    Chapter 5 Components and Functions (Theory) 5-61
    5-16-4 Service Home Page
    When the Service Interface is started, the Service Home Page appears. The Home Page contains the 
    software revision along with the hardware inventory and the results of the latest System Health 
    Information. The navigation bar at the top of the screen allows the user to select from several tools and 
    utilities.
    Figure 5-45   GE Service Home Page 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    5-62 Section 5-16 - Service Platform
    Figure 5-46  Customer Service Home Page 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 5 Components and Functions (Theory) 5-63
    5-16-5 Error Logs Tab
    From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit 
    selection. The Service Interface allows scanner logs to be viewed by all service users. 
    The Filter Error log is keyed to GE Service login access permissions and is not available to customer 
    level analysis.
    The log entries are color-coded to identify the error level severity at a glance.
    The Service Interface supports the transfer of these logs to local destinations such as the MOD, and 
    CD-ROM drive.  Table 5-35 Log Entry Key
    SeverityError LevelColor Code
    1
    InformationGreen
    2
    WarningBlue
    3
    ErrorRed
    Figure 5-47   Log Viewer / Logs / Log Entries 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    5-64 Section 5-16 - Service Platform
    5-16-5-1 Logs
    The seven sub-menus of the Logs category are System, Power, Infomatics, Temperature, Probe, 
    Board, and DICOM.
    NOTE: Figure 5-47 provides a graphical example of the log entries for the System Logs.
    Log table headings for the differrent logs are as follows:
    •System
    Log entry headings include Time Stamp; Error Level; Package; and Error Message.
    •Power
    Log entry headings include Time Stamp; Error Level; Package; and Error Message.
    • Infomatics
    Log entry headings include TimeStamp, Revision, PtID, PtDOB, PtSex, PtWeight, PtHeight, 
    ExamID, Exam Category, ExamCurDate, and ExamStartTime.
    •Temperature
    Log entry headings include Time Stamp; Error Level; Package; Upper FEC Sensor; and Lower FEC 
    Sensor.
    • Probe
    Log entry headings include Time Stamp; Error Level; Package; Error Message; Severity; Revision; 
    and three (3) new labels that have not yet been named.
    •Board
    Log entry headings include Time Stamp; Error Level; Package; Board; Severity; and two (2) new 
    labels that have yet been named.
    •DICOM
    Log entry headings include Time Stamp; Error Level; Package; and Error Message. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 5 Components and Functions (Theory) 5-65
    5-16-5-2 Utilities
    The two sub-menus of the Utilities category are Plot Log, and Plot Page.
    • Plot Log
    Allows for the color coded plot of all Log contents with the package on the ‘x’ axis and incident count 
    on the ‘y’ axis.
    •Plot Page
    Allows for the color coded plot of all Page contents with the package on the ‘x’ axis and incident 
    count on the ‘y’ axis.
    5-16-5-3 Search
    On the Text Search sub-menu of the Search category, users enter case-sensitive text they wish to find. 
    This filter field works well for filtering the Sys log file for the word “fail”.Figure 5-48   Utilities Sub-Menus
    Figure 5-49   Search Sub-Menu
    Plot Log Plot Page 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    5-66 Section 5-16 - Service Platform
    5-16-5-4 Filter
    Filtering is used to extract select data from the Error Logs. The capability to Filter is keyed to the login 
    permissions of the user.
    NOTE: Only GE Service personnel may Filter the Logs.
    5-16-5-5 Exit
    The sub-menu, Exit Log Viewer, returns the user to the Service Desktop home page.
    Figure 5-50   Exit Log Sub-Menu 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 5 Components and Functions (Theory) 5-67
    5-16-6 Diagnostics
    Detailed Diagnostic information is found in Chapter 7 - Diagnostics/Troubleshooting.
    Figure 5-51   GE Service Diagnostic Page
    Figure 5-52  User Diagnostic Page 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    5-68 Section 5-16 - Service Platform
    5-16-6-1 Diagnostics Execution
    Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level 
    diagnostic selection based on the user’s level and login access permissions. Remote access will require 
    disruptive diagnostic permissions to run Acquisition diagnostics.
    5-16-6-2 Diagnostic Reports
    Diagnostic tests return a report to the Service Platform. The platform retains the report and allows for 
    future viewing of the diagnostic logs.
    5-16-6-3 Proactive Diagnostics
    A system of self-monitoring is largely supported with the integration of iLinq. The scheduler, executive, 
    user interfaces, and some of the proactive diagnostic functions are provided by iLinq. Other tasks 
    should be provided by the product team. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 5 Components and Functions (Theory) 5-69
    5-16-7 Image Quality
    The Image Quality page is intended to contain tools for troubleshooting image quality issues.
    5-16-8 Calibration
    The Calibration page is intended to contain the tools used to calibrate the system.Figure 5-53   Image Quality Page
    Figure 5-54   Calibration Page
    FIELD IS NOT YET POPULATED
    FIELD IS NOT YET POPULATED 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    5-70 Section 5-16 - Service Platform
    5-16-9 Configuration
    The Configuration page is intended to be used to setup various configuration files on the system.
    The Service Platform is the access and authorization control for remote access to the configuration 
    subsystem.
    The enable/disable of software options can be done from this Configuration page.
    5-16-10 Utilities
    The Utilities page contains several miscellaneous tools.Figure 5-55   Configuration Page
    Figure 5-56   GE Service Utilities Page 
    						