GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-13 Section 7-6 Diagnostics As described in Chapter 5 - Components and Functions (Theory), the service platform uses a web- based user interface (UI) to provide access to common service components. A service dongle (part number 066E0703) is required to access areas proprietary to GE. The Diagnostic home page displays a listing of test categories consisting of Acquisition Diagnostics, I/ O Devices and Common Diagnostics. Expand the desired main category to select groups or individual tests. Figure 7-13 Global Service User Interface Diagnostic Screen
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-14 Section 7-7 - Acquisition Diagnostics Section 7-7 Acquisition Diagnostics Acquisition Diagnostics is one of three main headings under the GSUI Diagnostics Tab. Figure 7-14 shows the top level menu choices under Acquisition Diagnostics. The subheadings and their respective diagnostic/troubleshooting menu choices are called out and described in the sections that follow. Figure 7-14 GE Service Acquisition Diagnostic Menu Selections
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-15 Section 7-7 Acquisition Diagnostics (cont’d) Individual selections run several diagnostics in a Bottom-up Confidence-Building Order and return status information identifying any problems. The Acquisition Diagnostic menu is structured as follows: 7-7-1 Automatic Troubleshooting - Runs all diagnostics and analyzes the results to isolate to the faulty FRU. - This is the BEST way to diagnose a problem with the system. - Takes less than 6 minutes for all tests to run. - Disconnect all probes from the scanner before executing the Automatic Troubleshooting sequence. The choices for Automatic Troubleshooting, shown in Figure 7-16 on page 7-16 , allows for running all acquisition diagnostics under the selection Troubleshoot All Acquisition Hardware. Figure 7-15 Customer Acquisition Diagnostics
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-16 Section 7-7 - Acquisition Diagnostics 7-7-1Automatic Troubleshooting (cont’d) Figure 7-16 Automatic Troubleshooting Selections Figure 7-17 User Acquisition Diagnostics
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-17 When you select a test to run, like Troubleshoot all Acquisition Hardware , the display changes as shown in Figure 7-18 on page 7-18 . 1.) Instructions Frame - Displays either a test-specific text or the default instructions. 2.) Status Frame - Initially displays the last known status for a selected diagnostic. Once the diagnostic starts, the frame displays the “current” status of all test results. - The Status Frame also contains the user interface elements used for Diagnostic Control and Operator Feedback. 3.) Execute Button - This switch has two modes - each with appropriate text: - Execute - to start the diagnostic - Abort - to stop a diagnostic
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-18 Section 7-7 - Acquisition Diagnostics 7-7-1Automatic Troubleshooting (cont’d) Figure 7-18 Automatic Diagnostics Selected Figure 7-19 Diagnostic Selected 1 2 3 45 6 7 1 2 345 6 7
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-19 4.) Loop Count - This is an editable text field that will only accept numeric values of 4 digits or less. When the switch is configured as an “execute” switch and pressed, the loop count field will be queried to determine the number of times to execute the diagnostic. 5.) Progress Indicator - Displays a graphical progress indicator for the user. 6.) Short Text Message - Displays brief messages about the test’s progress during execution. 7.) Status Frame Background Color - Initially gray, the Status Frame background color changes upon completion of a diagnostic to indicate completion status. - Code Status Fail = Red - Code Status Pass = Green - Code Status neither Pass nor Fail (Example: Aborted) = Gray (default color). 7-7-2 Reliability Testing Overnight relialibility test. - Runs all diagnostics. Logs ALL results. Results accumulate in the log each time the tests are run. The log file continues to grow. NOTE: This is not intended for field service use. Typically a manufacturing test.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-20 Section 7-7 - Acquisition Diagnostics 7-7-3 Diagnostic Listings Table 7-2 Acquisition Diagnostics and Utility Listing Diagnostic/UtilityFunction Troubleshoot All Acquisition DiagnosticsRuns all Acquisition Hardware diagnostics and isolates faults to a FRU. Overnight Reliability Test Runs all diagnostics and records a running log. SCB PCI Interface Diag Tests the host’s ability to communicate with the SCB via the secondary PCI bus SCB Memory Diag Tests all of the on-board memory, including registers. SCB DSP Diag Tests the Scan Sequencer’s DSP and associated components. SCB Scan Bus Diag Tests the local Scan Bus and the associated system timing. SCB Image Port Diag Tests the Image Port’s functionality. BMP PCI Interface Diag Tests the host’s ability to communicate with the BMP via the PCI bus. BMP Memory Diag Tests all of the on-board memory, including registers. BMP Signal Path Diag Tests the entire BMP signal path. EQ PCI Interface DiagTests the host’s ability to communicate with the registers and memory on the EQ via the PCI bus. EQ Memory Diag Tests all of the on-board memory, including registers. EQ Signal Path Diag Tests the entire EQ signal path. TD Memory Diag Tests all of the on-board memory, for all installed TD boards, including registers. TD Access DiagTests all of the Access Modes utilized for Host/TD communication as well as Channel RAM to Rigel transfers. XDIF Interface DiagTests all of the XDIF functionality that is testable without a probe or probe simulator attached to the system, including the IIC communication. Comprehensive Signal Path DiagTests the signal path of every channel from the TD Rigel Transmitter Memories to the SCB, via both the EQ and the BMP. TD Analog Reference Diag Tests the AD_REF and VREF voltages on the TD. Analog TGC Diag Tests the generation of the analog TGC voltage. Power Supply Diag Tests the system voltages available on the EQ. Temperature Diag Tests the temperature of the system, wherever available. Noise Floor DiagMeasures the noise floor of the system. The noise floor will be calculated form IQ data received without a signal source. Loopback Diag Tests the analog front-end of the system. Acquisition Hardware Information Utility Writes a file for each Acquisition Hardware FRU containing all Vital Product Data. Transmit Utility - Start Transmit Utility - Change Channel Transmit Utility - Stop Receive Utility - Single Frame Receives signal on one channel at a time. Receive Utility - 8 Frames Visual Channel Utility
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-21 High Voltage Reset Utility Tests whether all TD boards can detect a high voltage fault. Set VRF Utility Set the analog VREF to a specified value. Set TGC Utility Set the analog TGC to a specified value. Probe Select Utility Select a probe connector to be active. Probe ID Read Utility Read the Probe ID of all installed probes. Front-End Calibration Utility Measures the Front End DC Offset, Gain, and Phase. Table 7-2 Acquisition Diagnostics and Utility Listing (Continued) Diagnostic/UtilityFunction
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-22 Section 7-7 - Acquisition Diagnostics 7-7-4 Diagnostic Groups NOTE: In the simplified block diagrams for the Diagnostic Groups, all unused system components (boards or signals) are ghosted (drawn in gray). 7-7-4-1 Board-Level SCB Diagnostics •All Board-Level SCB Diags: For this board-level Diagnostic Group, the setup simply includes the SCB and the Host. NOTE: The basic wellness of the i960 Intelligent I/O processor is tested indirectly via system start-up and the diagnostics listed below.Figure 7-20 Diagnostic Group Selections - SCB Tests Figure 7-21 Board-Level SCB Block Diagram XDIF/RFTD sEQBMP Scan Control BoardHostPCI PCI PCI IQ, RxSYNCBM IQ IIC RFProbeRF TD CTRLTxSYNC RxSYNC TxSYNCTxSYNC