GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-53 7-12-1System Does Not Power On / Boot Up (cont’d) Figure 7-46 System does not Boot (contd.) Turning off the LOGIQ 9.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-54 Section 7-12 - Troubleshooting Trees Section 7-12Troubleshooting Trees (cont’d) Figure 7-47 BEP1 Connector Reference
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-55 Section 7-12Troubleshooting Trees (cont’d) Figure 7-48 Boot Sequence Illustrations
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-56 Section 7-12 - Troubleshooting Trees 7-12-3 Noise in Image Figure 7-49 Noise in Image Noise is shown in Image. YES NOIs there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices. Check probes.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-57 7-12-4 Unable to Scan Figure 7-50 Unable to Scan Unable to scan! Check the probe connection.YESNO Are probes connected in right way? Restart the scanner. Are other operations possible?YES Check the HV power supply unit. NO END
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-58 Section 7-12 - Troubleshooting Trees 7-12-5 Trackball Figure 7-51 Trackball Troubleshooting The trackball is low sensitive! Remove the ball from the Trackball, and clean it. YES NODoes the trackball move at all? Replace the trackball unit.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-59 7-12-6 Minimum Cable Configuration for System to Boot The following cables must be connected between the INT I/O board and the rest of the system: • Cables to Hard Disk (inside Back End Processor) • Patient I/O cable to Patient I/O box • I/O RS-232 serial cable to Front Panel • ACCTRL cable between INT I/O and ACCTRL (power-on signal) • DC Power cable to Front Panel • Cables to the monitor All other cables can be removed and the system will still boot. 7-12-7 Minimum Cable Configuration for System to Scan The following cables must be connected between the INT I/O board and the rest of the system: • Cables to Hard Disk (inside Back End Processor) • Patient I/O cable to Patient I/O box • I/O RS-232 serial cable to Front Panel • ACCTRL cable between INT I/O and ACCTRL (power-on signal) • DC Power cable to Front Panel • Cables to the monitor • PCI Cable from BEP to SCB
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-60 Section 7-12 - Troubleshooting Trees 7-12-8 System Does Not Power Off / Shutdown Figure 7-52 Power Off / Shutdown Flowchart Is the ON/OFF switch illuminated with a green color? Yes If the ON/OFF switch is illuminated with an amber color, the scanner is in stand-by mode which indicates that it has been Shut Off, but not completely Powered Down. If the ON/OFF switch is not illuminated at all, the scanner has been Shut Off and Powered Down. No Shutdown the scanner from a software screen or depress the ON/OFF switch. Did the scanner Shut Off? No Yes Switch the breaker to the OFF position to completely Power Down the scanner. Unplug the scanner from the wall outlet if it needs to be relocated. Depress and hold the ON/OFF switch on the scanner for approximately 10 seconds. If the scanner fails to Shut Off, follow the step below. Depress the reset button on the back of the scanner. The scanner will step through an ordered Shut Off sequence, ending with the scanner in standby mode. If the scanner still does not Shut Off, follow the two steps below. Switch the circuit breaker to the OFF position. A tone should begin to sound in one-second intervals. This indicates the Backup Power Supply in the Back End Processor is functioning. Depress the ON/OFF switch for approximately 4 seconds. The ON/OFF switch should illuminate with an amber color indicating the scanner is Shut Off and in a standby mode.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-61 7-12-9 Monitor Troubleshooting NOTE: Reset the monitor settings to the factory defaults prior to troubleshooting the monitor. NOTICEComments to remember about the 17” monitor. The monitor is the diagnostic monitor and is only connected to 115 VAC. The monitor powers down with the front end card rack. Power output from AC Control Box: When the monitor is not connected to the AC Control Box there is a leakage current which when measured with a DVM can indicate high voltage. When a load is applied, the voltage collapses to low voltage if the monitor is not switched on.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-62 Section 7-12 - Troubleshooting Trees 7-12-9Monitor Troubleshooting (cont’d) Figure 7-53 Error on the monitor display! Set the Monitor setting to the default.NOYES Do the colors seem abnormal? Does the Contrast or the Brightness adjustment seem improper? Adjust Contrast and Brightness. YESNO Does distortion occur on the monitor? Does the display area seems abnormal? NO Does the monitor display nothing? Replace the monitor. YES Check the PCVIC, and AGP video tape. YES