Home > GE > Ultrasound System > GE Logiq 9 Service Manual

GE Logiq 9 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 9 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 570
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-123
    8-31-4 Fan Assembly Removal Procedure (cont’d)
    4.) Remove Modem and it’s power supply.
    5.) Disconnect all cables from the External I/O Assembly
    6.) Remove External I/O. Unscrew the two Phillips screws fastening the External I/O to the chassis leg. 
    (see External I/O Replacement Procedure on page 8-126 for more details).
    7.) Unscrew the Phillips screw from the Internal I/O rear wall and pull the External I/O free from the 
    socket on the Internal I/O.
    8.) Remove the AC Controller (Refer to AC Controller Replacement Procedure on page 8-106) for 
    more details).
    9.) Disconnect the Front End Card Rack Input Voltage cable and the Fan Speed Control cable.
    10.) Remove the four screws holding the AC Controller.
    11.) Turn the AC Controller to gain better access and unplug the remaining cables. Ensure each cable 
    and connector are properly marked fopr installation later.
    12.) Pull out the AC Controller.
    13.) Disconnect Fan Cable. See Figure 8-131.
    . 
    14.) Unscrew one (1) Phillips screw that fasten Fan Assembly. See Figure 8-132.
    15.) Free Fan Assembly from the Board Rack sidewall hitch-up. See Figure 8-131.
    16.) Push disconnected cables and connectors down and out of the way.
    17.) Lower the front side of Fan Assembly and remove the Assembly carefully out the back. See Figure 
    8-133.
    NOTE: Be careful of the External I/O Connector when removing the Fan Assembly.Figure 8-132   Cover Airduct
    Filter Screws
    Filter Frame
    Cover Airduct
    Vibration bumperFan Assy Screw 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-124 Section 8-31 - Fan Assembly Replacement Procedure
    8-31-4 Fan Assembly Removal Procedure (cont’d)
     
    8-31-5 Fan Assembly Installation Procedure
    1.) Remove the shroud from the replacement Fan Assembly
    2.) Thread Fan assembly in the same way as it was taken out. Be careful with the External I/O 
    connector, located on the side of the Internal I/O. 
    3.) Replace the Fan Assembly Shroud.
    4.) Hitch up Fan Assembly on Board Rack sidewall.
    5.) Fasten the assembly to the chassis with one (1) screw.Figure 8-133   Fan Assembly Removal
    Figure 8-134   Fan Assembly 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-125
    .
    6.) Replace AC Controller as described in AC Controller Replacement Procedure on page 8-106.
    7.) Turn the AC Controller as necessary to facilitate connecting the cables.
    8.) Secure the AC Controller to the chassis.
    9.) Replace External I/O as described in Internal I/O Replacement Procedure on page 8-128.
    10.) Plug the External I/O into it’s socket on the Internal I/O. Replace the screw that attaches the external 
    I/O to the Internal I/O rear wall.
    11.) Replace the two screws that attach the External I/O to the chassis leg.
    12.) Connect the telephone and interface cables to the rear of teh External I/O Assembly
    13.) Replace the Modem and it’s power supply.
    14.) Reconnect the power cable from the Front End Card Rack along with the Fan Speed Control cable
    15.) Install Air Duct cover. Slide it in over the Fan Assembly then reach in and force the front down over 
    the fan. 
    16.) Secure the cove with four (4) screws. See Figure 8-135.
    17.) Replace Filter Frame and filter media.
    18.) Connect Fan cable. See Figure 8-131.
    19.) Replace Lower Rear Cover, including different amount of washers and a magnetic lock for Filter 
    Cover on the upper screws, described in Lower Rear Cover Replacement Procedure on page 8-10.
    20.) Replace Rear Handle as described in Rear Handle Installation procedure on page 8-12.
    21.) Replace Top Cover as described in Top Cover Replacement Procedure on page 8-8.
    22.) Replace Right Side Cover as described in Side Covers (with Bumper) Replacement Procedure on 
    page 8-5.
    23.) Turn Mains Circuit Breaker ON and boot up the system as described in Section 4-3-1 on page 4-2. 
    Put a hand on the Filter and feel that the Fan draws air into the System.Figure 8-135   Fan Assembly in Position
    Cover Air Duct
    Vibration bumper
    Top of Fan Assembly 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-126 Section 8-32 - External I/O Replacement Procedure
    Section 8-32
    External I/O Replacement Procedure
    8-32-1 Manpower
    1 person, 1 hour
    8-32-2 Tool
    Long Phillips screwdriver, size 2.
    8-32-3 Preparations (click here to view the cover removal video)
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
    2.) Remove the Right Side Cover as described in Section 8-2-2 on page 8-5.
    3.) For full control during installation procedure: Remove the Upper Rear Cover as described in 
    Section 8-2-3 on page 8-7 and the Lower Rear Cover in Section 8-2-6 on page 8-10.
    8-32-4 External I/O Removal procedure
    1.) Disconnect all cables to and from the External I/O Assembly.
    2.) Unscrew two Phillips screws fastening it to the chassis leg. 
