6/02 2-251 DC1632/2240 15-362, 15-367Status Indicator RAPs Prelaunch Training/Review 15-362 X Hard FailureCommunication Failure.ProcedureSwitch the power off then on. The problem continues. YN Return to Service Call Procedures. Check the connectors on the IIT/IPS PWB (PL 18.3). If the check ia OK, replace the IIT/IPS PWB (PL 18.3). 15-367 X PIO FailureCommunication Failure.ProcedureSwitch the power off then on. The problem continues. YN Return to Service Call Procedures. Check the connectors on the IIT/IPS PWB (PL 18.3). If the check ia OK, replace the IIT/IPS PWB (PL 18.3).
6/02 2-252DC1632/2240 15-370, 15-371 Prelaunch Training/Review Status Indicator RAPs 15-370 X PIO Initialization Failure 1Communication Failure.ProcedureSwitch the power off then on. The problem continues. YN Return to Service Call Procedures. Check the connectors on the IIT/IPS PWB (PL 18.3) and the ESS PWB (PL 13.1). If the check ia OK, replace the ESS PWB (PL 13.1). If the problem continues, replace the IIT/IPS PWB (PL 18.3). 15-371 X PIO Initialization Failure 2Communication Failure.ProcedureSwitch the power off then on. The problem continues. YN Return to Service Call Procedures. Check the connectors on the IIT/IPS PWB (PL 18.3) and the ESS PWB (PL 13.1). If the check ia OK, replace the ESS PWB (PL 13.1). If the problem continues, replace the IIT/IPS PWB (PL 18.3).
6/02 2-253 DC1632/2240 15-372, 15-375Status Indicator RAPs Prelaunch Training/Review 15-372 X PIO Initialization Failure 3Communication Failure.ProcedureSwitch the power off then on. The problem continues. YN Return to Service Call Procedures. Check the connectors on the IIT/IPS PWB (PL 18.3) and the ESS PWB (PL 13.1). If the check ia OK, replace the ESS PWB (PL 13.1). If the problem continues, replace the IIT/IPS PWB (PL 18.3). 15-375 X PIO Before Scan Failure Communication Failure.ProcedureSwitch the power off then on. The problem continues. YN Return to Service Call Procedures. Check the connectors on the IIT/IPS PWB (PL 18.3) and the ESS PWB (PL 13.1). If the check ia OK, replace the ESS PWB (PL 13.1). If the problem continues, replace the IIT/IPS PWB (PL 18.3).
6/02 2-254DC1632/2240 15-376, 15-377 Prelaunch Training/Review Status Indicator RAPs 15-376 X PIO Non-match Failure 1Communication Failure.ProcedureSwitch the power off then on. The problem continues. YN Return to Service Call Procedures. Check the connectors on the IIT/IPS PWB (PL 18.3) and the ESS PWB (PL 13.1). If the check ia OK, replace the ESS PWB (PL 13.1). If the problem continues, replace the IIT/IPS PWB (PL 18.3). 15-377 X PIO Non-match Failure 2 Communication Failure.ProcedureSwitch the power off then on. The problem continues. YN Return to Service Call Procedures. Check the connectors on the IIT/IPS PWB (PL 18.3) and the ESS PWB (PL 13.1). If the check ia OK, replace the ESS PWB (PL 13.1). If the problem continues, replace the IIT/IPS PWB (PL 18.3).
