Home > GE > Ultrasound System > GE Logiq 9 Service Manual

GE Logiq 9 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 570
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 1 Introduction 1-3
    Section 1-2
    Important Conventions
    1-2-1 Conventions Used in Book
    Icons
    Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and 
    conventions used on the product and in the service information are described in this chapter.
    Safety Precaution Messages
    Various levels of safety precaution messages may be found on the equipment and in the service 
    information. The different levels of concern are identified by a flag word that precedes the precautionary 
    message. Known or potential hazards are labeled in one of following ways:
    NOTE: Notes provide important information about an item or a procedure. 
    Information contained in a NOTE can often save you time or effort.
    DANGERDANGER IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL 
    CAUSE SEVERE PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE 
    IGNORED.
    WARNINGWARNINGWARNING IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT CAN CAUSE 
    SEVERE PERSONAL INJURY AND PROPERTY DAMAGE IF INSTRUCTIONS ARE 
    IGNORED.
    CAUTIONCaution is used to indicate the presence of a hazard that will or can cause minor personal injury 
    and property damage if instructions are ignored.
    NOTICEEquipment Damage Possible
    Notice is used when a hazard is present that can cause property damage but has absolutely no 
    personal injury risk. 
    Example: Disk drive will crash. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    1-4 Section 1-2 - Important Conventions
    1-2-2 Standard Hazard Icons
    Important information will always be preceded by the exclamation point contained 
    within a triangle, as seen throughout this chapter. In addition to text, several 
    different graphical icons (symbols) may be used to make you aware of specific types 
    of hazards that could cause harm.
    Other hazard icons make you aware of specific procedures that should be followed. Table 1-3    Standard Hazard Icons
    ELECTRICALMECHANICALRADIATION
    LASERHEATPINCH
    Table 1-4    Standard Icons Indicating a Special Procedure Be Used
    AVOID STATIC ELECTRICITYTAG AND LOCK OUTWEAR EYE PROTECTION
    LASER
    LIGHT
    6LJQHGDWH
    7$ *
     
    /2&.287EYE
    PROTECTION 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 1 Introduction 1-5
    1-2-3 Product Icons
    The following table describes the purpose and location of safety labels and other important information 
    provided on the equipment.
    Table 1-5    Product Icons 
    LABEL/SYMBOLPURPOSE/MEANINGLOCATION
    Identification and Rating PlateManufacturers name and address
    Date of manufacture
    Model and serial numbers
    Electrical ratingsRear of console near power inlet
    Under monitor
    On each probe
    Device Listing/Certification LabelsLaboratory logo or labels denoting 
    conformance with industry safety standards 
    such as UL or IEC.Rear of console
    Under front of monitor
    Type/Class Label
    Used to indicate the degree of safety or protection.
    IP Code (IPX8)Indicates the degree of protection provided by 
    the enclosure per IEC 529. IPX8 indicates drip 
    proof.Footswitch
    Equipment Type BF (man in the box symbol) 
    IEC 878-02-03 indicates B Type equipment 
    having even more electrical isolation than 
    standard Type B equipment because it is 
    intended for intimate patient contact.Probe connectors 
    PCG connector
    or Rear of Console
    Equipment Type CF (heart in the box symbol) 
    IEC 878-02-05 indicates equipment having a 
    high degree of protection suitable for direct 
    cardiac contact.ECG connector
    Surgical probes
    “CAUTION
    This unit weighs...
    Special care must be used to avoid...
    This precaution is intended to prevent injury 
    that may result if one person attempt to move 
    the unit considerable distances or on an incline 
    due to the weight of the unit.On the console where easily seen during transport
    CAUTION The equilateral triangle is usually 
    used in combination with other symbols to 
    advise or warn the user.Various 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    1-6 Section 1-2 - Important Conventions
    ATTENTION - Consult accompanying 
    documents  is intended to alert the user to 
    refer to the operator manual or other 
    instructions when complete information cannot 
    be provided on the label.Various
    CAUTION - Dangerous voltage (the lightning 
    flash with arrowhead in equilateral triangle) is 
    used to indicate electric shock hazards.Various
    Mains OFF Indicates the power off position of 
    the mains power switch.Rear of system adjacent to mains switch
    OFF/Standby Indicates the power off/
    standby position of the power switch.
