Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Vivid 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 858
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 37
    4-3-6-6 2D Soft Menu Controls
      
    4-3-6-7 Check Compress, Contour, Reject and Tilt 
       
    Table 4-9    4-way rocker used for Soft Menu control
    KEYDESCRIPTION
    A 4-way rocker (4-way multi direction switch) used to;
    - access mode-specific menus on the screen.
    - select a menu option.
    - adjust option-related values.
    When using the Rocker;
    - the vertical arrows are used to select the menu options.
    - the horizontal arrows are used to adjust the values.
    Table 4-10    2D Mode: Check Compress, Contour, Reject and Tilt 
    StepTaskExpected Result(s)
    1. Use the vertical arrows to select Compress. 
    Use the horizontal arrows to adjust Compress. Compress controls the amount of contrast in the 2D 
    image. 
    An index number is displayed in the status window to 
    indicate the relative level of compression.
    2. Use the vertical arrows to select Contour.
    Use the horizontal arrows to adjust Contour. Contour controls image processing related to the 
    extent of edge enhancement applied to an image.
    3. Use the vertical arrows to select Reject.
    Use the horizontal arrows to adjust Reject. When Reject is increased, low-level echoes and noise 
    are rejected and appear darker in the 2D image. 
    An index number is displayed in the status window to 
    indicate the relative level of rejection.
    4.
    Use the vertical arrows to select Tilt.
    Use the horizontal arrows to adjust Tilt.
    Tilts the axis of the 2D image to the left or right. 
    By using this control in combination with Angle control, 
    the image can be “aligned” to the direction of interest, 
    and frame rates be optimized. 
    By default the axis of symmetry of a 2D image is 
    vertical. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 38 Section 4-3 - Functional Checks
    4-3-6-8 Check DDP, Diff, Power and Dynamic Range
         
    Table 4-11    2D Mode: Check DDP, Diff, Power and Dynamic Range
    StepTaskExpected Result(s)
    1. Use the vertical arrows to select DDP (Data 
    Dependent Processing).
    Use the horizontal arrows to adjust DDP. Performs temporal processing which reduces random 
    noise without affecting the motion of significant tissue 
    structures. 
    An index number is displayed in the status window 
    (under Proc) to indicate the relative DDP level.
    2. Use the vertical arrows to select Diff.
    Use the horizontal arrows to turn Diff On/Off. Affects the level of reverberations in the image.
    When turned ON, the frame rate (or the number of 
    focal zones) will decrease, 
    while the reverberations 
    will be attenuated.
    3.
    Use the vertical arrows to select Power.
    Use the horizontal arrows to adjust Power. Controls the amount of acoustic power applied in all 
    modes.
    When power is set to maximum, it is equal to or less 
    than the maximum acoustic power permitted by the 
    FDA.
    The Thermal Index (TI) and the Mechanical Index (MI) 
    are displayed on the screen.
    4.
    Use the vertical arrows to select Dynamic Range. 
    Use the horizontal arrows to adjust Dynamic Range.
    Enables control of the dynamic range or contrast of 
    the image.
    When Dynamic Range is set high, the image is softer 
    and more low-level data is visible. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 39
    4-3-7 M Mode Checks
    4-3-7-1 Introduction
    4-3-7-2 Preparations Use a phantom when doing these tests.
    1.) Connect one of the probes, to the  scanner’s left-most probe connector.
    -See  3-6-5 Probe Connection on page 3-21  for info about connecting the probes.
    -See  Section 9-26 Probes on page 9-80  for available probes.
    2.) Turn ON the scanner. -See see  4-2-2 Power ON/ Boot UP on page 4-3  for info about booting the scanner.
    The 2D Mode window is displayed (default mode).
    3.) Press  M-MODE
     to bring up an M-Mode picture on the screen.Figure 4-29   M-Mode Screen Example 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 40 Section 4-3 - Functional Checks
    4-3-7-2      Preparations (cont’d) Figure 4-30   M Mode, Freeze and Cine Replay controls
    Exit
    Gain * Mode Dependant 
    Keys
    Horizontal Sweep*
    Frequency
    Focus
    Contour
    Up/Down
    Color Maps
    Soft menu
    Compress
    Reject
    Dynamic Range
    Power
    Freeze
    Controls marked with * are also 
    available in freeze and cine replay
    SteCurZoom
    Alt. PW CW
    TVI
    2D MM
    Color
    Tra
    ck
    B al
    lU
    pd
    at
    e
    M e
    nu
    Act. Win
    Act. Mode Img. Size1/2/4
    Menu Measure
    Exit
    Caliper
    Img. Store
    2DAlt.
    R
    P
    Zoom
    Display zoom
    HR zoom
    Image size
    Small/large 2D sector 
    top/bottom display
    side by side display
    M-Mode 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 41
    4-3-7-3 Check Horizontal Sweep, Frequency, Focus and Contour 
        
