Home > GE > Ultrasound System > GE Logiq 9 Service Manual

GE Logiq 9 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 570
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 10 Periodic Maintenance 10-17
    10-7-5 Chassis Leakage Current Test
    10-7-5-1 Definition
    This test measures the current that would flow in a grounded person who touched accessible metal 
    parts of the bedside station if the ground wire should break. The test verifies the isolation of the power 
    line from the chassis. The meter is connected from accessible metal parts of the case to ground. 
    Measurements should be made with the unit On and Off, with the power line polarity Normal and 
    Reversed. Record the highest reading.
    .
    10-7-5-2 Generic Procedure
    The test verifies the isolation of the power line from the chassis. The testing meter is connected from 
    accessible metal parts of the case to ground. Measurements should be made with the unit ON and OFF, 
    with the power line polarity Normal and Reversed. Record the highest reading of current.
    When using the Microguard or a similar test instrument, its power plug may be inserted into the wall 
    outlet and the equipment under test is plugged into the receptacle on the panel of the meter. This places 
    the meter in the grounding conductor and the current flowing from the case to ground will be indicated 
    in any of the current ranges. The maximum allowable limit for chassis source leakage is shown in 
    Table 10-14.
    CAUTIONElectric Shock Hazard. When the meters ground switch is OPEN, dont touch the unit!
    CAUTIONEquipment damage possibility. Never switch the Polarity and the status of Neutral when the unit 
    is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY 
    switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged.
    Figure 10-5   Set Up for Chassis Source Leakage Current, 
    IEC 601-1 Clause 19 - Continuos Leakage Currents and 
    Patient, Auxiliary Currents 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    10-18 Section 10-7 - Electrical Safety Tests
    10-7-5-3 Dale 600 Meter Procedure
    When measuring system chassis currents with the Dale 600, always use the CHASSIS selection of the 
    external/chassis function switch. This requires the ground clip lead and changing the meters switches 
    in accordance with the IEC 601-1.1. Refer to the Dale 600 Instruction Manual for meter self-test and 
    operation. Record the highest leakage current measured.
    Follow these steps to test the unit for leakage current.
    1.) Turn the LOGIQ™ 9 unit OFF.
    2.) Plug the unit into the meter, and the meter into the tested AC wall outlet
    3.) Plug the black chassis cable into the meters “CHASSIS” connector and attach the black chassis 
    cable clamp to an exposed metal part of the LOGIQ™ 9.
    4.) Set the testers “FUNCTION” switch to CHASSIS position.
    5.) Follow the test conditions described for respective test points shown in Table 10-17. 
    6.) Keep a record of the results with other hard copies of PM data kept on site.Figure 10-6   Ground and Chassis Leakage Current Test
    Table 10-17 Chassis Leakage Current Test Condition
    TESTCONDITION
    1
    Mounting screw for probe receptacle
    2
    caster support
    3
    Mounting screw for CRT housing
    4Mounting screw for peripheral plugged into unit
    5Mounting screw for other peripheral powered by unit
    Chassis
    Cable
    Clamp Normal
    Reverse
    Polarity
    Neutral
    Open/Closed Functions
    Selector Outlet Test
    Lamps 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 10 Periodic Maintenance 10-19
    10-7-5-4 Data Sheet for Chassis Source Leakage Current
    The test passes when all readings measure less than the value shown in Table 10-14. Record all data 
    on the PM Inspection Certificate.
    10-7-6Isolated Patient Lead (Source) Leakage–Lead to Ground
    10-7-6-1 Definition
    This test measures the current which would flow to ground from any of the isolated ECG leads. The 
    meter simulates a patient who is connected to the monitoring equipment and is grounded by touching 
    some other grounded surface. Measurements should be made with the ground open and closed, with 
    power line polarity normal and reversed, and with the ultrasound console Off and On. For each 
    combination the operating controls, such as the lead switch, should be operated to find the worst case 
    condition. Table 10-18 Typical Data Sheet for Chassis Source Leakage Current
    Unit Power
    Tester Polarity 
    Switch
    Tester Neutral 
    or Ground 
    SwitchTest 1 
    Probe 
    Connector
    Test 2 
    casterTest 3 
    CRTOptional 
    Test 4 Optional 
    Test 5
    Enter Name of tested peripheral here:
    ON NORM OPEN
    ON NORM CLOSED
    ONREVOPEN
    ONREVCLOSED
    OFFNORMOPEN
    OFF NORM CLOSED
    OFF REV OPEN
    OFF REV CLOSED
    CAUTIONEquipment damage possibility. Never switch the Polarity when the unit is powered ON. Be sure 
    to turn the unit power OFF before switching the polarity using the POLARITY switch. Otherwise, 
    the unit may be damaged. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    10-20 Section 10-7 - Electrical Safety Tests
    10-7-6-2 Generic Procedure
    Measurements should be made with the ground open and closed, with power line polarity normal and 
    reversed, and with the unit Off and On. For each combination, the operating controls such as the lead 
    switch should be operated to find the worst case condition.
