    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 27
    3-6-7 Switching OFF the Unit
    See  4-2-3 Power Shut Down on page 4-7  for a detailed description.
    1.) Press the  ON/STANDBY
     button on the top left of the Control Panel to display the Exit dialog 
    window.
    2.) Select  SHUTDOWN
    .
    The shutdown process takes a few seconds and is completed when the Operator Panel 
    illumination is turned off.
    To completely switch off the unit before disconnec ting the Mains Power Cable, follow the additional 
    steps below:
    3.) Switch the Circu it Breaker (on rear of unit) to  OFF
    .
    4.) Remove the plug from the mains power outlet. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 28 Section 3-7 - .  Configuration
    Section 3-7. 
    Configuration
    3-7-1 Vivid 7 Configuration 
    3-7-1-1 Select System Settings Screen 1.) Press CONFIG
     (F2) and log on as  ADM, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    NOTE: For software version v5.x and earlier the  default password for ADM (or: adm) was left blank. 
    Starting with software vers ion v6.x, the password is: ulsadm
    2.) Select  System and then select  Settings, if needed.
    3-7-1-2 Enter Location Figure 3-18   Hospital and Department Name
    Table 3-11    Location Name
    STEPTASKEXPECTED RESULT(S)
    1 Select the Hospital field, see 
    Figure 3-18, and type the name of 
    the hospital (max 64 characters).  The 24 first characters of this
     name are displayed on the scanning 
    screen’s title bar (after restart)
    All 64 are displayed on the image properties on saved images (after 
    restart).
    2 Select in the Department field, see 
    Figure 3-18, and type the 
    name of the department (max 64 characters). This name will be displayed on t
    he image properties on saved images as 
    soon as the unit has been restarted.
    HOSPITAL NAME
    DEPARTMENT 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 29
    3-7-1-3 Date and Time Adjustments Figure 3-19   Date and Time Adjustments
    Table 3-12    Date and Time adjustments
    STEPTASKEXPECTED RESULT(S)
    1 Open the 
    System (Configuration) Window, see  4-2-4 Log On to 
    the System as ‘ADM’ on page 4-12 .
    Select  System , if needed. The 
    System Settings  window is displayed.
    2 Adjust the  date, see ( a) in  Figure 3-19 . New date  is displayed
    3  Adjust the  time, see ( b) in  Figure 3-19 . New time is displayed
    4Select the preferred  Time Format, see (c) in  Figure 3-19 .24: the 24 hour format is used
    12:  the 12 AM/PM hour format is used
    5
    Select the preferred  Date Format, see (d) in  Figure 3-19 .
    DD = Date (two digits)
    MM = Month (two digits)
    YYYY = Year (four digits)
    EU:  the European/International “DD.MM.YYYY” format is used
    US:  the American “MM.DD.YYYY” format is used
    6Select  Default Century  (1900, 2000 or None), see ( e) in 
    Figure 3-19 .
    1900: the number  19 is automatically displa yed when entering the year 
    in the patient date of birth.
    To edit century, press  BACKSPACE
     twice.
    2000:  the number  20 is automatically displa yed when entering the year 
    in the patient date of birth. 
    To edit century, press  BACKSPACE
     twice.
    None:  the four digits have to be typed when entering the year in the 
    patient date of birth.
    The selected setting will be used as soon as the unit has been restarted.
    c b
    a
    de 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 30 Section 3-7 - .  Configuration
    3-7-1-4 Language Selection
     
    Figure 3-20   Language Selection
    Table 3-13    Language Adjustments
    STEPTASKEXPECTED RESULT(S)
    1. Open the Configuration Window, see  4-2-4 Log On to the 
    System as ‘ADM’ on page 4-12 . 
    Select  System , if needed. The 
    System Settings  window is displayed.
    2. Use the  Language  drop down dialog, see  Figure 3-20, to select 
    your preferred language for the on-screen interface. 
    The following languages are available;
    - ENG - English
    - NOR - Norwegian
    - DEU - German
    - ITA - Italian
    - FRA - French
    - ESP - Spanish
    - RUS - Russian
    - GRE - Greek
    - DAN - Danish
    - SWE - Swedish The selected language will be us
    ed as soon as the unit has been 
    restarted.
    LANGUAGE 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 31
    3-7-1-5 Online Manual Language Selection
     
    Figure 3-21   Online Manual Language Selection
    Table 3-14    Online Manual Language Selection
    STEPTASKEXPECTED RESULT(S)
    1. Open the Configuration Window, see  4-2-4 Log On to the 
    System as ‘ADM’ on page 4-12 . 
    Select  SYSTEM
    , if needed. The 
    System Settings  window is displayed.
    2. Use the  Manual Language  drop down dialog, see  Figure 3-21, 
    to select your preferred language for the online manual.  The selected language will be used as soon as the unit has been 
    restarted.
    MANUAL LANGUAGE 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 32 Section 3-7 - .  Configuration
    3-7-1-6 Units of Measure
     
