Home > GE > Ultrasound System > GE Logiq 9 Service Manual

GE Logiq 9 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 570
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 10 Periodic Maintenance 10-7
    Section 10-5
    System Periodic Maintenance
    10-5-1 Preliminary Checks
    The preliminary checks take about 15 minutes to perform. Refer to the system user documentation 
    whenever necessary.
     
    Table 10-6    System Preliminary Checks
    StepItemDescription
    1 Ask & Listen
    Ask the customer if they have any problems or questions about the equipment.
    2 PaperworkFill in the top of the Periodic Maintenance (PM) Inspection Certificate. Note all probes and system 
    options.
    3 Power upTurn the system power on and verify that all fans and peripherals turn on. Watch the displays during 
    power up to verify that no warning or error messages are displayed.
    4ProbesVerify that the system properly recognizes all probes.
    5DisplaysVerify proper display on the monitor and touch panel.
    6PresetsBackup all customer presets on an CD-RW.  
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    10-8 Section 10-5 - System Periodic Maintenance
    10-5-2 Functional Checks (See Also Chapter 4)
    The functional checks take about 60 minutes to perform. Refer to the system user documentation 
    whenever necessary.
    10-5-2-1 System Checks
     
    NOTE: * Some software may be considered standard depending upon system model configuration.Table 10-7    System Functional Checks
    Chec
    k
    StepDescription
    B-ModeVerify basic B-Mode (2D) operation. Check the basic system controls that affect this mode 
    of operation.
    CF-ModeVerify basic CF-Mode (Color Flow Mode) operation. Check the basic system controls that 
    affect this mode of operation.
    Doppler ModesVerify basic Doppler operation (PW and CW if available). Check the basic system controls 
    that affect this mode of operation.
    M-ModeVerify basic M-Mode operation. Check the basic system controls that affect this mode of 
    operation.
    *Applicable Software 
    OptionsVerify the basic operation of all optional modes such as Multi-Image, 3D, Contrast, 
    Harmonics, Cine, Stress Echo,... etc. Check the basic system controls that affect each 
    options operation.
    Xmit/Recv ElementsUse the Visual Channel Utilityon the 10L probe to verify that all system xmit/recv channels 
    are functional.
    Keyboard Test
    Perform the Keyboard Test Procedure to verify that all keyboard controls are OK. 
    Monitor
    Verify basic Monitor display functions. Refer to Chapter 3 of the User Manual.
    Touch Panel
    Verify basic Touch Panel display functions. Refer to Chapter 3 of the User Manual.
    MeasurementsScan a gray scale phantom and use the measurement controls to verify distance and area 
    calculation accuracy. Refer to the User Manual, Chapter 18, for measurement accuracy 
    specifications. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 10 Periodic Maintenance 10-9
    10-5-2-2 Peripheral/Option Checks
    If any peripherals or options are not part of the system configuration, the check can be omitted. Refer 
    to the User Manual for a list of approved peripherals/options.
    10-5-3 Input Power
    10-5-3-1 Mains Cable Inspection
     
    10-5-4 Cleaning
    10-5-4-1 General Cleaning
     
    Table 10-8    GE Approved Peripheral/Hardware Option Functional Checks
    StepItemDescription
    1VCR
    Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
    2 B/W Printer
    Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
    3 Color Printer
    Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
    4DICOMVerify that DICOM is functioning properly. Send an image to a DICOM device.
    5InSite/iLinqVerify that InSite is functioning properly. Ensure two-way remote communications.
    (Warranty & Contract Customers only)
    6CameraVerify hardcopy output of the film camera. Clean as necessary.
    7 Footswitch
    Verify that the footswitch is functioning as programed. Clean as necessary.
    8ECG
    Verify basic operation with customer
    Table 10-9    Mains Cable Inspection
    StepItemDescription
    1 Unplug Cord
    Disconnect the mains cable from the wall and system.
    2 Inspect
    Inspect it and its connectors for damage of any kind.
    3VerifyVerify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no 
    strands may cause a short circuit.
    4VerifyInlet connector retainer is functional.
    Table 10-10   General Cleaning
    StepItemDescription
    1 ConsoleUse a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful 
    not to get the cloth too wet so that moisture does not enter the console.
    2 Probe Holder
    Clean probe holders (they may need to be soaked to remove excess gel).
    3Monitor and 
    Touch Panelwaiting for procedure from engineering 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    10-10 Section 10-5 - System Periodic Maintenance
    10-5-4-2 Air Filter Cleaning
    NOTE: For your convenience or of the air filter is too dirty, replacement filters are available. Refer to 
    Chapter 9 for the air filter replacement part number.
    10-5-5 Physical Inspection
     
