Hitachi F7000 Instruction Manual
Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
A - 18 E.2 Advantages of Fluorometry As contrasted with fluorometry, absorptiometry for a low-concentration sample is explained in the following: A sample having 99% transmittance to blank is taken as an example. In the absorbance measurement of such a substance, inaccuracies must always be taken into consideration. Here, the inaccuracy is assumed to be 0.1%. Since it has an effect on both the blank and sample, Percent transmittance of blank 100.0 ± 0.1% Percent transmittance of sample 99.0 ± 0.1% Difference (proportional to sample concentration) 1.0 ± 0.2% In this example, the uncertainty in concentration measurement is ±20%. While in the fluorometry, a difference from zero level corresponds to the concentration of sample. Accordingly, the measurement accuracy is as follows: Output signal level at sample measurement 100 ± 0.1 Value corresponding to blank 0 ± 0.1 Difference (proportional to sample concentration) 100 ± 0.2 As can be seen from the above, the fluorometry is very advantageous for analyzing a low-concentration sample since its uncertainty is in most cases theoretically independent of the sample concentration. Although actual practice may involve some error factors which increase as the sample concentration becomes extremely low, the fluorometry is capable of measuring low concentrations with an accuracy 100 times higher than in absorptiometry. Figure E-2 is an explanatory illustration of the foregoing description. In absorptiometry, a difference between the quantity of incident radiation I o and the quantity of transmitted radiation It is represented by signal Is. A level at which the signal I s becomes almost equal to a noise level is used as a detection limit. In fluorometry, however, since the quantity of fluorescence I F itself is represented as a signal, just a small amount of fluorescence can be amplified electrically for enabling detection.
A - 19 Still more, since a fluorescence wavelength of a substance is different from its excitation wavelength (incident light wavelength), the fluorescence wavelength is not readily affected by the exciting radiation, thereby contributing to ensuring high sensitivity. Fig. E-2 Comparison between Absorptiometry and Fluorometry In addition to high sensitivity, the fluorometry is advantageous in that more information is attainable. An emission spectrum is also available besides an excitation spectrum which corresponds to an absorption spectrum in absorptiometry. The two kinds of wavelengths can be selected as desired, and a fluorescence spectrum can be recorded using a properly selected excitation wavelength (or vice versa). Thus, quantitative and qualitative analyses can be made for a sample containing plural components. High concentration Low concentrationHigh concentration Low concentration (a) Absorptiometry (b) Fluorometry Amplified
A - 20 Figure E-3 shows a simplified spectral graph of measurement of a sample containing multiple components. In absorptiometry, since only the absorption spectrum can be measured, two or more component wavelengths are presented. If the absorption wavelengths are similar to each other, each component cannot be separated in measurement. In fluorometry, even if the absorption wavelengths are similar, a difference in fluorescence makes it possible to select each fluorescence wavelength properly. Thus, each component can be separated in measurement. Fig. E-3 Measurement of Multi-component Sample Table E-1 compares information attainable in absorptiometry and that in fluorometry. Table E-1 Comparison of Information Attainable in Absorptiometry and Fluorometry Absorptiometry Fluorometry Absorption spectrum only (corresponding to excitation spectrum in fluorometry) • Excitation spectrum • Fluorescence spectrum Absorption spectra Component A Component BExcitation spectra Fluorescence spectra Component A Component B (a) Absorptiometry (b) Fluorometry
A - 21 E.3 Remarks on Measurement in Fluorescence Analysis For most kinds of samples, an increase of 1 °C in sample temperature causes the fluorescence intensity to decrease by 1 to 2%. It is also reported that for some kinds of biochemical samples, the fluorescence intensity decreases as much as 10% as the temperature increases by 1 °C. When analyzing a sample having a temperature- dependent property, it is advisable to use the constant-temperature cell holder (P/N 650-0150). Constant-temperature measurement can be carried out by circulating constant-temperature water through this cell holder. Some kinds of samples may be susceptible to a chemical change due to exciting radiation. In analysis of such a sample, keep the shutter closed to cut off an excitation beam until measurement is started, and then open the shutter immediately before measurement. If any chemical change due to exciting radiation is observed still, determine a signal level at the start time through extrapolation according to variation in signal level. In fluorescence measurement, spectra having different natures from that of fluorescence may be observed. These are called Rayleigh scattering spectrum and Raman scattering spectra; the former appearing at the same wavelength position as the excitation spectrum, and the latter appearing at the longer-wavelength side near Rayleigh scattering. In a fluorescence spectrum, when the excitation wavelength is shifted, only the peak height is changed while the peak wavelength position remains intact. In a Raman scattering spectrum, when the excitation wavelength is shifted, the peak wavelength position is also changed accordingly. Both the Rayleigh scattering and Raman scattering are caused by a solvent which may be contained in the sample. When examining the spectral plot, be careful not to mistake these scattering effects for the fluorescence peak of interest. Table E-2 presents the Raman spectral peak position at each excitation wavelength for the purpose of reference. E.3.1 Temperature Dependency of Fluorescence Intensity E.3.2 Chemical Change in Sample due to Radiation E.3.3 Raman Scattering
A - 22 Fig. E-4 Raman Spectrum of Water Table E-2 Raman Peak Positions at Respective Excitation Wavelengths (excitation wavelength) Water Ethanol CyclohexaneCarbon Tetrachloride Chloroform 248 271 267 267 ⎯ ⎯ 313 350 344 344 320 346 365 416 405 408 375 410 405 469 459 458 418 461 Excitation wavelength and Raman peak position (nm) 436 511 500 499 450 502 Relative intensity Excitation wavelength Raman scattering
