    Chapter 8 Replacement Procedures 8-113
    Section 8-28
    Keypad Replacement Procedure
    8-28-1 Manpower
    One person, 
    8-28-2 Tools
    Fingernail or a small screwdriver blade
    8-28-3 Preparations
    This procedure describes how worn out keypads or keypads of one language can be replaced with new 
    keypads or keypads of other language.
    Shut down the system as described in Power Off / Shutdown on page 4-3.
    8-28-4 Keypad Replacement Procedure
    1.) Flip off keypads with your fingernail or a small screwdriver blade.
    2.) Install replacement pads.Figure 8-122   Flip off keypad 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-114 Section 8-29 - Alpha-Numeric (A/N) Keyboard Replacement Procedure
    Section 8-29
    Alpha-Numeric (A/N) Keyboard Replacement Procedure
    The A/N Keyboard can be replaced without removal of the operator panel from the frame.  The screws 
    for removal are hidden under the “return” and left “shift” keys.
    8-29-1 Manpower
    8-29-2 Tools
    A key removal tool is required. Acquire a commerical key removal tool or fashion two loops made from 
    paperclips as shown in Figure 8-123 on page 8-114 .
    8-29-3 Preparations
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3
    8-29-4 A/N Keyboard Removal
    1.) Place the loops under opposite corners of the key to be removed as shown in Figure 8-
    124 on page 8-115 .  To remove the AN keyboard, the “return” and left “shift” keys must be 
    removed. 
    2.) Pull up with the tool but also slightly to the top edge (number row) of the keyboard.  The slight pull 
    towards the top forces the white plunger to come off with the key.  If using paperclips, it works best 
    if you use one hand on each clip and pull together.
    3.) If the white plunger stays in the keyboard, you can reinstall the key and try again. Or, the plunger 
    can be removed from the keyboard needle nose pliers grabbing between the same points as where 
    the key inserts.Figure 8-123  Key Removal Tools 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-115
    4.) After the key is removed, push the plunger into the keytop to make sure it is seated all the way.Figure 8-124  Removal Tool in Place
    Figure 8-125  Keytop and Plunger Assembly
    Keytop
    Plunger 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-116 Section 8-29 - Alpha-Numeric (A/N) Keyboard Replacement Procedure
    8-29-4 A/N Keyboard Removal (cont’d)
    5.) With the “return” and left “shift” keys are removed, you can see the screws that secure the keyboard 
    to the plastic bezel.  Loosen these screws so that they pop up but do not remove them as they are 
    captive screws that remain with the AN keyboard.
    6.) Using the key removal tool, lift up on a key to help pull the AN keyboard out of the plastic bezel. 
    Choose a key to the  far right and lift the right side of the keyboard out first. 
    7.) Lift out the bottom edge of the keyboard bar to gain access to the interconnect cables.  Detach the 
    2 cables from the keyboard and lift  the keyboard out.Figure 8-126  Keyboard Screw Locations
    Screw
    Screw 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-117
    8-29-4 A/N Keyboard Removal (cont’d)
    8.) Install the removed keys back onto the removed keyboard.  If a key does not seem to seat down 
    fully, it is probably because the plunger is not pressed onto the keytop all the way.Figure 8-127  Interconnect Cable Locations 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-118 Section 8-29 - Alpha-Numeric (A/N) Keyboard Replacement Procedure
    8-29-5 A/N Keyboard Installation
    1.) Using the key removal tool, remove the “return” and left “shift” keys from the replacement keyboard. 
    use the same procedure as before. Make sure that the plungers come off with the keys and that 
    they are fully seated into the keytop after removal.
    2.) Check the replacement keyboards rubber gasket. When seated correctly, it should wrap around the 
    edge of the circuit board.  If the gasket is not seated correctly, the keyboard will be difficult to install.
    3.) Connect the cables to the replacement keyboard.  Check that the cables are not shifted by one row 
    of pins as this is possible to do.
