GE Logiq 9 Service Manual
Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-41 3-15-6 If the System is NOT on a Network If the system is connected to a network, skip this section. If the system is NOT on a network, a static dummy IP-Address must be install in Windows for DICOM and Daignostics to function. To install a Static IP-Address in windows: 1.) Boot up the system with the service dongle attached to the USB port. 2.) From the Start Applications dialog box, like the one shown in Figure 3-33, select MAINTENANCE Mode to enter the Windows desktop. 3.) The Maintenance Password screen will appear. Enter the current Service Password. 4.) Select Exit to Windows. 5.) From the task bar at the bottom of the Touch Panel select START >SETTINGS>NETWORK DIALUP CONNECTIONS. 6.) From the window shown in Figure 3-34, double click on LOCAL AREA CONNECTIONS . Figure 3-33 Select Maintenace Mode Figure 3-34 Network and Dialup Connections Window
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-42 Section 3-15 - Optional Software Procedures 7.) IF the local Area Connections Status Window is displayed, as shown in Figure 3-35, click on PROPERTIES . 8.) Hightlight INTERNET PROTOCOL (TCPIP) and select PROPERTIES as shown in Figure 3- 36. 9.) Select Use this following IP address. Fill in: * IP-Address: 3.28.28.28 * Subnet Mask : 255.255.252.0Figure 3-35 Connection Status Window Figure 3-36 TCPIP Properties Selection
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-43 Figure 3-37 Static Network Settings in Windows
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-44 Section 3-16 - Installing USB Line Printer - HP970/990 Section 3-16 Installing USB Line Printer - HP970/990 1.) Install the printer after the system has been ghosted, but before application software has been installed. Just plugging in the USB cable is enough to install the printer. 2.) If a service key is available, enter maintenance mode and exit to the desktop. Plug in the USB cable and the printer will install correctly. 3.) If the printer is installed while in application mode, excessive CPU time will be consumed and the system will become very slow to unusable. Shutting down and rebooting will clear the condition. The printer will function correctly, after this. 3-16-1 Optional USB Line Printer - HP970/990 This procedure assumes that the LOGIQ™ 9 base system software load image is installed and application software has been installed and is functioning properly. It also assumes that the printer was not installed directly after the Base Image Load. Disconnect any external printers connected to the service key port (USB port) on the rear of the LOGIQ™ 9 system. 1.) Power off the LOGIQ™ 9 ultrasound system. 2.) Insert a valid service key (dongle) into the service key port (USB port) on the rear of the scanner. 3.) Power up the scanner. 4.) When the “Start Application” dialog box appears, click “Maintenance” and type the current password and press enter. 5.) When the “Maintenance” screen appears, click “Exit to Windows”. 6.) Press CTRL-ALT-DEL to bring up the Task Manager and ensure that the “Applications” tab is the highlighted tab. 7.) Click on the “New Task...” button in the lower right hand corner of this window. 8.) In the “Create New Task” dialog box, type: “USERINIT” and press enter. 9.) Minimize the Task Manager. 10.) Click “Start” > “Settings” > “Printers”. 11.) Remove the service key (dongle) now that you’ve gotten to the desktop. 12.) Attach the USB cable from the printer to the USB port on the back of the ultrasound system. 13.) If you’re attaching a HP Diskette 990Cxi or 970Cxi printer, the system should immediately recognize the newly attached printer and automatically install and configure the printer for you without any additional user intervention. Print Key Assignment and Checkout 1.) Ensure that you have one of the front panel “Print” keys (P1-4) assigned to print on this new printer. Select Utility > Adman > Connectivity > Buttons >. Choose one of the print buttons and under “Destinations”, select “Standard Printer@MyComputer” from the drop-down list box. 2.) Press the key you just configured and you should get a printout of the screen you printed from.Figure 3-38 Found 990Cxi Printer
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-45 Section 3-17 Paperwork NOTE: During and after installation, the documentation (i.e. User Manuals, Installation Manuals...) for the peripheral units must be kept as part of the original system documentation. This will ensure that all relevant safety and user information is available during the operation and service of the complete system. 3-17-1 Product Locator Installation NOTE: The Product Locator Installation Card shown may not be same as the provided Product Locator card. 3-17-2 User Manual(s) Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability.Figure 3-39 Product Locator Installation Card
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-46 Section 3-17 - Paperwork This page was intentionally left blank.
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-1 Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ 9 scanner diagnostics by using the built-in service software, and power supply adjustments. Section 4-2 Required Equipment • An empty (blank) MO Disk. • At least one transducer. See “Probes” on page 1-41. for an overview. (normally you should check all the transducers used on the system) Table 4-1 Chapter 4 Contents SectionDescriptionPage Number 4-1 Overview4-1 4-2 Required Equipment4-1 4-3 General Procedure4-2 4-4 Software Configuration Checks4-20 4-5 Peripheral Checks4-21 4-6 Mechanical Function Checks4-22 4-7 Site Log4-24 NOTICEMost of the information pertaining to this Functional Checks chapter is found in the LOGIQ™ 9 Quick Guide (Direction number 2309820-100).
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-2 Section 4-3 - General Procedure Section 4-3 General Procedure 4-3-1 Power On / Boot Up NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. 4-3-1-1 Scanner Power On 1.) Connect the Power Cable to the back of the system. 2.) Ensure the Cable Clip slips securely over the shoulders on the molded plug. 3.) Connect the Power Cable to an appropriate mains power outlet. 4.) Switch ON the Circuit Breaker at the rear of the system. When power is applied to the scanner and the rear Circuit Breaker is turned ON, power is distributed to the Cooling Unit, Control Panel, Monitor, Internal and External I/O Boards, Peripherals and the Back- end Processor. CAUTIONSYSTEM REQUIRES ALL COVERS Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes. NOTICEEnergy Control and Power Lockout When servicing parts of the system where there is exposure to voltage greater than 30 Volts: Unplug the system Maintain control of the system power plug Wait for at least 20 seconds for capacitors to discharge Beware that some parts of the system may be energized even if the power is turned off but the cord is still plugged into the AC Outlet. Figure 4-1 Circuit Breaker and Power Cable on Back of Scanner NOTICEWhen AC power is applied to the scanner, the Standby switch on the Control panel illuminates amber, indicating the Backend Processor is in standby mode. 6LJQHGDWH 7$ * /2&.287 Circuit Breaker Mains Power Cable Cable Clip
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-3 4-3-1-2 Back-end Processor Boot Up Press the Standby switch on the Control Panel once. When the Standby switch on the Control Panel is pressed once, the Back-end Processor starts and the software code is distributed to initiate the scanner. No status messages are displayed during this process. 4-3-2 Power Off / Shutdown NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. 4-3-2-1 Back End Processor Power Off 1.) Press the Standby switch on the Control Panel once to display the SYSTEM - EXIT menu. 2.) Select Shutdown from the SYSTEM - EXIT menu. 3.) Wait for the system to properly close all files and shutdown.Figure 4-2 Standby Switch for Back-end Processor Boot Up Figure 4-3 System Exit Menu for Back-end Processor Power Down Standby Switch
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-4 Section 4-3 - General Procedure 4-3-2-2 Scanner Shutdown 1.) From the rear of the system. switch OFF the Circuit Breaker. 2.) Disconnect the Mains Power Cable by unfastening the metal wire fastener.Figure 4-4 Circuit Breaker and Power Cable on Back of Scanner NOTICEDisconnecting the Mains Power Cable before switching OFF the Circuit Breaker will activate the uninterruptible power system (UPS) in the Back End Processor, forcing an ordered shutdown of the system. Circuit Breaker Mains Power Cable Cable Clip