GE Vivid 7 Service Manual
Have a look at the manual GE Vivid 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 107 3-10-10-3 Query/Retrieve Verification (cont’d) Follow the Steps Below to Test th e DICOM Query/Retrieve Workflow NOTE: This check uses both Ping and DICOM Ping. - Ping is used to verify the TCP/IP connection to the server. - DICOM Ping is used to verify that the DICOM application is answering 1.) Select QueryRetrieve from the Selected devices list 2.) Select Check. - If the test pass, a pop-up dialog is displayed, see left illustration in Figure 3-114. NOTE: If the Check button fails directly, the AE Title is probably wrong. If the Check button fails after a long time (corres ponding to the timeout), the IP address or Port Number is probably wrong. - Select OK to continue. A green check mark to th e left of the selected devices indicates that the test was successfully passed. Figure 3-113 Check the DICOM Query/Retrieve workflow Figure 3-114 Test (Check) Passed QUERYRETRIEVE CHECK
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 108 Section 3-10 - Connectivity Setup - Software v3.x 3-10-10-3 Query/Retrieve Verification (cont’d) Follow the Steps Below to Test the DICOM Query/Retrieve Workflow (cont’d) - If the test fail, a pop-up warning is displayed, see left illustration in Figure 3-115. Select OK to continue. A red “X” to the left of the Selected devices indicates that the test failed. If Retrieve is Failing ... If Retrieve is failing but Qu ery is functioning, verify if the scanner has been registered as an receiver on the server. Figure 3-115 Test (Check) Failed
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 109 Section 3-11 Connectivity Setup - Software v2.x NOTE: FMI 76037 updates softwa re versions v2.3.0 and below to Software version v2.3.1. NOTE: If connected as a stand-alone network, only scanner, EchoPAC PC Workstation and an optional network printer in a separate network, you should use default delivery settings. 3-11-1 Introduction To be able to use the network functions when con nected to a hospital network, the scanner must have a proper network address. • Before you can set up the scanner, you need to collect some information. The Pre-quote Worksheet for DICOM Network Information on page 2-14 can be used for gathering this information. Typical source for this in formation is the network administrator. • Follow the steps below to prepare the scanner for use on the network. 3-11-2 Contents in this Section • 3-11-3 Compatibility on page 3-109 • 3-11-4 Select TCP/IP Set-up Screen - Software v2.x on page 3-110 • 3-11-5 Set the Scanner’s Network Information - Software Version v2.x on page 3-112 • 3-11-6 Set the Remo te Archive’s Network Information - Software v2.x on page 3-113 • 3-11-7 Save the New Settings - Software v2.x on page 3-113 • 3-11-8 Verification of a Connection Set Up to EchoPAC PC - Software v2.x on page 3-114 • 3-11-9 HL7 Communication Setup for Software v2.x on page 3-122 • 3-11-10 Set Up Connection to a DICOM Server in a Network - Software v2.x on page 3-126 3-11-3 Compatibility 3-11-3-1 Vivid 7 with So ftware Version v2.4.x Vivid 7 with Software Version v2.4.x (Syst em Software v1.5.1) can communicate with: • EchoPAC PC with software version v2.4.x • EchoPAC PC with software version v3.x • EchoPAC PC with software version v4.x • EchoPAC PC with software version v5.x • EchoPAC PC with software version v6.x • EchoPAC PC with software version v7.x • Image Vault 3 • Image Vault 4 • EchoServer 7
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 110 Section 3-11 - Connectivity Setup - Software v2.x 3-11-4 Select TCP/IP Set-up Screen - Software v2.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . Figure 3-116 Measuring and Analyzes - Global Level Connectivity
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 111 3-11-4 Select TCP/IP Set-up Screen - Software v2.x (cont’d) 2.) Select CONNECTIVITY (in the lower part of the window). 3.) Select the TCP/IP TAB (it is named Tcpip). (See example in Figure 3-118.) Figure 3-117 Connectivity Opening Screen Figure 3-118 TCP/IP Set-up Screen, Overview Connectivity TCP/IP Tab IP settings area for the scanner Remote Archive Setup area
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 112 Section 3-11 - Connectivity Setup - Software v2.x 3-11-5 Set the Scanner’s Network Information - Software Version v2.x In the IP settings area of the screen, enter the; • IP Address for scanner. (Default IP Address from factory: 10.0.0.3). • Subnet Mask for scanner. (Default Subn et Mask from factory: 255.255.255). • IP address for Default Gateway. (Default Subnet Mask from factory: 0.0.0.0). NOTE: For software v2.x, the Port Number has been set to 104. It can not be adjusted. . Figure 3-119 TCP/IP Set-up for Vivid 7 IP Address Subnet Mask Default Gateway
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 113 3-11-6 Set the Remote Archive’s Ne twork Information - Software v2.x In the Remote Archive Setup area of the screen (see example in Figure 3-118.), enter the; 1.) Remote Archive IP address. (Default IP Address from factory: 10.0.0.4). 2.) Remote Archive Name. (Default Remote Ar chive Name from factory: ECHOPAC7-000001). . 3-11-7 Save the New Settings - Software v2.x 1.) Press SAVE SETTINGS to save the new settings. The new settings are saved to a common settings file. After a restart, the settings are also included in other screens. 2.) Restart Vivid 7 to make the changes take place. Figure 3-120 TCP/IP Set-up fo r Remote Archive (EchoPAC PC) Remote Archive IP address Remote Archive Name Remote Archive Setup area Save settings
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 114 Section 3-11 - Connectivity Setup - Software v2.x 3-11-8 Verification of a Connection Set Up to EchoPAC PC - Software v2.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 2.) Select CONNECTIVITY (in the lower part of the window). 3.) Select the VIEWS tab, located in upper part of window, to display the Current Dataflow Views screen, see Figure 3-121 . 4.) Select the DATAFLOW button to see all available data flows, see Figure 3-122. 5.) Select the + beside the dataflow you want to check (here: Remote Archive - Remote HD) to see its contents. Figure 3-121 Current Dataflow Views screen Figure 3-122 Pre-configured Data flows in Vivid 7 VIEWS tab CONNECTIVITY tab DATAFLOW button DATAFLOW button Select +
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 115 3-11-8 Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d) . The Remote Archive - Remote HD includes an EchoPAC PC with one Service also called Remote Archive - Remote HD . The name of this service indicates that the archive is located on a hard disk on a remote server (EchoPAC PC is used as an server). 6.) Select the EchoPAC PC’s name to highlight it, see Figure 3-123. 7.) Select CHECK , see Figure 3-124 , to verify if there is a connecti on between the ultrasound unit and the EchoPAC PC Workstation. Figure 3-123 Remote HD dataflow contents Figure 3-124 Verify dataflow connection Select (highlight) the server’s name Select CHECK
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 116 Section 3-11 - Connectivity Setup - Software v2.x 3-11-8 Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d) • In the example in Figure 3-125, the red “X” next to the computer sign ( ), means that the check failed. The cause may be; - the setup of the network information, either on Vivid 7 or on the EchoPAC PC, or - the network connection is broken somewhere between the Vivid 7 and the EchoPAC PC. NOTE: If the Check button fails directly, the AE Title is probably wrong. If the Check button fails after a long time (correspo nding to the timeout), the IP address or Port Number is probably wrong. • A green light next to the computer sign me ans that a network connection is in place. Figure 3-125 Verification of connection failed Red “X”