    Chapter 4 Functional Checks 4-15
    4-3-11 Backup and Restore Database, Preset Configurations and Images
    4-3-11-1 Formatting Media
    1.) To format the backup media, CD-ROM or MOD, select the UTILITY
     tab on the Touch Panel.
    2.) Select CONNECTIVITY
    , then TOOLS. Properly label and Insert the backup media. 
    3.) Select the media type from the drop down menu.
    4.) Enter the label for the media as shown in Figure 4-11. It is best to use all capital letters with no 
    spaces or punctuation marks.
    5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-12. When the formatting 
    has been completed, press OK
     to continue.
    6.) If desired, verify that the format was successful by returning to Utility>Connectivity>Tools and 
    selecting VERIFY
     as shown in Figure 4-11.  Figure 4-11  Format and Verify Media
    Figure 4-12  Format Successful Pop-up Menu
    Verify Format
    Format
    Media Selection
    Media Label 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    4-16 Section 4-3 - General Procedure
    4-3-11-2 Backup System Presets and Configurations
    NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need 
    to be reloaded, after the software update, they will be the same ones the customer was using prior to 
    service.
    1.) Insert a formatted MOD (Magneto-Optical Disk) or CD-R into the drive.
    2.) On the Touch Panel, press UTILITY
    .
    3.) On the Utility Touch Panel, press SYSTEM
    .
    4.) On the monitor display, select BACKUP/RESTORE
    .
    NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login 
    window is displayed. Log on with administrator privileges.
    5.) In the Backup list, select Patient Archive, Report Archive, System Configuration and User Defined 
    Configuration.
    6.) In the Media field, select MO or CD-RW.
    7.) Select BACKUP
    .
    The system performs the backup. As it proceeds, status information is displayed on the Backup/Restore 
    screen.
    Figure 4-13  Backup/Restore Menu
    Check here to 
    backup presets 
    and configurations 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 4 Functional Checks 4-17
    4-3-11-3 Restore System Presets and Configurations
    1.) Insert the Backup/Restore MOD (Magneto-Optical Disk) or CD-R into the drive. 
    2.) On the Touch Panel, press UTILITY
    .
    3.) On the Utility Touch Panel, press SYSTEM
    .
    4.) On the monitor display, select BACKUP/RESTORE
    .
    NOTE: If you are not logged in with administrator privileges, the Operator Login window is displayed. 
    Log on with administrator privileges.
    5.) In the Restore list, select Patient Archive, Report Archive, System Configuration and User Defined 
    Configuration.
    6.) In the Media field, select the Backup/Restore MOD or CD-RW.
    7.) Select RESTORE
    .
    The system performs the restore. As it proceeds, status information is displayed on the Backup/Restore 
    screen.
    CAUTIONThe restore procedure overwrites the existing database on the local hard drive. Make sure to insert 
    the correct MOD or CD-R.
    Figure 4-14  Backup/Restore Menu
    Check here to 
    restore presets 
    and configurations 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    4-18 Section 4-3 - General Procedure
    4-3-11-4 Archiving Images
    1.) Insert the archive media. To format the archive media, CD-ROM or MOD, select the Utility tab 
    on the Touch Panel. 
    2.) Select Connectivity, then Tools. 
    3.) Format the MOD/CD-ROM. Verify the format if desired.
    4.) Images will be moved from the hard drive by date. Therefore, the best way is to label media 
    by date. 
    NOTE: Images will be moved from the hard drive by date. Therefore, the best way to label media is by 
    date. When images are moved to the archive media, they will be deleted from the system hard 
    drive. However, the patient database (backed up earlier) maintains pointers to the location of 
    the images on the archive media.
    5.) Press PATIENT
     and set the Dataflow to store images directly to MOD or CD-ROM.
    6.) From the image screen, press MORE
    , then select Move Images. The Move Images pop-up 
    appears.Figure 4-15  Format MOD/CD-ROM Screen
    Verify Format
    Format
    Media Selection
    Media Label 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 4 Functional Checks 4-19
    4-3-11-4Archiving Images (cont’d)
    7.) Fill in the From Date and To Date. Specify to Keep days together.
    8.) Press RECALCULATE
    .
    9.) Press OK. An in-progress message appears. The archive operation is complete when you 
    receive the message shown in Figure 4-17.
    10.) Repeat the image move steps until all images have been archived.
    All databases, presets and images should now be saved to removable media.Figure 4-16  Image Archive Move Pop-up Menu
    Figure 4-17  Archive Operation Complete Message 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    4-20 Section 4-4 - Software Configuration Checks
    4-3-12 ECG Check Out (this option is currently not available on the LOGIQ™ 9
    Connect the ECG Harness and check:
    Section 4-4
    Software Configuration Checks
    Table 4-5     ECG Harness Check
    StepTask to doExpected Result(s)
    1.Connect the ECG at the Connector on the Front of the 
    scanner.It will display a curve along the bottom edge of the image 
    sector
    Table 4-6     Software Configuration Checks
    StepTask to doExpected Result(s)
    1.
