    Chapter 8 Replacement Procedures 8-133
    • FEPS2 Power Supply Differences
    a.) There is more room to remove the PCI cable from the Scan Control Board and pull it through 
    the hole in the FEPS2.
    b.) Two AC connections are still required but are mounted on front of the FEPS2 for easy access.
    c.) The FEPS2 has two card ejectors so the screw in the bottom corner is not required to secure 
    the supply.
    d.) A special copper ground tab and ground plane grabber are incorporated in the FEPS2 to 
    ensure good contact to the FEP card rack.
    e.) A different FEP cover is needed, when an FEPS2 is installed, to add additional shielding 
    around the Top Plane Board connecting the XDIF to the RF Amp.
    f.) It is important that ALL screws be used when replacing the Front End Processor Cover.
    • FEPS3 Differences
    a.) FEPS3 Power Supplies must be used with TD5 and RF AMP2 Front End card racks. The +/- 
    6V and 13V produced are required for these boards. Refer to Table 5-2 on page 5-6.
    b.) Therefore, the FEPS3 can not be used with Front End card racks prior to R3.0.0 Software 
    (BT’03).Figure 8-143   FEPS2 Grounding Points
    Copper Grounding 
    Tab Ground Plane 
    Grabber 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-134 Section 8-34 - Front End Processor Boards Replacement Procedure
    • Installing the FEPS2/FEPS3
    When installing the FEPS2/;FEPS3 in a LOGIQ™ 9 Front End Card Rack, special attention needs to 
    be paid to the grounding points.
    1.) Slide the FEPS2/FEPS3 in the card rack rails. See Figure 8-143 on page 8-133 .
    - Ensure that the ground plane separating the SCB and FEPS2/;FEPS3 boards will slide into 
    the ground plane grabber on the edge of the FEPS2/;FEPS3.
    - Ensure that the copper grounding tab will make contact with the ground plane separating the 
    SCB and FEPS2/;FEPS3 boards.
    2.) Feed the PCI cable through the FEPS2/;FEPS3 and connect it to the SCB.
    3.) Connect the two AC cables to the FEPS2/;FEPS3.
    4.) Use the card ejectors to seat the boards pressing firmly on the top and bottom of the cards.
    5.) Double check all ground plane contacts.
    6.) When installing the FEP cover, ensure you use the new cover with the shielding riveted in the top 
    plane area.
    NOTE: Check the routing of the PCI cable through the Power Supply board. Ensure there are no loops 
    in the PCI cable that could short to test points or any part of the Power Supply.
    c.) The RF gasket on the inside of the Front End Processor cover MUST make good contact with 
    the Front End Processor card cage. To get a tight fit with the gasket, loosely fit all screws 
    through the cover into the card cage. Apply pressure to the bottom of the cover forcing it 
    upward as you tighten all screws from the bottom up. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-135
    Section 8-35
    Verify and Update Vital Product Data
    8-35-1 Introduction
    It is essential that Vital Product Data (VPD) is verified or updated after any hardware change to the Back 
    End or Front End Processors.
    NOTE: Failure to properly program Vital Product Data could cause adverse affects on system 
    operation and image quality.
    The items of concern are:
    - Verify that the Board or BEP Vital Product Gata is programmed
    - Program the new Board or BEP Vital Product data. (BEP VPD is stored on the FEP 
    backplane).
    8-35-2 Collecting Vital Product Data
    For the Back End Processor, remove the system left side cover and read the label on the front of the 
    processor. The Part Number is there. The Functional Revision is the PSP# (number) on the label. Get 
    the date for the BEP from this BEP rating plate. All other dates are taken from the Console rating plate.
    If the VPD Read Utility (8-35-3-3 Acquisition Diagnostics - Read EEPROM Data on page 8-138) shows 
    no data for the Equalization Board (EQ), obtain that data by removing the system right side cover, 
    remove the FEP cover, remove the EQ Board and obtain the data indicated in Table 8-9.
