GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 53 8-8-12-3 Verify TCP/IP Setti ngs - Software Version v3.x 1.) From the top of the con nectivity tab, select the Tcpip tab. 2.) Verify that all settings on the TC PIP tab matches those recorded in Table 8-11 Record settings from Tcpip screen (v3.x) on page 8-44 . 3.) Select Save settings . 4.) Select Ok to continue. 8-8-12-4 DC Offset Calibration 1.) Disconnect all probes. 2.) From the upper section of the window, select Test. 3.) Then select Go to System Test. 4.) From the Module selection block, select DC Offs. Cal. 5.) Reboot the System when the DC O ffset Calibration has been completed. 8-8-12-5 Verify System Software Version 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 2.) Select about tab. 3.) Verify that the SW versions are the same as what were recorded in Table 8-13 Record Software and Hardware versions (v3.x) on page 8-46 . 8-8-13 Verification - Functional Checks • Please perform the tests specified in Section 8-20 Verification - Functional Checks on page 8-105 . Figure 8-48 TCP/IP Example Settings
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 54 Section 8-9 - Software Loading Procedure - Software Version v2.x Section 8-9 Software Loading Procedur e - Software Version v2.x 8-9-1 Introduction This procedure describes how to perform a complete installation of the software on the Vivid 7 hard disk. 8-9-2 Preparations To Be Done Be fore Software Re-installation • Export Patient Images from the internal hard disk to removable MO disks. • Make backup of Customer Presets, Patient Databas e (Patient Archive, Report Archive, System Configuration and User Defined Configuration) to MO disks. • TCP/IP Addresses, Computer Name and Software Option Password (alphanumeric string) settings must be manually recorded. NOTICEDo not use this procedure when upgrading a unit’s software. Additional software and procedures, not described here, may be needed for such upgrades. Use this procedure if you need to reinst all the existing software on a Vivid 7. NOTICEIn order to complete a successful restore of the Pa tient Database, if needed, the images must be moved to MO disks before doing backup of Patient Database. If the backup procedure is not completed correctly, the images and database information could be lost.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 55 8-9-3 Customer Provided Prerequisite • Formatted and labelled MO Disks for Images storage • Formatted and labelled MO Disk for Patient Archive and User Defined Settings. 8-9-4 Tools Provided Wi th Unit or After a Software Upgrade 8-9-4-1 Units with BT’02 Software NOTE: Units with software version v2.3 .0 or below and all version v1.x should be upgraded to software version v2.3.1 as described in FMI 76037. 8-9-5 Preparations 1.) Switch ON the Mains Power Circuit Breaker and power up the unit, see 4-2-2 Power ON/ Boot UP on page 4-3 . 2.) Continue with 8-9-6 Prepare MO Disks for Image Storage - Software Version v2.x on page 8-56 . Table 8-14 CDs with BT’02 Software, Version v2.x ITEMTITLEPART NUMBERDESCRIPTION 1 V7 SYSTEM SW CD, BEP1 FC200563 or FC200165 FC200563_02 is system software version 1.5.1 (for BEP1) FC200165_05 is system software version 1.3.3 (for BEP1) V7 SYSTEM SW CD, BEP2.0 FC200564 or FC200265 FC200564_02 is system software version 1.5.1 (for BEP2) FC200265_04 is system software version 1.3.3 (for BEP2) V7 SYSTEM SW CD, BEP4 as BEP2 (and BEP1) replacement FC200825USED for BEP4 as BEP2 (and BEP1) replacement FC200824_01 is system software version 1.5.x 2 VIVID 7 & ECHOPAC PC APPL. SWFC200566 APPLICATION SOFTWARE, COMMON TO Vivid 7 AND ECHOPAC PC FC200566_03 IS APPLICATION SOFTWARE VERSION v2.4.2 VIVID 7 & ECHOPAC PC APPL. SWFC200220Application software, common to Vivid 7 and EchoPAC PC FC200220-14 is application sw. 2.3.5
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 56 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-6 Prepare MO Disks for Image Storage - Software Version v2.x NOTE: When formatting the MO disks, give them logi cal names and make sure to physically label the disks. - A labelling convention should be followed so that each MO disk gets a unique label. - A maximum numbers of eleven (11) label characters are allowed in the label name. - Use different labels on the MO disk’s A side and B side. 1.) Label the MO Disks sequentia lly.i.e Image 1A (Side A), Image 1B (side B), Image 2A etc. 2.) If the MO Disks is already form atted and labelled continue with 8-9-7 Move Images - Software Version v2.x on page 8-57 . 3.) Insert a MO disk, side A, into the empty MO drive. 4.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12. 5.) Select Connectivity in the Image and Analysis menu. 6.) Select Tools in the Connectivity menu. 7.) In the Label window, type the same name as labelled on the MO disk side A. 8.) Select Format. 9.) A warning “Are you sure you want to format H: and remove all data from the media” appears, select OK. 10.) Wait until the information dialogue window appears on screen. 11.) Select OK to continue. 12.) Verify the disk formattin g status has been updated. 13.) Press ALT-E. 14.) Select the 5 1/4 MO Disk, to eject Disk. 15.) Insert the MO disk, side B and repeat step 7 to 14. Figure 8-49 Prepare MO Disk - Image Storage
