Home > Hitachi > Fluorescence Spectrophotometer > Hitachi F7000 Instruction Manual

Hitachi F7000 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							A - 18 
    E.2  Advantages of Fluorometry 
     
    As contrasted with fluorometry, absorptiometry for a low-concentration 
    sample is explained in the following: 
    A sample having 99% transmittance to blank is taken as an example.   
    In the absorbance measurement of such a substance, inaccuracies 
    must always be taken into consideration.    Here, the inaccuracy is 
    assumed to be 0.1%.    Since it has an effect on both the blank and 
    sample, 
     
    Percent transmittance of blank  100.0 ± 0.1% 
    Percent transmittance of sample  99.0 ± 0.1% 
    Difference (proportional to sample concentration)  1.0 ± 0.2% 
     
    In this example, the uncertainty in concentration measurement is ±20%.   
    While in the fluorometry, a difference from zero level corresponds to the 
    concentration of sample.    Accordingly, the measurement accuracy is 
    as follows: 
     
    Output signal level at sample measurement  100 ± 0.1 
    Value corresponding to blank  0 ± 0.1 
    Difference (proportional to sample concentration)  100 ± 0.2 
     
    As can be seen from the above, the fluorometry is very advantageous 
    for analyzing a low-concentration sample since its uncertainty is in most 
    cases theoretically independent of the sample concentration. 
    Although actual practice may involve some error factors which increase 
    as the sample concentration becomes extremely low, the fluorometry is 
    capable of measuring low concentrations with an accuracy 100 times 
    higher than in absorptiometry. 
     
     
    Figure E-2 is an explanatory illustration of the foregoing description.     
    In absorptiometry, a difference between the quantity of incident radiation 
    I
    o and the quantity of transmitted radiation It is represented by signal Is.  
    A level at which the signal I
    s becomes almost equal to a noise level is 
    used as a detection limit. 
    In fluorometry, however, since the quantity of fluorescence I
    F itself is 
    represented as a signal, just a small amount of fluorescence can be 
    amplified electrically for enabling detection.  
    						
    							A - 19 
    Still more, since a fluorescence wavelength of a substance is different 
    from its excitation wavelength (incident light wavelength), the 
    fluorescence wavelength is not readily affected by the exciting radiation, 
    thereby contributing to ensuring high sensitivity. 
     
     
     
     
    Fig. E-2    Comparison between Absorptiometry and Fluorometry 
     
     
    In addition to high sensitivity, the fluorometry is advantageous in that 
    more information is attainable.    An emission spectrum is also available 
    besides an excitation spectrum which corresponds to an absorption 
    spectrum in absorptiometry. 
    The two kinds of wavelengths can be selected as desired, and a 
    fluorescence spectrum can be recorded using a properly selected 
    excitation wavelength (or vice versa).    Thus, quantitative and qualitative 
    analyses can be made for a sample containing plural components. 
    High 
    concentration Low 
    concentrationHigh 
    concentration Low 
    concentration
    (a)  Absorptiometry (b)  Fluorometry 
    Amplified  
    						
    							A - 20 
    Figure E-3 shows a simplified spectral graph of measurement of a 
    sample containing multiple components.    In absorptiometry, since only 
    the absorption spectrum can be measured, two or more component 
    wavelengths are presented.    If the absorption wavelengths are similar 
    to each other, each component cannot be separated in measurement.   
    In fluorometry, even if the absorption wavelengths are similar, a 
    difference in fluorescence makes it possible to select each fluorescence 
    wavelength properly.    Thus, each component can be separated in 
    measurement. 
     
     
     
    Fig. E-3    Measurement of Multi-component Sample 
     
     
    Table E-1 compares information attainable in absorptiometry and that in 
    fluorometry. 
     
    Table E-1  Comparison of Information Attainable in 
    Absorptiometry and Fluorometry 
     
    Absorptiometry Fluorometry 
    Absorption spectrum only 
    (corresponding to excitation spectrum 
    in fluorometry) • Excitation spectrum 
    • Fluorescence spectrum 
     
     
     
    Absorption spectra
    Component A
    Component BExcitation spectra Fluorescence spectra
    Component A Component B
    (a)  Absorptiometry (b)  Fluorometry  
    						
    							A - 21 
    E.3  Remarks on Measurement in   
    Fluorescence Analysis 
     
    For most kinds of samples, an increase of 1 °C in sample temperature 
    causes the fluorescence intensity to decrease by 1 to 2%. 
    It is also reported that for some kinds of biochemical samples, the 
    fluorescence intensity decreases as much as 10% as the temperature 
    increases by 1 °C.    When analyzing a sample having a temperature-
    dependent property, it is advisable to use the constant-temperature cell 
    holder (P/N 650-0150).    Constant-temperature measurement can be 
    carried out by circulating constant-temperature water through this cell 
    holder. 
     
