GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-113 Section 8-28 Keypad Replacement Procedure 8-28-1 Manpower One person, 8-28-2 Tools Fingernail or a small screwdriver blade 8-28-3 Preparations This procedure describes how worn out keypads or keypads of one language can be replaced with new keypads or keypads of other language. Shut down the system as described in Power Off / Shutdown on page 4-3. 8-28-4 Keypad Replacement Procedure 1.) Flip off keypads with your fingernail or a small screwdriver blade. 2.) Install replacement pads.Figure 8-122 Flip off keypad
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-114 Section 8-29 - Alpha-Numeric (A/N) Keyboard Replacement Procedure Section 8-29 Alpha-Numeric (A/N) Keyboard Replacement Procedure The A/N Keyboard can be replaced without removal of the operator panel from the frame. The screws for removal are hidden under the “return” and left “shift” keys. 8-29-1 Manpower 8-29-2 Tools A key removal tool is required. Acquire a commerical key removal tool or fashion two loops made from paperclips as shown in Figure 8-123 on page 8-114 . 8-29-3 Preparations 1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3 8-29-4 A/N Keyboard Removal 1.) Place the loops under opposite corners of the key to be removed as shown in Figure 8- 124 on page 8-115 . To remove the AN keyboard, the “return” and left “shift” keys must be removed. 2.) Pull up with the tool but also slightly to the top edge (number row) of the keyboard. The slight pull towards the top forces the white plunger to come off with the key. If using paperclips, it works best if you use one hand on each clip and pull together. 3.) If the white plunger stays in the keyboard, you can reinstall the key and try again. Or, the plunger can be removed from the keyboard needle nose pliers grabbing between the same points as where the key inserts.Figure 8-123 Key Removal Tools
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-115 4.) After the key is removed, push the plunger into the keytop to make sure it is seated all the way.Figure 8-124 Removal Tool in Place Figure 8-125 Keytop and Plunger Assembly Keytop Plunger
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-116 Section 8-29 - Alpha-Numeric (A/N) Keyboard Replacement Procedure 8-29-4 A/N Keyboard Removal (cont’d) 5.) With the “return” and left “shift” keys are removed, you can see the screws that secure the keyboard to the plastic bezel. Loosen these screws so that they pop up but do not remove them as they are captive screws that remain with the AN keyboard. 6.) Using the key removal tool, lift up on a key to help pull the AN keyboard out of the plastic bezel. Choose a key to the far right and lift the right side of the keyboard out first. 7.) Lift out the bottom edge of the keyboard bar to gain access to the interconnect cables. Detach the 2 cables from the keyboard and lift the keyboard out.Figure 8-126 Keyboard Screw Locations Screw Screw
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-117 8-29-4 A/N Keyboard Removal (cont’d) 8.) Install the removed keys back onto the removed keyboard. If a key does not seem to seat down fully, it is probably because the plunger is not pressed onto the keytop all the way.Figure 8-127 Interconnect Cable Locations
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-118 Section 8-29 - Alpha-Numeric (A/N) Keyboard Replacement Procedure 8-29-5 A/N Keyboard Installation 1.) Using the key removal tool, remove the “return” and left “shift” keys from the replacement keyboard. use the same procedure as before. Make sure that the plungers come off with the keys and that they are fully seated into the keytop after removal. 2.) Check the replacement keyboards rubber gasket. When seated correctly, it should wrap around the edge of the circuit board. If the gasket is not seated correctly, the keyboard will be difficult to install. 3.) Connect the cables to the replacement keyboard. Check that the cables are not shifted by one row of pins as this is possible to do. 4.) Insert the keyboard left side first making sure the gasket remains seated around the circuit board.Figure 8-128 Keyboard Gasket
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-119 5.) Continue to work the keyboard down into the plastics slightly pushing to the left so that the gasket on the right side goes into the plastics without getting disturbed. 6.) Seat the keyboard fully by pushing down on the whole keyboard with both hands. 7.) Check the function of the keys around the keyboard edge. If keys stick or have a bad feel. the gasket probably got disturbed and is not around the edge of the circuit board. Remove the keyboard check the gasket and try again. 8.) If the keyboard is seated correctly, secure it with the screws. 9.) Reinstall the keytops. If a keytop seem too high, the white plunger was probably not seated all the way into the keytop. The keytop and plunger must be removed to fully seat the plunger into the keytop. 10.) Energize the system and type with the keyboard to check the function of each key.Figure 8-129 Replacing Keyboard
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-120 Section 8-30 - Trackball Assembly Replacement Procedure Section 8-30 Trackball Assembly Replacement Procedure 8-30-1 Manpower One person, 1 hour 30 minutes 8-30-2 Tools • Phillips screwdriver size 1 • Socket 2,5 mm 8-30-3 Preparations 1.) Shut down the system as described in Section 4-3-2 on page 4-3. 2.) Remove the Operator Panel as described in Section Section 8-27 on page 8-111. 3.) Put the panel upside down on a soft cushion. Figure 8-130 Trackball Trackball assembly Screws Trackball unitConnectors Trackball housing Screws Operator Panel with Trackball assembly
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-121 8-30-4 Trackball Removal Procedure 1.) Disconnect two (2) connectors at the trackball unit and 1 at the print between trackball unit and operator panel. Remove trackball assembly. 2.) Unscrew four (4) Phillips screws on the print. 3.) Unscrew two (2) screws/nuts and remove trackball unit from print. 4.) Unscrew two (2) Phillips screws on the trackball housing. 5.) Open trackball housing, and remove trackball. 8-30-5 Trackball Installation Procedure 1.) Install trackball in housing. Replace cover and fasten removed screws. 2.) Install the trackball unit on the print and fasten removed screws/nuts. 3.) Install trackball assembly on the Operator Panel and fasten 4 screws. 4.) Connect three (3) connectors. 5.) Install Operator Panel as described in Section Section 8-27 on page 8-111.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-122 Section 8-31 - Fan Assembly Replacement Procedure Section 8-31 Fan Assembly Replacement Procedure 8-31-1 Manpower One person, 1 hour 8-31-2 Tools Phillips screwdriver, size 2. 8-31-3 Preparations (click here to view the cover removal video) 1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3 2.) Remove the Right and Left Covers as described on Section 8-2-2 on page 8-5. 3.) Remove Upper Rear Cover and Filter as shown in Section 8-2-3 on page 8-7. 4.) Remove the Lower Rear Cover as shown in Section 8-2-6 on page 8-10. NOTE: There are washers and magnetic lock for Upper Rear Cover on the upper screws. 5.) Remove the Top Cover as shown in Section 8-2-4 on page 8-8. 6.) Remove the Rear Handle ashown in Section 8-2-8 on page 8-12. 7.) Remove the filter media. . 8-31-4 Fan Assembly Removal Procedure 1.) Unscrew four (4) Phillips screws that fasten Cover Air Duct. See Figure 8-132. 2.) Push down the lower side of Cover Air Duct to free lower ends of the Filter Frame. Remove Filter Frame downwards. 3.) Becarful of sharp edges. Move Cover Air Duct up in the front, then backwards, off the Fan Assembly and out of the System. Select the movie camera icon to view the video of the monitor replacement procedure. 6 minutes and 23 seconds Figure 8-131 Fan Assembly, location and hitch up Fan assembly External I/O Fan Cover Air Duct AC Controller Module Modem Fan Cable Hitch up