Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Vivid 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 858
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 9 - Renewal Parts 9 - 35
    9-15-2 CRT Monitor
    Figure 9-14   CRT Monitor Assembly
    Table 9-27    CRT Monitor Parts   
    ITEMPART NAMEPART NUMBERDESCRIPTIONQTYFRU
    1 P9534WA CRT MONITOR ASSY 2283334-4
    OR
    2283334-5 OPEN FRAME 17” MONITOR 1 Y
    2 MON CABLE ASSY 2304171 MONITOR CABLE ASSEMBLY
    1Y
    3 P9536AL MON-REAR-NSP-ASSY 2303931MONITOR REAR COVER ASSEMBLY (WITHOUT 
    SPEAKERS) 1Y
    4MON CAP ASSY2303932MONITOR SCREW CAPS.2Y
    5MON FRONT ASSY2303929
    MONITOR FRONT BEZEL ASSEMBLY
    INCLUDES THREE LOGOS, Vivid 7, LOGIQ 7 and 
    LOGIQ 9.
    - PLEASE CHOOSE THE APPROPRIATE LABEL 
    AND ATTACH IT TO THE MONITOR FRONT 
    BEZEL. 
    - DISCARD THE OTHER LABELS.
    1Y
    6P9534WK USER SW ASSY OF 
    MONITOR 2297050MONITOR SWITCH ASSEMBLY1Y
    STEREO GLASSES FOR 3D 
    VISUALIZATION FD200123 USED TOGETHER WITH THE 4D LV FUNCTION 1 Y
    Logo Label
    NOT 
    ILLUSTRATED
    NOT 
    ILLUSTRATED
    1
    2
    3
    4
    5
    6 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    9 - 36 Section 9-16 - Casters
    Section 9-16
    Casters
     
    Figure 9-15   Casters
    Table 9-28    Casters 
    ITEMPART NAME
    PART 
    NUMBER
    DESCRIPTIONQTYFRU
    1 CASTER FRONT 2293008-2FRONT CASTER WITHOUT SWIVEL LOCK ASSEMBLY.
    2Y
    2 CASTER REAR 2 293008-3REAR CASTER WITH SWIVEL LOCK ASSEMBLY.
    2Y
    3 CASTER HUB CAPS 2293008-5 Need to order 2 per caster, adhesive mounting 1 Y
    12
    3 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 9 - Renewal Parts 9 - 37
    Section 9-17
    Brake Assembly
    Figure 9-16   Brake Assembly and Console Lock
    Table 9-29    Brake Assembly 
    ITEMPART NAME
    PART 
    NUMBER
    DESCRIPTIONQTYFRU
    1 PEDAL ASSEMBLY
    FB200595Brake Pedal Assembly
    1Y
    PEDAL ASSEMBLY FD200002Brake Pedal Assembly
    REPLACEMENT FOR FB200595 1Y
    2 ROD CONNECTION BRAKE LEFT  FB307471 1 Y
    3ROD CONNECTION BRAKE RIGHT FB3074721Y
    4SCREW, PT FASTENER M6 X12MM 
    WN1411 080X0612PARKING BRAKE PEDAL SCREWS 
    FRONT BUMPER SCREWSY
    1 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    9 - 38 Section 9-18 - Front-End Processor Card Cage
    Section 9-18
    Front-End Processor Card Cage
    9-18-1 Introduction
    Two versions of the Front-End Processor have been used;
    • Front-End Processor Version 1 (FEP1), see page  9-42.
    • Front-End Processor Version 2 (FEP2), see below.