    Figure 3-22   Select Units of Measure
    Table 3-15    Select Units of Measure
    STEPTASKEXPECTED RESULT(S)
    1. Open the Configuration Window, see  4-2-4 Log On to the 
    System as ‘ADM’ on page 4-12 .
    Select  SYSTEM
    , if needed. The 
    System Settings  window is displayed.
    2. Use the drop down dialog to select Metric or US Units. The selected units (Metric or US) will
     be used for measurements as soon 
    as the unit has been restarted.
    3. Restart the scanner. All the changed setti ngs will be used after the restart.
    UNITS 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 33
    3-7-2 Service Screen Setup
    3-7-2-1 Overview The Service Screen gives you access to: 
    • Select Video Format to be used by DVD or VCR
    • Select DVD or VCR type
    • Adjust the Operat or Panel LCD’s Contrast and Backlight  Intensity 
    • Alphanumeric Keyboard Setups 
    • Printer Setup
    3-7-2-2 Open Service Screen 1.) Press CONFIG
     (F2) and log on as  ADM, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    2.) Select  SERVICE
     to view the Service Screen, see  Figure 3-23 on page 3-33.
     
     
    Figure 3-23   Service Screen 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 34 Section 3-7 - .  Configuration
    3-7-2-3 Select Video Format, PAL or NTSC This selection must correspond to the Video Stan dard (PAL or NTSC) used by the Video Recorder. 
    • From the Video Settings drop-down menu, select the correct video format (NTSC or PAL).
    3-7-2-4 Select Video Recorder (DVD or VCR) Type •Use the VCR drop down menu to select the type. See  Approved Internal Peripherals on page 3-
    36  for approved video recorders for use with Vivid 7.
    3-7-2-5 Adjust Operator Panel LCD’s Contrast and Backlight Intensity •Use the LCD Contrast  and LCD Backlight  controls to adjust Contrast  and Backlight Intensity for 
    the Operator Panel’s (OP’s) LCD displays.  Figure 3-24   Select Video Format
    Figure 3-25   Select Video Player
    Figure 3-26   Adjust LCD’s Contrast and Backlight Intensity
    Select Video 
    Format 
    SELECT VIDEO PLAYER
    Adjust LCD Contrast and 
    Backlight Intensity 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 35
    3-7-2-6 Add Printer NOTE: This function may be unavailabl e for some software versions and it will not always function due 
    to that usually, a special Installation Wizard is  to be used. Please follow instructions in the 
    respective printer installation manual for correct printer installation. See  3-7-3 Optional 
    Peripherals/Peripheral Connection on page 3-36  for more information.
    1.) Select  ADD PRINTER
     to start the Add Printer (Installation) Wizard.
    2.) Follow the instructions in the  Wizard to install a new printer.
    Figure 3-27   Select Add Printer
    Figure 3-28   Add Printer Wizard
    Select Add Printer 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 36 Section 3-7 - .  Configuration
    3-7-3 Optional Peripherals/Peripheral Connection
    3-7-3-1 Approved Internal Peripherals This list covers the internal peripherals approved for Vivid 7:
    • Printer, Black & White, Video, SONY.
    • Printer, Black & White, Digital, SONY (need s software version v3.0 or above and BEP2). 
    • Printer, Monochrome, Dig ital, SONY UP-D897SYN
    • Printer, Monochrome, Digital, SO NY UP-D895MD (needs software version v3.0 or above and 
    BEP2).
    • Printer, Color, Video, SONY, both PAL and NTSC versions are available.
    • Printer, Color, Digital, SONY UP-D21MD
    • Printer, Color, Digital, SONY UP-D23MD
    • DVD Recorder, JVC BD-X201M (introduced for software v6.0)
    • VCR, SONY, both PAL and NTSC versions are available.
    • VCR, PANASONIC, both PAL and NTSC versions are available.
    • VCR, MITSUBISHI HS-MD3000E
    See the  Vivid 7 Peripherals Installation Manual , Direction Number FC294048, for installation 
    instructions for the peripherals listed above.
    • MO Drive (option) (mounted inside Back-End Processor). 
    • CD-R Drive (part of Back-End Pr ocessor) (BEP1, BEP2 and BEP2.2)
    • CD/DVD Drive (part of Back-End  Processor) (BEP3.x and BEP4.x)
    • USB Flash Card, 256 MB (needs software version v4.0  or higher). Install in USB port on front of 
    Vivid 7.
    NOTICEVivid 7 with BEP1, BEP2 and BEP2.2 suppor ts CD-R disks. CD-RW disks are not supported.
    Vivid 7 with BEP3, BEP3.2 and BEP4.x su pports both CD-R and DVD-R disks. 
    						