    Chapter 7 Diagnostics/Troubleshooting 7-63
    7-12-9Monitor Troubleshooting (cont’d)
    • Monitor does not work 
    - Check AC circuit breaker
    - Verify power is present at monitor outlet
    - Check if video is present at backplane
    • Colored streaks appear in image
    - Check for presence of magnetic sources near the monitor 
    - Eliminate the sources and then degauss the monitor
    - The monitor should automatically degauss itself each time power is applied if you wait at least 
    10 seconds before you turn power back ON
    • Prints do not match monitor
    - Verify factory default settings. Chapter 18 in the User Manual calls out suggested settings for 
    various exam and lighting conditions.
    • Video test patterns are not clear, bright, parallel or square
    - Replace the monitor 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-64 Section 7-12 - Troubleshooting Trees
    7-12-10 Unable to Record to VCR
    Figure 7-54   VCR Troubleshooting
    Unable to record to VCR!
    Check that all signal, and power 
    supply cable connections to the VCR 
    unit are OK.YESNO
    Is the VCR properly 
    connected?
    Is the VCR tape put into 
    the device?
    Put the tape into the device and 
    rewind it. NOYES
    Replace VCR.
    Is recording unable even 
    by manual operation?
    YES- Check the signal cable connections 
    between VCR and Internal I/O.
    - Check the cables between BEP and 
    Internal I/O NO 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-65
    7-12-11 Printer Troubleshooting
    Figure 7-55   Printer Troubleshooting
    Unable to use Printer(s)!
    Check that all signal, and power 
    supply cable connections between 
    Printer and Scanner are OK.YESNO
    Is the printer(s) properly 
    connected?
    Are there print papers left?
    Insert the printer papers. NOYES
    Is proper configuration for 
    the Print key(s) set?
    Configure the Print key(s). NO
    See next page. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-66 Section 7-12 - Troubleshooting Trees
    7-12-11Printer Troubleshooting (cont’d)
    Figure 7-56   Printer Troubleshooting (contd.)
    Still unable to use the Printer(s)!
    Check the signal cables Between 
    BEP and Internal I/O.YESNO
    Is printing unavailable 
    even by manual 
    operation?
    Is there an error message 
    on the monitor display?
    Replace BEP. NO
    YES
    Replace the Printer. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-67
    7-12-12 CD-RW Troubleshooting
    Figure 7-57   CD-RW Troubleshooting
    Unable to record to CD-R!
    Insert a new, un-used CD-R disc.YESNO
    Is a new, unused CD-R 
    disc inserted to the drive?
    Replace BEP.
    Are CD-Rs rated at the proper write 
    speed?
    YESNOAcquire CD-Rs rated at the proper write 
    speed. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-68 Section 7-12 - Troubleshooting Trees
    7-12-13 MOD Troubleshooting
    Figure 7-58   
    Unable to record to MOD!
    Replace MOD.
    YESNO
    Does Replay function?
    Replace the BEP. NO
    Is CD-R able to record the 
    data?
    YES
    Replace the MOD device. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-69
    7-12-14 Network Troubleshooting
    7-12-14-1 No Connection to the Network at All
    1.) Check that the network cable between the scanner and the wall network is connected and well 
    seated in both ends.
    2.) Try a network cable that is known to be OK.
    3.) Check the cable between the Back End Processor and Internal I/O.
    4.) Connect a Crossed network cable between the Scanner and your PC. Try to ping from the scanner 
    to the IP address on the PC. If OK, the hardware connection inside the scanner is OK.
    7-12-14-2 No Contact With Server, But Connection to Network is OK
    Check Network setup on the scanner. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-70 Section 7-12 - Troubleshooting Trees
    7-12-14Network Troubleshooting (cont’d)
    Figure 7-59   
    No Connection to the Network
    Connect the cable and verify that it is 
    well seated in both ends.YESNO
    Is the 
    TCP/IP Cable between 
    LOGIQ™ 9 and network 
    outlet connected and well 
    seated?
    Replace the network cable. YES
    Is the connection OK now?
    NO
    Connect a Crossed Network Cable 
    between the Scanner and your PC/
    Laptop. 
    Try to ping from the PC/Laptop to 
    LOGIQ™ 9.
     If OK, the hardware connection 
    inside LOGIQ™ 9 is OK.
    Try a network cable that is known to 
    be OK
    NO
    Is the 
    cable between BEP and 
    Internal I/O OK?
    YESReplace the network cable between 
    BEP and Internal I/O. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-71
    Section 7-13
    Troubleshooting TiPs
    Use Troubleshooting TiPs Table 7-3 on page 71 to access the troubleshooting TiPs listed. Each entry 
    in the table will hyper-link to the issue and resolution
    Table 7-3    Troubleshooting TiPs Table
    Sub-SystemIssue
    Power SupplyNoise on FEPS Fans
    HV Errors
    No Boot
    System Will Not Boot
    Image ArtifactsImage Arifacts Caused by Front End Boards
    Connectivity
    No Ping
    No Verify
    System pings and verifies but doesn’t send
    Back End Processor
    System halt errors – Lock ups or intermittent problems
    Drive failures (CD or MOD)
    Video problems on main monitor
    Simulator mode
    Operator Panel
    No Audio
    No remote control for peripherals
    No video on LCD display
    Wrong key activated on the Touch panel
    Touch panel not responding
    I2C Buss 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-72 Section 7-13 - Troubleshooting TiPs
    7-13-1 Soft Keys
    There are several Soft keys to be utilized on the LOGIQ™ 9:
    •Alt + F4 = Shut Down
    •Alt + F10 = is the SW Shut Down.
    This is very important tool to shut-down the system. If there is something problem with the I
    2C bus 
    you can shut-down the system by using this Alt + F10 keys. If the systems can easily be shut-down 
    by using these keys it means that SW is working fine so the problem could be the I
    2C bus, ON/OFF 
    switch or the wire that carriers shut-down signal.
    •Alt +F5 or F11 for the echo simulator.
    7-13-2 Power Supply
    7-13-2-1 Noise on FEPS Fans
    • Check the part number of your FEPS. It should be 2260203 Rev H or Rev J. If not, the site may be 
    eligible for FMI 70168 (Part Number 2340861)
    • Replace the power supply.
    7-13-2-2 HV Errors
    • HV faults will show up whenever the TD boards have HV out of specification. Check FEPS voltages. 
    You can visually check the LED status on the FEPS and on EQ (for HV). If any of the Power 
    supplies is missing, there will not be HV output. If this is the case the Power Supply needs to be 
    replaced.
    • Run Diagnostics on the Power Supply to check the voltages and their tolerances. Refer to 7-7-4-8 
    System-Level Analog Diagnostics on page 7-29. It will show you all the voltages including the 
    Pulser HV (~ +120V) and HVP (+100V) and HVN (-100V). Pulser HV is generated on the FEPS and 
    goes straight to the TD boards; HVN and HVP are generated on the EQ board from the Pulser HV. 
    If HVN and/or HVP are the only ones missing, replace the EQ board. If Pulser HV is missing, 
    replace FEPS.
    • Remember that Pulser HV won’t be present if you don’t have any probe connected.
    NOTE: You may find older power supplies that have a –15V supply. This power supply is not used on 
    the system; newer FEPS don’t even have it present. 
    						