GE Logiq 9 Service Manual
Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 6 Service Adjustments 6-9 6-5-5-2 Procedure NOTE: The operator panel must be at or near its highest position. 1.) From underneath the handle use a 10mm open-end (small “ignition” style) wrench to screw out the cable adjustment screw so the underside of the screw head is about half way between the casting and the unsqueezed lever. See Figure 6-5 on page 6-8 . 2.) There is a Wire Clamp nestled in the Forked Retainer. Refer to Figure 6-6. Using 2 slotted screwdrivers (one on each side of the clamp) loosen the clamp. 3.) Using needle nose pliers, hold the cable end which is above the clamp and pull upward to take up any cable slack. 4.) Using one of the two slotted screwdrivers, push the Clamp down along the cable so that it is nestled back in the Forked Retainer. 5.) Tighten the clamp with one screwdriver to hold the cable. Let go of the pliers and then use the two screwdrivers to fully tighten the clamp. 6.) Without squeezing the release lever, check that the forked retainer is in its fully up position. 7.) Squeeze the release lever and watch the forked retainer move down immediately as the lever is squeezed, there should be very little play in the cable. Small amounts of play can be taken up (or slackened) with the cable adjustment screw per step 1.Figure 6-6 Gas Spring Mechanism WARNINGWARNINGDo NOT attempt to pinch together or pry apart the forked retainer. If one of the tines should break off, the gas spring piston may have to be replaced at the factory. Wire Clamp Cable End Forked Retainer
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 6-10 Section 6-5 - Z-Release Cable and Gas Spring Adjustment 6-5-6 Gas Spring Adjustment 1.) Using a 17mm wrench, loosen the Jamb Nut. 2.) Using pliers, grab the Piston Shaft just below the nut and turn the shaft based on the your situation: - If the operation panel can move without squeezing the release lever: Turn so that the shaft un-threads from the release mechanism - If the operation panel does not move even when squeezing the release lever: Turn so the shaft threads into the release mechanism further NOTE: Slight turning makes a large difference, turn only ¼ of a turn at a time. The nut will probably turn with the shaft, use it as a guide as to how much you are turning. 3.) Check the operation, push down or pull up on the operator panel handle while slowly squeezing the release lever, motion should start with the lever at 1/4 to 2/3 of the full range of movement. Repeat step 2 as necessary. 4.) Tighten the Jamb nut.Figure 6-7 Gas Spring Adjustment Release Mechanism Jamb Nut Piston Shaft
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 6 Service Adjustments 6-11 6-5-7 Emergency Release Procedure When the operator panel is at or near its highest position it is possible to see the gas spring mechanism with the Black and White Printer or the VCR removed. If the operator panel is stuck in the down position, it is not possible to see the gas spring mechanism as it is down behind the card rack. To avoid removing the Card Rack to make adjustments, try this procedure to raise the operator panel. 1.) Using 3mm Allen Wrench, remove the shoulder bolt holding the release lever on the handle. 2.) Tilt the release lever down and pull on the cable as shown in the figure below. This will give unlimited travel and hopefully release the gas spring. Push up on the operator panel while you are pulling on the cable. 3.) If you have no cable to pull (cable broken) or you are unable to release the gas spring in this manner, you must remove the Card Rack to gain access to the gas spring mechanism. 6-5-8 Procedure Completion 1.) Install the VCR. 2.) Install the Black-and-White printer. 3.) Replace the front cover as described in Section 8-2-5 “Front Cover Replacement Procedure” on page 8-9. 4.) Replace the side covers as described in Section 8-2-2 “Side Covers (with Bumper) Replacement Procedure” on page 8-5. 5.) If required, Power On / Bootup the scanner as described in Section 4-3-1 “Power On / Boot Up” on page 4-2. Select the movie camera icon to view the video of the emergency release procedure. 1 minute and 58 seconds CAUTIONDo NOT use excessive force! Figure 6-8 Release Lever
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 6-12 Section 6-6 - Brake Arm Adjustment Section 6-6 Brake Arm Adjustment Adjustments to the brake arm may be necessary if the pedals do not depress readily nor engage so the front wheels lock in place and do not swivel. 6-6-1 Tools Flat-head screwdriver 6-6-2 Manpower One person, 1 hour + travel 6-6-3 Preparations 1.) Remove the front bumper as described in Front Bumper Replacement Procedure on 8-11. Select the movie camera icon to view the video of the brake arm adjustment procedure. 1 minute and 58 seconds Figure 6-9 Brake Adjustment TOP VIEW UNDERSIDE VIEW Swivel Lock Pedal / Brake Pedal Release Brake Pedal / Swivel Lock Release Brake Arm Adjustment Rod Front Brake Arm Adjustment Nuts Rear Brake Arm Adjustment Nut
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 6 Service Adjustments 6-13 6-6-4 Brake Arm Adjustment Procedure To adjust the brake arm it is recommended to separate both front adjustment nuts from the brake assembly. 2.) Remove the brake clamps that are inserted through a small hole in the front adjustment nuts and slipped over the nut at its collar. 3.) Loosen the front retaining nuts to allow room for adjustments. 4.) With a screwdriver you can pry the front adjustment nuts away from their connectors in the brake assembly. 5.) Set the main brake - the brake assembly should be pulled to the left. 6.) Seat the right adjustment rod into the right adjustment nut under the right front wheel as far as it will go. 7.) Seat the right front adjustment nut onto the adjustment rod as far as it will go. 8.) In order to insure that the same length of adjustment rod is seated into the two adjustment nuts you will turn the adjustment rod one turn out of the rear adjustment nut under the front wheel and turn the front adjustment nut off the adjustment rod a similar turn. 9.) Repeat this alternating rotation until the front adjustment nut lines up easily with the connector in the brake assembly. 10.) Push the adjustment nut onto the brake assembly connector and secure the retaining nut tight against the adjustment nut. NOTE: Repeat the same procedure to adjust the swivel lock. The only difference would be to set the swivel lock prior to the adjusments. 6-6-5 Procedure Completion 1.) Install the front bumper.Figure 6-10 Front Adjustment Hardware Brake ClampScrewdriver Adjustment Nuts Retaining NutsBrake Assembly