    9-18-2 Hardware/Software Versions,  Front-End Processor Version 2 (FEP2)
    This overview is valid for the following Vivid 7 models:
    • BT’03-RFI: FC000410, FC000420, FC000430, FC000440
    • BT’04: FC000610, FC000620, FC000630,  FC000640, FC000650, FC000660, FC000699
    • BT’05: FC000760, FC000770, FC000780, FC 000790, FC000800, FC000810, FC000820, 
    FC000830, FC000840, FC000850 
    • BT’06:FC000890, FC000900, FC000910, FC 000920, FC000930, FC000940, FC000950, 
    FC000960, FC000970, FC000980
    • BT’08: FD000010, FD000020, FD000120, FD 000130, FD000140, FD000150, FD000160, 
    FD000170, FD000180, FD000190
    Use  Ta b l e 9 - 3 0  to verify the correct/needed revision on each card in the FEP2.
     
    Table 9-30    Required Revisions for Cards and Modules, FEP2   sheet 1 of 2
    2nd VERSION OF FRONT-END CARD CAGE (FEP2)
    PART 
    NUMBER
    NAME
    BT’03 
    (SW. v3.2.x/ v3.4.x/
    v3.6.x)
    BT’04 
    (SW. v4.x)BT’05
    (SW. v5.x)BT’06
    (SW. v6.x)BT’08
    (SW. v7.x)
    COMMENTS
    MCDMCDMCDMCDMCD
    FC200230 BACK-
    PLANE 02 02 02 02 02
    FB200060 RLY-3 07 07 07 07 07 RELAY BOARD VERSION 3
    - MCD REV 14 Value of internal fuse increased 
    to solve issue that Pedof probe is not 
    recognized due to defective fuse.
    - MCD REV 13 includes a noise fix for noise in 
    Color Flow that was a result of changes on 
    MCD rev 12.
    - MCD REV 11 includes a noise-fix to avoid 
    noise in CW duplex with M3S probe.
    - MCD REV 09 includes a noise-fix for M3S 
    probe- MCD REV 08 has LEDs for +5V and 
    +12V.
    FC200314 RLY-4 N/A 02 02 02 02 RELAY BOARD VERSION 4 
    TO BE USED IN UNITS WITH 3D
    FC200057RX128-40202020202RECEIVER BOARD VERSION 4 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 9 - Renewal Parts 9 - 39
    FC200459TX128-503
    03 
    (requires 
    software 
    v4.2.0 or  later)
    030303
    TRANSMITTER BOARD VERSION 5
    2X TO BE USED IN UNITS WITH RT3D/4D
    NOTE! Do not install one FC200459 and one  FC200375 in the same Vivid 7.
    - MCD REV 07 DOC. ERROR FIX.
    - MCD REV 06 Improved Manufacturing 
    process. 
    - MCD REV 05 Improved tolerances. 
    (Manufacturing).
    - MCD REV 04 Removed unused components.
    - MCD REV 03 is the first released version of 
    this card.
    FC200375TX128-40103030303
    TRANSMITTER BOARD VERSION 4
    2X TO BE USED IN UNITS WITH RT3D/4D.
    NOTE! Do not install one FC200459 and one  FC200375 in the same Vivid 7.
    FC200239 XDBUS-3 01 01 01 01 01 TRANSDUCER BUS BOARD VERSION 3 
    2X USED 
    FC200364 XDBUS-4 N/A 01 01 01 01 TRANSDUCER BUS BOARD VERSION 4
    2X USED
    TO BE USED IN UNITS WITH RT3D/4D
    FC200100 BF64 WITH 
    4 MLA 01 01 01 01 01
    2X USED
    FC200054RFI04N/AN/AN/AN/A
    INTRODUCED 2003. 
    FOR UNITS WITHOUT SUPPORT FOR 3D 
    OPTION.
    OBSOLETE, PLEASE ORDER FC200507
    FC200507RFI0102020202CARD WITH SUPPORT FOR 3D OPTION
    FA200945POWER, DC0202020202
    Issues on some systems when using 3V 
    PROBE in CW. mode.
    OBSOLETE, PLEASE ORDER FD200041
    FD200041 POWER, DC 01 01 01 01 01 Increased +6VDC output current for use with 
    3V PROBE in CW mode.