    Table 4-12    M-Mode: Check Horizontal Sweep, Frequency, Focus and Contour
    StepTaskExpected Result(s)
    1. Adjust  HORIZONTAL SWEEP 
    Adjusts the horizontal refresh rate of the M-Mode area 
    of the display.
    Horizontal sweep does not  change the acquisition 
    resolution, so you can change the horizontal sweep in 
    replay without any loss of quality.
    2. Adjust  FREQUENCY
    Rotate the knob clockwise to increase the probe’s 
    operating frequency. The probe’s operating frequency is increased.
    The selected frequency is displayed in the status 
    window.
    For some probes/applications the lowest frequency 
    settings will be Octave imaging settings.
    3. Adjust  FOCUS
    Changes the location of the focal point(s). 
    A triangular focus marker indicates the depth of the 
    focal point.
    4.Adjust CONTOUR
    Controls image processing related to the extent of edge 
    enhancement applied to an image. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 42 Section 4-3 - Functional Checks
    4-3-7-4 Check Compress, Reject, Power and Dynamic Range
     
    Table 4-13    M-Mode Soft menu Controls check
    StepTask Expected Result(s)
    1. Adjust Compress  Controls the amount of contrast in the image.
    An index number is displayed in the status window to 
    indicate the relative level of compression.
    2. Adjust Reject Adjust reject level.
    When Reject is increased, low-level echoes are 
    rejected and appear darker in the image. 
    An index number in the status window indicates the 
    relative level of rejection.
    3. Adjust Power Controls acoustic power in all modes. 
    When power is set to maximum, it is equal to or less 
    than the maximum acoustic power permitted by the 
    FDA.
    Both Thermal Index (TI) and Mechanical Index (MI) are 
    displayed on the screen.
    4.Adjust Dynamic Range
    Enables control of the dynamic range or contrast of the 
    image. 
    When dynamic range is set to High, the image is softer 
    and more low-level data is visible. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 43
    4-3-8 Color Mode Checks
    4-3-8-1 Introduction Color Flow screens is 2D or M Mode screens with  colors representing blood or tissue movement.
    Color Flow may be selected both from 2D mode or from M mode or a combination of these.
    4-3-8-2 Preparations Use a phantom when doing these tests.
    1.) Connect one of the probes, to the  scanner’s left-most probe connector.
    -See  Probe Connection on page 3-21  for info about connecting the probes.
    -See  Chapter 9 -  , section 9-26 for available probes.
    2.) Turn ON the scanner. -See  4-2-2 - Power ON/ Boot UP  on page 4-3 for info about booting the scanner
    The 2D Mode window is displayed (default mode).
    4-3-8-3 Color 2D Mode Press COLOR
    .
    4-3-8-4 Color M Mode 1.) Press MM
     (M Mode) to bring up an M Mode display on the screen.
    2.) Press  COLOR
    . 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 44 Section 4-3 - Functional Checks
    4-3-8-4      Color M Mode (cont’d) Figure 4-31   Color Flow Mode Controls
    Exit
    Depth
    Gain * Mode Dependant Keys
    Horiz. sweep (Color M-Mode only) 
    *
    Scale
    Baseline  *
    LVR
    Invert  *
    Variance  *
    Color maps  *
    Simultaneous (Color 2D only)
    Angio (Color 2D only)
    Cineloop (in Freeze, Color 2D mode only)
    Soft menu
    Sample Volume
    Tissue Priority *
    DDP (Color 2D only)  *
    Frequency
    Frame rate
    Radial avg.
    Lateral avg.
    Power
    Freeze 
    Controls marked with * are also  available in freeze and cine replay
    ColorZoom
    TVI
    Zoom
    Display zoom
    HR zoom
    Image size
    (Color M-Mode)
    Small/large 2D sector 
    top/bottom display
    side by side display 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 45
    4-3-8-5 Check Horizontal Sweep, PRF, Baseline, LVR and Invert
     
    Table 4-14    Color Flow Mode Controls Check
    StepTask Expected Result(s)
    1. Adjust Horizontal Sweep (Color M-Mode only)  Adjusts the horizontal refresh rate of the M-Mode area 
    of the display.
    2. Adjust Baseline  Adjusts the color map to emphasize flow either toward 
    or away from the probe. 
    Baseline is available in both Live and Freeze.
    3. Change PRF (Pulse Repetition Frequency) Adjusts the repetition rate of the Doppler pulses 
    transmitted to acquire the data for color flow mapping. 
    The PRF (Nyquist limit) should be adjusted so that no 
    aliasing occur, while still having a good resolution of 
    velocities, meaning the 
    Nyquist limit should be 
    somewhat above the maximum velocity found in the 
    data.
    4.Adjust LVR (Low Velocity Rejection
    LVR, also called Wall Moti on Filter, enables the extent 
    of low velocity removal to be adjusted.
    5.Invert
    Enables the color scheme assigned to positive and 
    negative velocities to be inverted. 
    Invert is available in  live and Cine Replay. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 46 Section 4-3 - Functional Checks
    4-3-8-6 Check Variance, Color Maps and Cineloop
     
    Table 4-15    Color Flow Mode: Check Variance, Color Maps and Cineloop
    StepTask Expected Result(s)
    6. Adjust Variance Controls the amount of variance data added to a color 
    display. 
    Variance enables computer-aided detection of non-
    luminary flow (e.g. jets or regurgitation). 
    Variance is available in live and cine replay.
    7. Select Color Maps Displays a menu of color map options. 
    Use the trackball to point to a color map and press 
    SELECT to activate the desired color map. 
    Each color map is assigning different color hues to 
    different velocities.
    8. Make Cineloop (in Freeze, Color 2D mode only)
    Starts cine loop acquisition. 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Vivid 7 Service Manual