Home > GE > Ultrasound System > GE Logiq 9 Service Manual

GE Logiq 9 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 570
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 3 Installation 3-1
    Chapter 3
    Installation
    Section 3-1
    Overview
    3-1-1 The Purpose of Chapter 3
    This chapter contains information needed to install the unit. Included are procedures to receive, unpack 
    and configure the equipment.
    Table 3-1    Contents in Chapter 3
    SectionDescriptionPage Number
    3-1
    Overview3-1
    3-2
    Installation Reminders3-2
    3-3
    Receiving and Unpacking the Equipment3-4
    3-4
    Preparing for Installation3-5
    3-5
    Completing the Installation3-6
    3-6
    System Configuration3-9
    3-7
    Available Probes3-15
    3-8
    Software/Option Configuration3-15
    3-9
    Connectivity Installation Worksheet3-16
    3-10
    Insite IP Address Configuration3-17
    3-11
    Before Starting a Software Upgrade3-17
    3-12
    Loading Base Image Software3-18
    3-13
    Loading Application Software3-24
    3-14
    System Setup3-28
    3-15
    Optional Software Procedures3-30
    3-16
    Installing USB Line Printer - HP970/9903-44
    3-17
    Paperwork3-45 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    3-2 Section 3-2 - Installation Reminders
    Section 3-2
    Installation Reminders
    3-2-1 Average Installation Time
     
    3-2-2 Installation Warnings
    1.) Since the LOGIQ™ 9 weighs approximately 225 kg (500 lb.) without options, preferably two people 
    should unpack it. Two people are also preferable for installing any additional items in excess of 35 
    pounds.
    2.) There are no operator serviceable components. To prevent shock, do not remove any covers or 
    panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service 
    personnel should carry out servicing and troubleshooting.
    3.) After being transported, the unit may be very cold or hot. If this is the case, allow the unit to 
    acclimate before you turn it on. It requires one hour for each 2.5°C increment its temperature is 
    below 10°C or above 40°C.
     
    3-2-2-1 Brake Pedal Operation
    3-2-2-2 Operator I/O Panel Position
    If weight is placed on the Operator I/O Panel in it’s extended position the console could tip over. Table 3-2    Average Installation Time
    DescriptionAverage Installation TimeComments
    Unpacking the scanner 0.5 hours
    Scanner wo/options 0.5 hours Dependant on the required configuration
    DICOM Option 0.5 - 2.0 hours Dependant on the configuration amount
    InSite Option 0.5 hours
    CAUTIONEquipment damage possibility. Turning the system on without acclimation after arriving at site 
    may cause the system to be damaged.
    Table 3-3    Acclimation Time
    °C60 55 50 45 40 35 30 25 20 15 10 5 0 -5 -10 -15 -20 -25 -30 -35 -40
    °F140131122113104968677685950413223145-4-13-22-31-40
    hrs864200000002468101214161820
    WARNINGWARNINGREMEMBER: If the front caster swivel lock is engaged for transportation, pressing the 
    release pedal once disengages the swivel lock. You must depress the release pedal a 
    second time to engage the brake.
    WARNINGWARNINGThe system should NOT be moved with the Operator I/O Panel extended. Move the 
    Operator I/O Panel to it’s centered and locked position. Lower the Operator I/O Panel 
    as much as possible before moving the system. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 3 Installation 3-3
    3-2-3 Safety Reminders
    DANGERWHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE 
    AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH 
    THE UNIT!
    CAUTIONTwo people should unpack the unit because of its weight. Two people are required whenever 
    a part weighing 19kg (35 lb.) or more must be lifted.
    CAUTIONIf the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate 
    to its operating environment. 
    CAUTIONTo prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use 
    a three to two prong adapter. This defeats safety grounding.
    CAUTIONDo NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is 
    present.
    CAUTIONDo not use a 20 Amp to 15 Amp adapter on the 120 Vac unit’s power cord. This unit requires 
    a dedicated 20 A circuit and can have a 15A plug if the on board peripherals do not cause the 
    unit to draw more than 14.0 amps.
    CAUTIONDo not operate this unit unless all board covers and frame panels are securely in place. 
    System performance and cooling require this.
    CAUTIONOPERATOR MANUAL(S) 
    The User Manual(s) should be fully read and understood before operating the LOGIQ™ 9 and 
    kept near the unit for quick reference.
    CAUTIONACOUSTIC OUTPUT HAZARD
    Although the ultrasound energy transmitted from the LOGIQ™ 9 probe is within FDA limits, 
    avoid unnecessary exposure. Ultrasound energy can produce heat and mechanical damage.
    Figure 3-1   Environmental Labels 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    3-4 Section 3-3 - Receiving and Unpacking the Equipment
    Section 3-3
    Receiving and Unpacking the Equipment 
     
