    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-53
    7-12-1System Does Not Power On / Boot Up (cont’d)
    Figure 7-46   System does not Boot (contd.)
    Turning off the LOGIQ 9. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-54 Section 7-12 - Troubleshooting Trees
    Section 7-12Troubleshooting Trees (cont’d)
    Figure 7-47  BEP1 Connector Reference 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-55
    Section 7-12Troubleshooting Trees (cont’d)
    Figure 7-48  Boot Sequence Illustrations 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-56 Section 7-12 - Troubleshooting Trees
    7-12-3 Noise in Image
    Figure 7-49   Noise in Image
    Noise is shown in Image.
    YES
    NOIs there any electrical 
    devices that affect the 
    Scanner?
    Keep the Scanner away from those 
    electrical devices.
    Check probes. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-57
    7-12-4 Unable to Scan
    Figure 7-50   Unable to Scan
    Unable to scan!
    Check the probe connection.YESNO
    Are probes connected in 
    right way?
    Restart the scanner.
    Are other operations 
    possible?YES
    Check the HV power supply unit. NO
    END 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-58 Section 7-12 - Troubleshooting Trees
    7-12-5 Trackball
    Figure 7-51   Trackball Troubleshooting
    The trackball is low sensitive!
    Remove the ball from the Trackball, 
    and clean it.
    YES
    NODoes the trackball move at 
    all?
    Replace the trackball unit. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-59
    7-12-6 Minimum Cable Configuration for System to Boot
    The following cables must be connected between the INT I/O board and the rest of the system:
    • Cables to Hard Disk (inside Back End Processor)
    • Patient I/O cable to Patient I/O box
    • I/O RS-232 serial cable to Front Panel
    • ACCTRL cable between INT I/O and ACCTRL (power-on signal)
    • DC Power cable to Front Panel
    • Cables to the monitor
    All other cables can be removed and the system will still boot.
    7-12-7 Minimum Cable Configuration for System to Scan
    The following cables must be connected between the INT I/O board and the rest of the system:
    • Cables to Hard Disk (inside Back End Processor)
    • Patient I/O cable to Patient I/O box
    • I/O RS-232 serial cable to Front Panel
    • ACCTRL cable between INT I/O and ACCTRL (power-on signal)
    • DC Power cable to Front Panel
    • Cables to the monitor
    • PCI Cable from BEP to SCB 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-60 Section 7-12 - Troubleshooting Trees
    7-12-8 System Does Not Power Off / Shutdown
    Figure 7-52   Power Off / Shutdown Flowchart
    Is the ON/OFF switch
    illuminated with a
    green color?
    Yes
    If the ON/OFF switch is illuminated with an amber color, the scanner is in
    stand-by mode which indicates that it has been Shut Off, but not completely
    Powered Down.
    If the ON/OFF switch is not illuminated at all, the scanner has been Shut Off
    and Powered Down.
    No
    Shutdown the scanner from a software
    screen or depress the ON/OFF switch.
    Did the scanner
    Shut Off?
    No
    Yes
    Switch the breaker to the OFF position
    to completely Power Down the scanner.
    Unplug the scanner from the wall outlet
    if it needs to be relocated.
    Depress and hold the ON/OFF switch on the scanner for approximately 10
    seconds.
    If the scanner fails to Shut Off, follow the step below.
    Depress the reset button on the back of the scanner. The scanner will step
    through an ordered Shut Off sequence, ending with the scanner in standby
    mode.
    If the scanner still does not Shut Off, follow the two steps below.
    Switch the circuit breaker to the OFF position. A tone should begin to sound in
    one-second intervals. This indicates the Backup Power Supply in the Back End
    Processor is functioning.
    Depress the ON/OFF switch for approximately 4 seconds. The ON/OFF switch
    should illuminate with an amber color indicating the scanner is Shut Off and in
    a standby mode. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-61
    7-12-9 Monitor Troubleshooting
    NOTE: Reset the monitor settings to the factory defaults prior to troubleshooting the monitor.
    NOTICEComments to remember about the 17” monitor.
    The monitor is the diagnostic monitor and is only connected to 115 VAC.
    The monitor powers down with the front end card rack.
    Power output from AC Control Box:
    When the monitor is not connected to the AC Control Box there is a leakage current 
    which when measured with a DVM can indicate high voltage.
    When a load is applied, the voltage collapses to low voltage if the monitor is not 
    switched on. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-62 Section 7-12 - Troubleshooting Trees
    7-12-9Monitor Troubleshooting (cont’d)
    Figure 7-53   
    Error on the monitor display!
    Set the Monitor setting to the default.NOYES
    Do the colors seem 
    abnormal?
    Does the Contrast or the 
    Brightness adjustment 
    seem improper?
    Adjust Contrast and Brightness. YESNO
    Does distortion occur on 
    the monitor? Does the 
    display area seems 
    abnormal?
    NO
    Does the monitor display 
    nothing?
    Replace the monitor. YES
    Check the PCVIC, and AGP video 
    tape. YES 
    						