GE Logiq 9 Service Manual
Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-7 Section 10-5 System Periodic Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-6 System Preliminary Checks StepItemDescription 1 Ask & Listen Ask the customer if they have any problems or questions about the equipment. 2 PaperworkFill in the top of the Periodic Maintenance (PM) Inspection Certificate. Note all probes and system options. 3 Power upTurn the system power on and verify that all fans and peripherals turn on. Watch the displays during power up to verify that no warning or error messages are displayed. 4ProbesVerify that the system properly recognizes all probes. 5DisplaysVerify proper display on the monitor and touch panel. 6PresetsBackup all customer presets on an CD-RW.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-8 Section 10-5 - System Periodic Maintenance 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks NOTE: * Some software may be considered standard depending upon system model configuration.Table 10-7 System Functional Checks Chec k StepDescription B-ModeVerify basic B-Mode (2D) operation. Check the basic system controls that affect this mode of operation. CF-ModeVerify basic CF-Mode (Color Flow Mode) operation. Check the basic system controls that affect this mode of operation. Doppler ModesVerify basic Doppler operation (PW and CW if available). Check the basic system controls that affect this mode of operation. M-ModeVerify basic M-Mode operation. Check the basic system controls that affect this mode of operation. *Applicable Software OptionsVerify the basic operation of all optional modes such as Multi-Image, 3D, Contrast, Harmonics, Cine, Stress Echo,... etc. Check the basic system controls that affect each options operation. Xmit/Recv ElementsUse the Visual Channel Utilityon the 10L probe to verify that all system xmit/recv channels are functional. Keyboard Test Perform the Keyboard Test Procedure to verify that all keyboard controls are OK. Monitor Verify basic Monitor display functions. Refer to Chapter 3 of the User Manual. Touch Panel Verify basic Touch Panel display functions. Refer to Chapter 3 of the User Manual. MeasurementsScan a gray scale phantom and use the measurement controls to verify distance and area calculation accuracy. Refer to the User Manual, Chapter 18, for measurement accuracy specifications.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-9 10-5-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. 10-5-3 Input Power 10-5-3-1 Mains Cable Inspection 10-5-4 Cleaning 10-5-4-1 General Cleaning Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks StepItemDescription 1VCR Verify record/playback capabilities of the VCR. Clean heads and covers if necessary. 2 B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary. 3 Color Printer Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary. 4DICOMVerify that DICOM is functioning properly. Send an image to a DICOM device. 5InSite/iLinqVerify that InSite is functioning properly. Ensure two-way remote communications. (Warranty & Contract Customers only) 6CameraVerify hardcopy output of the film camera. Clean as necessary. 7 Footswitch Verify that the footswitch is functioning as programed. Clean as necessary. 8ECG Verify basic operation with customer Table 10-9 Mains Cable Inspection StepItemDescription 1 Unplug Cord Disconnect the mains cable from the wall and system. 2 Inspect Inspect it and its connectors for damage of any kind. 3VerifyVerify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no strands may cause a short circuit. 4VerifyInlet connector retainer is functional. Table 10-10 General Cleaning StepItemDescription 1 ConsoleUse a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful not to get the cloth too wet so that moisture does not enter the console. 2 Probe Holder Clean probe holders (they may need to be soaked to remove excess gel). 3Monitor and Touch Panelwaiting for procedure from engineering