    Table 10-11    Air Filter Cleaning - frequency varies with your environment
    StepItemDescription
    1Remove Filter 
    CoverRefer to Chapter 8 for air filter location and removal instructions.
    2 Clean FilterThe filters can be cleaned in sprinkling water, or they can be dusted with a vacuum cleaner. If the filter 
    is metal wash and/or vacuum. If the filter is fiber or plastic vacuum or replace.
    3 Install Filter
    Install the clean filter.
    Table 10-12   Physical Checks
    StepItemDescription
    1 LabelingVerify that all system labeling is present and in readable condition. Refer to the User Manual,..... for 
    details.
    2 Scratches & Dents
    Inspect the console for dents, scratches or cracks.
    3 Control Panel
    Inspect keyboard and control panel. Note any damaged or missing items.
    4Control Panel 
    MovementVerify ease of control panel (Operator I/O Panel) movement in all acceptable directions. Ensure that 
    it latches in position as required.
    5Wheels & BrakesCheck all wheels and casters for wear and verify operation of foot brake, to stop the unit from moving, 
    and release mechanism. Check all caster locks and caster swivel locks for proper operation.
    6Cables & 
    ConnectorsCheck all internal cable harnesses and connectors for wear and secure connector seating. Pay 
    special attention to footswitch assembly and probe strain or bend reliefs.
    7Shielding & 
    CoversCheck to ensure that all EMI shielding, internal covers, air flow panels and screws are in place. 
    Missing covers and hardware could cause EMI/RFI problems while scanning.
    8 External I/O
    Check all connectors for damage and verify that the labeling is good.
    9 Op Panel Lights
    Check for proper operation of all operator panel and TGC lights.
    10Monitor LightCheck for proper operation of any monitor lights if available.
    11External 
    MicrophoneCheck for proper operation of any external microphones by recording an audio test. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 10 Periodic Maintenance 10-11
    10-5-6 Optional Diagnostic Checks
    To complete the PM checks, access the diagnostic software as described in Chapters 5 or 7. View the 
    error logs and run desired diagnostics.
    10-5-6-1 View the Logs
    1.) Review the system error log for any problems.
    2.) Check the temperature log to see if there are any trends that could cause problems in the future.
    3.) Check the Configuration Log; update if needed.
    10-5-7 Probe Maintenance
    10-5-7-1 Probe Related Checks
     
    10-5-7-2 Basic Probe Care
    The system user manuals and various probe handling cards provide a complete description of probe 
    care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper 
    care of GE probes.
    Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care 
    cards for more details. Failure to follow these precautions can result in serious injury and equipment 
    damage. Failure to properly handle or maintain a probe may also void its warranty.
    Any evidence of wear indicates the probe cannot be used.
    Do a visual check of the probe pins and system sockets before plugging in a probe.
    TEE and Interoperative probes often have special considerations and individual probe user manuals. 
    For TEE and Interoperative probes also refer to their separate user manuals.
    10-5-7-3 Basic Probe Cleaning
    Refer to the User’s Manual for details on probe cleaning.
    NOTE: To help protect yourself from blood borne diseases, wear approved disposable gloves. These are made 
    of nitrile derived from vegetable starch to prevent allergic latex reactions.
    NOTE: Failure to follow the prescribed cleaning or disinfection procedures will void the probe’s 
    warranty. DO NOT soak or wipe the lens with any product not listed in the User Manual. Doing 
    so could result in irreparable damage to the probe. Follow care instructions that came with the 
    probe.
    NOTE: Disinfect a defective probe before you return it. Be sure to tag the probe as being disinfected.Table 10-13   Probe Related Checks
    StepItemDescription
    1 Probe Holder
    Clean probe holders (they may need to be soaked to remove excess gel).
    2ProbesThoroughly check the system probe connectors and remove dust from inside the connector sockets 
    if necessary. Visually check for bent, damaged or missing pins 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    10-12 Section 10-6 - Using a Phantom
    Section 10-6
    Using a Phantom
    See the Basic User Manual “Customer Maintenance” for information on using a phantom and quality 
    assurance tests.
    Section 10-7
    Electrical Safety Tests
    10-7-1 Safety Test Overview
    The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC 
    60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety evaluation of 
    cord-connected, electrically operated, patient care equipment. If additional information is needed, refer 
    to the NFPA 99 (For USA) and IEC 60601-1 documents.
     
    Test the system, peripherals and probes for leakage current. Excessive leakage current can cause 
    injury or death in sensitive patients. High leakage current can also indicate degradation of insulation and 
    a potential for electrical failure. Do not use probes or equipment having excessive leakage current.
    To minimize the risk that a probe may shock someone the customer should:
    • Not use a probe that is cracked or damaged in any way
    .i.Probe Leakage Current, Frequency;
    • Check probe leakage current:
    * Based on your facilities QA program for surface probes
    * Based on your facilities QA program for endocavitary probes
    * whenever probe damage is suspected
    WARNINGWARNINGTHE USER MUST ENSURE THAT THE SAFETY INSPECTIONS ARE PERFORMED AT 
    LEAST EVERY 12 MONTHS ACCORDING TO THE REQUIREMENTS OF THE PATIENT 
    SAFETY STANDARD IEC-EN 60601-1. ONLY TRAINED PERSONS ARE ALLOWED TO 
    PERFORM THE SAFETY INSPECTIONS MENTIONED ABOVE.
    CAUTIONTo avoid electrical shock, the unit under test must not be connected to other electrical 
    equipment. Remove all interconnecting cables and wires. The unit under test must not be 
    contacted by users or patients while performing these tests.
    CAUTIONPossible risk of infection. Do not handle soiled or contaminated probes and other components 
    that have been in patient contact. Follow appropriate cleaning and disinfecting procedures 
    before handling the equipment. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 10 Periodic Maintenance 10-13
    10-7-2 GEMS Leakage Current Limits
    The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment 
    Safety Standards. These limits are GEMS standards and in some cases are lower than the above 
    standards listed.
     
