    Chapter 4 - Functional Checks 4 - 57
    4-3-11 Tissue Velocity Imaging (TVI) Checks
    4-3-11-1 Introduction TVI calculates and color codes the velocities in tissue. The tissue velocity information is acquired by 
    sampling of tissue Doppler veloci ty values at discrete points.
    4-3-11-2 Preparations Use a phantom when doing these tests.
    1.) Connect one of the probes, to the  scanner’s left-most probe connector. 
    -See  Probe Connection , page 3-21 for info about connecting the probes
    -See  Probes, page 9-80 for available probes
    2.) Turn ON the scanner -See see  4-2-2 - Power ON/ Boot UP  on page 4-3 for info about booting the scanner.
    The 2D Mode window is displayed (default mode).
    3.) If needed, adjust the Display’s  Brightness and Contrast setting. 
    4.) Press  TVI
    . 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 58 Section 4-3 - Functional Checks
    4-3-11-2      Preparations (cont’d) Figure 4-34   Doppler Mode Controls
    Mode Dependant Keys
    2D Width
    Scale
    Transparency
    Threshold
    Invert *
    Tissue Tracking
    Simultaneous *
    Cineloop (Freeze only)
    Color Maps
    Soft menu
    Tissue Priority *
    Frame Rate
    Compress *
    Reject *
    Frequency
    Lateral Average
    Radial Average
    Baseline *
    Power
    TVI
    Freeze
    2D
    Alt.
    Alternatives
    Anat. M-Mode
    Curved Anat. M-Mode
    Zoom
    Display zoom
    HR zoom
    Controls marked with * (asterisk) are  also available in freeze and cine replay 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 59
    4-3-11-3 Check 2D Width, Scale,  Transparency and Threshold
      
    Table 4-23    TVI Mode Controls Checks
    StepTaskExpected Result(s)
    1. Adjust 2D WIDTH
    Controls the size or angular width of the 2D sector. 
    2. Adjust  SCALE
    Controls the repetition rate of the Doppler pulses 
    transmitted to acquire the data for color mapping. 
    The scale value influences the Nyquist limit (the ability 
    to detect maximal velocity without color aliasing.
    3. Adjust  TRANSPARENCY
    Controls the degree of transparency of the TVI color.
    4.Adjust THRESHOLD.
    Controls the level of greyscal e intensity that is used as 
    a threshold for color. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 60 Section 4-3 - Functional Checks
    4-3-11-4 Check Invert, Tissue Tracking, Simultaneous and Color Maps  
      
    Table 4-24    TVI Mode Controls Checks
    St e pTa s kExpected Result(s)
    5. Press  INVERT
     
    Enables the color scheme assigned to positive and 
    negative tissue velocities to be inverted. 
    Invert is available in live and cine replay.
    6. Press  TISSUE TRACKING
     
    Starts Tissue Tracking mode.
    7. Choose  COLOR MAPS
    Displays a drop down menu of different colorization 
    maps.
    Use the trackball to select the desired map and press 
    Select to activate the map. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 61
    4-3-11-5 Check Tissue Priority, Frame Rate, Compress, Reject, Frequency, Lateral & Radial Average,  Baseline and Power
        
    Table 4-25    TVI Mode: Soft Menu Controls Check 
    StepTask Expected Result(s)
    8. Adjust  TISSUE PRIORITY 
    Emphasizes either the color of the TVI or the greyscale 
    tissue detail of the 2D image.
    Tissue priority is avail able in both Live and Freeze.
    9. Adjust  FRAME RATE
     
    Controls the line density. When adjusting frame rate, 
    there is a trade off between spatial and temporal 
    resolution.
    10. Adjust  COMPRESS
     
    Controls the amount of color compression. The color 
    bar is adjusted accordingly.
    11.Adjust REJECT 
    Adjust the cut-off level for the low velocity of TVI to be 
    discarded when generating the color image.
    12.Adjust  FREQUENCY 
    Enables the adjustment of  the transmission frequency 
    to control the sensitivity or the level of penetration. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 62 Section 4-3 - Functional Checks
    13.Adjust  LATERAL AVERAGE 
    Smooths the image by averaging collected data along 
    the same horizontal line. An increase of the lateral 
    averaging will reduce noise, but  this will also reduce the 
    lateral resolution.
    14. Adjust  RADIAL AVERAGE
     
    Smooths the image by averaging collected data along 
    the same radial line. An increase of the radial 
    averaging will reduce noise, but  this will also reduce the 
    radial resolution.
    15. Adjust  BASELINE
     
    Adjusts the color map to emphasize tissue motion 
    either toward or away from the probe.
    Baseline is available in both Live and Freeze.
    16. Adjust  POWER
     