6/02 2-255 DC1632/2240 15-380, 15-381Status Indicator RAPs Prelaunch Training/Review 15-380 CCD AGCAutomatic Gain Control (AGC) for CCD Channel 1 red failed.Initial Actionsï Verify that Platen Glass is installed correctly. ï Verify that NVM 715-400 = 1ProcedureNOTE: Refer the BSD 6.1, 6.2 and 6.7 as required. Enter dC330 [6-002] and press Start. The Exposure Lamp illuminates. YN Go to the 6-371 Exposure Lamp RAP. The Exposure Lamp illumination is bright and steady. YN Go to p/j722 and p/j724 on BSD 6.1 and check the wires for an open or short circuit. Verify that j742 on the Lamp Ballast PWB and p/j724 on the IIT/IPS PWB are seated. If the wires and connectors are good, replace the Exposure Lamp (PL 18.6). If the problem continues, replace the Lamp Ballast PWB (PL 18.6). If the problem continues, replace the IIT/IPS PWB (PL 18.3). Remove the Platen Glass (REP 6.2). The white reference strip on the bottom of the Platen Glass is clean and undamaged. YN Clean the white reference strip or replace the Platen Glass (PL 18.3). Go to p/j721 on BSD 6.2 and check the flat cable on the IIT/IPS PWB and j741 on the CCD PWB for damage. Reseat the flat cable. If the problem continues, replace the Lens Kit (PL 18.4). If the problem continues, replace the IIT/IPS PWB (PL 18.3). 15-381 CCD AGCAutomatic Gain Control (AGC) for CCD Channel 2 red failed.Initial Actionsï Verify that Platen Glass is installed correctly. ï Verify that NVM 715-400 = 1ProcedureNOTE: Refer the BSD 6.1, 6.2 and 6.7 as required. Enter dC330 [6-002] and press Start. The Exposure Lamp illuminates. YN Go to the 6-371 Exposure Lamp RAP. The Exposure Lamp illumination is bright and steady. YN Go to p/j722 and p/j724 on BSD 6.1 and check the wires for an open or short circuit. Verify that j742 on the Lamp Ballast PWB and p/j724 on the IIT/IPS PWB are seated. If the wires and connectors are good, replace the Exposure Lamp (PL 18.6). If the problem continues, replace the Lamp Ballast PWB (PL 18.6). If the problem continues, replace the IIT/IPS PWB (PL 18.3). Remove the Platen Glass (REP 6.2). The white reference strip on the bottom of the Platen Glass is clean and undamaged. YN Clean the white reference strip or replace the Platen Glass (PL 18.3). Go to p/j721 on BSD 6.2 and check the flat cable on the IIT/IPS PWB and j741 on the CCD PWB for damage. Reseat the flat cable. If the problem continues, replace the Lens Kit (PL 18.4). If the problem continues, replace the IIT/IPS PWB (PL 18.3).
6/02 2-256DC1632/2240 15-382, 15-383 Prelaunch Training/Review Status Indicator RAPs 15-382 CCD AGCAutomatic Gain Control (AGC) for CCD Channel 3 green failed.Initial Actionsï Verify that Platen Glass is installed correctly. ï Verify that NVM 715-400 = 1ProcedureNOTE: Refer the BSD 6.1, 6.2 and 6.7 as required. Enter dC330 [6-002] and press Start. The Exposure Lamp illuminates. YN Go to the 6-371 Exposure Lamp RAP. The Exposure Lamp illumination is bright and steady. YN Go to p/j722 and p/j724 on BSD 6.1 and check the wires for an open or short circuit. Verify that j742 on the Lamp Ballast PWB and p/j724 on the IIT/IPS PWB are seated. If the wires and connectors are good, replace the Exposure Lamp (PL 18.6). If the problem continues, replace the Lamp Ballast PWB (PL 18.6). If the problem continues, replace the IIT Driver PWB (PL 18.3). Remove the Platen Glass (REP 6.2). The white reference strip on the bottom of the Platen Glass is clean and undamaged. YN Clean the white reference strip or replace the Platen Glass (PL 18.3). Go to p/j721 on BSD 6.2 and check the flat cable on the IIT/IPS PWB and j741 on the CCD PWB for damage. Reseat the flat cable. If the problem continues, replace the Lens Kit (PL 18.4). If the problem continues, replace the IIT/IPS PWB (PL 18.3). 15-383 CCD AGCAutomatic Gain Control (AGC) for CCD Channel 4 green failed.Initial Actionsï Verify that Platen Glass is installed correctly. ï Verify that NVM 715-400 = 1ProcedureNOTE: Refer the BSD 6.1, 6.2 and 6.7 as required. Enter dC330 [6-002] and press Start. The Exposure Lamp illuminates. YN Go to the 6-371 Exposure Lamp RAP. The Exposure Lamp illumination is bright and steady. YN Go to p/j722 and p/j724 on BSD 6.1 and check the wires for an open or short circuit. Verify that j742 on the Lamp Ballast PWB and p/j724 on the IIT/IPS PWB are seated. If the wires and connectors are good, replace the Exposure Lamp (PL 18.6). If the problem continues, replace the Lamp Ballast PWB (PL 18.6). If the problem continues, replace the IIT Driver PWB (PL 18.3). Remove the Platen Glass (REP 6.2). The white reference strip on the bottom of the Platen Glass is clean and undamaged. YN Clean the white reference strip or replace the Platen Glass (PL 18.3). Go to p/j721 on BSD 6.2 and check the flat cable on the IIT/IPS PWB and j741 on the CCD PWB for damage. Reseat the flat cable. If the problem continues, replace the Lens Kit (PL 18.4). If the problem continues, replace the IIT/IPS PWB (PL 18.3).