    Table 8-9    BEP VPD Data
    Circuit Board NamePart Number
    Functional Revision 
    PWA LetterFab Revision 
    PWB NumberBar Code or
    Serial NumberDate
    YYYYMMDD
    BEPBEP Rating Plate PSP Revison
    N/ANo Dash Allowed BEP Rating Plate
    2 0 0 _  _ _  01
    Table 8-10    Typical Circuit Board VPD Data
    Circuit Board NamePart Number
    Functional Revision 
    PWA LetterFab Revision 
    PWB NumberBar Code or
    Serial NumberDate
    YYYYMMDD
    Equalization 22602102 0 0 _  _ _  01 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-136 Section 8-35 - Verify and Update Vital Product Data
    8-35-2 Collecting Vital Product Data (cont’d)
    Table 8-11    VPD Data Collection
    Circuit Board NamePart Number
    Functional Revision 
    PWA LetterFab Revision 
    PWB NumberBar Code or
    Serial NumberDate
    YYYYMMDD
    BM Processor   2260207 Do Not Update 2 0 0 _  _ _  01
    Backplane 2260219 A 1 012345 2 0 0 _  _ _  01
    ConsoleFrom Rating Plate
    Only -2 can be enteredN/A N/A 2 0 0 _  _ _  01
    BEPBEP Rating Plate
    Only -3 can be enteredPSP Revison
    -2 after upgradeN/ANo Dash AllowedBEP Rating Plate
    2 0 0 _  _ _  01
     Equalization22602102 0 0 _  _ _  01
    EBM22736392 0 0 _  _ _  01
    Power Supply2260203 / 2273641 / 
    2376843 / 22602032 0 0 _  _ _  01
    RF Amplifier2260212 / 2341985 / 
    23576212 0 0 _  _ _  01
    Scan Control 2260214 / 23657392 0 0 _  _ _  01
    Time Delay 02260194 / 23628632 0 0 _  _ _  01
    Time Delay 12260194 / 23628632 0 0 _  _ _  01
    Time Delay22260194 / 23628632 0 0 _  _ _  01
    Time Delay 3 2260194 / 23628632 0 0 _  _ _  01
    Time Delay 4 2260194 / 23628632 0 0 _  _ _  01
    Time Delay 5 2260194 / 23628632 0 0 _  _ _  01
    Time Delay 62260194 / 23628632 0 0 _  _ _  01
    Time Delay 72260194 / 23628632 0 0 _  _ _  01
    Transducer Interface22602212 0 0 _  _ _  01
    PC2IP Not a FRU Part of BEP. Not Updated 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-137
    8-35-3 How to Update Vital Product Data on LOGIQ™ 9
    8-35-3-1 Initial Steps
    The System should be in the scan mode.  If not:
    1.) Properly shut down the system.
    2.) Plug in the service dongle to the USB Port on the External I/O.
    3.) Power up system
    8-35-3-2 Start Service Platform
    1.) Right Click on the Wrench icon to start the service platform.
    2.) A dialog box appears Service Platform is starting ...
    3.) It will take a few seconds for the Service Login to appear. Select login as GE Service
    4.) Enter current Service Password for Service Diagnostics.
    NOTE: Remember that the Caps Lock must be OFF.Figure 8-144   Service Login Screen 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-138 Section 8-35 - Verify and Update Vital Product Data
    8-35-3-3 Acquisition Diagnostics - Read EEPROM Data
    Navigate to the following menu location:
    1.) Acquisition Diagnostics>Diagnostic Groups>Diagnostic Utilities>Acquisition Hardware 
    Information Utilities>Acquisition Hardware Information Read Utilities.
