Home > GE > Ultrasound System > GE Logiq 9 Service Manual

GE Logiq 9 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 570
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 6 Service Adjustments 6-9
    6-5-5-2 Procedure
    NOTE: The operator panel must be at or near its highest position.
    1.) From underneath the handle use a 10mm open-end (small “ignition” style) wrench to screw out the 
    cable adjustment screw so the underside of the screw head is about half way between the casting 
    and the unsqueezed lever. See Figure 6-5 on page 6-8 .
    2.) There is a Wire Clamp nestled in the Forked Retainer. Refer to Figure 6-6. Using 2 slotted 
    screwdrivers (one on each side of the clamp) loosen the clamp.
    3.) Using needle nose pliers, hold the cable end which is above the clamp and pull upward to take up 
    any cable slack.
    4.) Using one of the two slotted screwdrivers, push the Clamp down along the cable so that it is nestled 
    back in the Forked Retainer.
    5.) Tighten the clamp with one screwdriver to hold the cable. Let go of the pliers and then use the two 
    screwdrivers to fully tighten the clamp.
    6.) Without squeezing the release lever, check that the forked retainer is in its fully up position.
    7.) Squeeze the release lever and watch the forked retainer move down immediately as the lever is 
    squeezed, there should be very little play in the cable. Small amounts of play can be taken up (or 
    slackened) with the cable adjustment screw per step 1.Figure 6-6   Gas Spring Mechanism
    WARNINGWARNINGDo NOT attempt to pinch together or pry apart the forked retainer. If one of the tines 
    should break off, the gas spring piston may have to be replaced at the factory.
    Wire Clamp Cable End
    Forked Retainer 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    6-10 Section 6-5 - Z-Release Cable and Gas Spring Adjustment
    6-5-6 Gas Spring Adjustment
    1.) Using a 17mm wrench, loosen the Jamb Nut.
    2.) Using pliers, grab the Piston Shaft just below the nut and turn the shaft based on the your situation:
    - If the operation panel can move without squeezing the release lever: Turn so that the shaft 
    un-threads from the release mechanism
    - If the operation panel does not move even when squeezing the release lever: Turn so the 
    shaft threads into the release mechanism further
    NOTE: Slight turning makes a large difference, turn only ¼ of a turn at a time. The nut will probably turn with 
    the shaft, use it as a guide as to how much you are turning.
    3.) Check the operation, push down or pull up on the operator panel handle while slowly squeezing the 
    release lever, motion should start with the lever at 1/4 to 2/3 of the full range of movement. Repeat 
    step 2 as necessary.
    4.) Tighten the Jamb nut.Figure 6-7   Gas Spring Adjustment
    Release Mechanism
    Jamb Nut
    Piston Shaft 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 6 Service Adjustments 6-11
    6-5-7 Emergency Release Procedure
    When the operator panel is at or near its highest position it is possible to see the gas spring mechanism 
    with the Black and White Printer or the VCR removed. If the operator panel is stuck in the down position, 
    it is not possible to see the gas spring mechanism as it is down behind the card rack. To avoid removing 
    the Card Rack to make adjustments, try this procedure to raise the operator panel.
    1.) Using 3mm Allen Wrench, remove the shoulder bolt holding the release lever on the handle.
    2.) Tilt the release lever down and pull on the cable as shown in the figure below. This will give 
    unlimited travel and hopefully release the gas spring. Push up on the operator panel while you are 
    pulling on the cable.
    3.) If you have no cable to pull (cable broken) or you are unable to release the gas spring in this 
    manner, you must remove the Card Rack to gain access to the gas spring mechanism.
    6-5-8 Procedure Completion
    1.) Install the VCR.
    2.) Install the Black-and-White printer.
    3.) Replace the front cover as described in Section 8-2-5 “Front Cover Replacement Procedure” on 
    page 8-9.
    4.) Replace the side covers as described in Section 8-2-2 “Side Covers (with Bumper) Replacement 
    Procedure” on page 8-5.
    5.) If required, Power On / Bootup the scanner as described in Section 4-3-1 “Power On / Boot Up” on 
    page 4-2.
    Select the movie camera icon to view the video of the emergency 
    release procedure.