     
    NOTE: *Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the 
    part to determine the amount of current that will pass (or sink) to ground if a patient contacted 
    mains voltage.
    The following tests are performed at the factory and should be performed at the site. These tests are: 
    grounding continuity, chassis leakage current, probe leakage current, and ECG leakage current. All 
    measurements are made with an electrical safety analyzer Model 600/600E built by Dale Technology 
    Corporation or equivalent device. Table 10-14   Chassis Leakage Current Limits—Accessible Metal Surfaces
    CountryNormal ConditionOpen GroundReverse PolarityOpen Neutral
    USAN/A
    0.3 mA 0.3 mA N/A
    Other0.1 mA
    0.5 mA 0.5 mA 0.5 mA
    Table 10-15   Type BF Applied Part Leakage Current Limits - Non-Conductive (Floating) Surface 
    and Cavity Probes
    Country
    Normal 
    Condition
    Open GroundReverse PolarityOpen Neutral*Mains Applied
    USA0.05 mA
    0.05 mA 0.05 mA 0.05 mA N/A
    Other0.1 mA
    0.5 mA 0.5 mA 0.5 mA 5.0 mA
    Table 10-16   Type CF Applied Part Leakage Current Limits - Surgical Probes and ECG 
    Connections
    Country
    Normal 
    Condition
    Open GroundReverse PolarityOpen Neutral*Mains Applied
    USA0.01 mA
    0.05mA 0.05 mA N/A 0.025 mA
    Other0.01 mA
    0.05 mA 0.05 mA 0.05 mA 0.05 mA 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    10-14 Section 10-7 - Electrical Safety Tests
    10-7-3 Outlet Test - Wiring Arrangement - USA & Canada
    Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet 
    tester and noting the combination of lights that are illuminated. Any problems found should be reported 
    to the hospital immediately and the receptacle should not be used.
    The Dale 600 has self-contained lamps designed for testing the outlet wiring arrangement. Plug the 
    Dale 600 into each outlet to be tested comparing the lamp status.
    NOTE: No outlet tester can detect the condition where the Neutral (grounded supply) conductor and the 
    Grounding (protective earth) conductor are reversed. If later tests indicate high leakage currents, this 
    should be suspected as a possible cause and the outlet wiring should be visually inspected.Figure 10-1   Typical Outlet Tester
    Figure 10-2   Dale 600 Outlet Test
    Combination 
    of 
    Lights
    Condition
    CORRECT
    WIRINGREVERSE
    POLARITYOPEN
    GROUNDOPEN
    HOTHOT/GROUND
    REVERSED 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 10 Periodic Maintenance 10-15
    10-7-4 Grounding Continuity  
    Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case. 
    The ground wire resistance should be less than 0.2 ohms. Reference the procedure in the IEC 601-1.1.
    10-7-4-1 Meter Procedure
    Follow these steps to test the ground wire resistance.
    1.) Turn the LOGIQ™ 9 unit OFF.
    2.) Plug the unit into the meter, and the meter into the tested AC wall outlet.
    3.) Plug the black chassis cable into the meters “CHASSIS” connector and attach the black chassis 
    cable clamp to an exposed metal part of the LOGIQ™ 9 unit.
    4.) Set the meters “FUNCTION” switch to the RESISTANCE position.
    5.) Set the meters “POLARITY” switch to the OFF (center) position.
    6.) Measure and record the ground wire resistance.
    CAUTIONElectric Shock Hazard. The patient must not be contacted to the equipment during this test
    Figure 10-3   Ground Continuity Test 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    10-16 Section 10-7 - Electrical Safety Tests
    10-7-4-2 Dale 600 - Ground Continuity
    The Dale 600 measures line cord resistance from the third pin of the attachment plug to the meter’s 
    Chassis Cable clamp. Test the grounding continuity of the system to all exposed metal parts in 
    accordance with the IEC 601-1.1 procedure as above. Refer to the Dale 600 Instruction Manual for 
    meter self tests and operation. Record measured resistance of the grounding continuity. The ground 
    wire resistance should be less than 0.2 (Use any safety analyzer.)
     
    Figure 10-4   Dale 600 Ground Continuity Test
    Chassis
    Cable
    Clamp Normal
    Reverse
    Polarity
    Neutral
    Open/Closed Functions
    Selector Outlet Test
    Lamps 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq 9 Service Manual