    CAUTIONHave two people available to unpack the LOGIQ™ 9. Attempts to move the unit considerable 
    distances or on an incline by one person could result in injury or damage or both.
    Table 3-4    Unpacking Procedure
    StepTask
    1.
    Open the four hinges on each door
    2.
    Remove the doors at both ends of the box
    3.
    Place the front door against the rear edge of the pallet. The front door can be used as a ramp. 
    4. Carefully remove the accessory box, and any other items, including the wooden shelf above the scanner’s keyboard and all the filling 
    material, from the wooden box. 
    5.
    Press once on the swivel lock to release the brake. 
    Caution: Two people are needed in the next step due to the weight of the equipment.
    6. Carefully move the scanner out of the wooden box, down the ramp.
    7. Remove the clear plastic (wrapped around the scanner) from the unit.
    Be sure to place the
    ramp (front door) as
    illustrated here.Front Door
    used as ramp.
    Swivel Lock 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 3 Installation 3-5
    Section 3-4
    Preparing for Installation
    3-4-1 Verify Customer Order
    Compare items received by the customer to that which is listed on the delivery order. Report any items 
    that are missing, back ordered or damaged.
    3-4-2 Physical Inspection
    3-4-2-1 System Voltage Settings
    Verify that the scanner is set to the correct voltage.
    The Voltage settings for the LOGIQ™ 9 Scanner is found on a label to the right of the Power switch and 
    External I/O, on the rear of the system.
    3-4-2-2 Video Formats
    Check that the video format is set to the locally used video standard, NTSC or PAL.
    3-4-3 EMI Protection
    This unit has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Many of the 
    covers, shields, and screws are provided primarily to protect the system from image artifacts caused by 
    this interference. For this reason, it is imperative that all covers and hardware are installed and secured 
    before the unit is put into operation.Figure 3-2   Rating Plate Example
    WARNINGWARNINGCONNECTING A LOGIQ™ 9 SCANNER TO THE WRONG VOLTAGE LEVEL WILL MOST 
    LIKELY DESTROY THE SCANNER.  
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    3-6 Section 3-5 - Completing the Installation
    Section 3-5
    Completing the Installation
    3-5-1 Power On / Boot Up
    NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be 
    able to boot if power is recycled too quickly.
    3-5-1-1 Scanner Power On
    1.) Connect the Power Cable to the back of the system.
    2.) Ensure the Cable Clip slips securely over the shoulders on the molded plug.
    3.) Connect the Main Power Cable to a hospital grade power receptacle with the proper rated voltage. 
    Never use an adapter that would defeat the safety ground. 
    4.) Switch ON the Circuit Breaker at the rear of the system.
    When power is applied to the scanner and the rear Circuit Breaker is turned ON, power is distributed to 
    the Cooling Unit, Control Panel, Monitor, Internal and External I/O Boards, Peripherals and the Back-
    end Processor.Figure 3-3   Circuit Breaker and Power Cable on Back of Scanner
    NOTICEWhen AC power is applied to the scanner, the Standby switch on the Control panel illuminates amber, 
    indicating the Back End Processor is in standby mode.
    Circuit Breaker
    Mains Power Cable
    Cable Clip 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 3 Installation 3-7
    3-5-1-2 Back-end Processor Boot Up
    Press the Standby switch on the Control Panel once.
    When the Standby switch on the Control Panel is pressed once, the Back-end Processor starts and the 
    software code is distributed to initiate the scanner.
    No status messages are displayed during this process.
    3-5-2 Power Off/ Shutdown
    NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be 
    able to boot if power is recycled too quickly.
    3-5-2-1 Back-end Processor Power Down
    1.) Press the Standby switch on the Control Panel once to display the SYSTEM - EXIT menu.
    2.) Select Shutdown from the SYSTEM - EXIT menu.Figure 3-4   Standby Switch for Back-end Processor Boot Up
    Figure 3-5   System Exit Menu for Back-end Processor Power Down
    Standby Switch 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    3-8 Section 3-5 - Completing the Installation
    3-5-2-2 Scanner Shutdown
    1.) Switch OFF the Circuit Breaker at the back of the system.
    2.) Disconnect the Mains Power Cable is necessary. For example: Relocating the scanner.
    3-5-3 Transducer Connection
    1.) Connect a transducer to one of the three rightmost transducer receptacle as follows:
    A.) Ensure that the transducer twist lock lever to the horizontal position.
    B.) Insert the transducer connector on the receptacle guide pin until it touches the receptacle 
    mating surface.
    C.) Twist the transducer twist lock lever to vertical position to lock it in place. Twist the lever 
    to the horizontal position to disconnect the transducer.
    NOTE: It is not necessary to turn OFF power to connect or disconnect a transducer.Figure 3-6   Circuit Breaker and Power Cable on Back of Scanner
    NOTICEDisconnecting the Mains Power Cable before switching OFF the Circuit Breaker will activate the 
    uninterruptible power system (UPS) in the backend processor, forcing an ordered shutdown of the 
    system.
    NOTICEThe UPS serves only as a battery backup that allows for a soft shut-down of the scanner to prevent file 
    corruption. IT DOES NOT REGULATE INCOMING VOLTAGE.
    Circuit Breaker
    Mains Power Cable
    Cable Clip 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 3 Installation 3-9
    Section 3-6
    System Configuration
    3-6-1 System Specifications
    3-6-1-1 Physical Dimensions Physical Dimensions of 
    The physical dimensions of the LOGIQ™ 9 unit are summarized in Ta b l e 3 - 1 and Ta b l e 3 - 2.
    Table 3-1 lists the size of LOGIQ™ 9, with monitor and on-board peripherals.
    3-6-1-2 Weight without Monitor and Peripherals
    3-6-1-3 Acoustic Noise Output:
    Less than 70dB(A) according to DIN 45635 - 19 - 01 - KL2. 
    3-6-2 Electrical Specifications
    3-6-3 On-Board Optional Peripherals
     