    Check Date and Time setting Date and Time are correct
    2.
    Check that Location (Hospital Name) is correct Location Name is correct
    3.
    Check Language settings Desired Language is displayed
    4.Check assignment of Printer KeysPrint1-4 Keys are assigned as desired by the customer
    5.Check that all of the customer’s options are set up correctAll authorized functions are enabled 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 4 Functional Checks 4-21
    Section 4-5
    Peripheral Checks
    Check that peripherals work as described below:
    Table 4-7    Peripheral Checks
    StepTask to do Expected Result(s)
    1.
    Press (FREEZE) Stop image acquisition.
    2.Press (PRINT 1
    ) or (PRINT 2) on the Control 
    Panel The image displayed on the screen is printed on B&W or Color printer, 
    depending on the key assignment configuration
    3.
    Press VIDEO
     on the Control Panel. To start the video counter at a different point:
    4.
    Press UPDATE MENU
     in the Trackball area. The Video Set Counter / Search Dialog window is displayed
    5.Use the alphanumeric keyboard to enter 
    the counter number in the counter field.Number appears
    6.
    Press SET COUNTER
     to save the change. Number is saved
    7.
    Press VIDEO
     on the Control Panel to return to the scanning mode
    8.
    Press REC/PAUSE
     on the Control panel.to Start Recording
    A red dot is displayed in the VCR status area on the Title bar to indicate 
    that recording has begun
    9.
    Press REC/PAUSE
     on the Control panel.To Stop recording
    The video status icon is changed to (Pause)
    10.Press VIDEO
     on the Control Panel and the 
    assignable playTo start, Play back an examination
    11.Use the Assignable keys on the Control 
    Panelto perform actions on the recorded session, such as stop, pause, rewind 
    or fast forward.
    The video status icon in updated accordingly.
    12.While in playback mode, use the 
    TRACKBALL
     to adjust the video playback 
    speed and scroll through the record.To search on the tape
    13.
    Press the assignable PAUSE
    to stop the tape at the desired frame.
    14.
    When playing back an examinationpart of it can be stored on the computer’s memory as a cineloop.The 
    cineloop enables the user to perform further operations on the stored 
    section (see for further information on cineloop operation).
    15.Press (FREEZE
    ) while playing back a 
    recorded session.To store a recorded sequence as a cineloop. The last few seconds are 
    stored as a cineloop. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    4-22 Section 4-6 - Mechanical Function Checks
    Section 4-6
    Mechanical Function Checks 
    4-6-1 Operator I/O Movement
    Figure 4-18  
    StepTask Expected Result(s)
    1. Pull handle up under the platform towards you to 
    release horizontal lockIt is possible to maneuver the Display Platform 
    horizontally at present level of height
    2.
    Pull handle at left hand grip towards you to release 
    the Vertical LockIt is possible to maneuver the Display Platform up and 
    down (between the vertical end stops) as long as you 
    pull the handle towards you
    3.
    Pull both handles towards you to release both 
    Vertical and Horizontal locksIt is possible to maneuver the Display Platform both 
    horizontally and vertically as long as you keep pulling 
    these handles towards you
    Table 4-8     Display Platform Maneuverability Checks 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 4 Functional Checks 4-23
    4-6-2 Brakes and Direction Locks Checks
    Check that the brakes and direction locks function as described below.
    NOTE: The swivel lock and brake cannot be engaged (pressed down) at the same time since pressing the main 
    brake releases the swivel lock and pressing the swivel lock releases the main brake. Therefore, if the 
    swivel lock is engaged you need to press the main brake pedal (#2) twice to lock the scanner from 
    rolling. Once will release the swivel lock and once again will engage the main brake in full lock as stated 
    in step 1. Table 4-9     Brakes and Direction Lock Instructions
    StepTaskExpected Result(s)If Not, Remedy
    1. Press on pedal no. 2  
    To engage the brakes in full lockTest Rods Bolts and nuts and wheel 
    rubber as described in Brake Adjustment 
    on page 6-14.
    2. Press on pedal no.1  
    To release the brake
    3. Press on pedal no.1  
    To engage swivel lock.
    You may need to move the system around a 
    little to get the wheels to lock into the locked 
    direction.
    4.Press on pedal no. 2  
    To release swivel lock 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    4-24 Section 4-7 - Site Log
    Section 4-7
    Site Log
    Table 4-10    Site Log 
    DateService personProblemComments 
    						