    FB200574 POWER, TX 
    (TXPS) 02 N/A N/A N/A N/A BT’03:
    USED IN ALL SYSTEMS 
    BT’04 and later:
    DON’T USE. 
    FC200386 POWER, TX 
    (TXPS)  N/A 03030303 BT’04 and later: 
    TO BE USED IN ALL SYSTEMS.
    Table 9-30    Required Revisions for Cards and Modules, FEP2 (cont’d) sheet 2 of 2
    2nd VERSION OF FRONT-END CARD CAGE (FEP2)
    PART 
    NUMBER
    NAME
    BT’03 
    (SW. v3.2.x/ v3.4.x/
    v3.6.x)
    BT’04 
    (SW. v4.x)BT’05
    (SW. v5.x)BT’06
    (SW. v6.x)BT’08
    (SW. v7.x)
    COMMENTS
    MCDMCDMCDMCDMCD 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    9 - 40 Section 9-18 - Front-End Processor Card Cage
    9-18-3 Front-End Processor  Version 2 FRU Parts
     
     
    Figure 9-17   PCB Boards FEP2
    12
    3
    4567
    89
    10
    12
    13
    1415
    16
    17 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 9 - Renewal Parts 9 - 41
    9-18-3 Front-End Processor Vers ion 2 FRU Parts (cont’d)
    Table 9-31    PCB Boards used in FEP- 2
    ITEMPART NAME
    PART 
    NUMBER
    DESCRIPTIONQTYFRU
    1 RLY-3_MODULE FB200060RELAY PCB, VERSION 3 1 Y
    2 RLY-4 FC200314RELAY PCB, VERSION 4. INTRODUCED 2004 1 Y
    3 RX 128 BOARD VER. 4 FC200057PCB RX-128-4, RECEIVER PCB, VERSION 4 1 Y
    4TX128-5FC200459
    PCB TX-128-5, VERSION 5 - USED ON FEP2 (RFI)
    WITHOUT 4D: 1X USED 
    Introduced 2005 for use on new systems.
    1Y
    5TX128-5FC200459PCB TX-128-5, VERSION 5 - USED ON FEP2 (RFI) WITH 4D: 2X USED 
    Introduced 2005 for use on new systems.2Y
    6TX128-4, TX-BOARD FOR RFIFC200375PCB TX-128-4, VERSION 4 - USED ON FEP2 (RFI) WITHOUT 4D: 1X USED1Y
    7 TX128-4, TX-BOARD FOR RFI FC200375PCB TX-128-4, VERSION 4 - USED ON FEP2 (RFI)
    WITH 4D: 2X USED 2Y
    8 XD BUS 3  FC200239TRANSDUCER BUS PCB, VERSION 3 - USED ON 
    FEP2 2Y
    9 XD BUS 4 FC200364TRANSDUCER BUS PCB, VERSION 4 - USED ON 
    FEP2 on units with 4D
    10MODULE BF64_4MLAFC200100BEAM FORMER PCB2Y
    12RADIO FREQUENCY INTERFACEFC200507RADIO FREQUENCY INTERFACE PCB - USED ON  FEP2-BT041Y
    13POWER SUPPLY DC
    FA200945DC POWER SUPPLY (OBSOLETE)1N
    FD200041 DC POWER SUPPLY WITH INCREASED CURRENT 
    OUTPUT ON +6VDC 1Y
    14 TRANSMITTER PS SWITCHED 
    MODE FB200574
    TRANSMITTER POWER SUPPLY SWITCHED MODE
    FOR UNITS WITH SOFTWARE VERSION v3.2.x 1Y
    15 POWER, TX FC200386TRANSMITTER POWER SUPPLY SWITCHED MODE
     FOR UNITS WITH SOFTWARE VERSION v4.x / v5.x 1Y
    16NUT PANEX STUDS 080B2000RETAINER RECEPT (7 places)7Y
    17CARD-RACK COMPLETEFC200308RACK W/ MOTHERBOARD
    FEP2 (INTRODUCED 2ND HALF 2003)1Y 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    9 - 42 Section 9-18 - Front-End Processor Card Cage
    9-18-4 Hardware/Software Versions , Front-End Processor Version 1
    This overview is valid for the following Vivid 7 models:
    • FC000030, FC000060
    • FC000180, FC000190, FC000200, FC000210
    • FC000310, FC000320, FC000330, FC000340
    Use Ta b l e 9 - 3 2  to verify the correct/needed revision on each card in the Card Rack (FEP1).