    4.) Insert the keyboard left side first making sure the gasket remains seated around the circuit board.Figure 8-128  Keyboard Gasket 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-119
    5.) Continue to work the keyboard down into the plastics slightly pushing to the left so that the gasket 
    on the right side goes into the plastics without getting disturbed. 
    6.) Seat the keyboard fully by pushing down on the whole keyboard with both hands.
    7.) Check the function of the keys around the keyboard edge.  If keys stick or have a bad feel. the 
    gasket probably got disturbed and is not around the edge of the circuit board. Remove the keyboard 
    check the gasket and try again.
    8.) If the keyboard is seated correctly, secure it with the screws.
    9.) Reinstall the keytops. If a keytop seem too high, the white plunger was probably not seated all the 
    way into the keytop. The keytop and plunger must be removed to fully seat the plunger into the 
    keytop.
    10.) Energize the system and type with the keyboard to check the function of each key.Figure 8-129  Replacing Keyboard 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-120 Section 8-30 - Trackball Assembly Replacement Procedure
    Section 8-30
    Trackball Assembly Replacement Procedure
    8-30-1 Manpower
    One person, 1 hour 30 minutes
    8-30-2 Tools
    • Phillips screwdriver size 1
    • Socket 2,5 mm
    8-30-3 Preparations
    1.) Shut down the system as described in Section 4-3-2 on page 4-3.
    2.) Remove the Operator Panel as described in Section Section 8-27 on page 8-111.
    3.) Put the panel upside down on a soft cushion.
    Figure 8-130   Trackball
    Trackball assembly Screws
    Trackball unitConnectors
    Trackball housing
    Screws
    Operator Panel with Trackball assembly 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-121
    8-30-4 Trackball Removal Procedure
    1.) Disconnect two (2) connectors at the trackball unit and 1 at the print between trackball unit and 
    operator panel. Remove trackball assembly.
    2.) Unscrew four (4) Phillips screws on the print.
    3.) Unscrew two (2) screws/nuts and remove trackball unit from print.
    4.) Unscrew two (2) Phillips screws on the trackball housing.
    5.) Open trackball housing, and remove trackball.
    8-30-5 Trackball Installation Procedure
    1.) Install trackball in housing. Replace cover and fasten removed screws.
    2.) Install the trackball unit on the print and fasten removed screws/nuts.
    3.) Install trackball assembly on the Operator Panel and fasten 4 screws.
    4.) Connect three (3) connectors.
    5.) Install Operator Panel as described in Section Section 8-27 on page 8-111. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-122 Section 8-31 - Fan Assembly Replacement Procedure
    Section 8-31
    Fan Assembly Replacement Procedure
    8-31-1 Manpower
    One person, 1 hour
    8-31-2 Tools
     Phillips screwdriver, size 2.
    8-31-3 Preparations (click here to view the cover removal video)
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3
    2.) Remove the Right and Left Covers as described on Section 8-2-2 on page 8-5.
    3.) Remove Upper Rear Cover and Filter as shown in Section 8-2-3 on page 8-7.
    4.) Remove the Lower Rear Cover as shown in Section 8-2-6 on page 8-10.
    NOTE: There are washers and magnetic lock for Upper Rear Cover on the upper screws.
    5.) Remove the Top Cover as shown in Section 8-2-4 on page 8-8.
    6.) Remove the Rear Handle ashown in Section 8-2-8 on page 8-12.
    7.) Remove the filter media.
    .
    8-31-4 Fan Assembly Removal Procedure
    1.) Unscrew four (4) Phillips screws that fasten Cover Air Duct. See Figure 8-132.
    2.) Push down the lower side of Cover Air Duct to free lower ends of the Filter Frame. Remove Filter 
    Frame downwards.
    3.) Becarful of sharp edges. Move Cover Air Duct up in the front, then backwards, off the Fan Assembly 
    and out of the System.
    Select the movie camera icon to view the video of the monitor 
    replacement procedure.
    6 minutes and 23 seconds
    Figure 8-131   Fan Assembly, location and hitch up
    Fan assembly
    External I/O
    Fan Cover Air Duct
    AC Controller Module
    Modem
    Fan Cable
    Hitch up 
    						