GE Logiq 9 Service Manual
Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-43 8-7-4 MOD Drive Removal Procedure (cont’d) NOTE: Mounting hardware may vary in size from M3 to 4-40 threads. Hardware length may seem the same but in fact may be different and very critical. ALWAYS keep track of EXACTLY where each screw was removed so that it is returned to the same place on re-assembly. Figure 8-46 MOD Case without MOD
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-44 Section 8-7 - BEP MOD Replacement Procedure (BEP 2.x and above only) 8-7-5 MOD Drive Installation Procedure Now reverse the removal procedure to install the new MOD. It cannot be stressed enough to use the EXACT same screws to mount the new drive in the SAME position as they were removed. If wrong screw, that is only slightly too long, is used to mount the MOD to the case the MOD will NOT function. 1.) Refer to Figure 8-45 on page 8-42 and Figure 8-46 on page 8-43 , go through the side of the MOD case to install the four screws mounting the MOD to the case. 2.) Attach the power and IDE Cables to the MOD. Refer to Figure 8-45 on page 8-42 . 3.) Install the four screws, two on each side, to attach the top cover and side rails from the MOD case. NOTE: The side rails may need to be adjusted slightly so the MOD face is flush to the front of the BEP when it is installed in the chassis. 4.) Slide the MOD into the front of the BEP as shown in Figure 8-42 on page 8-41 . 5.) Install the two screws in the front of the MOD drive as shown in Figure 8-42 on page 8-41 . 6.) Place the adhesive trim back on the BEP front face as shown in Figure 8-41 on page 8-40 . 7.) Install the one screw securing the side of the MOD to the rack as shown in Figure 8-40 on page 8-40 . 8.) Plug in the power connectors previously removed. Plug the IDE Cable into the CD-RW Drive and the Motherboard. See Figure 8-43 on page 8-41 for reference. Pay attention to the orientation of the guide pin on the IDE Cable connector and the missing pin on the IDE connection.Figure 8-47 IDE Ribbon Connector IDE Guide Pin
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-45 8-7-5 MOD Drive Installation Procedure (cont’d) 9.) Connect the cover fan power cable and replace the side cover from the BEP. NOTE: When replacing the BEP Cover, ensure that you start it at the bottom to get the cover lip between the BEP chassis and AC transformer cover. Slide it forward and secure it. 10.) Replace the console front cover as shown in Section 8-2-5 on page 8-9. 11.) Replace the left- and/or right-side covers as shown in Section 8-2-2 on page 8-5. 12.) If required, Power On/Boot-up the system as described in Section 4-3-1 on page 4-2. 13.) Check MOD operation.Figure 8-48 BEP Cover Installation Tip
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-46 Section 8-8 - BEP Patient I/O Replacement Procedure (BEP 2.x and above only) Section 8-8 BEP Patient I/O Replacement Procedure (BEP 2.x and above only) 8-8-1 Manpower One person, 1 hour + travel 8-8-2 Tools Phillips screwdriver 8-8-3 Preparations (click here to view the cover removal video) 1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3. 2.) Remove the left- and/or right-side covers as shown in Section 8-2-2 on page 8-5. 3.) Remove the front cover as shown in Section 8-2-5 on page 8-9. 8-8-4 Patient I/O Removal Procedure NOTE: Mounting hardware may vary in size from M3 to 4-40 threads. Hardware length may seem the same but in fact may be different and very critical. ALWAYS keep track of EXACTLY where each screw was removed so that it is returned to the same place on re-assembly. 1.) Remove the side cover from the BEP. Carefully disconnect the cover fan power cable and set the cover aside. 2.) Disconnect the cable from the back of the patient I/O. It may be easier to disconnect this cable at the I/O VIC Tray because access is limited due to space behind the Patient I/O. 3.) Remove the one screw attaching the side of the Patient I/O as shown in Figure 8-50 on page 8-47 . Figure 8-49 Patient I/O Connection
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-47 8-8-4 Patient I/O Removal Procedure (cont’d) 4.) Peel back the adhesive trim as shown in Figure 8-51 on page 8-47 . Figure 8-50 CD-RW Side Screw Figure 8-51 BEP Front Adhesive Trim
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-48 Section 8-8 - BEP Patient I/O Replacement Procedure (BEP 2.x and above only) 8-8-4 Patient I/O Removal Procedure (cont’d) 5.) Remove the two screws in the front of the Patient I/O as shown in Figure 8-51 on page 8-47 . 6.) Slide the Patient I/O out the front of the BEP. 7.) Remove the interface cable from the back of the Patient I/O. 8-8-5 Patient I/O Replacement Procedure 1.) Attach the interface cable to the back of the new Patient I/O. 2.) Insert the new Patient I/O and cable into the drive bay and secure it in the front with the two screws previously removed. refer to Figure 8-51 on page 8-47 . 3.) Attach the one screw to the side of the Patient I/O through the rack. refer to Figure 8- 50 on page 8-47 . 4.) Attach the Patient I/O interface cable to the BEP IO VIC Tray. Ensure all cables are properly arranged and secure.Figure 8-52 IDE Ribbon Connector IDE Guide Pin
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-49 8-8-5 Patient I/O Replacement Procedure (cont’d) 5.) Connect the cover fan power cable and replace the side cover from the BEP. NOTE: When replacing the BEP Cover, ensure that you start it at the bottom to get the cover lip between the BEP chassis and AC transformer cover. Slide it forward and secure it. 6.) Replace the front cover as shown in Section 8-2-5 on page 8-9. 7.) Replace the left- and/or right-side covers as shown in Section 8-2-2 on page 8-5. 8.) If required, Power On/Boot-up the system as described in Section 4-3-1 on page 4-2. 9.) Check Patient I/O operation.Figure 8-53 BEP Cover Installation Tip