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-10 Section 10-5 - System Periodic Maintenance 10-5-4-2 Air Filter Cleaning NOTE: For your convenience or of the air filter is too dirty, replacement filters are available. Refer to Chapter 9 for the air filter replacement part number. 10-5-5 Physical Inspection Table 10-11 Air Filter Cleaning - frequency varies with your environment StepItemDescription 1Remove Filter CoverRefer to Chapter 8 for air filter location and removal instructions. 2 Clean FilterThe filters can be cleaned in sprinkling water, or they can be dusted with a vacuum cleaner. If the filter is metal wash and/or vacuum. If the filter is fiber or plastic vacuum or replace. 3 Install Filter Install the clean filter. Table 10-12 Physical Checks StepItemDescription 1 LabelingVerify that all system labeling is present and in readable condition. Refer to the User Manual,..... for details. 2 Scratches & Dents Inspect the console for dents, scratches or cracks. 3 Control Panel Inspect keyboard and control panel. Note any damaged or missing items. 4Control Panel MovementVerify ease of control panel (Operator I/O Panel) movement in all acceptable directions. Ensure that it latches in position as required. 5Wheels & BrakesCheck all wheels and casters for wear and verify operation of foot brake, to stop the unit from moving, and release mechanism. Check all caster locks and caster swivel locks for proper operation. 6Cables & ConnectorsCheck all internal cable harnesses and connectors for wear and secure connector seating. Pay special attention to footswitch assembly and probe strain or bend reliefs. 7Shielding & CoversCheck to ensure that all EMI shielding, internal covers, air flow panels and screws are in place. Missing covers and hardware could cause EMI/RFI problems while scanning. 8 External I/O Check all connectors for damage and verify that the labeling is good. 9 Op Panel Lights Check for proper operation of all operator panel and TGC lights. 10Monitor LightCheck for proper operation of any monitor lights if available. 11External MicrophoneCheck for proper operation of any external microphones by recording an audio test.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-11 10-5-6 Optional Diagnostic Checks To complete the PM checks, access the diagnostic software as described in Chapters 5 or 7. View the error logs and run desired diagnostics. 10-5-6-1 View the Logs 1.) Review the system error log for any problems. 2.) Check the temperature log to see if there are any trends that could cause problems in the future. 3.) Check the Configuration Log; update if needed. 10-5-7 Probe Maintenance 10-5-7-1 Probe Related Checks 10-5-7-2 Basic Probe Care The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details. Failure to follow these precautions can result in serious injury and equipment damage. Failure to properly handle or maintain a probe may also void its warranty. Any evidence of wear indicates the probe cannot be used. Do a visual check of the probe pins and system sockets before plugging in a probe. TEE and Interoperative probes often have special considerations and individual probe user manuals. For TEE and Interoperative probes also refer to their separate user manuals. 10-5-7-3 Basic Probe Cleaning Refer to the User’s Manual for details on probe cleaning. NOTE: To help protect yourself from blood borne diseases, wear approved disposable gloves. These are made of nitrile derived from vegetable starch to prevent allergic latex reactions. NOTE: Failure to follow the prescribed cleaning or disinfection procedures will void the probe’s warranty. DO NOT soak or wipe the lens with any product not listed in the User Manual. Doing so could result in irreparable damage to the probe. Follow care instructions that came with the probe. NOTE: Disinfect a defective probe before you return it. Be sure to tag the probe as being disinfected.Table 10-13 Probe Related Checks StepItemDescription 1 Probe Holder Clean probe holders (they may need to be soaked to remove excess gel). 2ProbesThoroughly check the system probe connectors and remove dust from inside the connector sockets if necessary. Visually check for bent, damaged or missing pins
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-12 Section 10-6 - Using a Phantom Section 10-6 Using a Phantom See the Basic User Manual “Customer Maintenance” for information on using a phantom and quality assurance tests. Section 10-7 Electrical Safety Tests 10-7-1 Safety Test Overview The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety evaluation of cord-connected, electrically operated, patient care equipment. If additional information is needed, refer to the NFPA 99 (For USA) and IEC 60601-1 documents. Test the system, peripherals and probes for leakage current. Excessive leakage current can cause injury or death in sensitive patients. High leakage current can also indicate degradation of insulation and a potential for electrical failure. Do not use probes or equipment having excessive leakage current. To minimize the risk that a probe may shock someone the customer should: • Not use a probe that is cracked or damaged in any way .i.Probe