Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Vivid 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 858
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 69
    7-8-3-4 System Test Screen Description
    7-8-3-5 Test Loops, System Figure 7-74   System Test Screen, Upper Part 
    Table 7-6    Test Loops, System - Test All, System - Break (FEP1)
    FIELDKEYDESCRIPTION
    TEST 
    LOOPS -IN SOME CASES YOU MAY WANT TO DO THE SAME TEST SEVERAL TIMES. 
    ENTER A NUMBER TO INDICATE HOW MANY TI
    MES YOU WANT THE TEST TO REPEAT ITSELF. 
    DEFAULT VALUE = 1..
    SYSTEM TEST ALL USE 
    TEST ALL
     TO START A SEQUENCE OF ALL THE  TESTS AVAILABLE IN THE TEST SCREEN. 
    THE DIAGNOSTIC TEST IS DONE IN THIS SEQUENCE ON UNITS WITH FEP1;
    1.PC2IP
    2.IMP (IMAGE PORT BOARD)
    3.VME (VME DATA BUS ON THE BACK PLANE)
    4.SDP (SPECTRUM DOPPLER PROCESSOR)
    5.RFT (RADIO FREQUENCY & TISSUE BOARD)
    6.FEC (FRONT-END CONTROLLER BOARD) 
    7.BF (BEAMFORMER BOARDS)
    8.RX (RECEIVER BOARD)
    9.TXP (TX POWER)
    NOTE: Relay Board (RLB) and Chain are not tested in this test. 
    NOTE: The Transmitter Board (TX) don’t have its own test, but some parts of  the TX board is tested during the FEC test. 
    NOTE: If the system must be rebooted,  do so. Continue with the testing from 
    the point at which you were required to stop.
    SYSTEM BREAK IF YOU HAVE STARTED A LOOPED TEST, YOU MAY SELECT 
    BREAK TO STOP THE TEST. 
    THE TEST WILL CONTINUE UNTIL THE CURRENT TEST HAS FINISHED.
    System
    Test Loops
    Mode 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 70 Section 7-8 - Acquisition Diagnostics
    7-8-3-5      Test Loops, System (cont’d)
    7-8-3-6 Mode (Calibration/Test Screen Selection) Table 7-7    Test Loops, System - Test All, System - Break (FEP2)
    FIELDKEYDESCRIPTION
    TEST 
    LOOPS -IN SOME CASES YOU MAY WANT TO 
    DO THE SAME TEST SEVERAL TIMES. 
    ENTER A NUMBER TO INDICATE HOW MANY TI MES YOU WANT THE TEST TO REPEAT ITSELF. 
    DEFAULT VALUE = 1..
    SYSTEM TEST ALL USE 
    TEST ALL
     TO START A SEQUENCE OF ALL THE  TESTS AVAILABLE IN THE TEST SCREEN. 
    THE DIAGNOSTIC TEST IS DONE IN THIS SEQUENCE ON UNITS WITH FEP2;
    1.) PC2IP
    2.) RFI (RADIO INTERFACE BOARD)
    3.) BF (BEAM FORMER BOARDS)
    4.) RX (RECEIVER BOARD)
    5.) TXP (TX POWER)
    NOTE: Relay Board (RLB) and Chain are not tested in this test. 
    NOTE: The Transmitter Board (TX) don’t have its own test, but some parts of  the TX board is tested during the FEC test. 
    NOTE: If the system must be rebooted,  do so. Continue with the testing from 
    the point at which you were required to stop.
    SYSTEM BREAK IF YOU HAVE STARTED A LOOPED TEST, YOU MAY SELECT 
    BREAK TO STOP THE TEST. 
    THE TEST WILL CONTINUE UNTIL THE CURRENT TEST HAS FINISHED.
    Table 7-8    Mode (Calibration/Test Screen selection)
    FIELDKEYDESCRIPTION
    MODE The buttons in this field are used to select the Calibration Screen or the Test Screen.
    Calibration Select this button to select the Calibration Screen.
    Test Select this button to select the Test Screen.  
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 71
    7-8-3-7 “Block Schematics” with Single Tests Select one of the buttons in the “block schematics”  in the upper part of the screen (or one of the 
    duplicate buttons in the lower row of the screen) to select a particular test. 
    Some of the tests are described after the figure below. 
    Figure 7-75   Diagnostics Test Screen, Upper Part
    BLOCK 
    SCHEMATICS, FEP2BLOCK 
    SCHEMATICS, FEP1 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 72 Section 7-8 - Acquisition Diagnostics
    7-8-3-8 Receiver Board (RX) Test
    7-8-3-9 Relay Board Test NOTE: A Test Attenuator, Part Number  AA200215 is needed for this test.
    Table 7-9    Receiver Board (RX) Test
    Test (Key)Sub-testsDescription
    RX
    Receiver Board (RX)
    Te s t  A l l Activates the following tests, in s
    equence, see description for each sub test;
    - RECV0
    - RECV1
    RECV.0 Max Gain test for RX using FEC signal generator.
    RECV.1Reduced Gain test for RX using FEC signal generator.
    Table 7-10    Relay Board (RLB) Test
    Test (Key)Sub-testsDescription
    RLB Relay Board (RLB)
    Te s t  A l l Activates the following tests, in sequence 
    (see the descriptions for the tests below);
    1. Conn. 1
    2. Conn. 2
    3. Conn. 3
    Conn. 1 Max Gain test for RX using the signal generator
     on the Front-End Controller (FEC) and a Test 
    Attenuator, Part Number AA200215, in left (active) probe connector.
    Conn. 2Max Gain test for RX using the signal generator  on the Front-End Controller (FEC) and a Test 
    Attenuator, Part Number AA200215, in mid (active) probe connector.
    Conn. 3Max Gain test for RX using the signal generator  on the Front-End Controller (FEC) and a Test 
    Attenuator, Part Number AA200215, in right probe connector. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 73
    7-8-3-10 Image Port Test When selecting the ImagePort Test, the scanner issues a test command to the local software running 
    on the I960RP processor on the Image Port board. 
    The test command will start the following tests;
    • test the interrupt based command/replay mechanism
    • reset the ImagePort board
    • test the PipeLink data fifo
    • test the video data fifos
    • test the PipeLink control fifo
    • test the video control fifo
    • test the video grabber chip
    Results;
    • Green light: - OK
    • Red light: - Failed
    7-8-3-11 Chain (Signal Chain, RLB/RX/BF Test) NOTE: A Test Attenuator, Part Number  AA200215 is needed for this test.
    Remember to restart the Vivid 7 after the diagnostics session.
    ,
      
