GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-31 3-15-2 Optional Check Devices 1.) On the desktop, right click on the MY COMPUTER icon, and select PROPERTIES. 2.) Select the Hardware tab, and then the DEVICE MANAGER . It may take up to a minute for the Device Manager screen to appear. Figure 3-22 System Properties
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-32 Section 3-15 - Optional Software Procedures 3-15-2Optional Check Devices (cont’d) 3.) Verify that the Device Manager appears as shown in Figure 3-23. Your display may vary. 4.) You should see no devices with question marks or other icons, except the VIA PCI Audio Controller (WDM). This one device will be shown as below with a red X through it. 5.) There are two audio controllers. The VIA PCI Audio Controller (WDM) is disabled so that the Doppler audio works correctly through the Creative Labs SB Live series sound card. 6.) Close the Device Manager window by selecting the close window X in the corner. 7.) Close the System Properties window by selecting the close window X in the corner. Figure 3-23 Device Manager Close Window
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-33 3-15-3 Optional Procedure for Changing (Re-mapping) Drive Letters An automated process was developed for this. It was implemented in Base Image Load -9 or later. NOTE: If you are using a Base Image Load of -8 or earlier, or the -9 and later Base Image Load fails to properly map the drives, prodeed with manually assigning the proper drive letters. 1.) First check to see if you need to change the drive letters on the CDROM drive and the Magneto Optical drives – they SHOULD be G: and H: respectively. 2.) Right-click on the “My Computer” icon on the Windows desktop. 3.) Select “Manage” you should see a screen like Figure 3-24. 4.) Click on the “Disk Management” icon on the left side of the screen under “Storage”. You’ll see this screen like Figure 3-25. Figure 3-24 Computer Management
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-34 Section 3-15 - Optional Software Procedures 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) 5.) Notice that “Disk 1 Removable” is shown as drive letter “G:” and “CD-ROM 0” is shown as drive letter “H:” - These are incorrect drive letters. This procedure will show you how to change them to the proper designations which is the reverse of how they are now. 6.) Left-click on the “Disk 1 Removable” and select “Change Drive Letter and Path…”. You’ll see a screen as shown in Figure 3-26. Figure 3-25 Disk Management
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-35 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) 7.) Right-click on “Edit…” and you’ll see a screen as shown in Figure 3-27. 8.) Now Right-click on the little down arrow and you’ll see the list box open with a list of available drive letters. Select the letter “I:”Figure 3-26 Change Drive Letter and Path Figure 3-27 Select Drive to Assign
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-36 Section 3-15 - Optional Software Procedures 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) 9.) Now Right-click “OK”. You’ll see the following message as shown in Figure 3-29. 10.) Right-click “OK” to acknowledge the message. 11.) Now follow the previous steps again and change the “CD-ROM 0” to “G:”. 12.) Follow the previous steps one more time to change the “Disk 1 Removable” to “H:”. 13.) When you’re all done the settings should look like those shown in Figure 3-30. Figure 3-28 Select Drive Letter I Figure 3-29 Confirm Drive Letter Change
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-37 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Congratulations! You’ve just successfully re-mapped the drive letters of the CDROM drive and Magneto Optical drive to their proper configuration. Exit the Computer Management screen.Figure 3-30 Confirm Drive Letter Assignment
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-38 Section 3-15 - Optional Software Procedures 3-15-4 Possible Behavior of the First Power Up After Application Load After installing the new Application Software, if a new Front Panel code is needed: The first time the Ultrasound application (echoloader) is run after the software upgrade, the application will detect that the Front Panel code is out of date. Because the Front Panel has to have power removed and re-applied before the Front Panel code update process can continue, the application boot up will automatically shut down the LOGIQ ™ 9 just before the application boot up is complete. In some cases, the system may not shutdown automatically. If the dialog box appears, Ending Process, Please wait..., please wait 2 minutes. If the system does not turn off, force the shut down by holding down the on/off key until the monitor turns off then let it complete the power down process. Simply ensure the system has powered down, and the power button is amber, then restart the system to load the Ultrasound application. The auto-power down will only occur on the first power up after the application upgrade. 3-15-5 GE USB Control Panel Driver Loss NOTE: While the loss of drivers should not happen with the current application software loading process, this section remains as a reference. Sometimes, after loading new software (or the first time load of software), the GE Front Panel may not function after Applications (echoloader) has brought the system up in scan mode. The alpha-numeric keyboard, trackball and two set keys will function but the touch panel and all other GE controls on the operator interface will NOT function. Watch the system closely as the system boots up. Early in the process, during the BIOS load, the front panel lights will flash yellow then green and go out. This basically means the front panel is healthy or functional. During the Applications loading process of boot up, when the blue bar on the image monitor is about 2/ 3rds of the way complete, the front panel lights will flash again. This basically confirms USB communication with the front panel. During the software load process, the USB driver for the GE Control Panel may get corrupted. A simple Windows Driver Upgrade (update) will cure the problem.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-39 3-15-5GE USB Control Panel Driver Loss (cont’d) 1.) Properly Power down the system. 2.) Ensure that your Service Dongle is installed on the USB port in the back of the system. 3.) Power up the system. 4.) Select MAINTENACE... to enter the maintenance mode. 5.) Leave the User name blank, enter the current service Password and select EXIT TO WINDOWS . 6.) Trackball to MY COMPUTER and left click. 7.) Right click on PROPERTIES 8.) Right click on the HARDWARE Tab 9.) Right Click on DEVICE MANAGER in the middle-right of the Hardware tab window. Look through the Device Manager menu items and hardware selections for big yellow question marks or exclamation marks in a yellow oval. You should find an unknown device or a USB hardware selection that is not functioning. 10.) Double right click on the yellow question mark or exclamation mark.Figure 3-31 Bad USB Driver
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-40 Section 3-15 - Optional Software Procedures 3-15-5GE USB Control Panel Driver Loss (cont’d) 11.) In the GENERAL Tab for the hardware device select REINSTALL DRIVER. 12.) Select NEXT in the Driver Upgrade Wizard. 13.) Ensure that the Search for a suitable driver for my device (recommended) button is highlighted and then click on NEXT . 14.) From the Locate Driver Files window select Specify a location. Then click on NEXT . 15.) In the Copy manufacturer’s files from: drop down box select C:\WINNT\System32\Drivers and select OK 16.) The Found Driver window will appear with the selected driver for the hardware device. Select NEXT to install the driver. 17.) Select FINISH from the last Driver Upgrade Wizard screen. There should not be other problems but, if there are other hardware devices that need correcting, repeat step number 10 through 17. NOTE: Double check in the Device Manager the under USB Controllers to ensure that the GE USB Control Panel has no yellow question marks or exclamation marks in a yellow oval. 18.) Close all windows and select the GO MUSASHI shortcut to start the echoloader application. The touch panel and GE control panel knobs, buttons and switches should now be fully functional.Figure 3-32 Good USB Drivers