GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 117 3-11-8 Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d) 8.) Select the DATAFLOW tab to view the Dataflow Services window. 9.) Select Remote Archive - Remote HD on the drop-down menu to see the contents of that dataflow. Figure 3-126 Dataflow Services Window Figure 3-127 Obtaining Dataflow Contents Overview Dataflow selection area Dataflow Services area DATAFLOW tab Locale Archive -Internal HD service DATAFLOW tab Dataflow Name drop- down menu Remote Archive -Remote HD
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 118 Section 3-11 - Connectivity Setup - Software v2.x 3-11-8 Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d) Figure 3-128 Remote Archive - Remote HD Dataflow Contents Overview Remote Archive - Remote HD
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 119 3-11-8 Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d) 10.) Select the SERVICES tab (located near the top of the window). 11.) Select the EchoPAC PC Workstation you want to connect to, from the “Name” drop-down menu. In this example, it is called “EchoPAC7-000000”. Figure 3-129 Dataflow Services Figure 3-130 Dataflow Services, Choices and Setup Name drop-down menu SERVICES tab Local Archive - Local Storage Select Server Name drop-down menu
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 120 Section 3-11 - Connectivity Setup - Software v2.x 3-11-8 Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d) 12.) Verify that the IP-address is correct, in dicating that you have selected the right unit. If the IP number is wrong, it means that yo u are addressing the wrong device (computer). Correct the address and do a new verification before you continue. Figure 3-131 Verify/change IP address to selected device. IP address for selected device
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 121 3-11-8 Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d) 13.) In the list of “Services”, select “Remote Archive - Remote Storage”. 14.) In the list of Services select Remote Archive - Remote Storage. 15.) Change the Corresponding sql.ini name to reflect the name of the ECHOPAC7-nnnnnn Workstation. The Corresponding sql.ini name is a combination of the host name plus a space and the two letters “DB”. Example: If the EchoPAC PC name is ECHOPAC7-987654, the Corresponding sql.ini name will be ECHOPAC7- 987654 DB. 16.) Change the Placement of Image Files and Placement of Report Files names so they reflect the name of the EchoPAC PC Workstation. Example: If the EchoPAC PC name is ECHOPAC7-987654, the Placement of Image Files and Placement of Report Files are both ECHOPAC7-987654\Archive. This completes the EchoPAC PC connection setup. NOTE: Remember that these settings have no effect until you have logged on to the dataflow. It may be necessary to log on and log off this dataflow before the settings work. Figure 3-132 Dataflow services Choice and Setup Remote Archive - Remote Storage Corresponding sql.ini name Storage location for Image Files Storage location for Report Files
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 122 Section 3-11 - Connectivity Setup - Software v2.x 3-11-9 HL7 Communication Se tup for Software v2.x 3-11-9-1 Introduction The procedures below tell you how to set up the Vivid 7 workstation so it can connect to the Hospital Information System (HIS), via th e Vivid HL7 Gateway. By connec ting to the HIS, demographic information i.e Patient ID, Name, Gender etc. can be pulled from the HIS, reducing “double-work” and typing errors. Patient Reports can be sent back to the HIS for storage together with the rest of the Patient Records. 3-11-9-2 Preparations 1.) Before you can set up the HL 7 Communicat ion, you need to collect some information. - The IP address for the Vivid HL7 Gateway - Port Number used by the Vivid HL7 Gateway (Default port: 3320) Typical source for this information is the Mitra support personnel. Figure 2-2 Pre-quote Wo rksheet for DICOM Network Information on page 2-14 can be used for gathering this information. 2.) Provide the computer name of each EchoPAC PC Workstation and Vivid 7 being configured to the Mitra support personnel. 3-11-9-3 Configuration of Modality Worklist Services Identify Dataflow to be Used 1.) Identify which dataflow is to be used at the si te. If they will be using DICOM Modality Worklist, then the dataflow must include a Worklist service. 2.) On the Connectivity screen, select the Dataflow tab. 3.) Select the name of the sites dataflow from the pull-down. 4.) In the Services section, locate the Worklist service. 5.) Record the services name and destination device. Configure the DICOM Worklist Service 1.) Select the Services tab. 2.) Select the Worklist destination device from the pull down menu. 3.) Enter the IP address for the Vivid HL7 Gateway. 4.) Enter a descriptive name in the Name field. Exam ple: Vivid HL7 Gateway. The use of this field is optional). 5.) Enter AE Title. Use the Computer Name for the unit you are configuring. 6.) Enter Port Number. Port 33 20 is default, but Mitra may have provided another port number. 7.) Set Time-out to 90. 8.) Leave the rest of the fields unchanged.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 123 3-11-9-3 Configuration of Modality Worklist Services (cont’d) Configure the Search Criterias NOTE: This configuration is required. 1.) Select the Search Criterias button. 2.) Select the criteria Modality and select the Remove button. 3.) Press the OK button. Configure Search Criterias: Constrain Query by AE Title NOTE: This configuration is optional. It should only be performed if the site is directing orders to the Vivid 7 that is scheduled to perform the study. Please contact the Mitra support personnel to verify that this step is needed. 