    3.) Unscrew the Phillips screw on the Internal I/O rear wall and pull the Assembly free from the socket 
    it is plugged into.
    WARNINGWARNINGDo not touch any boards with integrated circuits prior to taking the necessary ESD 
    precautions: 
    1.) Always connect yourself, via an arm-wrist strap, to the advised ESD connection 
    point located on the Rear of the scanner (to the right of the power connector). 
    2.) Follow general guidelines for handling of electrostatic sensitive equipment.
    Figure 8-136   Side Views of the External I/O Assembly
    Screws (2) External I/O Plug
    Flange for fastening the external I/O to the
    internal I/O just below the socket/plug
    connection. 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-127
    8-32-5 External I/O Installation Procedure
    1.) Install the new External I/O Assembly by pushing it into the socket on the rear wall of the Internal I/
    O and fastening it with the Phillips screw.
    2.) Fasten it to the chassis leg with the two screws.
    3.) Mount the Right side as described Section 8-2-2 on page 8-5.
    4.) Mount the Upper Rear Cover as described in Section 8-2-3 on page 8-7 and the Lower Rear Cover 
    in Section 8-2-6 on page 8-10.
    5.) Connect all cabling to and from the External I/O. Remember: There are two connectors on the 
    bottom side of the box. These two are identical in size and pinout. The left one is manufacture 
    power cycle testing and the one on the right is for the modem connection. If they are mixed up the 
    system will experience power up problems as soon as the modem powers on.
    6.) Turn ON the Main Breaker and boot up the system as described in Section 4-3-1 on page 4-2.
    7.) Do functional check-out as described.Figure 8-137  External I/O Location
    Figure 8-138  External I/O Installation
    Screws on chassis leg
    Screw on Internal I/O
    External I/O rear side
    Internal I/O
    External I/O
    Connector on External I/O
    Flange for fastening External I/O to 
    Internal I/O. Centering pins 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-128 Section 8-33 - Internal I/O Replacement Procedure
    Section 8-33
    Internal I/O Replacement Procedure
    8-33-1 Manpower
    1 person, 1 hour 30 minutes
    8-33-2 Tool
    Long Phillips screwdriver, size 2.
    8-33-3 Preparations (click here to view the cover removal video)
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
    2.) Remove the Side Covers as described in Section 8-2-2 on page 8-5.
    3.) Remove Rear Covers as described on Section 8-2-6 on page 8-10 and Section 8-2-3 on page 8-7.
    4.) Remove the External I/O as described in Section Section 8-32 on page 8-126.
    5.) Disconnect the cables connected to the I/O on both sides.
    WARNINGWARNINGDo not touch any boards with integrated circuits prior to taking the necessary ESD 
    precautions: 
    1.) Always connect yourself, via an arm-wrist strap, to the advised ESD connection 
    point located on the Rear of the scanner (to the right of the power connector). 
    2.) Follow general guidelines for handling of electrostatic sensitive equipment.
    Select the movie camera icon to view the video of the cover 
    removal and installation procedures.
    Figure 8-139   Internal I/O Removal
    Screw
    Internal I/O Screws
    Backend Processor side
    External I/O side 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-129
    8-33-4 Removal Procedure
    1.) From Back End Processor (BEP) side (left side of the system), unscrew two (2) screws. These 
    fasten the upper edge of the Internal I/O Assembly to the chassis.
    2.) From External I/O side (right side), unscrew one (1) screw. This screw fasten the lower edge of 
    Internal I/O to the chassis. 
    3.) Pull the Internal I/O Assembly backwards, free from its socket in the Card Rack sidewall, and out 
    of the system.
    8-33-5  Installation Procedure
    1.) Align Internal I/O in position and push it into its card rack socket.
    2.) Fasten two (2) screws on BEP side and one (1) screw on External Power side.
    3.) Connect the cables to both sides of the Internal I/O.
    4.) Fasten the two fastening screws.
    5.) Mount the External I/O as described on Section Section 8-32 on page 8-126.
    6.) Mount the Rear Cover as described on Section 8-2-6 on page 8-10 and Section 8-2-3 on page 8-7.
    7.) Mount the two Side Covers as described Section 8-2-2 on page 8-5. Figure 8-140  Internal I/O
    Front end
    External I/O side
    Connector to Card Rack socket BEP side
    Card Rack socket
    Hitch up in Card Rack 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-130 Section 8-34 - Front End Processor Boards Replacement Procedure
    Section 8-34
    Front End Processor Boards Replacement Procedure
    8-34-1 Manpower
    1 persons, 0.5 hours
    8-34-2 Tools
    8-34-3 Preparations 
    (click here to view the cover removal video)
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
    2.) Remove Right Side Cover as shown on Section 8-2-2-4 on page 8-5.
    3.) Remove the Front End Processor Cover as shown in Figure 8-141.
    It is essential that Vital Product Data (VPD) is verified or updated after any hardware change to the Front 
    End Processors. Refer to Section 8-35 Verify and Update Vital Product Data on page 8-135 for details.
