Hitachi F7000 Instruction Manual
Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
2.2 Creating an Analysis Method 2 - 4 7 Load Click this button to read out saved analytical conditions. Upon clicking, the file opening window appears. So, select an analytical condition file. 8 Save When you click this button after setting the parameters, the analytical conditions can be saved by overwriting. In case of a new analysis method, clicking this button displays the “Save As” window. You must enter a file name for saving in this window. 9 Save As Click this button for renaming the set conditions for saving. The “Save As” window appears, so enter a file name.
2.2.2 2 - 5 When you click the Instrument tab, the window in Fig. 2-4 appears. For the settable range of each parameter, refer to Appendix G. Fig. 2-4 Instrument Tab 1 Scan mode Select a scan mode here. In case of Synchronous, the abscissa of the graph stands for excitation wavelength. The unit indication of the abscissa becomes ‘EX/EM(nm).’ 2 Data mode Select a data mode. The Luminescence mode is for measuring luminescence in the biological/chemical field. In this mode, only the sample side signal is acquired without performing ratio photometry. The Phosphorescence mode is for acquiring only phosphorescence after extinction of excitation beam while turning a chopper on the excitation side. 2.2.2 Instrument Tab 5 6 7 8 3 2 1 9 10 11 15 16 17 18 19 14 13 12 20 4 21
2.2 Creating an Analysis Method 2 - 6 3 Chopping speed Select a chopping speed. Setting is allowed when Phosphorescence is selected for Data mode. The Phosphorescence mode is for acquiring only phosphorescence after extinction of excitation beam while turning a chopper on the excitation side. An increase in chopping speed makes the chopper turns faster. 4 EM WL For the Excitation scan mode, enter the fixed wavelength on the emission side . For the Synchronous scan mode, enter the start wavelength on the emission side. 5 EX Start WL This is the start wavelength on the excitation side in wavelength scan. Enter the shorter wavelength side of measuring wavelength range. 6 EX End WL This is the end wavelength on the excitation side in wavelength scan. Enter the longer wavelength side of measuring wavelength range. 7 EX WL For the Emission scan mode, enter the fixed wavelength on the excitation side. 8 EM Start WL This is the start wavelength on the emission side in wavelength scan. Enter the shorter wavelength side of measuring wavelength range. 9 EM End WL This is the end wavelength on the emission side in wavelength scan. Enter the longer wavelength side of measuring wavelength range. 10 Scan speed Set a wavelength scan speed. 11 Delay After pressing the Measure button, measurement is started following the delay time set here. It is used for temperature stabilization, etc. In repeat measurement, it is the time until the start of the first measurement. The delay time is invalid for the second and subsequent measurements.
2.2.2 2 - 7 12 EX Slit Select a slit width on the excitation side. 13 EM Slit Select a slit width on the emission side. 14 PMT Voltage A function for controlling the voltage of the photomultiplier detector (photomultiplier tube). As a reference, changing from 400 V to 700 V increases the data value about two digits and changing from 700 V to 950 V increases it about 1 digit. When a check mark is put for PMT Voltage 0-1000V, the photomultiplier voltage is settable within a range from 0 to 1000 V. 15 PMT Voltage 0-1000V When a check mark is put in this box, the photomultiplier voltage is settable within a range from 0 to 1000 V. 16 Response When “Auto” is selected, response will be set automatically depending on scan speed. Select “Auto” usually. 17 Corrected spectra A function for determining the spectrum inherent to a sample by correcting the photometer wavelength characteristic using the saved instrument parameters, following measurement with the instrument parameters for photometer control. If there are no instrument parameters for the wavelength range to be measured, only the wavelength range that allows correction is scanned. And if there is no wavelength area measurable, an error will occur. The Instrument Response command is provided in the Utility menu for operating this function. When this setting is at ON, “Corrected spectra” appears on the monitor window. (See Fig. 1-43 in 1.6.) 18 Shutter control The shutter can automatically be closed in other than measurement for slowing down sample deterioration due to the energy of excitation beam and opened when measurement starts. When you put a check mark at the head, the shutter will close, and open at start of measurement. The shutter will close again when measuring wavelength begins returning to the start wavelength after measurement in the wavelength scan mode.