1.) In the Search Criterias dialog box, select 00400001 Scheduled Station AE Title from the Select Tag pull down menu. 2.) Enter the Vivid 7 computer name for the Value. 3.) Press the Add to List button. 4.) Press the OK button. Configure Search Criterias: Constrain Query by MRN NOTE: This configuration should be per formed if the site will need to constrain the Worklist queries by MRN (patient id) 1.) In the Search Criterias dialog box, select 00101 000 Other Patient IDs from the Select Tag pull down menu. 2.) Enter a 9 digit number for the Value. 3.) Select the Add to List button. 4.) Select the OK button. Assign Dicom Worklist name to Dataflow 1.) Select the Dataflow tab. 2.) From the Dataflow name pull down menu, select the Dataflow which will be used. 3.) Remove the Dicom Worklist service currently used in the Dataflow. 4.) From the Select Service pull down menu, select the Gateway Modality Worklist service. 5.) Select the Add button. 6.) Ensure that the service is displaying the value In for the Dir parameter and that its role is Primary.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 124 Section 3-11 - Connectivity Setup - Software v2.x 3-11-9-3 Configuration of Modality Worklist Services (cont’d) Test Connectivity with the Vivid HL7 Gateway The purpose of this test is to verify that the Vivid 7 has connectivity with the Vivid HL7 Gateway. 1.) From either the Current or Dataflow view, expand the Dataflow incorporating the Gateway Modality Worklist service. 2.) Expand the Gateway destination device. 3.) Select the Gateway Modality Worklist service. 4.) Select the Check button. - If connectivity was achieved, then there should be a green check next to the name of the Gateways Modality Worklist service. - If connectivity was not achieved, then there will be a red X next to the name of the Gateways Modality Worklist service. NOTE: If the Check button fails directly, the AE Title is probably wrong. If the Check button fails after a long time (corres ponding to the timeout), the IP address or Port Number is probably wrong. Troubleshoot Connectivity with the Gateway Typical causes if connection between Vivid 7 and HIS failed: 1.) No network connection. 2.) IP address and/or port number was incorrectly entered when configuring the Gateway Modality Worklist service. 3.) Mitra support personnel have not configured the Gateway yet to accept Modality Worklist queries from the Vivid 7. 4.) Verify that the Vivid 7 computer name matc hes the value specified for the Gateway Modality Worklist service and that it also matches the value supplied to the Mitra support personnel. Configure the Export to HL7 Path 1.) Enter the configuration screen for the EchoPAC PC. 2.) Press the Connectivity button. 3.) Select the Tools tab. 4.) In the section labeled Export Paths, make the following changes: a.) For the Export To HL7 path, enter the following information: \\ < Name or IP Address of the Vivid HL7 Gateway>\POLL_DIR Example: \\Gateway\POLL_DIR b.) Verify that Text is select for the Export file format.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 125 3-11-9-3 Configuration of Modality Worklist Services (cont’d) Test Connectivity with the Vivid HL7 Gateway 1.) Start a new study and acquire an image, or use an existing study. 2.) In the Measurements mode, make some of the measurements. 3.) In the Report screen, cr eate and store a report. 4.) In the Patient screen, select the study that you created the report in from the Examination List. 5.) From the More pull down menu, select HL7 Results. 6.) If the export was successful, then you should get the message: Export exam status: OK. 7.) If the export was not successful, then you should get the message, Export exam status: Unable to copy file or something similar. Troubleshoot Connectivity with the Vivid HL7 Gateway Typical causes if Connectivity with the Vivid HL7 Gateway failed: 1.) Path provided in Export to HL7 export path was incorrect. 2.) Mitra support personnel did not crea te the user account for the Vivid 7. NOTE: Required setup on the remote share: User: E1c2h3o4C5l6i7e8n9t Password: u1l3t5r7a 3.) The POLL_DIR directory on the Vivid HL7 Gateway was not shared or the permissions are incorrect. Verify with Mitra support personnel. 4.) Domain or other network issue. 5.) If Report files (ReportXXX.chm) are not exported to the Vivid HL7 Gatewa y, then the reports may not have been stored when they were created on the EchoPAC PC.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 126 Section 3-11 - Connectivity Setup - Software v2.x 3-11-10 Set Up Connection to a DICOM Se rver in a Network - Software v2.x No special service tools are required to do this set-up. 3-11-10-1 Preparations Obtain the following information from the network administrator: • DICOM server’s IP address, subnet mask (and gateway, if required). • Server’s Port Number. • Server’s AE Title. • IP address for the Vivid 7. 3-11-10-2 Establish Network Connection to Hospital’s Network Connect an Ethernet cable from the Vivid 7 to the wall network outlet. 3-11-10-3 Set up TCP/IP address on Vivid 7 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 2.) Select CONNECTIVITY . 3.) Select TCPIP tab. 4.) Enter IP-Address, Subnet Mask (and gateway, if required). 5.) Select SAVE SETTINGS . You will be asked to reboot, but wait with that until later. 3-11-10-4 DICOM Server’s Set Up It may be necessary for the administrator to enter in formation about Vivid 7 into the DICOM server’s set- up. This information is typical Vivid 7’s AE Title an d maybe also the IP address which administrator provides. (See 3-11-10-5 Vivid 7’s AE Title on page 3-126 for a description on how to obtain Vivid 7’s AE Title.) 3-11-10-5 Vivid 7’s AE Title 1.) Select CONNECTIVITY . 2.) Select the SERVICES tab. Vivid 7’s AE Title is listed below My Computer (here it is VIVID7-000000). Figure 3-133 Vivid 7’s AE Title Vivid 7’s AE Title