    1 minute and 58 seconds
    CAUTIONDo NOT use excessive force!
    Figure 6-8   Release Lever 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    6-12 Section 6-6 - Brake Arm Adjustment
    Section 6-6
    Brake Arm Adjustment
    Adjustments to the brake arm may be necessary if the pedals do not depress readily nor engage so the 
    front wheels lock in place and do not swivel.
    6-6-1 Tools
    Flat-head screwdriver
    6-6-2 Manpower
    One person, 1 hour + travel
    6-6-3 Preparations
    1.) Remove the front bumper as described in Front Bumper Replacement Procedure on 8-11.
    Select the movie camera icon to view the video of the brake arm 
    adjustment procedure.
    1 minute and 58 seconds
    Figure 6-9   Brake Adjustment 
    TOP VIEW
    UNDERSIDE VIEW Swivel Lock Pedal /
    Brake Pedal Release
    Brake Pedal /
    Swivel Lock Release
    Brake Arm
    Adjustment Rod
    Front Brake Arm
    Adjustment Nuts
    Rear Brake Arm
    Adjustment Nut 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 6 Service Adjustments 6-13
    6-6-4 Brake Arm Adjustment Procedure
    To adjust the brake arm it is recommended to separate both front adjustment nuts from the brake 
    assembly.
    2.) Remove the brake clamps that are inserted through a small hole in the front adjustment nuts and 
    slipped over the nut at its collar.
    3.) Loosen the front retaining nuts to allow room for adjustments.
    4.) With a screwdriver you can pry the front adjustment nuts away from their connectors in the brake 
    assembly.
    5.) Set the main brake - the brake assembly should be pulled to the left.
    6.) Seat the right adjustment rod into the right adjustment nut under the right front wheel as far as it will 
    go.
    7.) Seat the right front adjustment nut onto the adjustment rod as far as it will go.
    8.) In order to insure that the same length of adjustment rod is seated into the two adjustment nuts you 
    will turn the adjustment rod one turn out of the rear adjustment nut under the front wheel and turn 
    the front adjustment nut off the adjustment rod a similar turn.
    9.) Repeat this alternating rotation until the front adjustment nut lines up easily with the connector in 
    the brake assembly.
    10.) Push the adjustment nut onto the brake assembly connector and secure the retaining nut tight 
    against the adjustment nut.
    NOTE: Repeat the same procedure to adjust the swivel lock. The only difference would be to set the swivel lock 
    prior to the adjusments.
    6-6-5 Procedure Completion
    1.) Install the front bumper.Figure 6-10   Front Adjustment Hardware
    Brake ClampScrewdriver Adjustment Nuts
    Retaining NutsBrake Assembly 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    6-14 Section 6-7 - Brake Adjustment
    Section 6-7
    Brake Adjustment
    6-7-1 Front Wheel Brakes
    Service adjustment to the front brakes involve setting the brake so it will properly “bite” the rear tire. The 
    default setting should provide proper braking. If not, follow the procedures below.
    6-7-1-1 Tools
    Allen Wrenches
    6-7-1-2 Manpower
    One person, 1 hour + travel
    6-7-1-3 Preparations
    1.) Set the rear brakes to prevent the scanner from rolling.
    2.) Remove the front bumper Section 8-2-7 “Front Bumper Replacement Procedure” on page 8-11.
    6-7-1-4 Front Brake Adjustment Procedure
    1.) Refer to Figure 6-11. Use the Allen wrench to add more pressure in order to get more braking action.
    2.) If the brakes are too tight, loosen the adjustment with the Allen wrench.
    3.) Check the brakes and readjust as necessary before replacing the front bumper.Figure 6-11   Front Brake Adjustment
    Adjustment HoleAllen Wrench to Adjust Brake
    Wheel Brake 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 6 Service Adjustments 6-15
    6-7-2 Rear Wheel Brakes Adjustment
    Service adjustment to the rear brakes involves setting the brake so it will properly “bite” the rear tire. 
    The default setting should provide proper braking. If not, follow the procedures below to remove the rear 
    wheel and adjust the brake.
      
    6-7-2-1 Tools
    Bottle Jack
    Allen Wrenches
    6-7-2-2 Manpower
    One person, 1 hour + travel
    6-7-2-3 Preparations (click here to view the cover removal video)
    1.) Remove the left- and right-side covers.