    CAUTION
    This Power Switch DOES NOT ISOLATE 
    Mains Supply
    Adjacent to On-Off/Standby Switch
    Mains ON Indicates the power on position of 
    the mains power switch.
    ON Indicates the power on position of the 
    power switch.
    CAUTION
    This Power Switch DOES NOT ISOLATE 
    Mains SupplyFront Panel Switch
    Protective Earth Indicates the protective 
    earth (grounding) terminal.Internal
    Indicates an earth GROUND potentialProduct schematics
    Peripherals
    Equipotentiality Indicates the terminal to be 
    used for connecting equipotential conductors 
    when interconnecting (grounding) with other 
    equipment.Rear of console
    Table 1-5    Product Icons (Continued)
    LABEL/SYMBOLPURPOSE/MEANINGLOCATION 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 1 Introduction 1-7
    Section 1-3
    Safety Considerations
    1-3-1 Introduction
    The following safety precautions must be observed during all phases of operation, service and repair of 
    this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this 
    manual, violates safety standards of design, manufacture and intended use of the equipment.
    1-3-2 Human Safety
    Operating personnel must not remove the system covers. 
    Servicing should be performed by authorized personnel only. 
    Only personnel who have participated in a LOGIQ™ 9 Training are authorized to service the equipment.
    1-3-3 Mechanical Safety
    WARNINGWARNINGWHEN THE UNIT IS RAISED FOR A REPAIR OR MOVED ALONG ANY INCLINE, USE 
    EXTREME CAUTION SINCE IT MAY BECOME UNSTABLE AND TIP OVER.
    WARNINGWARNINGULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT 
    CAN EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN 
    HANDLING AND PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A 
    DAMAGED OR DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS 
    CAN RESULT IN SERIOUS INJURY AND EQUIPMENT DAMAGE.
    WARNINGWARNINGNEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR. EVEN IF IT LOOKS OK, 
    IT MAY BE DAMAGED.
    CAUTIONAlways lower and center the Opertaor I/O Panel before moving the scanner.
    CAUTIONThe LOGIQ™ 9 weighs 225 kg or more, depending on installed peripherals, (500 lbs., or more) 
    when ready for use. Care must be used when moving it or replacing its parts. Failure to follow 
    the precautions listed could result in injury, uncontrolled motion and costly damage.
    ALWAYS:
    Be sure the pathway is clear.
    Use slow, careful motions.
    Use two people when moving on inclines or lifting more than 16 kg (35 lbs). 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    1-8 Section 1-3 - Safety Considerations
    1-3-3Mechanical Safety (cont’d)
    NOTE: Special care should be taken when transporting the unit in a vehicle:
    • Secure the unit in an upright position.
    • Lock the wheels (brake)
    • DO NOT use the Control Panel as an anchor point.
    • Place the probes in their carrying case.
    • Eject any Magneto Optical disk or CD from their drive. 
    1-3-4 Electrical Safety
    To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The 
    system is equipped with a three-conductor AC power cable. This must be plugged into an approved 
    electrical outlet with safety ground. If an extension cord is used with the system, make sure that the total 
    current rating of the system does not exceed the extension cord rating. 
    The power outlet used for this equipment should not be shared with other types of equipment.
    Both the system power cable and the power connector meet international electrical standards.
    1-3-5 Labels Locations
    Refer to the LOGIQ™ 9 Basic User Manual, Chapter 2, for the label locations.
    CAUTIONKeep the heat venting holes on the monitor unobstructed to avoid overheating of the 
    monitor. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 1 Introduction 1-9
    1-3-6 Dangerous Procedure Warnings
    Warnings, such as the examples below, precede potentially dangerous procedures throughout this 
    manual. Instructions contained in the warnings must be followed.
    1-3-7 Lockout/Tagout Requirements (For USA Only)
    Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the electrical Mains 
    plug.
    1-3-8 Returning/Shipping Probes and Repair Parts
    Equipment being returned must be clean and free of blood and other infectious substances.
    GEMS policy states that body fluids must be properly removed from any part or equipment prior to 
    shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment 
    have been properly decontaminated prior to shipment. Under no circumstance should a part or 
    equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or 
    an ultrasound probe).
