GE Logiq 9 Service Manual
Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-17 10-7-5 Chassis Leakage Current Test 10-7-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis. The meter is connected from accessible metal parts of the case to ground. Measurements should be made with the unit On and Off, with the power line polarity Normal and Reversed. Record the highest reading. . 10-7-5-2 Generic Procedure The test verifies the isolation of the power line from the chassis. The testing meter is connected from accessible metal parts of the case to ground. Measurements should be made with the unit ON and OFF, with the power line polarity Normal and Reversed. Record the highest reading of current. When using the Microguard or a similar test instrument, its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter. This places the meter in the grounding conductor and the current flowing from the case to ground will be indicated in any of the current ranges. The maximum allowable limit for chassis source leakage is shown in Table 10-14. CAUTIONElectric Shock Hazard. When the meters ground switch is OPEN, dont touch the unit! CAUTIONEquipment damage possibility. Never switch the Polarity and the status of Neutral when the unit is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged. Figure 10-5 Set Up for Chassis Source Leakage Current, IEC 601-1 Clause 19 - Continuos Leakage Currents and Patient, Auxiliary Currents
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-18 Section 10-7 - Electrical Safety Tests 10-7-5-3 Dale 600 Meter Procedure When measuring system chassis currents with the Dale 600, always use the CHASSIS selection of the external/chassis function switch. This requires the ground clip lead and changing the meters switches in accordance with the IEC 601-1.1. Refer to the Dale 600 Instruction Manual for meter self-test and operation. Record the highest leakage current measured. Follow these steps to test the unit for leakage current. 1.) Turn the LOGIQ™ 9 unit OFF. 2.) Plug the unit into the meter, and the meter into the tested AC wall outlet 3.) Plug the black chassis cable into the meters “CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the LOGIQ™ 9. 4.) Set the testers “FUNCTION” switch to CHASSIS position. 5.) Follow the test conditions described for respective test points shown in Table 10-17. 6.) Keep a record of the results with other hard copies of PM data kept on site.Figure 10-6 Ground and Chassis Leakage Current Test Table 10-17 Chassis Leakage Current Test Condition TESTCONDITION 1 Mounting screw for probe receptacle 2 caster support 3 Mounting screw for CRT housing 4Mounting screw for peripheral plugged into unit 5Mounting screw for other peripheral powered by unit Chassis Cable Clamp Normal Reverse Polarity Neutral Open/Closed Functions Selector Outlet Test Lamps
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-19 10-7-5-4 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-14. Record all data on the PM Inspection Certificate. 10-7-6Isolated Patient Lead (Source) Leakage–Lead to Ground 10-7-6-1 Definition This test measures the current which would flow to ground from any of the isolated ECG leads. The meter simulates a patient who is connected to the monitoring equipment and is grounded by touching some other grounded surface. Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the ultrasound console Off and On. For each combination the operating controls, such as the lead switch, should be operated to find the worst case condition. Table 10-18 Typical Data Sheet for Chassis Source Leakage Current Unit Power Tester Polarity Switch Tester Neutral or Ground SwitchTest 1 Probe Connector Test 2 casterTest 3 CRTOptional Test 4 Optional Test 5 Enter Name of tested peripheral here: ON NORM OPEN ON NORM CLOSED ONREVOPEN ONREVCLOSED OFFNORMOPEN OFF NORM CLOSED OFF REV OPEN OFF REV CLOSED CAUTIONEquipment damage possibility. Never switch the Polarity when the unit is powered ON. Be sure to turn the unit power OFF before switching the polarity using the POLARITY switch. Otherwise, the unit may be damaged.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-20 Section 10-7 - Electrical Safety Tests 10-7-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition. 