GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-63 7-12-9Monitor Troubleshooting (cont’d) • Monitor does not work - Check AC circuit breaker - Verify power is present at monitor outlet - Check if video is present at backplane • Colored streaks appear in image - Check for presence of magnetic sources near the monitor - Eliminate the sources and then degauss the monitor - The monitor should automatically degauss itself each time power is applied if you wait at least 10 seconds before you turn power back ON • Prints do not match monitor - Verify factory default settings. Chapter 18 in the User Manual calls out suggested settings for various exam and lighting conditions. • Video test patterns are not clear, bright, parallel or square - Replace the monitor
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-64 Section 7-12 - Troubleshooting Trees 7-12-10 Unable to Record to VCR Figure 7-54 VCR Troubleshooting Unable to record to VCR! Check that all signal, and power supply cable connections to the VCR unit are OK.YESNO Is the VCR properly connected? Is the VCR tape put into the device? Put the tape into the device and rewind it. NOYES Replace VCR. Is recording unable even by manual operation? YES- Check the signal cable connections between VCR and Internal I/O. - Check the cables between BEP and Internal I/O NO
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-65 7-12-11 Printer Troubleshooting Figure 7-55 Printer Troubleshooting Unable to use Printer(s)! Check that all signal, and power supply cable connections between Printer and Scanner are OK.YESNO Is the printer(s) properly connected? Are there print papers left? Insert the printer papers. NOYES Is proper configuration for the Print key(s) set? Configure the Print key(s). NO See next page.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-66 Section 7-12 - Troubleshooting Trees 7-12-11Printer Troubleshooting (cont’d) Figure 7-56 Printer Troubleshooting (contd.) Still unable to use the Printer(s)! Check the signal cables Between BEP and Internal I/O.YESNO Is printing unavailable even by manual operation? Is there an error message on the monitor display? Replace BEP. NO YES Replace the Printer.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-67 7-12-12 CD-RW Troubleshooting Figure 7-57 CD-RW Troubleshooting Unable to record to CD-R! Insert a new, un-used CD-R disc.YESNO Is a new, unused CD-R disc inserted to the drive? Replace BEP. Are CD-Rs rated at the proper write speed? YESNOAcquire CD-Rs rated at the proper write speed.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-68 Section 7-12 - Troubleshooting Trees 7-12-13 MOD Troubleshooting Figure 7-58 Unable to record to MOD! Replace MOD. YESNO Does Replay function? Replace the BEP. NO Is CD-R able to record the data? YES Replace the MOD device.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-69 7-12-14 Network Troubleshooting 7-12-14-1 No Connection to the Network at All 1.) Check that the network cable between the scanner and the wall network is connected and well seated in both ends. 2.) Try a network cable that is known to be OK. 3.) Check the cable between the Back End Processor and Internal I/O. 4.) Connect a Crossed network cable between the Scanner and your PC. Try to ping from the scanner to the IP address on the PC. If OK, the hardware connection inside the scanner is OK. 7-12-14-2 No Contact With Server, But Connection to Network is OK Check Network setup on the scanner.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-70 Section 7-12 - Troubleshooting Trees 7-12-14Network Troubleshooting (cont’d) Figure 7-59 No Connection to the Network Connect the cable and verify that it is well seated in both ends.YESNO Is the TCP/IP Cable between LOGIQ™ 9 and network outlet connected and well seated? Replace the network cable. YES Is the connection OK now? NO Connect a Crossed Network Cable between the Scanner and your PC/ Laptop. Try to ping from the PC/Laptop to LOGIQ™ 9. If OK, the hardware connection inside LOGIQ™ 9 is OK. Try a network cable that is known to be OK NO Is the cable between BEP and Internal I/O OK? YESReplace the network cable between BEP and Internal I/O.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-71 Section 7-13 Troubleshooting TiPs Use Troubleshooting TiPs Table 7-3 on page 71 to access the troubleshooting TiPs listed. Each entry in the table will hyper-link to the issue and resolution Table 7-3 Troubleshooting TiPs Table Sub-SystemIssue Power SupplyNoise on FEPS Fans HV Errors No Boot System Will Not Boot Image ArtifactsImage Arifacts Caused by Front End Boards Connectivity No Ping No Verify System pings and verifies but doesn’t send Back End Processor System halt errors – Lock ups or intermittent problems Drive failures (CD or MOD) Video problems on main monitor Simulator mode Operator Panel No Audio No remote control for peripherals No video on LCD display Wrong key activated on the Touch panel Touch panel not responding I2C Buss
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-72 Section 7-13 - Troubleshooting TiPs 7-13-1 Soft Keys There are several Soft keys to be utilized on the LOGIQ™ 9: •Alt + F4 = Shut Down •Alt + F10 = is the SW Shut Down. This is very important tool to shut-down the system. If there is something problem with the I 2C bus you can shut-down the system by using this Alt + F10 keys. If the systems can easily be shut-down by using these keys it means that SW is working fine so the problem could be the I 2C bus, ON/OFF switch or the wire that carriers shut-down signal. •Alt +F5 or F11 for the echo simulator. 7-13-2 Power Supply 7-13-2-1 Noise on FEPS Fans • Check the part number of your FEPS. It should be 2260203 Rev H or Rev J. If not, the site may be eligible for FMI 70168 (Part Number 2340861) • Replace the power supply. 7-13-2-2 HV Errors • HV faults will show up whenever the TD boards have HV out of specification. Check FEPS voltages. You can visually check the LED status on the FEPS and on EQ (for HV). If any of the Power supplies is missing, there will not be HV output. If this is the case the Power Supply needs to be replaced. • Run Diagnostics on the Power Supply to check the voltages and their tolerances. Refer to 7-7-4-8 System-Level Analog Diagnostics on page 7-29. It will show you all the voltages including the Pulser HV (~ +120V) and HVP (+100V) and HVN (-100V). Pulser HV is generated on the FEPS and goes straight to the TD boards; HVN and HVP are generated on the EQ board from the Pulser HV. If HVN and/or HVP are the only ones missing, replace the EQ board. If Pulser HV is missing, replace FEPS. • Remember that Pulser HV won’t be present if you don’t have any probe connected. NOTE: You may find older power supplies that have a –15V supply. This power supply is not used on the system; newer FEPS don’t even have it present.