     
    10-7-6-3 Dale 600 Meter Procedure
    The Dale 600 provides five snap type ECG buttons for testing patient leads. Snap on all patient leads 
    to the meter and assure that the ground clip is connected to the system’s ground terminal. Select the 
    meter’s LEAD-GND function. Select and test each ECG lead positions (except “ALL”) of the LEAD 
    selector, testing each to the power condition combinations found in “PATIENT LEAD LEAKAGE” table 
    in the “PM CHECKLIST”. Record the highest leakage current measured for each Power selection.
    Follow these steps to test the ECG module for leakage current.
    1.) Turn the LOGIQ™ 9 unit OFF.
    2.) Plug the unit into the meter, and the meter into the tested AC wall outlet.
    3.) Plug the black chassis cable into the meters “CHASSIS” connector and attach the black chassis 
    cable clamp to an exposed metal part of the LOGIQ™ 9 unit.
    4.) Connect the patient leads to the corresponding snaps located at the upper front of the Dale 600/
    600E. Lead nomenclature for this test is not important.Figure 10-7   Test Circuit for Measuring Non-Isolated 
    Patient Leads
    CAUTIONLine voltage is applied to the ECG leads during this test. To avoid possible electric shock 
    hazard, the system being tested must not be touched by patients, users or anyone while the ISO 
    TEST switch is depressed. When the meters ground switch is OPEN, dont touch the unit!
    POWER
    OUTLET
    H   (Black)
    N   (White)
    G   (Green)Internal
    circuitry
    Insulating surface
    Polarity reversing switch
    (use both positions) Appliance power switch
    (use both off and on positions)
    Patient connected leads The connection is at service entrance
    or on the supply side of a separately
    derived system
    Building
    ground
    Grounding contact switch (use in both
    open and closed positions)Current meter
    H = Hot
    N = Neutral (grounded)
    G = Grounding conductor H 
    N
    G
    Patient lead selector switch (if any)
    (activated as required)
    Appliance
    Between each patient 
    lead and ground 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 10 Periodic Maintenance 10-21
    10-7-6-3Dale 600 Meter Procedure (cont’d)
    5.) Set the meters “FUNCTION” switch to LEAD TO GROUND position to measure the patient lead to 
    ground leakage current.
    6.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the 
    power condition combinations.Figure 10-8   ECG Leakage Current Test
    Table 10-19  Testing Power Conditions
    ECG PowerMeter’s Polarity SwitchMeter’s Neutral Switch
    ON NORM CLOSED
    ON NORM OPEN
    ON REVERSE CLOSED
    ONREVERSEOPEN
    OFFNORMCLOSED
    OFFNORMOPEN
    OFF REVERSE CLOSED
    OFF REVERSE OPEN
    ECG
    Patient
    Cable
    ISO
    Rocker
    Switch
    ECG
    Lead
    Selector 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    10-22 Section 10-7 - Electrical Safety Tests
    10-7-6-3Dale 600 Meter Procedure (cont’d)
    7.) Record the patient lead to ground leakage current measured on the data sheet.
    8.) Set the meters “FUNCTION” switch to LEAD TO LEAD position to measure the lead to lead leakage 
    current.
    9.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the 
    power condition combinations.
    10.) Record the lead to lead leakage current measured on the data sheet.
    11.) Set the meters “FUNCTION” switch to LEAD ISO position to measure the patient lead isolation 
    current.
    12.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the 
    power condition combinations.
    13.) Depress the rocker switch to ISO TEST and read the isolation current. To apply the voltage to the 
    lead safely, the voltage is only applied when the rocker switch is depressed to ISO TEST.
    14.) Record the patient lead isolation current measured on the data sheet.
    10-7-7 Isolated Patient Lead (Source) Leakage–Lead to Lead
    Reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function 
    selector to the LEAD-LEAD position. Select and test each of the five ECG lead positions (except ALL) 
    on the LEAD selector, testing each to the power condition combinations found in the table. Record the 
    highest leakage current measured.
    10-7-7-1 Dale 600 Patient Lead Tests
    NEUTRAL POLARITY
    1.) Closed Normal
    2.) Open Normal
    3.) Closed Reversed
    4.) Open Reversed
    10-7-8 Isolated Patient Lead (Sink) Leakage-Isolation Test
    Reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function 
    selector to the LEAD-ISO. Select the ALL position on the lead selector. Depress the rocker switch to 
    ISO TEST to test lead isolation.
    NOTE: It is not necessary to test each lead individually or power condition combinations as required in 
    previous tests.
    CAUTIONLine voltage is applied to the ECG leads during this test. To avoid possible electric shock 
    hazard, the system being tested must not be touched by patients, users or anyone while the ISO 
    TEST switch is depressed. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 10 Periodic Maintenance 10-23
    10-7-8-1 Data Sheet for ECG Leakage Current
    The test passes when all readings measure less than the value shown in the table below. Record all 
    data on the PM Inspection Certificate.