10-7-6-3 Dale 600 Meter Procedure The Dale 600 provides five snap type ECG buttons for testing patient leads. Snap on all patient leads to the meter and assure that the ground clip is connected to the system’s ground terminal. Select the meter’s LEAD-GND function. Select and test each ECG lead positions (except “ALL”) of the LEAD selector, testing each to the power condition combinations found in “PATIENT LEAD LEAKAGE” table in the “PM CHECKLIST”. Record the highest leakage current measured for each Power selection. Follow these steps to test the ECG module for leakage current. 1.) Turn the LOGIQ™ 9 unit OFF. 2.) Plug the unit into the meter, and the meter into the tested AC wall outlet. 3.) Plug the black chassis cable into the meters “CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the LOGIQ™ 9 unit. 4.) Connect the patient leads to the corresponding snaps located at the upper front of the Dale 600/ 600E. Lead nomenclature for this test is not important.Figure 10-7 Test Circuit for Measuring Non-Isolated Patient Leads CAUTIONLine voltage is applied to the ECG leads during this test. To avoid possible electric shock hazard, the system being tested must not be touched by patients, users or anyone while the ISO TEST switch is depressed. When the meters ground switch is OPEN, dont touch the unit! POWER OUTLET H (Black) N (White) G (Green)Internal circuitry Insulating surface Polarity reversing switch (use both positions) Appliance power switch (use both off and on positions) Patient connected leads The connection is at service entrance or on the supply side of a separately derived system Building ground Grounding contact switch (use in both open and closed positions)Current meter H = Hot N = Neutral (grounded) G = Grounding conductor H N G Patient lead selector switch (if any) (activated as required) Appliance Between each patient lead and ground
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-21 10-7-6-3Dale 600 Meter Procedure (cont’d) 5.) Set the meters “FUNCTION” switch to LEAD TO GROUND position to measure the patient lead to ground leakage current. 6.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the power condition combinations.Figure 10-8 ECG Leakage Current Test Table 10-19 Testing Power Conditions ECG PowerMeter’s Polarity SwitchMeter’s Neutral Switch ON NORM CLOSED ON NORM OPEN ON REVERSE CLOSED ONREVERSEOPEN OFFNORMCLOSED OFFNORMOPEN OFF REVERSE CLOSED OFF REVERSE OPEN ECG Patient Cable ISO Rocker Switch ECG Lead Selector
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-22 Section 10-7 - Electrical Safety Tests 10-7-6-3Dale 600 Meter Procedure (cont’d) 7.) Record the patient lead to ground leakage current measured on the data sheet. 8.) Set the meters “FUNCTION” switch to LEAD TO LEAD position to measure the lead to lead leakage current. 9.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the power condition combinations. 10.) Record the lead to lead leakage current measured on the data sheet. 11.) Set the meters “FUNCTION” switch to LEAD ISO position to measure the patient lead isolation current. 12.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the power condition combinations. 13.) Depress the rocker switch to ISO TEST and read the isolation current. To apply the voltage to the lead safely, the voltage is only applied when the rocker switch is depressed to ISO TEST. 14.) Record the patient lead isolation current measured on the data sheet. 10-7-7 Isolated Patient Lead (Source) Leakage–Lead to Lead Reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function selector to the LEAD-LEAD position. Select and test each of the five ECG lead positions (except ALL) on the LEAD selector, testing each to the power condition combinations found in the table. Record the highest leakage current measured. 10-7-7-1 Dale 600 Patient Lead Tests NEUTRAL POLARITY 1.) Closed Normal 2.) Open Normal 3.) Closed Reversed 4.) Open Reversed 10-7-8 Isolated Patient Lead (Sink) Leakage-Isolation Test Reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function selector to the LEAD-ISO. Select the ALL position on the lead selector. Depress the rocker switch to ISO TEST to test lead isolation. NOTE: It is not necessary to test each lead individually or power condition combinations as required in previous tests. CAUTIONLine voltage is applied to the ECG leads during this test. To avoid possible electric shock hazard, the system being tested must not be touched by patients, users or anyone while the ISO TEST switch is depressed.