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 57 8-9-7 Move Images - Software Version v2.x The Move Images function enables the movement of archived images from one volume/media to another and the deletion of the moved images from the original volume. The pointers to the database are restored in the new volume/media. NOTE: This function copies image data on ly, no database information is copied. Move Images from Local Archive to a Removable Media 1.) Insert a formatted MO Disk in the MO Drive. 2.) Press ARCHIVE on the Operator Panel to open the SEARCH / CREATE PATIENT screen. 3.) Select “Local Archive - Int HD” if it isn’t already selected as default. Figure 8-50 Search/Create Patient Window - Select “Local Archive - Int HD”
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 58 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-7Move Images - Software Version v2.x (cont’d) 4.) Select Move Images from the More pull-down menu ( Figure 8-51) to display the MOVE IMAGES window ( Figure 8-52 ). 5.) Adjust From date and To d a t e to select the images fr om a defined time period. 6.) Check Keep days together if images from examinations performed on the same day are to be kept on the same MO disk. 7.) Press Recalculate . The Free disk space and Needed space information is updated. Make sure that the space needed does not exceed the space available. NOTE: If the space needed is bigger than the availabl e space on the destination media, the time period selected must be decreased by selecting an earlier date in the To date field. Press Recalculate again and ensure that there is enough space. 8.) Select OK. A progress indicator is displayed. 9.) After completion, Select OK. 10.) Remove MO-Disk with moved Images from MO-Drive. 11.) In case archive HD exceed MO Disk media capacity; a.) Insert a new MO-Disk. b.) Repeat step 5 through 11 as needed.Figure 8-51 Select Move Images from the More pull-down menu Figure 8-52 The Move Images Window Move Images
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 59 8-9-8 Prepare MO Disk for Patien t Archive - Software Version v2.x NOTE: A separate MO disk is required for backup of Patient Database, System Configuration and User Defined Configurations (customer presets). This MO disk must be a separate disk with no images stored. 1.) Label the MO Disk “Backup”. 2.) If the MO Disks is already formatted and labelled, move to section 8-9-9 on page 8-60 3.) Insert a MO disk, side A, into the empty MO drive. 4.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12. 5.) Select Connectivity in the Image and Analysis menu. 6.) Select Tools in the Connectivity menu. 7.) In the Label window, type the same name as labelled on the MO disk side A. 8.) Select Format. 9.) A warning “Are you sure you want to format H: and remove all data from the media” appears, select OK. 10.) Wait until the information dialogue window appears on screen. 11.) Select OK to continue. Figure 8-53 Prepare MO Disk - Patient Archive
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 60 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-8 Prepare MO Disk fo r Patient Archive - Software Version v2.x (cont’d) 12.) Verify the disk formattin g status has been updated. 13.) Press ALT-E. 14.) Select the 5 1/4 MO Disk, to eject Disk. 15.) Optional: Insert the MO disk, side B and repeat step 7 to 14. 8-9-9 Backup of Patient Database & User Presets - Software Version v2.x This procedure describes how to backup Patient database and user presets (user defined configuration). 1.) Insert the formatted MO disk prepared for Backup in the MO Drive. 2.) Go to CONFIG -> Admin -> Backup . 3.) Select all check boxes. 4.) In the Destination field, select the MO disk from the drop-down menu. Figure 8-54 The Backup Screen Admin Function Backup Tab Check Boxes Destination Field Backup Now
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 61 8-9-9 Backup of Patient Data base & User Presets - Software Version v2.x (cont’d) 5.) Select Backup now . A Warning will appear on th e screen, as shown in Figure 8-55 on page 8-61 . . 6.) Select OK. NOTE: No process indicator will appear on screen. Veri fy that the “Finished, OK” Status appears in the Result column, and that the date appearing in th e Last Successful Backup column is the current date and time. The backup is completed. 7.) Remove the Backup MO Disk when the backup is completed. Figure 8-55 Backup Warning
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 62 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-10 Record SW Option K eys - Software Version v2.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12. 2.) From the Admin tab, select SYSTEM ADMIN . 3.) Record SW Options Keys from SW Options Key window. Figure 8-56 System Admin WARNINGWARNINGMake sure that the SW Option Keys (passwords) have been recorded correctly. If the key is incorrect, you will not be able to log on after the SW installation has been completed. The password is case sensitive. Hyphens must also be recorded. There may be more than one password. Table 8-15 SW Options Keys SW OPTIONS KEYS