GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-173 8-48-5 Black and White Video Printer Installation Procedure 1.) Replace printer and connect cables. 2.) Fasten the strap to secure the printer. 3.) Install Left Side Cover as described in Side Covers (with Bumper) Replacement Procedure on page 8-5. (click here to view the cover installation video) 4.) Boot up and test the system. 5.) Ensure that the print keys still function as they did before. 8-48-6 Differences in Replacing the Digital B/W Printer The Sony UPD-895 has only Power and USB connections (BEP USB#6 - C21) on the back panel.Figure 8-179 The B&W Video Printer location at the Front of the Scanner Figure 8-180 Sony UP-D895 Digital B/W Printer Disconnect cables at rear side of printer Fastening belt
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-174 Section 8-49 - VCR Replacement Procedure Section 8-49 VCR Replacement Procedure 8-49-1 Manpower One person, 0.5 hours + travel. 8-49-2 Tools Long Phillips driver. 8-49-3 Preparations(click here to view the cover removal video) 1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3. 2.) Remove the Left and Right Side Covers as shown on Section 8-2-2 on page 8-5. 3.) Raise the Control Console to its upper position and as far forwards as possible. 4.) Remove the Air Filter Cover and Top Cover as described on Section 8-2-3 on page 8-7. 5.) Disconnect probes and remove Front Cover as described on Section 8-2-5 on page 8-9. 8-49-4 Removal Procedure 1.) Loosen the fastening belts for the VCR and disconnect all the cables from the VCR. 2.) Remove the VCR from the frame. 8-49-5 VCR Mounting Procedure 1.) Place the VCR in the scanners frame and connect all the cables to the rear of the VCR. 2.) Fasten and tighten the straps to secure the VCR. 3.) Boot Up and test the system. 4.) Replace the covers (Top Cover, Front Cover and Side Covers). (click here to view the cover installation video) Click on the movie camear icon to view the video illustrating the removal and replacement of the VCR. 2 minutes and 9 seconds Figure 8-181 The VCR’s Placement in the Scanner Fastening belt Disconnect cables at rear side of VCR
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-175 Section 8-50 Wireless Voice Scan Option 8-50-1 Charging Batteries Unpack the battery charger kit and inspect the parts. NOTE: Initially, batteries take several hours to charge. A working battery is required to complete the installation. If you did not start the batteries charging earlier. Do It Now. 8-50-1-1 Battery Charger and Batteries 1.) Locate the battery charger and 9 Volt rechargeable batteries. The battery charger model may vary. 2.) Remove the cellophane wrap from the batteries if present.Figure 8-182 9V Batteries and Charger
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-176 Section 8-50 - Wireless Voice Scan Option 8-50-1-1Battery Charger and Batteries (cont’d) 3.) Unwrap the battery charger power adapter. The adapter for North American sites (120Vac only) comes with the charger. An AC Adapter for international sites (100-240 Vac) is packaged separately. Also included is a set of AC plug adapters. Plug the power adapter into the battery charger and AC outlet. Plug a 9 Volt battery into the charger. The first time new batteries are charged, it may take approximately 200 minutes to reach a full charge. Recharging batteries will take from 75 to 120 minutes. NOTE: A partially charged battery may be used to test the system, but available run time will be reduced.Figure 8-183 Power Adapter and Charger with Batteries WARNINGWARNINGDo Not attempt to charge batteries other than the type included in the kit (NiMH). Standard alkaline batteries CANNOT be recharged. READ all safey precautions that come from the battery and battery charger manufacturers.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-177 8-50-2 Microphone/Transmitter Assembly 8-50-2-1 Assemble Transmitter 1.) Screw the antenna into the antenna jack on the transmitter. 2.) Put a fully Charged Battery into the Transmitter a.) Open the battery door by sliding b.) To remove a discharged battery , Hold the transmitter with one hand so that your thumb or finge can catch the battery. c.) User the other hand to pull on the red lever and let gravity pull the battery out and fall to your thumb. d.) Push the new battery in, make sure it is installed with the correct orientation of + and - e.) Close the door and slide to lock.Figure 8-184 Transmitter and Microphone Figure 8-185 Changing the Battery Transmitter with antenna
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-178 Section 8-50 - Wireless Voice Scan Option 8-50-2-2 Assemble Microphone 1.) Locate the transmitter and microphone with accessories. 2.) Unpack the microphone. Figure 8-186 Transmitter and Microphone Figure 8-187 Microphone and Accessories Microphone with accessories Microphone Microphone Clip and Wind Screen
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-179 3.) With the windscreen off, note the taper from the microphone wire to the microphone element. 4.) When putting on the wind screen, slightly twirl the foam in your fingers while threading it onto the microphone element. Once the element has entered the foam, you must push further to fit the plastic cap that is inside the foam onto the microphone element. This takes a little force and rocking motion. 5.) When fitted correctly, the taper should be totally covered by the windscreen. If you see the taper as in the above picture, the cap inside the foam is not pushed on all the way and the wind screen will easily fall off during use.Figure 8-188 Microphone with Wind Screen Figure 8-189 Tarnsmitter Microphone Figure 8-190 Improper Wind Screen Placement MicrophoneWind Screen (May already be installed) Microphone Element Wire Taper Taper still visible – windscreen not fully installed!!
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-180 Section 8-50 - Wireless Voice Scan Option 6.) Plug the microphone assembly into the transmitter and secure it by tightening the outer collar. Refer to Figure 8-192 on page 8-180 . 7.) Use the transmitter clip to place it on your belt or clothing. Place the microphone on your ear and position. 8.) Attach the cord clip to the microphone cord and attach the clip to your clothing to avoid changing the microphone postion during scanning.Figure 8-191 Proper Wind Screen Placement Figure 8-192 Plug Microphone into Transmitter Figure 8-193 Microphone User Instructions Taper fully covered – windscreen correctly installed
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-181 Figure 8-194 Cord Retention Clip Place the cord firmly into the clip
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-182 Section 8-50 - Wireless Voice Scan Option 8-50-3 Receiver Setup Remove Side Covers, Filter Cover and Top Cover if necessary. You will see the receiver and its associated cables nested in a tray with foam. 1.) Lift off the top foam so you can turn the receiver and look at its control panel. 2.) Install the Service Dongle and turn on the LOGIQ™ 9 .Figure 8-195 Voice Scan Receiver Tray Assembly