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-23 10-7-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate. Table 10-20 Maximum Allowance Limit for ECG Leakage Current AC Power Source Maximum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Leakage Current Test and Patient Lead to Lead Leakage Current Test115V 10uA 10uA 220/240V 500uA 10uA Table 10-21 Maximum Allowance Limit for ECG Leakage Current AC Power SourceMaximum Allowance Limit Patient Lead Isolation Current Test115V 20uA 220/240V 5mA Table 10-22 Typical Data Sheet for ECG Leakage Current ECG Power Tester Polarity SwitchTester Ground SwitchTester Lead Selector RLRALALLC ON NORM CLOSED ON REVERSE CLOSED ON NORM OPEN ONREVERSEOPEN OFFNORMCLOSED OFFREVERSECLOSED OFF NORM OPEN OFF REVERSE OPEN
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-24 Section 10-7 - Electrical Safety Tests 10-7-9 Probe Leakage Current Test 10-7-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface. 10-7-9-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the probe must be active to find the worst case condition. NOTE: Each probe will have some amount of leakage current, dependent on its design. Small variations in probe leakage currents are normal from probe to probe. Other variations will result from differences in line voltage and test lead placement. 10-7-9-3 Meter Procedure Using Probe Adapter The Dale 600/600E provides a method for testing probes independently from the system. The meter utilizes a probe adapter to apply a test potential commonly to all connector pins. The probes imaging area is immersed in a saline solution along with a grounding probe from the meter to complete the current path. Saline solution is a mixture of water and salt. The salt adds a free ion to the water, making it conductive. Normal saline solution is 0.9% salt or 1/2 gram salt per 1 liter of water. If saline is not available, a mixture of 1 quart water with one or more grams of table salt, mixed thoroughly, will substitute. Follow these steps to test each transducer for leakage current. 1.) Turn the LOGIQ™ 9 unit OFF. 2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet. 3.) Connect the probe for test with the meters appropriate adapter. 4.) Plug the saline probe into the meters “CHASSIS” connector. 5.) Plug the probe adapter into the meters connector marked “EXTERNAL”.Figure 10-9 Set Up for Probe Leakage Current CAUTIONTo avoid probe damage and possible electric shock, do not immerse probes into any liquid beyond the level indicated in the probe users manual. Do not touch the probe, conductive liquid or any part of the unit under test while the ISO TEST switch is depressed. POWER OUTLETH (BLACK)POLARITY REVERSING SWITCH MOMENTARY SWITCH N (WHITE) G (GREEN)LEAKAGE TEST METER CONSOLE PROBE
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-25 10-7-9-3Meter Procedure Using Probe Adapter (cont’d) 6.) Set the meters “FUNCTION” switch to EXTERNAL position. 7.) Add the saline probe and the imaging area of the probe into the saline bath. 8.) Have unit power ON for the first part; turn it OFF for the second half. 9.) Depress the ISO TEST rocker switch and record the highest current reading. 10.) Follow the test conditions described in Table 10-23 for every transducer. 11.) Keep a record of the results with other hand copies of PM data.Figure 10-10 Transducer Source Leakage Current Test
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-26 Section 10-7 - Electrical Safety Tests 10-7-9-4 No Meter Probe Adapter Procedure Follow these steps to test each transducer for leakage current. 1.) Turn the LOGIQ™ 9 unit OFF. 2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet. 3.) Plug the external probe into the meters (Dale 600) “EXTERNAL” connector. 4.) Set the meters “FUNCTION” switch to EXTERNAL position. 5.) Connect the probe for test with the connector of the console. 6.) Add the saline probe and the imaging area of the probe into the saline bath. 7.) Have unit power ON for the first part; turn it OFF for the second half. 8.) Depress the ISO TEST rocker switch and record the highest current reading. 9.) Follow the test conditions described in Table 10-23 for every transducer. 10.) Keep a record of the results with other hand copies of PM data.Figure 10-11 Check Without Probe Adapter