    Controls the amount of acoustic power applied in all 
    modes. When power is set to maximum, it is equal to or 
    less than the maximum acoustic power level permitted 
    by regulatory standards.
    The Thermal Index (TI) and the Mechanical Index (MI) 
    are displayed on the screen.
    Table 4-25    TVI Mode: Soft Menu Controls Check 
    StepTask Expected Result(s) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 63
    4-3-12 Contrast Checks
    Refer to Chapter 5 in Vivid 7 User Manual.
    4-3-13 Stress Echo
    Refer to Chapter 4 in Vivid 7 User Manual. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 64 Section 4-3 - Functional Checks
    4-3-14 Measurements and Multi Image Checks
    4-3-14-1 Introduction to Measurements & Analysis (M&A) • A complete M&A software package for  Cardiac is included in the Vivid 7. 
    • The M&A software runs on the Back-End Processor.
    • In this test, we just check one measurement, Calipers.
    4-3-14-2 Preparations, M&A Figure 4-35   Measurement Keys
    Table 4-26     Measurement Key Checks
    StepTask Expected Result(s)
    1. Generate a 2D image 
    (see 4-3-6 - 2D Mode (B mode) Checks  for details).A 2D image is active on the screen
    2. Press FREEZE
     on the Operator Panel The last scanned picture is on the screen.
    3. Scroll back until you find the picture you want to do the 
    measurement on. The picture of interest is on the screen.
    F
     1F  2F  3F  4
    Measure
    Assign and Measure
    Measure and Assign
    Caliper
    Measure and Assign
    (when in measure 
    mode))
    Menu
    Fast access to the 
    Measurement menu
    (when in measure mode) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 65
    4-3-14-3 Check Caliper Measurement, Distance
    4-3-14-4 Check Post-measurement assignment labels
       
    Table 4-27     Measurement Key Checks
    StepTask Expected Result(s)
    4. Press MEASURE
     on the Operator Panel.
    Select  Caliper  in the Measurement menu.
    or
    Press  CALIPER
     on the Operator Panel.
    Press the  CALIPER
     mode dependant button The Measurement menu with a list of studies is 
    displayed in the Parameters window on the screen. 
    When entering the Measurement mode for the first 
    time, the Caliper tool is selected by default. 
    When re-entering the Measurement mode, the first 
    measurement in the actual study that has not been 
    performed is selected by default.
    5. Use the Trackball to move the cursor to the start point 
    of the measurement. The cursor points to the start point for the 
    measurement.
    6. Press SELECT
     to anchor the start point of the 
    measurement. The start point of the measurement is marked, and a 
    line between the mark and the active cursor indicates 
    the distance.
    7.Use the trackball to move the cursor to the end point of 
    the measurement.The current distance value is displayed in the 
    Measurement result table and is instantaneously 
    updated when moving the cursor.
    8.
    Press 
    SELECT to mark the end point for the 
    measurement.
    The measurement display color on the 2D image 
    changes from green to red after completion of the 
    measurement.
    The measurement result is displayed in the 
    Measurement result table.
    9.Trackball to the measurement result in the 
    Measurement result table.The highlighted parameter label is assigned to the 
    measurement. 
    Table 4-28     Post-measurement assignment labels
    StepTask Expected Result(s)
    10. Trackball to the actual measurement in the 
    Measurement Result Window.
    Press SELECT
     on the Operator Panel. A Parameter menu is displayed.
    11. Trackball through the Parameter label menu to 
    highlight the desired label.
    Press  SELECT
     on the Operator Panel. The highlighted parameter label is assigned to the 
    measurement. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 66 Section 4-3 - Functional Checks
    4-3-15 Multi Image Checks
    4-3-15-1 Introduction to Multi Image This unit has a special function for displaying tw o or four examination pictures at the same time
     
    4-3-15-2 Preparations
    Use a phantom when doing these tests.
    1.) Connect one of the probes, to the  scanner’s left-most probe connector. 
    -See  Probe Connection , page 3-21 for info about connecting the probes
    -See  Probes, page 9-80 for available probes
    2.) Turn ON the scanner -See  4-2-2 - Power ON/ Boot UP  on page 4-3 for info about booting the scanner.
    The 2D Mode window is displayed (default mode).
    If needed, adjust the Display’s Brightness and Contrast setting.
    3.) Store four cineloops.
    4-3-15-3 1,2,4
     Key Check
       
    Table 4-29    Use of Multi Image Key
    KeyDescription
    Enables multiple image display windows in which tw o or four images can be viewed simultaneously. 
    When reducing the number of images, the  active window will always be kept.
    Table 4-30     Multi Image Checks
    StepTaskExpected Result(s)
    1.
    Press the 1,2,4 Key once to display two images on the 
    screen. 2 Images are displayed on the screen.
    2. Press the 1,2,4
     Key once more to display four images 
    on the screen. 4 Images are displayed on the screen.
    3. Press the 1,2,4
     Key again to reduce the number of 
    images on the screen. 1 image is displayed on the screen.
    1/2/4 
    						