    Table 7-11    Signal Chain (TX/RLB/RX/BF) Test
    Test (Key)Sub-testsDescription
    Chain Signal Chain Test (Chain)
    The Signal Chain/ Data Flow test checks for er
    rors in the TX, RLY, RX and BF chain of data 
    flow.
    Te s t  A l l Not Used.
    TX/RLB/RX/ BF Tests data flow in Front-End Transmit Receive signal chain.
    If an error occurs with this test, testing 
    should be continued in the following order;
    1. BF digital and analog test
    2. RX test 
    3. RLY test. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 74 Section 7-8 - Acquisition Diagnostics
    7-8-4 General Recommendations  to Troubleshoot Noise Issues
    Ultrasound machines are susceptible to Electromagne tic Interference (EMI) from radio frequencies, 
    magnetic fields, and transients in the air or wiring.  They also generate EMI. The Vivid 7 complies with 
    limits as stated on the EMC label. Ho wever there is no guarantee that interference will not occur in a 
    particular installation. 
    Possible EMI sources should be identified before the unit is installed. 
    Electrical and electronic equipment may produce EMI un intentionally as the result of a defect. Some of 
    these sources include:
    • medical lasers
    • scanners
    • cauterizing guns
    • computers
    •monitors
    •fans
    • gel warmers
    • microwave ovens
    • light dimmers 
    • portable phones
    The presence of a broadcast station or br oadcast van may also cause interference. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 75
    7-8-5 EMI Prevention/abatement
         
    Table 7-12    EMI Prevention/abatement
    EMI RULEDETAILS
    BE AWARE OF RF 
    SOURCES
    - Keep the unit at least 5 meters (15 feet) away from other emi sources. 
    - Special shielding may be requi
    red to eliminate interference problems caused by high frequency, 
    - High powered radio or video broadcast signals.
    GROUND THE UNIT - Poor grounding is the most likely 
    reason a unit will have noisy images. 
    - Check grounding of the power cord and power outlet.
    REPLACE ALL SCREWS, 
    RF GASKETS, COVERS, 
    CORES - After you finish repairing or updating the s
    ystem, replace all covers and tighten all screws. 
    - Any cable with an exter nal connection requires a magnet wrap at each end. 
    - Install the card rack cover over the card rack. 
    Loose or missing covers or RF gaskets allow  radio frequencies to interfere with the ultrasound 
    signals.
    REPLACE BROKEN RF 
    GASKETS- If more than 20% or a pair of the fingers on an rf gasket are broken, replace the gasket. 
    - Do not turn on the unit until any  loose metallic part is removed.
    DO NOT PLACE LABELS 
    WHERE RF GASKETS 
    TOUCH METALNever place a label where rf gaskets meet the uni t. otherwise, the gap created will permit rf leakage. 
    or, if a label has been found in such a position, move the label.
    USE GE SPECIFIED 
    HARNESSES AND 
    PERIPHERALS- The interconnect cables are grounded and r equire ferrite beads and other shielding. 
    - Also, cable length, material, and routing are all  important; do not change from what is specified.
    TAKE CARE WITH 
    CELLULAR PHONES Cellular phones may transmit a 5 v/m signal
    ; that could cause image artifacts. 
    PROPERLY DRESS 
    PERIPHERAL CABLES Do not allow cables to lie across the top of t
    he card rack or hang out of the peripheral bays. Loop 
    the excess length for peripheral cabl es inside the peripheral bays. attach the monitor cables to the 
    frame. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 76 Section 7-8 - Acquisition Diagnostics
    This page was intentionally left blank. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 1
    Chapter 8 
    Replacement Procedures
    Section 8-1
    Overview
    8-1-1 Purpose of Chapter 8
    This chapter holds replacement procedures for the following modules/subsystems.
    NOTE: Use the ESD-kit during replacing boards. 
     
     
    WARNINGWARNINGDO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE 
    NECESSARY ESD PRECAUTIONS: 
    ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD 
    CONNECTION POINT LOCATED AT THE REAR  OF THE SCANNER (TO THE RIGHT OF 
    THE POWER CONNECTOR). 
    FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE 
    EQUIPMENT.
    WARNINGWARNINGTHE WASTE OF ELECTRICAL AND ELEC TRONIC EQUIPMENT MUST NOT BE 
    DISPOSED AS UNSORTED MUNICIPAL WASTE AND MUST BE COLLECTED 
    SEPARATELY.
    PLEASE CONTACT THE MANUFACTURER OR OTHER AUTHORIZED DISPOSAL 
    COMPANY TO DECOMMISSION YOUR EQUIPMENT. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 2  - 
    8-1-2 Definitions of Le ft, Right, Front and Back
    Figure 8-1   Definition of Left, Right, Front and Back of Vivid 7
    RIGHT SIDE
    REAR / BACK
    LEFT SIDE FRONT 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Vivid 7 Service Manual