GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 73 8-9-14-2 Vivid 7 Syst em Software Loading - Software Version v2.x (cont’d) 7.) The Software loader program starts and the below Time Bar Info Screen appears. 8.) Wait for the SW Installation to be completed. (Typic al installation time: 5 minutes). Status bar on the screen indicates process. 9.) You will now be prompted to remove the first CD and restart th e unit when the SW Installation is completed. Remove the CD, but don’t restart the System this time . The next step is to load the application software. Figure 8-69 Software Loading Underway (Picture distorted here due to photo technical limitations, it should look OK on screen) WARNINGWARNINGDo not interrupt the SW loading at any time.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 74 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-14-3 Application SW Installation - Software Version v2.x 1.) Insert the applicable CD labelled Vivid7 & EchoPAC PC Application SW , see Table 8-14 CDs with BT’02 Software, Version v2.x on page 8-55 . 2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the scanner. (This is called Forced Shut Down). 3.) Wait until the LCD’s on the Oper ator Panel is switched OFF and the ON/Standby Switch on the Operator Panel has turned yellow. 4.) Power ON the System, see 4-2-2 Power ON/ Boot UP on page 4-3 . 5.) When the dialogue box named Start Application appear, select Install SW. 6.) A dialogue box, named Star tloader, will appear, select OK to continue. 7.) A dialogue box will appear, al so named Startloader stating “New System SW installation will start now”, select OK to continue. 8.) A Command prompt window will ap pear, hit any key on the alphanumeric keyboard to continue. 9.) A message in the Command prompt window may prompt you that copying application software takes three minutes. This may ta ke as much as five minutes. 10.) Several dialogue boxe s will appear on the screen during the SW load. 11.) Remove the CD from the CD drive when prompted. 12.) Press any key to continue. 13.) The Scanner will now pow er down and reboot. WARNINGWARNINGDO NOT INTERRUPT THE SOFTWARE LOADING AT ANY TIME.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 75 8-9-15 Verifications - Software Version v2.x NOTE: Software Option Strings are software versio n dependant. If a new software version has been installed, please follow installa tion instructions for that softwar e. Usually, new software option strings are needed. 8-9-15-1 Verify Software Option Key Settings - Software Version v2.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12. 2.) Select Admin (lower part of window), 3.) Select the SYSTEM ADMIN tab. 4.) Verify that the SW Options Keys (alphanumeric strings) in the SW Options Key window are the same as recorded in 8-9-10 Record SW Option Keys - Software Version v2.x on page 8-62. 5.) Select New to open the New Key dialog where you can type any missing Software Option Key(s) (alphanumeric string(s)). Figure 8-70 System Admin screen Figure 8-71 Type Software Option Key System Admin Tab New Admin Field for Software Option Key (Serial String)
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 76 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-15-2 Verify Video and VCR Settings - Software Version v2.x 1.) Select Service (lower, right corner) . 2.) From Video Settings pull down menu, verify Format as recorded in Table 8-19 Video Settings and VCR Type on page 8-68 . Correct the selection if needed. 3.) From the VCR pull down menu, verify VCR type as recorded in Table 8-19 on page 8-68 . Correct the selection if needed. 8-9-15-3 Verify TCP/IP Setti ngs - Software Version v2.x 1.) From the top of the con nectivity tab, select the Tcpip tab. 2.) Verify that all settings on the TC PIP tab matches those recorded in Table 8-16 Record settings from Tcpip screen (v2.x) on page 8-63 . 3.) Select Save settings . 4.) Select Ok to continue. Figure 8-72 Video Settings Figure 8-73 TCP/IP Example Settings
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 77 8-9-16 Restore Standard Printer IP and Port Settings - Software Version v2.x 1.) Select Connectivity . 2.) Select Service. 3.) Select Standard Print . 4.) Select Open. Figure 8-74 Choose Standard Printer Connectivity Standard Print Open Services Scroll down to “Standard Print”
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 78 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-16 Restore Standard Printer IP and Port Settings - Software Version v2.x (cont’d) 5.) Select Printer. 6.) Select Properties. 7.) Select Standard TCP\IP Port . 8.) Select Configure Port .Figure 8-75 Properties Figure 8-76 Configure Port Printer Properties Standard TCP\IP Port. Configure Port
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 79 8-9-16 Restore Standard Printer IP and Port Settings - Software Version v2.x (cont’d) 9.) Verify Printer Name or IP Address as recorded in Table 8-18 on page 8-67 . 10.) Select Raw. 11.) Verify Port Number as recorded in Table 8-18 on page 8-67 . 12.) Select ok. 13.) Close the rest of the Windows by selecting ok. 8-9-16-1 DC Offset Calibration 1.) Disconnect all probes. 2.) From the upper section of the window, select Test. 3.) Then select Go to System Test. 4.) From the Module selection block, select DC Offs. Cal. 5.) Reboot the System when the DC O ffset Calibration has been completed. 8-9-16-2 Verify System Software Version 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 2.) Select About tab. 3.) Verify that the SW versions are the same as what were recorded in Table 8-20 Record Software and hardware versions on page 8-69 . 8-9-17 Verification - Functional Checks • Please perform the tests specified in Section 8-20 Verification - Functional Checks on page 8-105 . Figure 8-77 Configure Port Raw Port Number ok Printer Name or IP Address
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 80 Section 8-10 - BEP Filter Assembly Replacement Procedure Section 8-10 BEP Filter Assembly Re placement Procedure 8-10-1 Manpower One person, 5 minutes 8-10-2 Tools No tool required for installation and removal of the Filter Assembly. 8-10-3 Preparations None 8-10-4 BEP Filter Removel Procedure Follow these steps to remove the Filter Assembly: 1.) Press down the upper rib and pull out the upper part of the assembly before removing it completely. 2.) Lift the filter assembly away. 8-10-5 BEP Filter Installation Procedure Follow these steps to remove the Filter Assembly: 1.) Place the two lower tags into the opening in the BEP. 2.) Push the upper part of the Filter Assembly against the BEP until the assembly is fixed by a click. Figure 8-78 BEP Filter ASSY removal Figure 8-79 BEP F ilter ASSY installation
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 81 8-10-6 Verification - Functional Checks • Please perform the tests specified in Section 8-20 Verification - Functional Checks on page 8-105 .
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 82 Section 8-11 - AC Power Replacement Procedure Section 8-11 AC Power Replacement Procedure 8-11-1 Manpower One person, 1 hour. 8-11-2 Tools #2 Phillips screwdriver 8-11-3 Preparations 1.) Shut down the system, see 4-2-3 Power Shut Down on page 4-7 . 2.) Remove Right Side Cover, see Side Covers Removal Procedure on page 8-4 . 3.) Remove Filter Cover, see Upper Rear Cover Removal Procedure on page 8-6 . 4.) Remove Lower Rear Cover, see Lower Rear Cover Removal Procedure on page 8-10 . NOTE: The Power Supply, as delivered in a Vivid 7, has been configured for either 115 VAC or 230 VAC. Do not interchange the two types and don’t try to reconfigure the Power Supply for use on another voltage; the Mains Circuit Breaker are different between the two types. 5.) Do this step if the intention is to change to another AC Power Supply: - Record the voltage for the AC Power Supply be ing replaced (see label on rear of system). ___ - 115VAC or ___ - 230VAC. DANGERDangerous voltages are present in this equipment. Completely power off and shut down the scanner before servicing.