     
    Some kinds of samples may be susceptible to a chemical change due to 
    exciting radiation.    In analysis of such a sample, keep the shutter 
    closed to cut off an excitation beam until measurement is started, and 
    then open the shutter immediately before measurement.    If any 
    chemical change due to exciting radiation is observed still, determine a 
    signal level at the start time through extrapolation according to variation 
    in signal level. 
     
     
    In fluorescence measurement, spectra having different natures from 
    that of fluorescence may be observed.    These are called Rayleigh 
    scattering spectrum and Raman scattering spectra; the former 
    appearing at the same wavelength position as the excitation spectrum, 
    and the latter appearing at the longer-wavelength side near Rayleigh 
    scattering. 
    In a fluorescence spectrum, when the excitation wavelength is shifted, 
    only the peak height is changed while the peak wavelength position 
    remains intact.    In a Raman scattering spectrum, when the excitation 
    wavelength is shifted, the peak wavelength position is also changed 
    accordingly.    Both the Rayleigh scattering and Raman scattering are 
    caused by a solvent which may be contained in the sample.     
    When examining the spectral plot, be careful not to mistake these 
    scattering effects for the fluorescence peak of interest.    Table E-2 
    presents the Raman spectral peak position at each excitation 
    wavelength for the purpose of reference. 
     
    E.3.1 Temperature 
    Dependency of 
    Fluorescence 
    Intensity 
    E.3.2 Chemical 
    Change in 
    Sample due to 
    Radiation 
    E.3.3 Raman 
    Scattering 
      
    						
    							A - 22 
     
     
     
    Fig. E-4    Raman Spectrum of Water 
     
     
    Table E-2    Raman Peak Positions at Respective Excitation Wavelengths 
     
     (excitation 
    wavelength) Water Ethanol CyclohexaneCarbon 
    Tetrachloride Chloroform
    248 271 267  267 ⎯ ⎯ 
    313 350 344  344  320  346 
    365 416 405  408  375  410 
    405 469 459  458  418  461 
    Excitation 
    wavelength 
    and Raman 
    peak 
    position (nm) 
    436 511 500  499  450  502 
     
    Relative intensity 
    Excitation wavelength Raman 
    scattering  
    						
    							A - 23 
    In measurement of a high-concentration sample, a variety of error 
    factors may be involved. 
    The most significant error factor consists in that an excitation beam is 
    absorbed at the entrance of a cell to prevent a sufficient level of 
    excitation at the center of the cell. 
    Figure E-5 illustrates an extreme case of this condition.     
    Although fluorescence is emitted in the vicinity of the entrance for the 
    excitation beam, it is not taken into the emission monochromator. 
     
     
     
     
     
     
     
     
     
    Fig. E-5    Sample Having an Extremely High Concentration 
     
     
    If only the incident point of excitation beam is bright, it is necessary to 
    dilute the sample properly for measurement. 
    The second significant error factor consists in extinction due to 
    concentration.    This condition is caused by preventing activation 
    through interaction of molecules. 
    The third significant error factor consists in re-absorption of 
    fluorescence.    As shown in Fig. E-6, this condition occurs due to 
    overlapping between the short-wavelength side of fluorescence 
    spectrum and the long-wavelength side of excitation spectrum.   
    Therefore, it seems that the fluorescence spectrum has been shifted 
    toward the long-wavelength side to some extent.    In measurement of 
    an ordinary kind of sample, however, this condition will not impede 
    quantitative determination significantly 
     
     
     
     
    Fig. E-6    Explanatory Illustration of Re-absorption 
    E.3.4 Handling of 
    High-
    concentration 
    Samples 
    Excitation beam 
    Fluorescence
    Fluorescence is 
    reflected here. 
    Relative intensity 
    Fluorescence 
    spectrum  Excitation 
    spectrum 
    Re-abso
    rption 
    occurs here. Wavelength  
    						
    							A - 24 
    In any case, if there is a possibility of a measurement error due to high 
    concentration of a sample, dilute the sample properly or carry out 
    surface fluorescence measurement using a solid sample holder. 
     