    NOTE: Vital Product Data (VPD) must be verified and/or updated when Front End Boards are replaced. 
    Failure to properly program Vital Product Data could cause adverse affects on system 
    operation and image quality.Figure 8-141  Front End Processor Location
    Table 8-8    Typical Circuit Board VPD Data
    Circuit Board NamePart Number
    Functional Revision 
    PWA LetterFab Revision 
    PWB NumberBar Code or
    Serial NumberDate
    YYYYMMDD
    Equalization 22602102 0 0 _  _ _  01
    2 0 0 _  _ _  01
    2 0 0 _  _ _  01
    Screws 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-131
    8-34-4 Front End Board Removal Procedure
    The following boards are removed from the card rack with the ejectors at the top and bottom of each 
    card. Slide the card out of the rack slowly with equal pressure on both the top and bottom of the card.
    NOTE: Observe all ESD proper precautions when handling any part of the Front End Processor.
    • TD Boards 0-7, EQ Board, BMP Board, EBM/EBM2 Board
    • XDIF and RF Amp/RF Amp2
    a.) Remove the three screws that fasten the XDIF board to the front of the card rack chassis.
    b.) Carefully eject and partially remove the XDIF and RF Amp boards simultaneously
    . Remove 
    the boards just enough in order to grab both sides of the Top Plane board.
    c.) Hold the XDIF and RF Amp boards. Carefully grab each SIDE of the Top Plane Board. Gently 
    rock the Top Plane board side to side while pulling it off of its connectors.
    NOTE: Rocking the Top Plane board top to bottom will afford a greater opportunity to bend pins during 
    its extraction. Only rock the board side-to-side while removing it.
    d.) Once the Top Plane Board is removed, either the XDIF or RF Amp can be removed from the 
    card rack.Figure 8-142   Side View of System with Card Cage Open
    CAUTIONThe following boards require special attention when removing them from the card rack. READ 
    the hints that follow to avoid damage to the system or personal injury.
    TD/TD5 Boards 0-7SCB/SCB2 EQ4
    BMPRF Amp/RF Amp2
    XDIF
    Top Plane
    Power 
    Supply
    3 Screws
    Screw
    EBM/EBM2Empty Slot
    FEP Version 1
    FEP Version BT’02/BT’03 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-132 Section 8-34 - Front End Processor Boards Replacement Procedure
    8-34-4 Front End Board Removal Procedure (cont’d)
    • SCB/SCB2 and Power Supply (FEPS, FEPS2, FEPS2.1, FEPS3)
    a.) For the old supplies, remove the one screw at the bottom of the Power Supply Board.
    b.) If necessary, unscrew the two thumb screws attaching a cover plate over the PCI cable as it 
    enters the card rack chassis.
    c.) Partially remove the SCB and Power Supply boards simultaneously. Remove them just 
    enough to gain access to the PCI cable from the Back End Processor that passes through the 
    Power Supply board and attaches to the SCB.
    d.) Disconnect the PCI cable from the SCB and pull it through the hole in the Power Supply board.
    e.) If you are removing just the SCB, completely slide it out of the card rack.
    f.) If you are removing the Power Supply board, pull the PCI cable out of the card rack chassis so 
    it will not interfere with the Power Supply board.
    g.) Pull the Power Supply board out of the rack to expose the black/white/green AC input wires.
    h.) Disconnect these wires from the Power Supply board and slide the board out of the card rack.
    8-34-5 Front End Board Installation Procedure
    • The TD Boards 0-7, EQ Board, BMP Board and EBM/EBM2
    a.) Installation is accomplished by sliding the top and bottom edges of the board into the guide 
    slots with firm even pressure to securely seat the board into the motherboard and secure it with 
    the ejectors.
    • XDIF, RF Amp/RF Amp2 and Top Plane
    a.) When inserting the XDIF board back into the card rack, ensure that the RF shield is behind the 
    card rails to avoid binding. 
    b.) Slide the XDIF in and secure it with the three screws.
    c.) Slide in the RF Amp.
    d.) Install the Top Plane board by first aligning the pins and then pushing straight in to seat the 
    board.
    • SCB and Power Supply (FEPS, FEPS2, FEPS2.1, FEPS3)
    a.) The SCB and Power Supply Boards should be started simultaneously but not pushed in all the 
    way.
    b.) Reconnect the black/white/green AC input wires on the Power Supply board.
    c.) Thread the PCI cable through the chassis and Power Supply board and connect it to the SCB.
    d.) Simultaneously seat the SCB and Power Supply boards into motherboard.
    e.) If removed, replace the RF shield over the PCI Cable chassis hole.
    f.) Since there is no ejector at the bottom of the Power Supply board, be sure to push firmly on 
    the bottom to completely seat the card. Secure the bottom of the Power Supply board with the 
    screw previously removed.
    WARNINGWARNINGEnsure that the system is turned off and unplugged. Wait at least 20 seconds for the 
    charge on the FEPS capacitors to bleed off before handling the power supply board. 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq 9 Service Manual