GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-15 4-3-11 Backup and Restore Database, Preset Configurations and Images 4-3-11-1 Formatting Media 1.) To format the backup media, CD-ROM or MOD, select the UTILITY tab on the Touch Panel. 2.) Select CONNECTIVITY , then TOOLS. Properly label and Insert the backup media. 3.) Select the media type from the drop down menu. 4.) Enter the label for the media as shown in Figure 4-11. It is best to use all capital letters with no spaces or punctuation marks. 5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-12. When the formatting has been completed, press OK to continue. 6.) If desired, verify that the format was successful by returning to Utility>Connectivity>Tools and selecting VERIFY as shown in Figure 4-11. Figure 4-11 Format and Verify Media Figure 4-12 Format Successful Pop-up Menu Verify Format Format Media Selection Media Label
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-16 Section 4-3 - General Procedure 4-3-11-2 Backup System Presets and Configurations NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need to be reloaded, after the software update, they will be the same ones the customer was using prior to service. 1.) Insert a formatted MOD (Magneto-Optical Disk) or CD-R into the drive. 2.) On the Touch Panel, press UTILITY . 3.) On the Utility Touch Panel, press SYSTEM . 4.) On the monitor display, select BACKUP/RESTORE . NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login window is displayed. Log on with administrator privileges. 5.) In the Backup list, select Patient Archive, Report Archive, System Configuration and User Defined Configuration. 6.) In the Media field, select MO or CD-RW. 7.) Select BACKUP . The system performs the backup. As it proceeds, status information is displayed on the Backup/Restore screen. Figure 4-13 Backup/Restore Menu Check here to backup presets and configurations
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-17 4-3-11-3 Restore System Presets and Configurations 1.) Insert the Backup/Restore MOD (Magneto-Optical Disk) or CD-R into the drive. 2.) On the Touch Panel, press UTILITY . 3.) On the Utility Touch Panel, press SYSTEM . 4.) On the monitor display, select BACKUP/RESTORE . NOTE: If you are not logged in with administrator privileges, the Operator Login window is displayed. Log on with administrator privileges. 5.) In the Restore list, select Patient Archive, Report Archive, System Configuration and User Defined Configuration. 6.) In the Media field, select the Backup/Restore MOD or CD-RW. 7.) Select RESTORE . The system performs the restore. As it proceeds, status information is displayed on the Backup/Restore screen. CAUTIONThe restore procedure overwrites the existing database on the local hard drive. Make sure to insert the correct MOD or CD-R. Figure 4-14 Backup/Restore Menu Check here to restore presets and configurations
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-18 Section 4-3 - General Procedure 4-3-11-4 Archiving Images 1.) Insert the archive media. To format the archive media, CD-ROM or MOD, select the Utility tab on the Touch Panel. 2.) Select Connectivity, then Tools. 3.) Format the MOD/CD-ROM. Verify the format if desired. 4.) Images will be moved from the hard drive by date. Therefore, the best way is to label media by date. NOTE: Images will be moved from the hard drive by date. Therefore, the best way to label media is by date. When images are moved to the archive media, they will be deleted from the system hard drive. However, the patient database (backed up earlier) maintains pointers to the location of the images on the archive media. 5.) Press PATIENT and set the Dataflow to store images directly to MOD or CD-ROM. 6.) From the image screen, press MORE , then select Move Images. The Move Images pop-up appears.Figure 4-15 Format MOD/CD-ROM Screen Verify Format Format Media Selection Media Label
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-19 4-3-11-4Archiving Images (cont’d) 7.) Fill in the From Date and To Date. Specify to Keep days together. 8.) Press RECALCULATE . 9.) Press OK. An in-progress message appears. The archive operation is complete when you receive the message shown in Figure 4-17. 10.) Repeat the image move steps until all images have been archived. All databases, presets and images should now be saved to removable media.Figure 4-16 Image Archive Move Pop-up Menu Figure 4-17 Archive Operation Complete Message