    Table 10-20 Maximum Allowance Limit for ECG Leakage Current
    AC Power 
    Source
    Maximum
    Allowance Limit
    GROUND OPEN
    GROUND 
    CLOSED
    Patient Lead to Ground Leakage Current Test
    and
    Patient Lead to Lead Leakage Current Test115V 10uA 10uA
    220/240V 500uA 10uA
    Table 10-21   Maximum Allowance Limit for ECG Leakage Current
    AC Power 
    SourceMaximum
    Allowance Limit
    Patient Lead Isolation Current Test115V 20uA
    220/240V 5mA
    Table 10-22 Typical Data Sheet for ECG Leakage Current
    ECG 
    Power
    Tester 
    Polarity 
    SwitchTester 
    Ground 
    SwitchTester Lead Selector
    RLRALALLC
    ON NORM CLOSED
    ON REVERSE CLOSED
    ON NORM OPEN
    ONREVERSEOPEN
    OFFNORMCLOSED
    OFFREVERSECLOSED
    OFF NORM OPEN
    OFF REVERSE OPEN 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    10-24 Section 10-7 - Electrical Safety Tests
    10-7-9 Probe Leakage Current Test
    10-7-9-1 Definition
    This test measures the current that would flow to ground from any of the probes through a patient who 
    is being scanned and becomes grounded by touching some other grounded surface.
    10-7-9-2 Generic Procedure
    Measurements should be made with the ground open and closed, with power line polarity normal and 
    reversed, and with the unit Off and On. For each combination, the probe must be active to find the worst 
    case condition.
    NOTE: Each probe will have some amount of leakage current, dependent on its design. Small variations in 
    probe leakage currents are normal from probe to probe. Other variations will result from differences in 
    line voltage and test lead placement.
    10-7-9-3 Meter Procedure Using Probe Adapter
    The Dale 600/600E provides a method for testing probes independently from the system. The meter 
    utilizes a probe adapter to apply a test potential commonly to all connector pins.
    The probes imaging area is immersed in a saline solution along with a grounding probe from the 
    meter to complete the current path. Saline solution is a mixture of water and salt. The salt adds a free 
    ion to the water, making it conductive. Normal saline solution is 0.9% salt or 1/2 gram salt per 1 liter of 
    water. If saline is not available, a mixture of 1 quart water with one or more grams of table salt, mixed 
    thoroughly, will substitute.
    Follow these steps to test each transducer for leakage current.
    1.) Turn the LOGIQ™ 9 unit OFF.
    2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet.
    3.) Connect the probe for test with the meters appropriate adapter.
    4.) Plug the saline probe into the meters “CHASSIS” connector.
    5.) Plug the probe adapter into the meters connector marked “EXTERNAL”.Figure 10-9   Set Up for Probe Leakage Current
    CAUTIONTo avoid probe damage and possible electric shock, do not immerse probes into any liquid 
    beyond the level indicated in the probe users manual. Do not touch the probe, conductive liquid 
    or any part of the unit under test while the ISO TEST switch is depressed.
    POWER
    OUTLETH   (BLACK)POLARITY REVERSING SWITCH
    MOMENTARY
    SWITCH N   (WHITE)
    G   (GREEN)LEAKAGE TEST
    METER
    CONSOLE
    PROBE 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 10 Periodic Maintenance 10-25
    10-7-9-3Meter Procedure Using Probe Adapter (cont’d)
    6.) Set the meters “FUNCTION” switch to EXTERNAL position.
    7.) Add the saline probe and the imaging area of the probe into the saline bath.
    8.) Have unit power ON for the first part; turn it OFF for the second half.
    9.) Depress the ISO TEST rocker switch and record the highest current reading.
    10.) Follow the test conditions described in Table 10-23 for every transducer.
    11.) Keep a record of the results with other hand copies of PM data.Figure 10-10   Transducer Source Leakage Current Test 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    10-26 Section 10-7 - Electrical Safety Tests
    10-7-9-4 No Meter Probe Adapter Procedure
    Follow these steps to test each transducer for leakage current.
    1.) Turn the LOGIQ™ 9 unit OFF.
    2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet.
    3.) Plug the external probe into the meters (Dale 600) “EXTERNAL” connector.
    4.) Set the meters “FUNCTION” switch to EXTERNAL position.
    5.) Connect the probe for test with the connector of the console.
    6.) Add the saline probe and the imaging area of the probe into the saline bath.
    7.) Have unit power ON for the first part; turn it OFF for the second half.
    8.) Depress the ISO TEST rocker switch and record the highest current reading.
    9.) Follow the test conditions described in Table 10-23 for every transducer.
    10.) Keep a record of the results with other hand copies of PM data.Figure 10-11   Check Without Probe Adapter 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq 9 Service Manual