GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-63 Section 8-12 Loading Base Image Software (cont’d) Once you have chosen to continue with Option 1 or Option 2, you will see a screen showing Norton Ghost. This automatically loads a base set of software onto the hard disk. Let it run to completion (Approximately 8-10 mins). A message, as shown in Figure 8-69, will be displayed when the process is complete. NOTE: IMPORTANT ... With the clinical version of Base Image Software you will see an error message “Not Ready Error Reading from Drive A” Ignore this message and continue. 4.) Remove the CD-ROM NOTE: Remove the CD-ROM from the drive; otherwise you will be repeating the Base System Software Load process when the system boots up. 5.) After removing the CDROM from the drive, properly turn off the scanner by holding down the Standby Switch until the light turns amber. 6.) Turn the scanner back on. It will now boot up and automatically log on to start checking hardware and mapping disk drives. This is a normal and should be allowed to run to completion. NOTE: While the script is running, several windows or dialog boxes will appear on the screen. At times it may look like the system is unresponsive. Wait for the “System Settings Change”dialog box “Do You Want to Restart computer now” to appear, approximately 3-4 minutes after the desktop is displayed. NOTE: Prior to the System Settings Change dialog box, Do NOT touch the system during this process. Activating the keyboard, mouse or front panel could corrupt the installation. 7.) When you get a message to restart Windows, select YES . 8.) Allow the system to reboot. 9.) The system will come up into the Windows mode.Figure 8-69 Base Load Complete
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-64 Section 8-12 - Loading Base Image Software Section 8-12 Loading Base Image Software (cont’d) 8-12-1 System Checks 1.) From the Windows mode display, double-click (RIGHT SET key) on the My Computer icon. 2.) Check the drive letter assignments for the Compact Disk and Removable Disk. They should be Compact Disk (G) and Removable Disk (H). 3.) If the Drive letter assignments are properly mapped, proceed to loading applications software. If the assignments are reversed, proceed to remap the drives manually. 4.) If the customer has a USB (Digital Priters) to be install, plug it in now. The system is still in Windows mode and should recognize the printer and install the necessary driver. If you do not install the printer until after the Applications software has been loaded, you will need a service dongle to get into windows and manually install the printers. Print a test page to check the printer installation. NOTE: If the printer is installed while in application mode, excessive CPU time will be consumed and the system will become very slow to unusable. Shutting down and rebooting will clear the condition. The printer will function correctly. 5.) Check your Service Dongle. Plug in your Service Dongle to ensure that it is recognized and lights up.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-65 Section 8-13 Loading Application Software 8-13-1 Loading Applications Software NOTE: If you have not done so previously, double-click on the My Computer icon. Check for the drive letters assignments of the Removable Drive (MOD) and the CD-ROM Drive. If the CD-ROM Drive is not drive “g” and the MOD drive “h” , manually remap the drives. NOTE: When installing R2.1.1 software you may see error messages about not being able to replace the following *.dll files: ginastub.dll and msvcr70.dll. These are known, non-critical issues that are being tracked and will be fixed in the next revision of software. 1.) Place the “Application Software” CDROM into the CDROM drive. 2.) Select the START button on the tool bar at the bottom of the LCD screen and select “Run…”. 3.) Enter “g:\LoadSoftware” into the dialog box as shown in Figure 8-70. 4.) You will see a screen that warns you that you are about to load software, press the Enter key or select OK to start the process. 5.) You will see a Command (CMD) window open as shown in Figure 8-71. Figure 8-70 Run Load Software
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-66 Section 8-13 - Loading Application Software 8-13-1 Loading Applications Software (cont’d) 6.) As indicated, press any key to continue. The process is completely automated and provides you with messages as to its status. R2.1.1 Software provides the message in Figure 8-72. At this point you just have to wait while the software loads (Approximately 7-8 minutes). NOTE: When the Information Window appears asking you to select OK, do nothing. This window will disappear automatically.Figure 8-71 Command Window Figure 8-72 R2.1.1 Software Message
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-67 8-13-1 Loading Applications Software (cont’d) 7.) When the process has completed, the window will turn dark red and you will see the following message in Figure 8-73. 8.) When it completes loading the system will reboot. If possible while the system is rebooting remove the CD from the drive. If the CD is remove the system will automatically reboot into the applications software. If you didn’t remove the CD, don’t worry you will get a dialog box as in Figure 8-74. 9.) REMOVE the Applications CD 10.) From the Start Applications dialog box, select START . 11.) Select CANCEL . Figure 8-73 Software Load Complete Figure 8-74 Start Software
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-68 Section 8-13 - Loading Application Software 12.) During the reboot process, the system may ask for the option string (valid license) to be entered. When this happens, enter the option string provided in the installation paperwork. NOTE: Starting with R3.x Software, all systems must have at least the basic option string in order for the software to run. Any single option string may enable multiple options (i.e. Basic & B-Flow). Demo option strings turn on all options for a limited period of time. Do Not leave a demo option string in a unit when it is sold to a customer. The current R3.x Demo Option String expires 31 January, 2005.Figure 8-75 Enter Option String
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-69 Section 8-14 Upgrading Application Software 8-14-1 Applications Software Upgrade Procedure NOTE: When installing software you may see error messages about not being able to replace the following *.dll files: ginastub.dll and msvcr70.dll. These are known, non-critical issues that are being tracked and will be fixed in the next revision of software. 1.) Place the “Application Software” CD-ROM into the CD-ROM drive. 2.) Power Down the scanner. 3.) Power up the scanner. Wait for a dialog box like the one in Figure 8-76. WARNINGWARNINGWhile the software install procedure is designed to preserve data, you should save any patient data, images, system setups to a MOD or hardcopy. CAUTIONRemember to remove the MOD before continuing with the software install. Figure 8-76 Start Application Window
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-70 Section 8-14 - Upgrading Application Software 8-14-1 Applications Software Upgrade Procedure (cont’d) 4.) Select INSTALL SW ... . 5.) There will be two dialog boxes popping up warning you that you are about to install new software as shown in Figure 8-77. In both cases click OK . 6.) Another window will be displayed asking you to “Press any key to continue…”.Figure 8-77 StartLoader Dialog Boxes Figure 8-78 Command Window
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-71 8-14-1 Applications Software Upgrade Procedure (cont’d) 7.) As indicated, press any key to continue. The process is completely automated and provides you with messages as to its status. R2.1.0 Software provides the message in Figure 8-79. At this point you just have to wait while the software loads (Approximately 10 minutes). When the process has completed, the window will turn dark red and you will see the following message in Figure 8-80. 8.) When it completes loading the system will reboot. 9.) Remove the CD from the drive If possible, while the system is rebooting remove the CD from the drive. If you didn’t do that don’t worry you will get a dialog box like the one shown in Figure 8-81. 10.) REMOVE the Applications CD 11.) From the Start Applications dialog box, select START . 12.) Select CANCEL . Figure 8-79 R2.1.0 Software Message Figure 8-80 Software Load Complete Figure 8-81 Start Software
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-72 Section 8-14 - Upgrading Application Software 8-14-1 Applications Software Upgrade Procedure (cont’d) 13.) During the reboot process, the system may ask for the option string (valid license) to be entered. When this happens, enter the option string provided in the installation paperwork. NOTE: Starting with R3.x Software, all systems must have at least the basic option string in order for the software to run. Any single option string may enable multiple options (i.e. Basic & B-Flow). Demo option strings turn on all options for a limited period of time. Do Not leave a demo option string in a unit when it is sold to a customer. The current R3.x Demo Option String expires 31 January, 2005.Figure 8-82 Enter Option String