    NOTE: There are no external peripheral options for the LOGIQ™ 9Table 3-1    Physical Dimensions of LOGIQ™ 9
    HeightWidthDepthMinimumMaximum
    140.3 cm / 55.25 inches 160.7 cm / 63.25 inches 64 cm / 25.2 inches 89.9 cm / 35.4 inches
    Table 3-2    Weight of LOGIQ™ 9 With Monitor and Without Other Peripherals
    ModelWeight [kg]Weight [lbs.]
    LOGIQ™ 9 196.8 434
    Table 3-3    Electrical Specifications for LOGIQ™ 9
    VoltageTolerancesCurrentFrequency
    100-120 VAC ±10% 9.5 A 50-60 Hz
    200-240 VAC ±10% 4.75 A 50-60 Hz
    Table 3-4    Approved Peripherals
    DeviceManufacturerModelVideo Signal
    B/W Video Printer SONY UP-895MDW NTSC/PAL
    Video Cassette Recorder SONYSVO-9500MD
    SVO-9500-MDPNTSC
    PAL
    Video Cassette Recorder PanasonicAG-MD835
    AG-MD835P
    AG-MD835ENTSC
    NTSC
    PAL 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    3-10 Section 3-6 - System Configuration
    3-6-4 External I/O Connector Panel
    .i.External I/O Connector Panel;
     
    Figure 3-7   External I/O Panel Labels
    Figure 3-8   External I/O Panel Connectors
    123467
    910111213148
    5 
    						
    All GE manuals Comments (0)