6/02 2-257 DC1632/2240 15-384, 15-385Status Indicator RAPs Prelaunch Training/Review 15-384 CCD AGCAutomatic Gain Control (AGC) for CCD Channel 5 blue failed.Initial Actionsï Verify that Platen Glass is installed correctly. ï Verify that NVM 715-400 = 1ProcedureNOTE: Refer the BSD 6.1, 6.2 and 6.7 as required. Enter dC330 [6-002] and press Start. The Exposure Lamp illuminates. YN Go to the 6-371 Exposure Lamp RAP. The Exposure Lamp illumination is bright and steady. YN Go to p/j722 and p/j724 on BSD 6.1 and check the wires for an open or short circuit. Verify that j742 on the Lamp Ballast PWB and p/j724 on the IIT/IPS PWB are seated. If the wires and connectors are good, replace the Exposure Lamp (PL 18.6). If the problem continues, replace the Lamp Ballast PWB (PL 18.6). If the problem continues, replace the IIT Driver PWB (PL 18.3). Remove the Platen Glass (REP 6.2). The white reference strip on the bottom of the Platen Glass is clean and undamaged. YN Clean the white reference strip or replace the Platen Glass (PL 18.3). Go to p/j721 on BSD 6.2 and check the flat cable on the IIT/IPS PWB and j741 on the CCD PWB for damage. Reseat the flat cable. If the problem continues, replace the Lens Kit (PL 18.4). If the problem continues, replace the IIT/IPS PWB (PL 18.3). 15-385 CCD AGCAutomatic Gain Control (AGC) for CCD Channel 6 blue failed.Initial Actionsï Verify that Platen Glass is installed correctly. ï Verify that NVM 715-400 = 1ProcedureNOTE: Refer the BSD 6.1, 6.2 and 6.7 as required. Enter dC330 [6-002] and press Start. The Exposure Lamp illuminates. YN Go to the 6-371 Exposure Lamp RAP. The Exposure Lamp illumination is bright and steady. YN Go to p/j722 and p/j724 on BSD 6.1 and check the wires for an open or short circuit. Verify that j742 on the Lamp Ballast PWB and p/j724 on the IIT/IPS PWB are seated. If the wires and connectors are good, replace the Exposure Lamp (PL 18.6). If the problem continues, replace the Lamp Ballast PWB (PL 18.6). If the problem continues, replace the IIT Driver PWB (PL 18.3). Remove the Platen Glass (REP 6.2). The white reference strip on the bottom of the Platen Glass is clean and undamaged. YN Clean the white reference strip or replace the Platen Glass (PL 18.3). Go to p/j721 on BSD 6.2 and check the flat cable on the IIT/IPS PWB and j741 on the CCD PWB for damage. Reseat the flat cable. If the problem continues, replace the Lens Kit (PL 18.4). If the problem continues, replace the IIT/IPS PWB (PL 18.3).
6/02 2-258DC1632/2240 15-790 Prelaunch Training/Review Status Indicator RAPs 15-790 X Detail Failure Communication Failure.ProcedureSwitch the power off then on. The problem continues. YN Check the connectors on the IIT/IPS PWB (PL 18.3). If the check ia OK, replace the IIT/ IPS PWB (PL 18.3). Return to Service Call Procedures.
6/02 2-259 DC1632/2240 16-311, 16-315Status Indicator RAPs Prelaunch Training/Review 16-311 ScannerScanner not detected during power up.ProcedureSwitch power off then on. The problem continues. YN Return to service call procedures. Check connections between Scanner and IOT. Service as required (PL 18.3) (PL 18.4). 16-315 IIT InterfaceIIT interface error.ProcedureSwitch power off then on. The problem continues. YN Return to service call procedures. Pivot down T5 T6 HVPS (REP 1.6) to access I/F PWB. Check that connections on I/F PWB are secure. If problem continues, replace I/F PWB (PL 9.1).
6/02 2-260DC1632/2240 16-316, 16-317 Prelaunch Training/Review Status Indicator RAPs 16-316 Page MemoryPage memory not detected.ProcedureSwitch power off then on. The problem continues. YN Return to service call procedures. Check installation of Page Memory of the Scanner. 16-317 Page MemoryPage Memory error.ProcedureSwitch power off then on. The problem continues. YN Return to service call procedures. Replace the Page Memory of the Scanner.