GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 39 7-6-9-1 Windows 2000 Utilities (cont’d) Network Status This screen displays all open TCP/IP Ports on Vivid 7. Figure 7-42 Network Status
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 40 Section 7-6 - Diagnostics Tools 7-6-9-1 Windows 2000 Utilities (cont’d) Windows 2000 Services This page displays a list of active Windows 2000 services. A Windows 2000 Service is a computer program that has been automatically started and is running in the background on the computer. Figure 7-43 Windows 2000 Services
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 41 7-6-9-1 Windows 2000 Utilities (cont’d) User Accounts This page lists the Users that has been given access to this Vivid 7. Figure 7-44 User Accounts
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 42 Section 7-6 - Diagnostics Tools 7-6-9-1 Windows 2000 Utilities (cont’d) Shared Resources This screen displays all shared network resources on Vivid 7. Figure 7-45 Shared Resources
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 43 7-6-9-1 Windows 2000 Utilities (cont’d) System Shutdown NOTE: System Shutdown is intended for use from remote computer. This menu gives you the ability to eith er Restart or Shutdown the Vivid 7. NOTE: Retain Disruptive Mode should be kept unchecke d if you are working locally on the scanner. It must be checked if working from an external computer. Figure 7-46 Restart or Shutdown System
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 44 Section 7-6 - Diagnostics Tools 7-6-9-2 Scanner Utilities System Shutdown See: System Shutdown on page 7-43 Set Debug UtilityThis function is only to be used on Request fr om GE Service. It is used for setting Debug Variables. Figure 7-47 Set Debug Utility
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 45 7-6-9-2 Scanner Utilities (cont’d) Resource File Utility From this page you can review the listed User Defined Resource File and save it to MO. -View Select View to examine the User Defined Resource file on-screen. -Save Select Save and navigate to the MO drive to save the User Defined Resource file. Figure 7-48 Resource File Utility
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 46 Section 7-6 - Diagnostics Tools 7-6-10 Replacement Field is not populated. Figure 7-49 Part Replacement Page
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 47 7-6-11 PM Field is not populated. 7-6-12 Home Select HOME to return to the Service Home Page. 7-6-13 Exit From Diagnostics Select the “X” in the upper right corner of the screen to exit Diagnostics. Figure 7-50 Planned Maintenance Page
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 48 Section 7-7 - Common Diagnostics Section 7-7 Common Diagnostics Common Diagnostics is a collection of diagnostics to ols that are common for several of GE Ultrasound’s products. The Common Diagnostics folder has two sub-folders: • Utilities • PC Diagnostics Figure 7-51 Common Diagnostics