Home > Hitachi > Fluorescence Spectrophotometer > Hitachi F7000 Instruction Manual

Hitachi F7000 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							2.2    Creating an Analysis Method 
    2 - 4 
    7 Load  
    Click this button to read out saved analytical conditions.     
    Upon clicking, the file opening window appears.    So, select an 
    analytical condition file.   
    8 Save  
    When you click this button after setting the parameters, the 
    analytical conditions can be saved by overwriting.    In case of a 
    new analysis method, clicking this button displays the “Save As” 
    window.    You must enter a file name for saving in this window.   
    9 Save As   
    Click this button for renaming the set conditions for saving.     
    The “Save As” window appears, so enter a file name.   
      
    						
    							2.2.2 
    2 - 5 
    When you click the Instrument tab, the window in Fig. 2-4 appears.   
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 2-4  Instrument Tab 
     
     
    1 Scan mode   
    Select a scan mode here.    In case of Synchronous, the abscissa 
    of the graph stands for excitation wavelength.     
    The unit indication of the abscissa becomes ‘EX/EM(nm).’   
    2 Data mode   
    Select a data mode.    The Luminescence mode is for measuring 
    luminescence in the biological/chemical field.     
    In this mode, only the sample side signal is acquired without 
    performing ratio photometry.    The Phosphorescence mode is for 
    acquiring only phosphorescence after extinction of excitation beam 
    while turning a chopper on the excitation side.   
    2.2.2 Instrument Tab 
     
     
    5 
    6 
    7 
    8 
    3 
    2 
    1 
    9 
    10 
    11 
    15
    16
    17
    18
    19
    14
    13
    12
    20
    4 
    21 
    						
    							2.2    Creating an Analysis Method 
    2 - 6 
    3 Chopping speed   
    Select a chopping speed.    Setting is allowed when 
    Phosphorescence is selected for Data mode.     
    The Phosphorescence mode is for acquiring only phosphorescence 
    after extinction of excitation beam while turning a chopper on the 
    excitation side.    An increase in chopping speed makes the 
    chopper turns faster.   
    4 EM WL   
    For the Excitation scan mode, enter the fixed wavelength on the 
    emission side .    For the Synchronous scan mode, enter the start 
    wavelength on the emission side.   
    5 EX Start WL   
    This is the start wavelength on the excitation side in wavelength 
    scan.    Enter the shorter wavelength side of measuring wavelength 
    range.  
    6 EX End WL   
    This is the end wavelength on the excitation side in wavelength 
    scan.    Enter the longer wavelength side of measuring wavelength 
    range.  
    7 EX WL   
    For the Emission scan mode, enter the fixed wavelength on the 
    excitation side.   
    8 EM Start WL   
    This is the start wavelength on the emission side in wavelength 
    scan.    Enter the shorter wavelength side of measuring wavelength 
    range.  
    9 EM End WL   
    This is the end wavelength on the emission side in wavelength 
    scan.    Enter the longer wavelength side of measuring wavelength 
    range.  
    10 Scan speed   
    Set a wavelength scan speed.   
    11 Delay  
    After pressing the Measure button, measurement is started 
    following the delay time set here.    It is used for temperature 
    stabilization, etc.    In repeat measurement, it is the time until the 
    start of the first measurement.    The delay time is invalid for the 
    second and subsequent measurements.    
    						
    							2.2.2 
    2 - 7 
    12 EX Slit   
    Select a slit width on the excitation side.   
    13 EM Slit   
    Select a slit width on the emission side.   
    14 PMT Voltage   
    A function for controlling the voltage of the photomultiplier detector 
    (photomultiplier tube).    As a reference, changing from 400 V to 
    700 V increases the data value about two digits and changing from 
    700 V to 950 V increases it about 1 digit.   
    When a check mark is put for PMT Voltage 0-1000V, the 
    photomultiplier voltage is settable within a range from 0 to 1000 V.   
    15 PMT Voltage 0-1000V   
    When a check mark is put in this box, the photomultiplier voltage is 
    settable within a range from 0 to 1000 V.   
    16 Response  
    When “Auto” is selected, response will be set automatically 
    depending on scan speed.    Select “Auto” usually.   
    17 Corrected spectra   
    A function for determining the spectrum inherent to a sample by 
    correcting the photometer wavelength characteristic using the 
    saved instrument parameters, following measurement with the 
    instrument parameters for photometer control.    If there are no 
    instrument parameters for the wavelength range to be measured, 
    only the wavelength range that allows correction is scanned.     
    And if there is no wavelength area measurable, an error will occur.   
    The Instrument Response command is provided in the Utility menu 
    for operating this function.   
    When this setting is at ON, “Corrected spectra” appears on the 
    monitor window.    (See Fig. 1-43 in 1.6.)   
    18 Shutter control   
    The shutter can automatically be closed in other than measurement 
    for slowing down sample deterioration due to the energy of 
    excitation beam and opened when measurement starts.     
    When you put a check mark at the head, the shutter will close, and 
    open at start of measurement.    The shutter will close again when 
    measuring wavelength begins returning to the start wavelength 
    after measurement in the wavelength scan mode.    
    						
