    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 73
    8-9-14-2              Vivid 7 Syst em Software Loading - Software Version v2.x (cont’d)
    7.) The Software loader program starts and the below Time Bar Info Screen appears.
      
    8.) Wait for the SW Installation to be completed. (Typic al installation time: 5 minutes). Status bar on the 
    screen indicates process.
    9.) You will now be prompted  to remove the first CD and restart th e unit when the SW Installation is 
    completed. Remove the CD, but  don’t restart the System this time
    . The next step is to load the 
    application software. Figure 8-69   Software Loading Underway 
    (Picture distorted here due to photo technical limitations, it should look OK on screen)
    WARNINGWARNINGDo not interrupt the SW loading at any time. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 74 Section 8-9 - Software Loading Procedure - Software Version v2.x
    8-9-14-3 Application SW Installation - Software Version v2.x
    1.) Insert the applicable CD labelled  Vivid7 & EchoPAC PC Application SW , see Table 8-14 CDs with 
    BT’02 Software, Version v2.x on page 8-55 . 
    2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the  scanner. (This is called Forced Shut Down).
    3.) Wait until the LCD’s on the Oper ator Panel is switched OFF and the ON/Standby Switch on the 
    Operator Panel has turned yellow.
    4.) Power ON the System, see  4-2-2 Power ON/ Boot UP on page 4-3 .
    5.) When the dialogue box named  Start Application appear, select Install SW.
    6.) A dialogue box, named Star tloader, will appear, select OK to continue.
    7.) A dialogue box will appear, al so named Startloader stating “New  System SW installation will start 
    now”, select  OK to continue.
    8.) A Command prompt window will ap pear, hit any key on the alphanumeric keyboard to continue.
    9.) A message in the Command prompt window may  prompt you that copying application software 
    takes three minutes. This may ta ke as much as five minutes.
    10.) Several dialogue boxe s will appear on the screen during the SW load.
    11.) Remove the CD from the CD drive when prompted.
    12.) Press any key to continue.
    13.) The Scanner will now pow er down and reboot. 
    WARNINGWARNINGDO NOT INTERRUPT THE SOFTWARE LOADING AT ANY TIME. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 75
    8-9-15 Verifications -  Software Version v2.x
    NOTE: Software Option Strings are software versio n dependant. If a new software version has been 
    installed, please follow installa tion instructions for that softwar e. Usually, new software option 
    strings are needed.
    8-9-15-1 Verify Software Option Key Settings - Software Version v2.x 1.) Press CONFIG
     (F2) and log on as  adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12.
    2.) Select  Admin (lower part of window), 
    3.) Select the  SYSTEM ADMIN  tab.
     
    4.) Verify that the SW Options Keys (alphanumeric  strings) in the SW Options Key window are the 
    same as recorded in  8-9-10 Record SW Option Keys - Software Version v2.x on page 8-62.
    5.) Select  New to open the New Key dialog where you can type any missing Software Option Key(s) 
    (alphanumeric string(s)).
        
