery of the machine. Carpet is not recommended 
    because it collects dust and creates static. Potent ial sources of EMI (electromagnetic interference) 
    should also be investigated before  delivery. Dirt, static, and EMI can negatively impact system reliability. 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Vivid 7 Service Manual