    Chapter 3 - System Setup 3 - 47
    3-9-5-3 Change AE Title and/or Port Number (Port No.) NOTE: You don’t need to use the Modify butt on when changing AE Title and/or Port No.
    1.) To change  AE Title and/or Port No ., edit the respective fields.
    2.) Select  Save settings  to store your changes. This  will bring up a new Warning screen.
    3.) Select  Ok to save your changes or  Cancel to return without saving any changes.
    4.) Reboot Vivid 7 to activate the settings  or continue with other Tcpip set-up tasks.
    Figure 3-36   AE Title and Port No.
    Figure 3-37   Warning 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 48 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-6 Set the Scanner’s Ne twork Information - S/W v4.2/v5.2/v6.1/v7.x and later
    NOTE: For software version v4.0/v4.1/v5.0/v5.1/v6.0, see:  3-9-7 Set the Scanner’s Network 
    Information - S/W v4.0/v4.1/v5.0/v5.1/v6.0 on page 3-50
    .
    1.) When in the  Tcpip (TCP/IP) screen, select  Advanced Settings to display the Network 
    Connections  screen.
    2.) Right-click  Local Area Connection  and select Properties from the pop-up menu . This will display 
    the  Local Area Connection Properties  screen.
    Figure 3-38   Select Advanced Settings
    Figure 3-39   Network Connections screen
    ADVANCED SETTINGS 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 49
    3-9-6 Set the Scanner’s Ne twork Information - S/W v4.2/v5.2/v6.1/v7.x and later 
    (cont’d)
    3.) Select (highlight)  Internet Protocol (TCP/IP)  and select Properties to display the  Internet 
    Protocol (TCP/IP) Properties  screen.
    The  Internet Protocol (TCP/IP) Properties  screen is used to set the following parameters: 
    If using DHCP
    Check the  Obtain an IP address automatically  check box.
    If not using DHCP 
    1.) Uncheck the  Obtain an IP address automatically  check box, and enter the:
    - IP Address for Vivid 7. (Default IP Ad dress setting from factory: 10.0.0.3).
    - Subnet Mask for Vivid 7. (Default Subnet Ma sk setting from factory: 255.255.255.0).
    - IP address for Default Gateway. (Default Gatewa y setting from factory IP address: 0.0.0.0).
    2.) Select  OK to keep your changes or  Cancel to close the  Internet Protocol (TCP/IP) Properties  
    screen without keeping any changes.
    3.) Select  OK to close (and save) the  Local Area Connection Properties  dialog.
    4.) Select the “x” in the upper right corner to close the  Network Connections window.
    Figure 3-40   Local Area Connection Properties
    Figure 3-41   Internet Protocol (TCP/IP) Properties screen 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 50 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-7 Set the Scanner’s Network Info rmation - S/W v4.0/v4.1/v5.0/v5.1/v6.0
    In the IP settings  area of the screen, enter the:
    1.) IP Address for scanner. (Default IP Ad dress setting from factory: 10.0.0.3).
    2.) Subnet Mask for scanner. (Default Subnet  Mask setting from factory: 255.255.255.0).
    3.) IP address for Default Gateway. (Default Gate way setting from factory IP address: 0.0.0.0).
    4.) Select  Save Settings  to save any changes you have entered.
    5.) In addition, AE Title for the scanner must  be entered in the DICOM server’s setup. 
    NOTE: The use of DHCP is not supported by the Servic e Platform, so if DHCP is enabled, the Common Service 
    Platform must be disabled.
    .
    Figure 3-42   TCP/IP Set-up for Vivid 7
    IP ADDRESS
    SUBNET MASK
    DEFAULT GATEWAY
    AE TITLE
    SAVE SETTINGS 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 51
    3-9-8 Advanced Settings
    NOTE: New for software V4.0 (BT’04). Advanced Settings  give you access to the  Network and Dial-up Connections  screen where you can 
    set up special network connections and ad just settings for network connectivity.
    On Vivid 7, Advanced Settings is used to enable/disable DHCP.
    .
    Figure 3-43   Advanced Settings
    ADVANCED SETTINGS
    TO BE USED FOR 
    ENABLING/DISABLING 
    DHCP. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 52 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-9 DHCP Configuration
    3-9-9-1 Pre-Requisite • Software version v4.0 or higher.
    NOTE: The remote connectivity fo r the Global Service Platform will not function if DHCP is active.
    3-9-9-2 Turn DHCP ON  Follow the instructions below to  configure the Vivid 7 to use DHCP
    .
    1.) When in the TCP/IP screen, select  ADVANCED SETTINGS to display the Network and Dial-up 
    Connections  screen, see Figure 3-45 Network Connections on page 3-53 .
    Figure 3-44   Advanced Settings
    ADVANCED SETTINGS 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 53
    3-9-9-2      Turn DHCP ON (cont’d) 2.) Select Local Area Connection
    3.) Select  Properties  to display the  Local Area Connection Properties
    4.) Scroll down to  Internet Protocol (TCP/IP)  and select it.
    Figure 3-45   Network Connections 
    Figure 3-46   Local Area Connection Properties 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 54 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-9-2      Turn DHCP ON (cont’d) 5.) Select Properties
    6.) Select  Obtain an IP address automaticallyFigure 3-47   Local Area Connection Status 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 55
    3-9-10 Set the Remote Arch ive’s Network Information
    To be able to connect to a remote archive, on a remote computer or server, you must configure Vivid 7 
    to communicate with it.
    In the Remote Archive Setup area of the Tcpip screen, see example in  Figure 3-48, enter the;
    1.) Remote Archive IP address. (Default  IP Address from factory: 10.0.0.4).
    2.) Remote Archive Name. (Default Remote Ar chive Name from factory: ECHOPAC7-000001).
    .
    3-9-11 Save the New Settings
    1.) Select SAVE SETTINGS to save the new settings. 
    .
    A Warning is displayed on the  screen, see illustration below.
    2.) The new settings are saved to a common settings file. After a restart, the settings are also included  in other screens.
    3.) Restart Vivid 7 to activate the changes. Figure 3-48   TCP/IP Set-up for Vivid 7
    Figure 3-49   Save New TCP/IP settings Figure 3-50   Warning
    IP ADDRESS FOR THE 
    REMOTE ARCHIVECOMPUTER NAME FOR 
    THE REMOTE ARCHIVE
    SAVE SETTINGS 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 56 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-12 Remote Path
    Remote Path is used for Save As, Export from Q-A nalysis, and for exporting error logs with the  
    command. 
    Follow this procedure to Configure Vivid 7 to use the  Remote Path.
    1.) Press  Config (F2 ) and log on as  ADM, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    2.) If not already selected, select  Connectivity from the bottom row of “buttons” on the screen.
    3.) Select the  Tools tab.
    4.) Enter the  Remote Path.
    Example:  \\3.187. 187.30\EXPORT\
    5.) Use the  Check button to verify the connection. Figure 3-51   Tools screen
    REMOTE PATH
    CONFIGURABLE 
    REMOTE PATH USER 
    						