    Figure 8-70   System Admin screen
    Figure 8-71   Type Software Option Key
    System Admin Tab New
    Admin
    Field for Software Option Key (Serial String) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 76 Section 8-9 - Software Loading Procedure - Software Version v2.x
    8-9-15-2 Verify Video and VCR Settings - Software Version v2.x
    1.) Select  Service (lower, right corner)
    .
    2.) From Video Settings pull down menu, verify  Format as recorded in  Table 8-19 Video Settings and 
    VCR Type on page 8-68 . Correct the selection if needed.
    3.) From the VCR pull down menu, verify  VCR type as recorded in  Table 8-19 on page 8-68 . Correct 
    the selection if needed.
    8-9-15-3 Verify TCP/IP Setti ngs - Software Version v2.x
    1.) From the top of the con nectivity tab, select the Tcpip tab.
    2.) Verify that all settings on the TC PIP tab matches those recorded in Table 8-16 Record settings from 
    Tcpip screen (v2.x) on page 8-63 .
    3.) Select  Save settings .
    4.) Select  Ok to continue. Figure 8-72   Video Settings
    Figure 8-73   TCP/IP Example
     Settings 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 77
    8-9-16 Restore Standard Printer IP and  Port Settings - Software Version v2.x
    1.) Select Connectivity .
    2.) Select  Service.
    3.) Select  Standard Print .
    4.) Select  Open.
    Figure 8-74   Choose Standard Printer
    Connectivity
    Standard Print
    Open
    Services
    Scroll down to “Standard 
    Print” 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 78 Section 8-9 - Software Loading Procedure - Software Version v2.x
    8-9-16 Restore Standard Printer IP and Port  Settings - Software Version v2.x (cont’d)
    5.) Select Printer.
    6.) Select  Properties.
    7.) Select  Standard TCP\IP Port .
    8.) Select  Configure Port .Figure 8-75   Properties
    Figure 8-76   Configure Port
    Printer
    Properties
    Standard TCP\IP Port.
    Configure Port 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 79
    8-9-16 Restore Standard Printer IP and Port  Settings - Software Version v2.x (cont’d)
    9.) Verify Printer Name or IP Address  as recorded in Table 8-18 on page 8-67 . 
    10.) Select  Raw.
    11.) Verify  Port Number  as recorded in  Table 8-18 on page 8-67 .
    12.) Select  ok.
    13.) Close the rest of the Windows by selecting  ok.
    8-9-16-1 DC Offset Calibration 1.) Disconnect all probes.
    2.) From the upper section of the window, select  Test.
    3.) Then select  Go to System Test.
    4.) From the Module selection block, select  DC Offs. Cal.
    5.) Reboot the System when the DC O ffset Calibration has been completed.
    8-9-16-2 Verify System Software Version 1.) Press CONFIG
     (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    2.) Select   About tab.
    3.) Verify that the SW versions are  the same as what were recorded in Table 8-20 Record Software 
    and hardware versions on page 8-69 .
    8-9-17 Verification - Functional Checks
    • Please perform the tests specified in  Section 8-20 Verification - Functional Checks on page 8-105 .
    Figure 8-77   Configure Port
    Raw
    Port Number
    ok
    Printer Name or IP Address 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 80 Section 8-10 - BEP Filter Assembly Replacement Procedure
    Section 8-10
    BEP Filter Assembly Re placement Procedure
    8-10-1 Manpower
    One person, 5 minutes
    8-10-2 Tools
    No tool required for installation and removal of the Filter Assembly. 
    8-10-3 Preparations
    None
    8-10-4 BEP Filter Removel Procedure
    Follow these steps to remove the Filter Assembly: 
    1.) Press down the upper rib and pull out the upper part  of the assembly before removing it completely.
    2.) Lift the filter assembly away.
    8-10-5 BEP Filter Installation Procedure
    Follow these steps to remove the Filter Assembly: 
    1.) Place the two lower tags into the opening in the BEP.
    2.) Push the upper part of the Filter  Assembly against the BEP until the assembly is fixed by a click. 
    Figure 8-78   BEP Filter ASSY removal
    Figure 8-79   BEP F ilter ASSY installation 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 81
    8-10-6 Verification - Functional Checks
    • Please perform the tests specified in  Section 8-20 Verification - Functional Checks on page 8-105 . 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 82 Section 8-11 - AC Power Replacement Procedure
    Section 8-11
    AC Power Replacement Procedure
    8-11-1 Manpower
    One person, 1 hour.
    8-11-2 Tools
    #2 Phillips screwdriver
    8-11-3 Preparations
    1.) Shut down the system, see  4-2-3 Power Shut Down on page 4-7 .
    2.) Remove Right Side Cover, see  Side Covers Removal Procedure on page 8-4 .
    3.) Remove Filter Cover, see  Upper Rear Cover Removal Procedure on page 8-6 .
    4.) Remove Lower Rear Cover, see  Lower Rear Cover Removal Procedure on page 8-10 .
    NOTE: The Power Supply, as delivered in a Vivid 7, has been configured for either 115 VAC or 230  VAC. Do not interchange the two types and don’t  try to reconfigure the Power Supply for use 
    on another voltage; the Mains Circuit Breaker are different between the two types.
    5.) Do this step if the intention is to  change to another AC Power Supply: 
    - Record the voltage for the AC Power Supply be ing replaced (see label on rear of system).
    ___ - 115VAC or ___ - 230VAC.
    DANGERDangerous voltages are present in this  equipment. Completely power off and 
    shut down the scanner before servicing. 
    						