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-20 Section 4-4 - Software Configuration Checks 4-3-12 ECG Check Out (this option is currently not available on the LOGIQ™ 9 Connect the ECG Harness and check: Section 4-4 Software Configuration Checks Table 4-5 ECG Harness Check StepTask to doExpected Result(s) 1.Connect the ECG at the Connector on the Front of the scanner.It will display a curve along the bottom edge of the image sector Table 4-6 Software Configuration Checks StepTask to doExpected Result(s) 1. Check Date and Time setting Date and Time are correct 2. Check that Location (Hospital Name) is correct Location Name is correct 3. Check Language settings Desired Language is displayed 4.Check assignment of Printer KeysPrint1-4 Keys are assigned as desired by the customer 5.Check that all of the customer’s options are set up correctAll authorized functions are enabled
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-21 Section 4-5 Peripheral Checks Check that peripherals work as described below: Table 4-7 Peripheral Checks StepTask to do Expected Result(s) 1. Press (FREEZE) Stop image acquisition. 2.Press (PRINT 1 ) or (PRINT 2) on the Control Panel The image displayed on the screen is printed on B&W or Color printer, depending on the key assignment configuration 3. Press VIDEO on the Control Panel. To start the video counter at a different point: 4. Press UPDATE MENU in the Trackball area. The Video Set Counter / Search Dialog window is displayed 5.Use the alphanumeric keyboard to enter the counter number in the counter field.Number appears 6. Press SET COUNTER to save the change. Number is saved 7. Press VIDEO on the Control Panel to return to the scanning mode 8. Press REC/PAUSE on the Control panel.to Start Recording A red dot is displayed in the VCR status area on the Title bar to indicate that recording has begun 9. Press REC/PAUSE on the Control panel.To Stop recording The video status icon is changed to (Pause) 10.Press VIDEO on the Control Panel and the assignable playTo start, Play back an examination 11.Use the Assignable keys on the Control Panelto perform actions on the recorded session, such as stop, pause, rewind or fast forward. The video status icon in updated accordingly. 12.While in playback mode, use the TRACKBALL to adjust the video playback speed and scroll through the record.To search on the tape 13. Press the assignable PAUSE to stop the tape at the desired frame. 14. When playing back an examinationpart of it can be stored on the computer’s memory as a cineloop.The cineloop enables the user to perform further operations on the stored section (see for further information on cineloop operation). 15.Press (FREEZE ) while playing back a recorded session.To store a recorded sequence as a cineloop. The last few seconds are stored as a cineloop.
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-22 Section 4-6 - Mechanical Function Checks Section 4-6 Mechanical Function Checks 4-6-1 Operator I/O Movement Figure 4-18 StepTask Expected Result(s) 1. Pull handle up under the platform towards you to release horizontal lockIt is possible to maneuver the Display Platform horizontally at present level of height 2. Pull handle at left hand grip towards you to release the Vertical LockIt is possible to maneuver the Display Platform up and down (between the vertical end stops) as long as you pull the handle towards you 3. Pull both handles towards you to release both Vertical and Horizontal locksIt is possible to maneuver the Display Platform both horizontally and vertically as long as you keep pulling these handles towards you Table 4-8 Display Platform Maneuverability Checks
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-23 4-6-2 Brakes and Direction Locks Checks Check that the brakes and direction locks function as described below. NOTE: The swivel lock and brake cannot be engaged (pressed down) at the same time since pressing the main brake releases the swivel lock and pressing the swivel lock releases the main brake. Therefore, if the swivel lock is engaged you need to press the main brake pedal (#2) twice to lock the scanner from rolling. Once will release the swivel lock and once again will engage the main brake in full lock as stated in step 1. Table 4-9 Brakes and Direction Lock Instructions StepTaskExpected Result(s)If Not, Remedy 1. Press on pedal no. 2 To engage the brakes in full lockTest Rods Bolts and nuts and wheel rubber as described in Brake Adjustment on page 6-14. 2. Press on pedal no.1 To release the brake 3. Press on pedal no.1 To engage swivel lock. You may need to move the system around a little to get the wheels to lock into the locked direction. 4.Press on pedal no. 2 To release swivel lock
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-24 Section 4-7 - Site Log Section 4-7 Site Log Table 4-10 Site Log DateService personProblemComments