    Chapter 9 Renewal Parts 9-9
     
    Section 9-7
    Mechanical Kits
    Table 9-6    Operator I/O Mechanical Parts
    ItemPart NamePart NumberDescriptionQtyFRU
    170
    Top Console Upper Casting Assembly2331665 Assembly for Service 1 2
    171
    Top Console Lower Casting Assembly2331858 Assembly for Service 1 2
    172
    Frogleg (complete)FB200340- Same as Vivid 7 -12
    173Frogleg Bumper KitFC200086goes on back/top half of Frogleg arms 
    - Same as Vivid 7 -11
    175Rear Handle AssemblyFB200406with cable wrap hooks - Same as Vivid 7 -11
    176Brake Lock PedalFB30740711
    177
    Directional Lock PedalFB307409Swivel Lock Pedal 1 1
    178
    Column Cover KitFC200183- Same as Vivid 7 -11
    179
    Top Console Lock AssemblyFB200350- Same as Vivid 7 -12
    180Lock ReleaseFB200920- Same as Vivid 7 -12
    181X-Y Release HandleFB307297- Same as Vivid 7 -12
    182Z-Release HandleFB307296- Same as Vivid 7 -12
    183
    Main Handle with LabelFC200131 Front Handle with Label in center 1 1
    184
    Pedal AssemblyFB200595- Same as Vivid 7 -11
    Table 9-7    Kits, Overview 
    ItemPART NAME
    PART 
    NUMBER
    DESCRIPTIONQTYFRU
    1 BUMPER KIT FROGLEGFC200086BUMPER KIT, FROGLEG 
    - Same as Vivid 7 -11
    2 COLUMN COVER KITFC200183COLUMN COVER KIT 
    - Same as Vivid 7 -11 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    9-10 Section 9-7 - Mechanical Kits
    9-7-1 Bumper Kit, Frogleg
     
    Figure 9-5   Bumper Kit, Frogleg, Base Seen From Above
    Table 9-8    Bumper Kit, Frogleg FC200086
    ITEMPART NAME
    PART 
    NUMBER
    DESCRIPTIONQTYFRU
    1BUMPER CRASH PAD - Bumper Crash Rear 1 N
    2BUMPER INNER BOSS  - - 1 N
    3BUMPER RIGHT FROGLEG  - - 1 N
    4BUMPER LEFT FROGLEG --1N
    5BUMPER INNER BOSS CRASH --2N
    1
    2
    3
    4
    5 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 9 Renewal Parts 9-11
    Section 9-8
    Operator Control Panel Brackets
    Figure 9-6   Operator Panel Brackets
    Table 9-9    Operator Panel Brackets
    ITEMPART NAME
    PART 
    NUMBER
    DESCRIPTIONQTYFRU
    190Storage Tray Bracket 2306018 1 2
    191Upper Casting Bracket 2291449 1 2
    192“U” Bracket 2330476 ‘U’ Shaped Bracket, Nut is Attached to Casting 1 2
    193TV/TR Front Bracket230602112
    194Bulkhead Bracket2305886Does Not Include Cables12
    190
    191
    192
    193
    194 Bracket on units built before Mid-2002. 
    Replaced by Item # 192. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    9-12 Section 9-9 - Operator Control Panel Assembly
    Section 9-9
    Operator Control Panel Assembly
     
    Figure 9-7   Operator Control Panel Assembly (Op I/O)
    Inside 
    Touch Panel 
    Assembly
    201
    201A
    202202A
    203204
    205
    206
    206A
    208
    209
    207
    210
    211
    212 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 9 Renewal Parts 9-13
     