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 6-14 Section 6-7 - Brake Adjustment Section 6-7 Brake Adjustment 6-7-1 Front Wheel Brakes Service adjustment to the front brakes involve setting the brake so it will properly “bite” the rear tire. The default setting should provide proper braking. If not, follow the procedures below. 6-7-1-1 Tools Allen Wrenches 6-7-1-2 Manpower One person, 1 hour + travel 6-7-1-3 Preparations 1.) Set the rear brakes to prevent the scanner from rolling. 2.) Remove the front bumper Section 8-2-7 “Front Bumper Replacement Procedure” on page 8-11. 6-7-1-4 Front Brake Adjustment Procedure 1.) Refer to Figure 6-11. Use the Allen wrench to add more pressure in order to get more braking action. 2.) If the brakes are too tight, loosen the adjustment with the Allen wrench. 3.) Check the brakes and readjust as necessary before replacing the front bumper.Figure 6-11 Front Brake Adjustment Adjustment HoleAllen Wrench to Adjust Brake Wheel Brake
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 6 Service Adjustments 6-15 6-7-2 Rear Wheel Brakes Adjustment Service adjustment to the rear brakes involves setting the brake so it will properly “bite” the rear tire. The default setting should provide proper braking. If not, follow the procedures below to remove the rear wheel and adjust the brake. 6-7-2-1 Tools Bottle Jack Allen Wrenches 6-7-2-2 Manpower One person, 1 hour + travel 6-7-2-3 Preparations (click here to view the cover removal video) 1.) Remove the left- and right-side covers. 2.) Remove the upper rear cover. 3.) Remove the lower rear cover. Select the movie camera icon to view the video of the rear brake adjustment procedure. 2 minutes and 57 seconds WARNINGWARNINGPrior to elevating scanner, verify that the floating keyboard is locked in its lowest position. Verify that the front brake is locked and the scanner is unable to swivel. Verify that the rear brakes are in the locked position. WARNINGWARNINGUse extreme caution when elevating scanner. Elevating scanner will make it unstable which could cause the scanner to tip over. Figure 6-12 Rear Brake Adjustment 2 screws removed 2 screws removed from underside from top BrakeRear Brake Set / Release Handle Area on either side of bottom shield used to elevate scanner and remove rear wheel.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 6-16 Section 6-7 - Brake Adjustment 6-7-2-4 Rear Wheel Removal Procedure 4.) Elevate the scanner by placing the bottle jack on the back frame on the side of the bottom shield of the wheel you will be removing. You can also use a block of wood as a lever to raise the scanner and place a second block of wood under the chassis to support the rear wheel off of the ground. 5.) Remove the four screws and remove the wheel from the scanner. 6.) Use an allen wrench to adjust the brake so it will properly “bite” and release the rear tire. 6-7-2-5 Rear Wheel Installation Procedure 1.) After adjusting the rear brake, install the rear wheel on the scanner. Hand tighten the four screws. 2.) Carefully lower the scanner and remove the bottle jack. 3.) Tighten the four screws using the weight of the scanner to provide resistance. 4.) Verify brake and scanner operation then replace all covers previously removed.Figure 6-13 Rear Brake Adjustment
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-1 Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation. Basic host, system, and board level diagnostics are run whenever power is applied. Some Service Tools may be run at the application level. Section 7-2 Network Connectivity Troubleshooting There is a troubleshooting tool available that the customer can use as a first step to investigate network connectivity issues. It gives the current status of network and provides the option to restart the network components in the Back End Processor. This ALT+N function can also be used to ensure that GE Service is not connected remotely before placing the system in Stand-by Mode. If the syetem is placed in Stand-by Mode with GE Service connected the system will not come out of Stand-by Mode later. The system will have to be power cycled (rebooted). Activate the tool by pressing ALT+N and follow the choices displayed. Table 7-1 Contents in Chapter 7 SectionDescriptionPage Number 7-1 Overview7-1 7-2 Network Connectivity Troubleshooting7-1 7-3 Gathering Trouble Data7-2 7-4Screen Captures7-4 7-5Virtual Console Observation7-8 7-6Diagnostics7-13 7-7 Acquisition Diagnostics7-14 7-8 Preferred Test Strategy for PCI 9054 Interface7-35 7-9 Preferred Test Strategy for Memory7-37 7-10I/O Devices7-40 7-11Common Diagnostics7-44 7-12Troubleshooting Trees7-50 7-13 Troubleshooting TiPs7-71
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-2 Section 7-3 - Gathering Trouble Data Section 7-3 Gathering Trouble Data 7-3-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition through remote diagnostics (InSite) or to be sent back to the manufacturer for analysis. There are different options to acquire this data that would give different results. 7-3-2 Collect Vital System Information The following information is necessary in order to properly analyze data or images being reported as a malfunction or being returned to the manufacturer: - Product Name = LOGIQ™ 9 From the Utility>System>About screen: Applications Software - Software Version - Software Part Number System Base Image Software - Base Image Revision - Image Part Number