     
    NOTE: If VPD Data is not present the proceed with updating VPD Data.Figure 8-145   Read EQ Board VPD
    Figure 8-146  Read Backplane VPD 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-139
    8-35-3-3Acquisition Diagnostics - Read EEPROM Data (cont’d)
    Under this menu you will see a listing for each board that has an EEPROM with a location for Vital 
    Product Data, see Figure 8-145 and Figure 8-146. The Board Data can be read individually or there is 
    also a Read All Boards Utility.
    2.) Click on the Backplane and Equalization items and select the Execute button to read the 
    contents of the EEPROM on each circuit board. This will create two files for each board in the 
    C:/Musashi/target/RW/Diagnostics directory. The files generated are a *.txt and *.dat file. The 
    *.txt file will be the working file to edit where the *.dat file will be the file showing the actual 
    hexadecimal data stored in the memory chip.
    8-35-3-4 Modify VPD Text Files
    Once data has been read back successfully and all files have been created, you are now ready to 
    modify the contents of the *.txt file and write the data back.
    1.) Open Windows Explorer via the CTRL-ALT-DEL method. Press CTRL-ALT-DEL 
    simultaneously, select Task Manager, select New Task - Run, type in explorer in the Create 
    New Task dialog box and select OK. If no Explorer window is open, click on the Explorer icon 
    in the task bar to open a window.
    2.) In Windows Explorer, navigate to C:/Musashi/target/RW/Diagnostics. Here you will see all the 
    *.txt and *.dat files you have just created via the Read Utility. Each of the VPD files will have 
    the prefix of “EEP-”. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-140 Section 8-35 - Verify and Update Vital Product Data
    8-35-3-4Modify VPD Text Files (cont’d)
    Hints for Working with the Text Files
    • Sort the files in the Explorer window, Figure 8-147, by clicking on the file TYPE
     heading. This will 
    sort the files and put the .txt files in one contigious group.
    • With the files sorted, to highlight all text files:
    - Cursor to the first text file and press the Right-SET
     key.
    - Cursor to the last file, hold down the SHIFT
     key and press the Right-SET key.
    - All files should be highlighted.
    - Ensure that the cursor is in the highlighted file area and press the Left-SET
     key.
    - In the drop down menu, cursor to OPEN
     and press the Right-SET key.
    - All text files will open up in Note PadFigure 8-147   Example of *.txt and *.dat File Listing 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-141
    Notes For VPD Text File Editing
    • There needs to be one (1) space between the colon (:) and the first character of the data string.
    •NO spaces, characters, tabs, etc. at the end of the data string. Do a Fn-End (Function-End) on the 
    LOGIQ
    ™ 9 keyboard to see the actual end of the line. Delete all spaces by back spacing to the text 
    string.
    • Double click on a word to select it for editing.
    • Copy a field like the date for multiple pasting by: 
    - Double clicking on the edited date field to select it.
    -CTRL - C
     to copy the field.
    - Double clicking on the next  field to edit.
    -CTRL-V
     to paste in the copied text.
    - You can do multiple CTRL-V
     operations for the text until you do another CTRL-C. 
    • Use the board descriptions listed in Table 8-10 recorded prior to replacing a board.
    • The Back End Processor data is edited in the Backplane text file.
    • The file defaults values may have ?????? in a field, however this must be changed to alpha or 
    alpha-numberic characters (whichever is valid for the field being updated). 
    • Some boards may already have data present (no ??????). If this is the case, simply verify the data 
    present with that recorded in the table and correct it if necessary.
    • The EEP-PC VPD Data.txt file is for the PC2IP Board which is part of the Back End Processor FRU. 
    This file does not need to be updated.
    NOTE:Figure 8-148 on page 8-142  and Figure 8-150 on page 8-143  provide an example of a circuit 
    board text file and backplane text file, with hints, for reference. Proceed to page 145 for further 
    text processing. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-142 Section 8-35 - Verify and Update Vital Product Data
    Figure 8-148   Example of a Circuit Board Text File with Hints
    Figure 8-149  EQ Board Text File
    Enter or verify data from actual circuit boards
    Hints for data from actual circuit boards 
    						