    2.) Remove the upper rear cover.
    3.) Remove the lower rear cover.
    Select the movie camera icon to view the video of the rear brake 
    adjustment procedure.
    2 minutes and 57 seconds
    WARNINGWARNINGPrior to elevating scanner, verify that the floating keyboard is locked in its lowest 
    position. Verify that the front brake is locked and the scanner is unable to swivel. Verify 
    that the rear brakes are in the locked position.
    WARNINGWARNINGUse extreme caution when elevating scanner. Elevating scanner will make it unstable 
    which could cause the scanner to tip over. 
    Figure 6-12   Rear Brake Adjustment
    2 screws removed
    2 screws removed
    from underside from top
    BrakeRear Brake
    Set / Release
    Handle Area on either side
    of bottom shield used
    to elevate scanner and
    remove rear wheel. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    6-16 Section 6-7 - Brake Adjustment
    6-7-2-4 Rear Wheel Removal Procedure
    4.) Elevate the scanner by placing the bottle jack on the back frame on the side of the bottom shield of 
    the wheel you will be removing. You can also use a block of wood as a lever to raise the scanner 
    and place a second block of wood under the chassis to support the rear wheel off of the ground.
    5.) Remove the four screws and remove the wheel from the scanner.
    6.) Use an allen wrench to adjust the brake so it will properly “bite” and release the rear tire.
    6-7-2-5 Rear Wheel Installation Procedure
    1.) After adjusting the rear brake, install the rear wheel on the scanner. Hand tighten the four screws.
    2.) Carefully lower the scanner and remove the bottle jack.
    3.) Tighten the four screws using the weight of the scanner to provide resistance.
    4.) Verify brake and scanner operation then replace all covers previously removed.Figure 6-13   Rear Brake Adjustment 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-1
    Chapter 7
    Diagnostics/Troubleshooting
    Section 7-1
    Overview
    7-1-1 Purpose of Chapter 7
    This section describes how to setup and run the tools and software that help maintain image quality and 
    system operation. Basic host, system, and board level diagnostics are run whenever power is applied. 
    Some Service Tools may be run at the application level.
    Section 7-2
    Network Connectivity Troubleshooting
    There is a troubleshooting tool available that the customer can use as a first step to investigate network 
    connectivity issues. It gives the current status of network and provides the option to restart the network 
    components in the Back End Processor.
    This ALT+N function can also be used to ensure that GE Service is not connected remotely before 
    placing the system in Stand-by Mode. If the syetem is placed in Stand-by Mode with GE Service 
    connected the system will not come out of Stand-by Mode later. The system will have to be power 
    cycled (rebooted).
    Activate the tool by pressing ALT+N and follow the choices displayed. Table 7-1    Contents in Chapter 7
    SectionDescriptionPage Number
    7-1
    Overview7-1
    7-2
    Network Connectivity Troubleshooting7-1
    7-3
    Gathering Trouble Data7-2
    7-4Screen Captures7-4
    7-5Virtual Console Observation7-8
    7-6Diagnostics7-13
    7-7
    Acquisition Diagnostics7-14
    7-8
    Preferred Test Strategy for PCI 9054 Interface7-35
    7-9
    Preferred Test Strategy for Memory7-37
    7-10I/O Devices7-40
    7-11Common Diagnostics7-44
    7-12Troubleshooting Trees7-50
    7-13
    Troubleshooting TiPs7-71 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-2 Section 7-3 - Gathering Trouble Data
    Section 7-3
    Gathering Trouble Data
    7-3-1 Overview
    There may be a time when it would be advantageous to capture trouble images and system data (logs) 
    for acquisition through remote diagnostics (InSite) or to be sent back to the manufacturer for analysis. 
    There are different options to acquire this data that would give different results. 
    7-3-2 Collect Vital System Information
    The following information is necessary in order to properly analyze data or images being reported as a 
    malfunction or being returned to the manufacturer:
    - Product Name = LOGIQ™ 9
    From the Utility>System>About screen:
    Applications Software
    - Software Version
    - Software Part Number
    System Base Image Software
    - Base Image Revision
    - Image Part Number 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq 9 Service Manual