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-50 Section 8-9 - Power Supply Battery Pack Replacement Procedure Section 8-9 Power Supply Battery Pack Replacement Procedure 8-9-1 Manpower One person, .5 hour + travel 8-9-2 Tools Phillips screwdriver 8-9-3 Preparations (click here to view the cover removal video) We anticipate,with the exception of a failure, batteries will only need to be replaced once every three years. 1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3. 2.) Remove the left- and/or right-side covers as shown in Section 8-2-2 on page 8-5. 3.) Remove the upper rear cover as shown in Section 8-2-3 on page 8-7. 4.) Remove the lower rear cover as shown in Section 8-2-6 on page 8-10. 5.) Remove the Back End Processor as shown in Section 8-3-4 on page 8-14. 8-9-4 Power Supply Battery Pack Removal Procedure 1.) Remove the cover of the Back End Processor. 2.) Remove the four screws from the power supply battery pack cover plate. 3.) Disconnect the battery pack connection wire. 4.) Remove the power supply battery pack from the Back End Processor. Figure 8-54 Removing / Installing the Power Supply Battery Pack 1 2 34
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-51 8-9-5 Power Supply Battery Pack Installation Procedure 1.) Slide the new power supply battery pack into its position in the Back End Processor. 2.) Attach the power supply battery pack cover plate and secure it with the four screws removed in Section 8-9-4 on page 8-50. 3.) Apply Power .... turn the system on. 4.) Connect the new battery pack to the power supply connector while power is applied to the Back End Processor. This will avoid a possible spark while connecting the battery. 5.) Attach the cover for the Back End Processor removed in Section 8-9-4 on page 8-50. 6.) Install the Back End Processor into the scanner as described in Section 8-3-6 on page 8-16. The UPS batteries will drain slightly when the system is in the off condition. The off condition is defined as: - Circuit Breaker in the OFF position - System unplugged from the wall outlet If the system is left in the OFF condition for an extended period of time (3 to 5 days or more), the system may not boot up or may beep when turned ON. Should this occur, the system needs to recharge the UPS batteries. This could take 15 minutes to as long as 24 hours, depending on the battery age, system input voltage and system temperature. To eliminate draining the UPS batteries, the system should remain plugged into the wall outlet with the circuit breaker in the ON position.Figure 8-55 Power Supply Battery Pack CAUTIONThe battery pack contains a 24V charge that will cause a spark when it is re-connected to the power supply without power applied to the Back End Processor. Be careful that this spark does not damage any equipment or cause any personal injury. To avoid the spark, connect the battery with power applied to the BackEnd Processor. In Line 20A Fuse Battery 1 Battery 2 Compression tape used to minimize battery pack movement while installed in chassis Heat-shrink holds batteries together and insulates terminals. Spacer seperates batteries +- - + Label to be attached to this surface
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-52 Section 8-10 - Before Starting a Software Upgrade Section 8-10 Before Starting a Software Upgrade 8-10-1 Printers In R2.1.0 software and greater, if the customer has an HP990, HP970, UP-D50, UP-D895, UP-D21 or Talley printers, any settings they have made in the device driver, such as Gray Scale Images, will be lost. These printer settings will need to be recorded. You need to go to Utilities>Periperals>Printers in order to record the printer settings. 8-10-2 Save Connectivity Settings NOTE: DICOM TCPIP settings should not be deleted with the Base Image Load due to the use of upgrade scripts. However, It may be helpful to record this settings before proceeding in case an error occurs in the loading of software. NOTE: DICOM Workflows and DICOM Device Configurations are on the “D:” drive and will be saved when User Defined Cohfigurations are saved. If you want to Import exams back on to your system after the FMI, you should record any vital DICOM data before proceeding with this FMI or upgrading software. 1.) Select Utilities>Connectivity>TCPIP tab and record all information on that tab with particular attention to: - The “Enable DHCP” box should NOT be checked - Computer Name: ______________________________ - IP-Address: ______/________/_________/_________ - Subnet Mask: ______/________/________/________ - Remote Archive IP-Address: ______/________/_________/_________ - Remote Archive Name: ______________________________________ If the system has been Insite enabled, open the Service Diagnostic Interface and record InSite IP Address settings. 2.) Select the “Wrench” Icon. Log in as GE Service using the current password. 3.) Select Utilities>iLinq Configuration Utility>Accept (3 times)>InSite Checkout tab 4.) Record the Insite IP Address: _________/__________/__________/_________ and Unlabeled IP Address: __________/___________/__________/_________ - In the unlabeled IP Address line enter the System IP Address. This may be either the one obtained from the hospital administrator or if the system is not on a network it would be the factory default address. This address can also be found in the System>Connectivity>TCP/IP Tab 5.) Select the Device Connection Tab and record: CPU Serial Port Name ____________________ CPU Serial Port Speed ____________________ 6.) Select Exit in the Device Connection screen. 7.) Select OK in the Exit Confirmation window. 8.) Close the Netscape Diagnostic window.