    Table 9-10    Operator Control Panel Assembly
    ItemPart NamePart NumberDescriptionQtyFRU
    201
    Lower Operator Panel Assembly2188902-2R3.X.X Software and above MUST USE 2188902-12
    Will cross to 2188902-12 when stock is depleted11
    201A
    Lower Operator Panel Assembly2188902-12With Standby. Backward compatible. 
    R3.X.X Software and above MUST USE11
    202
    A/N Keyboard - English2188902-4Alpha Numeric Keyboard
    R3.X.X Software and above MUST USE 2188902-1411
    202AA/N Analog Keyboard - English2188902-14With Standby. Backward compatible. 
    R3.X.X Software and above MUST USE this part11
    203Russian A/N Keyboard2188902-20As required11
    204Greek A/N Keyboard2188902-21As required11
    205
    Trackball2188902-511
    206
    Upper Operator Panel Assembly2188902-3Will cross to a 2188902-13 when stock is depleted. 1 1
    206A
    Upper Operator Panel Assembly2188902-13 1 1
    207LCD Backlight Inverter2188902-611
    208TGC PWA2188902-711
    209Main Handle with LabelFB200131Front Handle with Label11
    210Operator Panel Mode select Knob Kit 
    and Key Ring2188902-8Includes: 4 Mode select knobs (Inner & Outer pieces) 
    and Key Ring11
    211
    Universal Knob Kit2384216 Contains all blue knobs for the LOGIQ 9 and Vivid 7
    212
    Programmable Button Keytop Kit2188902-9Key caps for customer to customize the P1 - P4 key 
    caps11 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    9-14 Section 9-10 - Monitor Assembly
    Section 9-10
    Monitor Assembly
     
     
    Figure 9-8   Monitor Assembly
    Table 9-11    Monitor Assembly
    ItemPart NamePart NumberDescriptionQtyFRU
    301
    Open Frame Monitor2283334- Same as Vivid 7 & LOGIQ 7-11
    302
    Monitor Cable Assembly2304171- Same as Vivid 7 & LOGIQ 7-12
    303
    Speaker and Mounting Bracket Kit229788222
    304Monitor Tilt/Swivel/Male Post2284225Neck Assembly12
    305Monitor Rear Cover Assembly2303930Assembly with Speaker Grills12
    306Monitor Left Speaker Grill2279675Labeled as the LEFT Speaker Grill12
    307
    Monitor Right Speaker Grill2279676 Labeled as the RIGHT Speaker Grill 1 2
    308
    Monitor Screw Caps (covers)2303932Covers the screws on each side in the back of the 
    cover - Same as Vivid 7 & LOGIQ 7-22
    309
    Monitor Front Bezel Assembly2303929 Includes three logos Vivid 7, LOGIQ 7 and LOGIQ 9 1 2
    310Monitor Switch Assembly2297050User Switch Assembly12
    301
    303
    305
    307
    308
    309
    310
    306 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 9 Renewal Parts 9-15
    Section 9-11
    Front End Processor Assembly
     
    Figure 9-9   Front End Processor Assembly
    Empty Card Slot
    BT’01 
    Card 
    Rack
    BT’02, 
    BTCard 
    Rack
    401
    402402A
    402
    411411A411B
    403404
    405
    405A405B
    406
    406A
    406B
    407
    408409
    409A409C
    409B
    410
    412 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    9-16 Section 9-11 - Front End Processor Assembly
    Section 9-11      Front End Processor Assembly (cont’d)
     
