    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 19
    7-6-4-8 DICOM Select LOGS > DICOM  to view the DICOM log.
    Figure 7-20   Select DICOM
    Figure 7-21   Example of DICOM Log
    SELECT DICOM 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 20 Section 7-6 - Diagnostics Tools
    7-6-4-9 DCPower Select LOGS > DCPower  to view the DCPower Log .
    Figure 7-22   Select DCPower
    Figure 7-23   Example of DC Power Log Screen
    SELECT DCPower 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 21
    7-6-4-10 Utilities The UTILITIES  pulldown menu has two choices, as shown in  Figure 7-24.
    The choices on the  UTILITIES pulldown menu are:
    • Plot Log (see:  7-6-4-11 Plot Log on page 7-22 )
    • Plot Page (see:  7-6-4-12 Plot Page on page 7-23 )
    Figure 7-24   Utilities Category 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 22 Section 7-6 - Diagnostics Tools
    7-6-4-11 Plot Log Select UTILITIES >  Plot Log  to make a plot of the active log on the screen.
    Figure 7-25   Utilities Category
    Figure 7-26   Utilities, Plot Log (example) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 23
    7-6-4-12 Plot Page Select UTILITIES > Plot Page  to make a plot of the active (log-) page on screen.
    Figure 7-27   Utilities Category
    Figure 7-28   Utilities, Plot Page (example) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 24 Section 7-6 - Diagnostics Tools
    7-6-4-13 Search On the Text Search sub-menu of the Search catego ry, users enter case-sensitive text they wish to 
    search for in the logs.
    7-6-4-14 Filter This function is available for GE Service only. Figure 7-29   Search Category 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 25
    7-6-4-15 Exit The sub-menu, Exit Log Viewer, returns the us er to the Common Service Desktop Home Page.
    Figure 7-30   Exit Log Viewer Category 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 26 Section 7-6 - Diagnostics Tools
    7-6-5 Diagnostics Screen Overview
    When the Diagnostics Screen has been selected, two main tools are available: - Acquisition Diagnostics, see  7-6-5-1 GSUI Acquisition Diagnostics - Test All on page 7-27.
    - Common Diagnostics, see  Section 7-7 Common Diagnostics on page 7-48.
    Figure 7-31   Diagnostics Screen Home Page 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 27
    7-6-5-1 GSUI Acquisition Diagnostics - Test All One test is available,  Test All. 
    NOTE: Disconnect all probes before running this test. Test All runs a sequential test as listed below.
    NOTE: Relay Board (RLB) and Chain are not tested in this test. Some parts of the Transmitter (TX)  Board  are tested during the FEC Board test, but  other parts of the TX Board aren’t tested at all.
    A.) Units with FEP1: 1.) PC2IP
    2.) IMP (Image Port Board)
    3.) VME (VME data bus on the Back Plane)
    4.) SDP (Spectrum Doppler Processor)
    5.) RFT (Radio Frequency & Tissue Board)
    6.) FEC (Front-End Controller Board) (Some parts of TX board is also tested)
    7.) BF (Beamformer Boards)
    8.) RX (Receiver Board)
    9.) TXP (TX Power)
    B.) Units with FEP2: 1.) PC2IP
    2.) RFI (Radio Frequency Interface Board)
    3.) BF (Beamformer Boards)
    4.) RX (Receiver Board)
    5.) TXP (TX Power) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 28 Section 7-6 - Diagnostics Tools
    7-6-5-1      GSUI Acquisition  Diagnostics - Test All (cont’d)
    7-6-5-2  Instructions Frame
    Displays a test-specific page or the default instructions page.
    7-6-5-3 Status Frame Initially displays the last known stat us for the particular diagnostic. Once the diagnostic starts, the frame 
    displays the “current” stat us of all test results.
    Figure 7-32   User Interface Diagnostic Screen, Test All 
    						