    							2.2    Creating an Analysis Method 
    2 - 8 
    19 Replicates  
    Set the number of repeat measurements.   
    An average spectrum can also be determined in repeat 
    measurement.    Setting is available on the Processing tab (shown 
    in Fig. 2-6).   
    20 Cycle time   
    Set a repetition interval.   
    21 Range of corrected spectra   
    Spectrum correction range is indicated.   
      
    						
    							2.2.3 
    2 - 9 
    When selecting the Monitor tab, the window in Fig. 2-5 appears.     
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 2-5  Monitor Tab 
     
     
    Assign the following parameters.   
     
    1 Y-Axis Max (maximum)   
    Enter a maximum value for Y-axis on the monitor window.   
    2 Y-Axis Min (minimum)   
    Enter a minimum value for Y-axis on the monitor window.   
    3 Open data processing window after data acquisition   
    Select whether or not to conduct data processing after sample 
    measurement.    When selected (check mark is applied), an icon 
    (as shown in 
    ) is displayed for the data 
    processing window at the end of measurement.    And by opening 
    this icon, data processing such as peak detection is available.   
    2.2.3 Monitor Tab 
     
    4 
    5 
    6 
    3 
    2 
    1  
    						
    							2.2    Creating an Analysis Method 
    2 - 10 
    NOTE:  When the remaining memory capacity of PC in operation is 
    very small (system resource is less than 20%), the data 
    processing window will not open even if this item is turned ON.   
    Either close the presently opened data processing window or 
    restart the FL Solutions program.   
     
    4 Open data processing window after acquisition (Overlay)   
    Select whether or not to conduct data processing with spectra 
    overlaid after sample measurement.    This setting is unavailable in 
    repeat measurement.   
    5 Print report after data acquisition   
    Select whether or not to print a report after sample measurement.   
    When selected (check mark is applied), printing will be made 
    automatically after measurement.     
    The items set at the Report tab will be printed.    When this item is 
    not selected, data processing is allowed by reading out a file via 
    [File]-[Open].  
    6 Overlay  
    Spectra can be overlaid in the monitor window.   
    This setting cannot be selected for repeat measurement.   
     
      
    						
    							2.2.4 
    2 - 11 
    On selection of the Processing tab, the window in Fig. 2-6 will be 
    displayed.  
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 2-6  Processing Tab 
     
     
    1 CAT (Computing for Averaging Transient)   
    Valid when 2 or larger value is entered for Replicates at the 
    Instrument tab.   
    When a check mark is put at the head, an average spectrum is 
    determined upon carrying out repeat measurement.   
    This is especially effective for measuring samples with which the 
    emission intensity varies with time.   
    2 Processing choices   
    A list of data processing is displayed.    Select a data processing 
    item, and click the Right arrow button (
    ).  Then, the selected 
    method appears in the Processing steps field.    For example, 
    select ‘Derivative’ and click 
    , then ‘Derivative’ is displayed in 
    the Processing steps field as shown in Fig. 2-7.   
    2.2.4 Processing Tab 
     
    4 
    56
    32 
    1  
    						
    							2.2    Creating an Analysis Method 
    2 - 12 
     
     
     
    Fig. 2-7   
     
     
    3 Processing steps   
    The processing sequence is displayed.    After measurement, data 
    processing will be carried out in the order set here.   
    How to Delete a Processing Method   
    Select a processing method to be deleted and click the Left 
    arrow button (
    ).    Then, the selected method disappears 
    from the Processing steps field.   
     
    Display and Change of Parameter   
    When you click [+] in the Processing steps field, a parameter list 
    is displayed.    Click a parameter name to be changed, and it 
    becomes changeable.    For changing a ‘Derivative’ parameter, 
    click [+] at the left of ‘Derivative’ in Fig. 2-7.    The present 
    derivative parameters (Derivative order, Smoothing order, 
    Number of points) are displayed.    For changing the derivative 
    order, click ‘Derivative order,” and it is highlighted as shown in 
    Fig. 2-8 to allow your change.   
     
     
     
    Fig. 2-8 
      
    						
    							2.2.4 
    2 - 13 
    4 Peak Finding:  Integration method  
    Select an integration method for the peaks to be displayed in the 
    data processing window.   
    5 Peak Finding:  Threshold  
    Enter the minimum photometric value difference (between peak 
    and valley) for definition as a peak.   
     
     
     
     
    6 Peak Finding:  Sensitivity  
    Select the number of data points in the horizontal axis direction 
    (abscissa).    For detecting sharp peaks, select a sensitivity value of 
    “1.”    Or select “8” for detecting broad peaks.   
     
     
    Peak
    Min. peak-valley 
    difference 
    Valley 
    						
    All Hitachi manuals Comments (0)