    Table 9-12    Front End Processor Assembly
    Applications SoftwareR2.X.XR3.X.X
    LOGIQ 9
    Model Number
    ItemPart Name
    Part 
    Number
    DescriptionQtyFRU
    401
    Back Plane2260219XXXXX Front End Back Plane 1 2
    402
    SCB12260214-2XScan Control Board
    Use 2260214-3 if stock is depleted11
    402A
    SCB1.12260214-3
    X
    1XScan Control Board - 
    Required if using EBM Board11
    402BSCB22365739XXXHas BT’03 Functionality11
    411EBM2273639-3X2XCombined functions of EQ4 & BMP4. Positioned in 
    the old EQ4 Slot. Needs SCB 2260214-311
    411AEBM1.82273639-5XXXEBM Board Upgraded to EBM2 functionality - 
    Replace with EBM21N
    411B
    EBM22273639-
    20X X X Has BT’03 Functionality 1 1
    403
    BMP42260207X B-Mode Processor 1 1
    404
    EQ42260210X Equalization Board 1 1
    405TD42260194-2XXXXTransmit/Delay Board81
    405ATD52362863XNew power requirements different from TD4. Can 
    not mix with TD4s. Requires FEPS381
    405BTD5.12386031XSame Function as TD5 but larger and fewer ASICs. 
    You can mix TD5s and TD5.1s81
    406
    RF AMP2260212XXRadio Frequency Amplifier Board
    Obsolete - Use 2341985-21N
    406A
    RF AMP1.22341985-2
    X
    3X
    X3X Design Improvement 1 1
    406B
    RF AMP22357621 XNew power requirements different from
    RF AMP1.X - Requires FEPS311
    407XDIF2260221XXXXXTransmit Differential Board11
    408Top Plane2291656XXXXXTop Plane connects XDIF to 
    RF Amp on front of cards11
    409FEPS2260203X
    Front End Power Supply - First Generation
    Obsolete
    When replaced with 2273641-3 or 2273641-4,
    requires replacement of the Front End Card Rack 
    Cover in Kit 2374426
    1N
    2188900-X
    (BT’01)
    2351100-X
    (BT’02)
    2375600-X
    (BT’03)
    2188900-X
    (BT’03 upgrade)
    2351100-X
    (BT’03 upgrade) 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 9 Renewal Parts 9-17
    1 SCB1 can be used on this model number also.
    2 EBM can be used on this model replacing the BMP4 and EQ4. However EBM requires SCB1.1 at a minimum.
    3 RF Amp1.2 can be used in this model number also.
    4 FEPS2 and FEPS2.1 can be used on this model number. However, must also use FEP Card Rack Cover Kit 2374426.
    5 FEPS2.1 can also be used on these model numbers
    409A
    FEPS22273641-3
    X4XSecond Generation FEPS
    Cross to a FEPS2.1 if stock is depleted
    May require Front End Card Rack cover kit 2374426 
    if replacing a 2260203 FEPS11
    409C
    FEPS2.12273641-4
    X
    4X5X X Second Generation with Standby Capabilities 1 1
    409B
    FEPS32376843 XAnalog Voltages are +/- 6V and 13V compared to
    +/-6v and 12V found in FEPS, FEPS2 and FEPS2.1
    Must be used in systems with TD5.X and RF Amp2
    (DO NOT use in R2.X.X systems or 
    Systems upgraded to BT’03)11
    410FEP AC Line 
    Filter2315960XXXXXAttached to side of FEP Card Rack11
    412FEP Card Cage 
    Front Cover Kit2374426XX
    Required if upgrading from FEPS to FEPS2/FEP2.1
    Has additional shielding around Top Plane area and 
    Power Supply areas.
    Labels included in kit to match FEP configuration
    11
    Table 9-12    Front End Processor Assembly
    Applications SoftwareR2.X.XR3.X.X
    LOGIQ 9
    Model Number
    ItemPart Name
    Part 
    Number
    DescriptionQtyFRU2188900-X
    (BT’01)
    2351100-X
    (BT’02)
    2375600-X
    (BT’03)
    2188900-X
    (BT’03 upgrade)
    2351100-X
    (BT’03 upgrade) 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    9-18 Section 9-12 - I/O Interfaces
    Section 9-12
    I/O Interfaces
     
     +
    Figure 9-10   I/O Interfaces
    Table 9-13    I/O Interfaces
    ItemPart NamePart NumberDescriptionQtyFRU
    451
    External I/O AssemblyFB200198- Same as Vivid 7 -11
    452
    Internal I/O AssemblyFB200197- Same as Vivid 7 -
    Obsolete - Use FC2004231N
    452A
    Internal I/O AssemblyFC200423- Same as Vivid 7 -
    Has gain increase for improved microphone 
    performance11
    451
    452452A 
    						