A - 23 In measurement of a high-concentration sample, a variety of error factors may be involved. The most significant error factor consists in that an excitation beam is absorbed at the entrance of a cell to prevent a sufficient level of excitation at the center of the cell. Figure E-5 illustrates an extreme case of this condition. Although fluorescence is emitted in the vicinity of the entrance for the excitation beam, it is not taken into the emission monochromator. Fig. E-5 Sample Having an Extremely High Concentration If only the incident point of excitation beam is bright, it is necessary to dilute the sample properly for measurement. The second significant error factor consists in extinction due to concentration. This condition is caused by preventing activation through interaction of molecules. The third significant error factor consists in re-absorption of fluorescence. As shown in Fig. E-6, this condition occurs due to overlapping between the short-wavelength side of fluorescence spectrum and the long-wavelength side of excitation spectrum. Therefore, it seems that the fluorescence spectrum has been shifted toward the long-wavelength side to some extent. In measurement of an ordinary kind of sample, however, this condition will not impede quantitative determination significantly Fig. E-6 Explanatory Illustration of Re-absorption E.3.4 Handling of High- concentration Samples Excitation beam Fluorescence Fluorescence is reflected here. Relative intensity Fluorescence spectrum Excitation spectrum Re-abso rption occurs here. Wavelength
A - 24 In any case, if there is a possibility of a measurement error due to high concentration of a sample, dilute the sample properly or carry out surface fluorescence measurement using a solid sample holder. Where the excitation and emission wavelengths are plotted near each other, care should be exercised not to mistake the Raman and Rayleigh scattering for the fluorescence spectrum as mentioned in E.3.3. Where the excitation and emission wavelengths are plotted apart from each other, care should be exercised not to mistake the second-order and third-order scattered radiation for the fluorescence spectrum. The second-order scattered radiation appears at a wavelength two times longer than the excitation wavelength, and the third-order scattered radiation occurs at a wavelength three times longer. For instance, if the excitation wavelength is 240 nm, the second-order and third-order radiation take place at 480 and 720 nm, respectively. For eliminating these scattered radiations, insert a short-wavelength cutoff filter in the path of fluorescing radiation (before emission monochromator). It is advisable to use the filter set (P/N 650-0157) which is available as an optional accessory. Since the fluorescence spectrophotometer provides high sensitivity, just a slight amount of contamination on a cell may have an adverse effect on results of measurement. To prevent this, treat the cell properly after its use. Do not leave the cell containing sample. In evaporation of a solvent, a residue of sample may adhere to the wall of the cell to cause contamination. In measurement of a very dilute sample, contamination on the inner and outer walls of the cell may cause a problem. If a droplet of sample solution is put on the outer wall of the cell in sample injection into it, wipe the cell with tissue paper and then set it into the cell holder. E.3.5 Second-order Scattered Radiation E.3.6 Contamination of Cell
A - 25 Figure E-7 shows a measurement example of fluorescence spectrum. ① Scattering of exciting radiation ② Raman spectrum of solvent ③ Fluorescence of impurities, solvent, etc. ④ Fluorescence of sample ⑤ Second-order spectrum of exciting radiation Fig. E-7 Measurement Example of Fluorescence Spectrum As shown in Fig. E-7, other peaks than a fluorescence peak of sample appear in measurement of fluorescence spectrum. With reference to this example, it is necessary to identify a fluorescence peak of sample. E.3.7 Measurement Example of Fluorescence Spectrum Relative intensity Wavelength
A - 26 APPENDIX F MEASUREMENT OF INSTRUMENTAL RESPONSE (CORRECTED SPECTRA) Spectrum correction is performed to enable measuring a true spectrum by eliminating instrumental response such as wavelength characteristics of the monochromator or detector (photomultiplier). The measurement of instrumental response is needed to perform spectrum correction. “Instrumental Response” is the function to measure and save the instrumental response. F.1 Measurement of Instrumental Response on Excitation Side This is the function to obtain the instrumental response on the excitation side such as wavelength characteristics of the excitation monochromator using Rhodamine B as a standard (quantum counter). The instrumental response is automatically read with a single wavelength scan operation. A spectrum is correctable within a range of 200 to 600 nm. WARNING Rhodamine B can cause injury if directly touched or accidentally ingested. When handling it, be sure to wear proper protective gear such as safety gloves and safety mask. If Rhodamine B adheres to the skin, wash it off with soap and plenty of water. Consult a physician when needed. If accidentally ingested, immediately consult physician. Pour Rhodamine B into a triangular cell in the procedure illustrated in Fig. F-1. The triangular cell filled with Rhodamine B should be in principle stored at a dark place. F.1.1 Handling of Rhodamine B
A - 27 Fig. F-1 Handling of Rhodamine B (1) Click the (analysis method) button on the Measurement toolbar. A box for setting your analysis method will appear. (2) Select the General tab. On the General tab page, specify “Wavelength scan” for the measurement mode. (3) Select the Instrument tab. (4) Set “Fluorescence” for the data mode, “400 V” for the photomultiplier voltage and “Excitation” for the scan mode. (5) Execute the Zero Adjust command from the Spectrophotometer menu to calibrate the zero point. Rhodamine B Cut the supplied ampoule of Rhodamine B with a cutter. Syringe (F649090) Suck the solution into a syringe. Open the cover of triangular cell and pour the solution into it. Fill the cell with the solution in a volume at least half the capacity and close the cover. F.1.2 Operating Procedures