    NOTE: Units with FEP1 can’t be upgraded to software  version 4.0.0 or higher. For software version 
    4.0.0 the RFI card in FEP2 is required.
     
    Table 9-32    Required Revisions for Card s and Modules, Front-End Processor V1 (FEP1)
    1st VERSION OF FRONT-END PROCESSOR CARD CAGE (FEP1)
    PART 
    NUMBER
    NAME
    v2.3.x/ v2.4.x
    v3.1.x/
    v3.3.x/ v3.5.x 
    COMMENTSMCDMCD
    FA200985 BACKPLANE 
    (MOTHERBOARD) 02 02
    FB200060 RLY-3 06 06 - MCD REV 07 INCLUDES A NOISEFIX FOR USE OF TEE PROBE.
    - MCD REV 08 has LEDs for +5V and +12V.
    - MCD REV 09 includes a noise-fix for M3S probe
    - MCD REV 11 includes a noise-fix to 
    avoid noise in CW duplex with M3S 
    probe.
    - MCD REV 14 Value of internal fuse increased to solve issue that Pedof 
    probe is not recognized due to defective fuse. 
    FB200170 TX128-2. B N/A FOR VIVID 7. Used before 2003.
    FC200022TX128-30101
    REPLACEMENT FOR FB200170. PHASED INTO MANUFACTURING  SECOND HALF OF 2002. 
    NEEDED FOR ALL BT’03 UNITS WITH SOFTWARE VERSION V3.1.x.
    FC200459TX128-5N/A03REQUIRES SOFTWARE VERSION V3.5.X
    FB200831RX128-3BA,BN/A
    FC200057 RX128-4 02 02 This card is compatible with Vivid 7 systems at Rev. v2.3.0/v2.4.0 and up.
    FB200158 XDBUS-2 02 N/A 2X USED 
    FB200900 BF64 C/D/E N/A 2X USED
    FC200100BF64 WITH 4 MLAN/A01 2X USED
    FB200165FEC-II03, 0403, 04
    FB200140RFT1B, C, DN/ABOTH MCD REV. B, C AND D MAY BE USED.
    FC200394 RFT1 N/A E IMPROVED GREYSCALE (COMPARED TO FB200140)
    REQUIRES V3.1.x SOFTWARE
    FB200865 SDP 02 02
    FB200991 IMP-2B A,B A,B
    FC200120IMP-30101REPLACES FB200991 REQUIRES BEP2
    FA200945POWER, DC0202OBSOLETE, Please order FD200041
    FD200041POWER, DC0101REPLACES FA200945 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 9 - Renewal Parts 9 - 43
    FB200574 POWER, TX (TXPS) 03 03
    Table 9-32    Required Revisions for Ca rds and Modules, Front-End Processor V1 
    1st VERSION OF FRONT-END PROCESSOR CARD CAGE (FEP1)
    PART 
    NUMBER
    NAME
    v2.3.x/ v2.4.x
    v3.1.x/
    v3.3.x/ v3.5.x 
    COMMENTSMCDMCD 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    9 - 44 Section 9-18 - Front-End Processor Card Cage
    9-18-5 Front-End Processor  Version 1 FRU Parts
     
     
    Figure 9-18   PCB Boards - FEP1
    1
    23
    6
    78
    9
    1011
    12
    1314
    15
    16
    17
    18
    4a4b4c 
    						
    All GE manuals Comments (0)