     
    Where the excitation and emission wavelengths are plotted near each 
    other, care should be exercised not to mistake the Raman and Rayleigh 
    scattering for the fluorescence spectrum as mentioned in E.3.3.   
    Where the excitation and emission wavelengths are plotted apart from 
    each other, care should be exercised not to mistake the second-order 
    and third-order scattered radiation for the fluorescence spectrum. 
    The second-order scattered radiation appears at a wavelength two 
    times longer than the excitation wavelength, and the third-order 
    scattered radiation occurs at a wavelength three times longer. 
    For instance, if the excitation wavelength is 240 nm, the second-order 
    and third-order radiation take place at 480 and 720 nm, respectively.   
    For eliminating these scattered radiations, insert a short-wavelength 
    cutoff filter in the path of fluorescing radiation (before emission 
    monochromator).    It is advisable to use the filter set (P/N 650-0157) 
    which is available as an optional accessory. 
     
     
    Since the fluorescence spectrophotometer provides high sensitivity, just 
    a slight amount of contamination on a cell may have an adverse effect 
    on results of measurement.    To prevent this, treat the cell properly after 
    its use.    Do not leave the cell containing sample.    In evaporation of a 
    solvent, a residue of sample may adhere to the wall of the cell to cause 
    contamination. 
    In measurement of a very dilute sample, contamination on the inner and 
    outer walls of the cell may cause a problem.    If a droplet of sample 
    solution is put on the outer wall of the cell in sample injection into it, wipe 
    the cell with tissue paper and then set it into the cell holder. 
     
    E.3.5 Second-order 
    Scattered 
    Radiation 
    E.3.6 Contamination 
    of Cell  
    						
    							A - 25 
    Figure E-7 shows a measurement example of fluorescence spectrum. 
     
     
     
     
    ①  Scattering of exciting radiation 
    ②  Raman spectrum of solvent 
    ③  Fluorescence of impurities, solvent, etc. 
    ④  Fluorescence of sample 
    ⑤  Second-order spectrum of exciting radiation 
     
    Fig. E-7    Measurement Example of Fluorescence Spectrum 
     
     
    As shown in Fig. E-7, other peaks than a fluorescence peak of sample 
    appear in measurement of fluorescence spectrum.    With reference to 
    this example, it is necessary to identify a fluorescence peak of sample. 
     
    E.3.7 Measurement 
    Example of 
    Fluorescence 
    Spectrum 
    Relative intensity 
    Wavelength 
    						
    							A - 26 
    APPENDIX F  MEASUREMENT OF 
    INSTRUMENTAL RESPONSE 
    (CORRECTED SPECTRA) 
     
    Spectrum correction is performed to enable measuring a true spectrum 
    by eliminating instrumental response such as wavelength characteristics 
    of the monochromator or detector (photomultiplier).    The measurement 
    of instrumental response is needed to perform spectrum correction.   
    “Instrumental Response” is the function to measure and save the 
    instrumental response. 
     
     
    F.1  Measurement of Instrumental Response on 
    Excitation Side 
     
    This is the function to obtain the instrumental response on the excitation 
    side such as wavelength characteristics of the excitation 
    monochromator using Rhodamine B as a standard (quantum counter).   
    The instrumental response is automatically read with a single 
    wavelength scan operation.    A spectrum is correctable within a range 
    of 200 to 600 nm. 
     
    WARNING 
    Rhodamine B can cause injury if directly touched or accidentally 
    ingested.    When handling it, be sure to wear proper protective 
    gear such as safety gloves and safety mask.    If Rhodamine B 
    adheres to the skin, wash it off with soap and plenty of water.   
    Consult a physician when needed.    If accidentally ingested, 
    immediately consult physician. 
     
     
    Pour Rhodamine B into a triangular cell in the procedure illustrated in 
    Fig. F-1.    The triangular cell filled with Rhodamine B should be in 
    principle stored at a dark place. 
    F.1.1 Handling of 
    Rhodamine B 
      
    						
    							A - 27 
     
     
     
     
     
     
    Fig. F-1    Handling of Rhodamine B 
     
     
    (1) Click the 
      (analysis method) button on the Measurement 
    toolbar.    A box for setting your analysis method will appear. 
     
    (2)  Select the General tab.    On the General tab page, specify 
    “Wavelength scan” for the measurement mode. 
     
    (3)  Select the Instrument tab. 
     
    (4)  Set “Fluorescence” for the data mode, “400 V” for the 
    photomultiplier voltage and “Excitation” for the scan mode. 
     
    (5)  Execute the Zero Adjust command from the Spectrophotometer 
    menu to calibrate the zero point. 
     
    Rhodamine B
    Cut the supplied 
    ampoule of 
    Rhodamine B with 
    a cutter. 
    Syringe 
    (F649090)
    Suck the 
    solution into a 
    syringe. Open the cover 
    of triangular cell 
    and pour the 
    solution into it. Fill the cell with the 
    solution in a volume 
    at least half the 
    capacity and close 
    the cover. 
    F.1.2 Operating 
    Procedures 
      
    						
    All Hitachi manuals Comments (0)