2.2 Creating an Analysis Method 2 - 8 19 Replicates Set the number of repeat measurements. An average spectrum can also be determined in repeat measurement. Setting is available on the Processing tab (shown in Fig. 2-6). 20 Cycle time Set a repetition interval. 21 Range of corrected spectra Spectrum correction range is indicated.
2.2.3 2 - 9 When selecting the Monitor tab, the window in Fig. 2-5 appears. For the settable range of each parameter, refer to Appendix G. Fig. 2-5 Monitor Tab Assign the following parameters. 1 Y-Axis Max (maximum) Enter a maximum value for Y-axis on the monitor window. 2 Y-Axis Min (minimum) Enter a minimum value for Y-axis on the monitor window. 3 Open data processing window after data acquisition Select whether or not to conduct data processing after sample measurement. When selected (check mark is applied), an icon (as shown in ) is displayed for the data processing window at the end of measurement. And by opening this icon, data processing such as peak detection is available. 2.2.3 Monitor Tab 4 5 6 3 2 1
2.2 Creating an Analysis Method 2 - 10 NOTE: When the remaining memory capacity of PC in operation is very small (system resource is less than 20%), the data processing window will not open even if this item is turned ON. Either close the presently opened data processing window or restart the FL Solutions program. 4 Open data processing window after acquisition (Overlay) Select whether or not to conduct data processing with spectra overlaid after sample measurement. This setting is unavailable in repeat measurement. 5 Print report after data acquisition Select whether or not to print a report after sample measurement. When selected (check mark is applied), printing will be made automatically after measurement. The items set at the Report tab will be printed. When this item is not selected, data processing is allowed by reading out a file via [File]-[Open]. 6 Overlay Spectra can be overlaid in the monitor window. This setting cannot be selected for repeat measurement.
2.2.4 2 - 11 On selection of the Processing tab, the window in Fig. 2-6 will be displayed. For the settable range of each parameter, refer to Appendix G. Fig. 2-6 Processing Tab 1 CAT (Computing for Averaging Transient) Valid when 2 or larger value is entered for Replicates at the Instrument tab. When a check mark is put at the head, an average spectrum is determined upon carrying out repeat measurement. This is especially effective for measuring samples with which the emission intensity varies with time. 2 Processing choices A list of data processing is displayed. Select a data processing item, and click the Right arrow button ( ). Then, the selected method appears in the Processing steps field. For example, select ‘Derivative’ and click , then ‘Derivative’ is displayed in the Processing steps field as shown in Fig. 2-7. 2.2.4 Processing Tab 4 56 32 1
2.2 Creating an Analysis Method 2 - 12 Fig. 2-7 3 Processing steps The processing sequence is displayed. After measurement, data processing will be carried out in the order set here. How to Delete a Processing Method Select a processing method to be deleted and click the Left arrow button ( ). Then, the selected method disappears from the Processing steps field. Display and Change of Parameter When you click [+] in the Processing steps field, a parameter list is displayed. Click a parameter name to be changed, and it becomes changeable. For changing a ‘Derivative’ parameter, click [+] at the left of ‘Derivative’ in Fig. 2-7. The present derivative parameters (Derivative order, Smoothing order, Number of points) are displayed. For changing the derivative order, click ‘Derivative order,” and it is highlighted as shown in Fig. 2-8 to allow your change. Fig. 2-8
2.2.4 2 - 13 4 Peak Finding: Integration method Select an integration method for the peaks to be displayed in the data processing window. 5 Peak Finding: Threshold Enter the minimum photometric value difference (between peak and valley) for definition as a peak. 6 Peak Finding: Sensitivity Select the number of data points in the horizontal axis direction (abscissa). For detecting sharp peaks, select a sensitivity value of “1.” Or select “8” for detecting broad peaks. Peak Min. peak-valley difference Valley