    The purpose of the regulation is to protect employees in the transportation industry, as well as the 
    people who will receive or open this package.
    NOTE: The US Department of Transportation (DOT) has ruled that “items that were saturated and/or 
    dripping with human blood that are now caked with dried blood; or which were used or intended 
    for use in patient care” are “regulated medical waste” for transportation purposes and must be 
    transported as a hazardous material.
    DANGERDANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT 
    IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING 
    AND ADJUSTING.
    WARNINGWARNINGEXPLOSION WARNING
    DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE. 
    OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT 
    CONSTITUTES A DEFINITE SAFETY HAZARD.
    WARNINGWARNINGDO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT
    BECAUSE OF THE DANGER OF INTRODUCING ADDITIONAL HAZARDS, DO NOT 
    INSTALL SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION 
    OF THE EQUIPMENT. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    1-10 Section 1-4 - EMC, EMI, and ESD
    Section 1-4
    EMC, EMI, and ESD
    1-4-1 Electromagnetic Compatibility (EMC)
    Electromagnetic compatibility describes a level of performance of a device within its electromagnetic 
    environment. This environment consists of the device itself and its surroundings including other 
    equipment, power sources and persons with which the device must interface. Inadequate compatibility 
    results when a susceptible device fails to perform as intended due interference from its environment or 
    when the device produces unacceptable levels of emission to its environment. This interference is often 
    referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through 
    space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy, 
    EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and 
    disturbances in the electrical power supply.
    1-4-2 CE Compliance
    The LOGIQ™ 9 unit conforms to all applicable conducted and radiated emission limits and to immunity 
    from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient 
    requirements.
    For applicable standards refer to the Safety Chapter in the Basic User Manual.
    NOTE: For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps, are in 
    good condition, installed tightly without skew or stress. Proper installation following all 
    comments noted in this service manual is required in order to achieve full EMC performance.
    1-4-3 Electrostatic Discharge (ESD) Prevention
    WARNINGWARNINGDO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING 
    THE NECESSARY ESD PRECAUTIONS: 
    1.ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED 
    ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE 
    RIGHT OF THE POWER CONNECTOR). 
    2.FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC 
    SENSITIVE EQUIPMENT. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 1 Introduction 1-11
    Section 1-5
    Customer Assistance
    1-5-1 Contact Information
    If this equipment does not work as indicated in this service manual or in the User Manual, or if you 
    require additional assistance, please contact the local distributor or appropriate support resource, as 
    listed below.
    Prepare the following information before you call:
    - System ID serial number.
    - Software version.
    Table 1-6    Phone Numbers for Customer Assistance
    LocationPhone Number
    USA/ Canada
    GE Medical Systems
    Ultrasound Service Engineering
    4855 W. Electric Avenue
    Milwaukee, WI 53219
    Customer Answer CenterService
    1-800–437–1171
    Applications
    1-800-682-5327
    1-262-524-5698
    Fax: +1-414-647-4125
    Latin America
    GE Medical Systems
    Ultrasound Service Engineering
    4855 W. Electric Avenue
    Milwaukee, WI 53219
    Customer Answer CenterService
    1-262-524-5300
    Applications
    1-262-524-5698
    Fax: +1-414-647-4125
    Europe
    GE Ultraschall Deutschland GmbH& Co. KG
    BeethovenstraBe 239
    Postfach 11 05 60, D-42665 Solingen
    GermanyTel: +49 212 2802 208
    +49 212 2802 207
    Fax: +49 212 2802 431
    Asia (Singapore/ Japan)
    GE Ultrasound Asia
    Service Department - Ultrasound
    298 Tiong Bahru Road #15-01/06
    Central Plaza
    Singapore 169730Tel: +65 291-8528
    +81 426-482950
    Fax: +65 272-3997
    +81 426-482902 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    1-12 Section 1-5 - Customer Assistance
    1-5-2 System Manufacturer
    Table 1-7    System Manufacturer
    ManufacturerFAX Number
    GE Medical Systems - Americas
    P.O. Box 414
    Milwaukee, Wisconsin 53201-0414414-647-4090 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq 9 Service Manual