Leakage Current, Frequency; • Check probe leakage current: * Based on your facilities QA program for surface probes * Based on your facilities QA program for endocavitary probes * whenever probe damage is suspected WARNINGWARNINGTHE USER MUST ENSURE THAT THE SAFETY INSPECTIONS ARE PERFORMED AT LEAST EVERY 12 MONTHS ACCORDING TO THE REQUIREMENTS OF THE PATIENT SAFETY STANDARD IEC-EN 60601-1. ONLY TRAINED PERSONS ARE ALLOWED TO PERFORM THE SAFETY INSPECTIONS MENTIONED ABOVE. CAUTIONTo avoid electrical shock, the unit under test must not be connected to other electrical equipment. Remove all interconnecting cables and wires. The unit under test must not be contacted by users or patients while performing these tests. CAUTIONPossible risk of infection. Do not handle soiled or contaminated probes and other components that have been in patient contact. Follow appropriate cleaning and disinfecting procedures before handling the equipment.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-13 10-7-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed. NOTE: *Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the part to determine the amount of current that will pass (or sink) to ground if a patient contacted mains voltage. The following tests are performed at the factory and should be performed at the site. These tests are: grounding continuity, chassis leakage current, probe leakage current, and ECG leakage current. All measurements are made with an electrical safety analyzer Model 600/600E built by Dale Technology Corporation or equivalent device. Table 10-14 Chassis Leakage Current Limits—Accessible Metal Surfaces CountryNormal ConditionOpen GroundReverse PolarityOpen Neutral USAN/A 0.3 mA 0.3 mA N/A Other0.1 mA 0.5 mA 0.5 mA 0.5 mA Table 10-15 Type BF Applied Part Leakage Current Limits - Non-Conductive (Floating) Surface and Cavity Probes Country Normal Condition Open GroundReverse PolarityOpen Neutral*Mains Applied USA0.05 mA 0.05 mA 0.05 mA 0.05 mA N/A Other0.1 mA 0.5 mA 0.5 mA 0.5 mA 5.0 mA Table 10-16 Type CF Applied Part Leakage Current Limits - Surgical Probes and ECG Connections Country Normal Condition Open GroundReverse PolarityOpen Neutral*Mains Applied USA0.01 mA 0.05mA 0.05 mA N/A 0.025 mA Other0.01 mA 0.05 mA 0.05 mA 0.05 mA 0.05 mA
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-14 Section 10-7 - Electrical Safety Tests 10-7-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used. The Dale 600 has self-contained lamps designed for testing the outlet wiring arrangement. Plug the Dale 600 into each outlet to be tested comparing the lamp status. NOTE: No outlet tester can detect the condition where the Neutral (grounded supply) conductor and the Grounding (protective earth) conductor are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected.Figure 10-1 Typical Outlet Tester Figure 10-2 Dale 600 Outlet Test Combination of Lights Condition CORRECT WIRINGREVERSE POLARITYOPEN GROUNDOPEN HOTHOT/GROUND REVERSED
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-15 10-7-4 Grounding Continuity Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case. The ground wire resistance should be less than 0.2 ohms. Reference the procedure in the IEC 601-1.1. 10-7-4-1 Meter Procedure Follow these steps to test the ground wire resistance. 1.) Turn the LOGIQ™ 9 unit OFF. 2.) Plug the unit into the meter, and the meter into the tested AC wall outlet. 3.) Plug the black chassis cable into the meters “CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the LOGIQ™ 9 unit. 4.) Set the meters “FUNCTION” switch to the RESISTANCE position. 5.) Set the meters “POLARITY” switch to the OFF (center) position. 6.) Measure and record the ground wire resistance. CAUTIONElectric Shock Hazard. The patient must not be contacted to the equipment during this test Figure 10-3 Ground Continuity Test
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-16 Section 10-7 - Electrical Safety Tests 10-7-4-2 Dale 600 - Ground Continuity The Dale 600 measures line cord resistance from the third pin of the attachment plug to the meter’s Chassis Cable clamp. Test the grounding continuity of the system to all exposed metal parts in accordance with the IEC 601-1.1 procedure as above. Refer to the Dale 600 Instruction Manual for meter self tests and operation. Record measured resistance of the grounding continuity. The ground wire resistance should be less than 0.2 (Use any safety analyzer.) Figure 10-4 Dale 600 Ground Continuity Test Chassis Cable Clamp Normal Reverse Polarity Neutral Open/Closed Functions Selector Outlet Test Lamps