Hitachi F7000 Instruction Manual
Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
3.3 Defining Your Samples 3 - 16 When “Use sample table” is selected in Analysis Method, samples must be defined in the following procedure before measurement. Open the monitor window and click the button. When the Sample Table window in Fig. 3-11 appears, enter the sample name and comment for each sample. 1 For creating a sample table newly, make entry for Number of Samples and click the Update button. For instance, if there are five samples, enter “5” for Number of Samples and click the Update button. Fig. 3-11 Sample Table 2 For adding a row at the end of sample table, click the Insert button. 3 To delete an unnecessary row, click its row No. to make it active as shown in Fig. 3-12. Then, click the Delete button. 3.3.2 With Sample Table 2 1
3.3.2 3 - 17 Fig. 3-12 4 Make entries for Sample and Comments of each sample. The sample names entered here appear on the monitor window. Maximum number of characters to be entered (half-size) Sample : 40 characters Comment : 255 characters Next, you must enter a file name for automatic saving after measurement. When you click a File Name column, the [...] button appears as shown in Fig. 3-13. Click this button. Upon opening of the window shown in Fig. 3-14, specify a saving location and enter a file name. Fig. 3-13 3 4 5 67 8
3.3 Defining Your Samples 3 - 18 Fig. 3-14 5 After entries, check them and then click the OK button. 6 Load button Used for reading out the existing sample table. Clicking the Load button displays the file opening dialog. Fig. 3-15 File Opening Dialog Select a sample table and click the Open button. The specified sample table will be read out.
3.3.2 3 - 19 7 Save button The contents of change can be saved by overwriting. In case of a new generation, the Save As dialog will open. 8 Save As button Used for saving the sample table with a name newly given. Click this button, and the Save As dialog will open. Fig. 3-16 Save As Dialog Select a saving location and enter a file name. Then, click the Save button.
3.4 Conducting Measurement 3 - 20 3.4 Conducting Measurement (1) When “Use sample table” is not selected in Analysis Method, you should click the button and make sure that a sample has already been defined. (2) Set a sample and click (measurement) button. Measurement starts and a spectrum is displayed. For halting the measurement midway, click the (stop) button. (3) Upon completion of the measurement, the measured data will be saved automatically at the specified saving location using the specified file name when auto saving is selected. If auto saving is not selected, select the “Save As” command from the File menu and save the data. Opening the saved data file will display the data processing window as shown in Fig. 3-17. This figure shows a spectrum and the result of rate calculation. Peak table can be displayed, but both the result of rate calculation and the peak table cannot be displayed simultaneously. Simultaneous display is selectable from the View menu. Fig. 3-17 Data Processing Window (time scan) 3.4.1 Without Using Sample Table
3.4.2 3 - 21 (1) Click the button and make sure a sample has already been defined. (2) Set a sample and click the button to start measurement. (3) Upon start of measurement, the window shown in Fig. 3-18 appears. Click Yes. Fig. 3-18 (4) A dialog box opens as shown below. Fig. 3-19 Click OK. Measurement starts and a spectrum is displayed. For halting the measurement midway, click the (stop) button. Measure all the samples set in the sample table according to the message guide. 3.4.2 With Sample Table
3.4 Conducting Measurement 3 - 22 (5) Upon completion of the measurement, the measured data will be saved automatically at the specified saving location using the specified file name. Opening the saved data file will display the data processing window as shown in Fig. 3-20. This figure shows a spectrum and the result of rate calculation. Peak table can be displayed, but both the result of rate calculation and the peak table cannot be displayed simultaneously. Simultaneous display is selectable from the View menu. Fig. 3-20 Data Processing Window (wavelength scan) ① Spectrum The spectrum of a measured sample is displayed. ② Result of rate calculation The result of rate calculation (slope, activity, R, R2) are displayed. Start time and end time are settable by numeric input or clicking Set with the line cursor aligned. After setting, click Update for recalculation. For rate calculation, refer to Appendix B. When peak table is selected, peak time, valley time, peak start/end times and peak area will be displayed. ①Spectrum ②Result of rate calculation
3.4.3 3 - 23 (1) Put a check mark for ‘Stopped Flow’ at the Instrument tab in the Analysis Method window. At the General tab in this window, remove a check mark from “Use sample table.” (2) Click the button to make sure a sample has already been defined. (3) Click the (measurement) button to start measurement. (4) When measurement starts, the window shown in 3-21 appears to await an external start signal. Fig. 3-21 Stopped Flow Measurement Window (external signal wait status) 3.4.3 Measurement in Stopped Flow Mode
3.4 Conducting Measurement 3 - 24 (5) On arrival of the external signal, measurement starts and a spectrum is drawn in the window. Fig. 3-22 Stopped Flow Measurement Window For stopping the measurement halfway, click the (stop) button. (6) Upon completion of the measurement, the measured data will be saved automatically at the specified saving location using the specified file name when auto saving is selected. If auto saving is not selected, select the “Save As” command from the File menu and save the data.
3.4.3 3 - 25 NOTICE: 1. The data measured under the following conditions will be saved automatically in Program Files\FL Solutions\Data. (Example: C:Program Files\FL Solutions\Data) (1) Measurement was carried out using a sample table with no file name specified in the File Name column. (2) Measurement was carried out with “Auto File” specified and a file name unspecified, avoiding use of a sample table. 2. The data measured under the following conditions will be saved automatically in Program Files\FL Solutions. (Example: C:Program Files\FL Solutions) (3) There is not an adequate free area in the specified storage medium (such as floppy disk or MO). (4) Writing is inhibited. A file automatically saved under such a condition as above is named according to the following rule. Rule 1: When a file name is given to analysis method: File name of analysis method + serial No. (001, 002, ...) Rule 2: When a file name is not given to analysis method: DEFLT + serial No. (001, 002, ...)