Home > Hitachi > Fluorescence Spectrophotometer > Hitachi F7000 Instruction Manual

Hitachi F7000 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							3.3  Defining Your Samples 
    3 - 16 
    When “Use sample table” is selected in Analysis Method, samples must 
    be defined in the following procedure before measurement.   
    Open the monitor window and click the 
     button.  
    When the Sample Table window in Fig. 3-11 appears, enter the sample 
    name and comment for each sample.   
     
    1 For creating a sample table newly, make entry for Number of 
    Samples and click the Update button.    For instance, if there are 
    five samples, enter “5” for Number of Samples and click the 
    Update button.   
     
     
     
    Fig. 3-11  Sample Table 
     
     
    2 For adding a row at the end of sample table, click the Insert button.   
    3 To delete an unnecessary row, click its row No. to make it active as 
    shown in Fig. 3-12.    Then, click the Delete button.   
    3.3.2 With Sample 
    Table 
     
    2
    1 
    						
    							3.3.2 
    3 - 17 
     
     
    Fig. 3-12 
     
     
    4 Make entries for Sample and Comments of each sample.     
    The sample names entered here appear on the monitor window.   
    Maximum number of characters to be entered (half-size)   
    Sample : 40 characters  
    Comment :  255 characters  
     
    Next, you must enter a file name for automatic saving after 
    measurement.  
    When you click a File Name column, the [...] button appears as 
    shown in Fig. 3-13.    Click this button.    Upon opening of the 
    window shown in Fig. 3-14, specify a saving location and enter a 
    file name.   
     
     
    Fig. 3-13 
    3
    4
    5
    67 8  
    						
    							3.3  Defining Your Samples 
    3 - 18 
     
     
     
    Fig. 3-14   
     
     
    5 After entries, check them and then click the OK button.   
    6 Load button   
    Used for reading out the existing sample table.   
    Clicking the Load button displays the file opening dialog.   
     
     
     
    Fig. 3-15    File Opening Dialog 
     
     
    Select a sample table and click the Open button.  The specified 
    sample table will be read out.   
      
    						
    							3.3.2 
    3 - 19 
    7 Save button   
    The contents of change can be saved by overwriting.    In case of a 
    new generation, the Save As dialog will open.   
    8 Save As button   
    Used for saving the sample table with a name newly given.   
    Click this button, and the Save As dialog will open.   
     
     
     
    Fig. 3-16    Save As Dialog 
     
     
    Select a saving location and enter a file name.    Then, click the 
    Save button.   
      
    						
    							3.4  Conducting Measurement 
    3 - 20 
    3.4  Conducting Measurement  
     
    (1)  When “Use sample table” is not selected in Analysis Method, you 
    should click the 
      button and make sure that a sample has 
    already been defined.   
     
    (2)  Set a sample and click 
     (measurement) button.  
    Measurement starts and a spectrum is displayed.   
    For halting the measurement midway, click the 
     (stop) 
    button.   
     
    (3)  Upon completion of the measurement, the measured data will be 
    saved automatically at the specified saving location using the 
    specified file name when auto saving is selected.     
    If auto saving is not selected, select the “Save As” command from 
    the File menu and save the data.   
    Opening the saved data file will display the data processing 
    window as shown in Fig. 3-17.    This figure shows a spectrum 
    and the result of rate calculation.    Peak table can be displayed, 
    but both the result of rate calculation and the peak table cannot 
    be displayed simultaneously.    Simultaneous display is selectable 
    from the View menu.   
     
     
     
    Fig. 3-17    Data Processing Window (time scan) 
     
     
    3.4.1 Without Using 
    Sample Table  
    						
    							3.4.2 
    3 - 21 
    (1) Click the   button and make sure a sample has already been 
    defined.  
     
    (2)  Set a sample and click the 
      button to start measurement.   
     
    (3)  Upon start of measurement, the window shown in Fig. 3-18 
    appears.  Click Yes.  
     
     
     
    Fig. 3-18 
     
     
    (4)  A dialog box opens as shown below.   
     
     
     
    Fig. 3-19 
     
     
    Click OK.    Measurement starts and a spectrum is displayed.   
    For halting the measurement midway, click the 
     (stop) 
    button.  
    Measure all the samples set in the sample table according to the 
    message guide.   
    3.4.2 With Sample 
    Table  
    						
    							3.4  Conducting Measurement 
    3 - 22 
    (5)  Upon completion of the measurement, the measured data will be 
    saved automatically at the specified saving location using the 
    specified file name.    Opening the saved data file will display the 
    data processing window as shown in Fig. 3-20.    This figure 
    shows a spectrum and the result of rate calculation.    Peak table 
    can be displayed, but both the result of rate calculation and the 
    peak table cannot be displayed simultaneously.     
    Simultaneous display is selectable from the View menu.   
     
     
     
    Fig. 3-20    Data Processing Window (wavelength scan) 
     
     
    ① Spectrum  
    The spectrum of a measured sample is displayed.   
     
    ②  Result of rate calculation   
    The result of rate calculation (slope, activity, R, R2) are 
    displayed.    Start time and end time are settable by numeric 
    input or clicking Set with the line cursor aligned.     
    After setting, click Update for recalculation.    For rate 
    calculation, refer to Appendix B.   
     
    When peak table is selected, peak time, valley time, peak start/end 
    times and peak area will be displayed.     
     
     
    ①Spectrum 
    ②Result of rate calculation  
    						
    							3.4.3 
    3 - 23 
    (1)  Put a check mark for ‘Stopped Flow’ at the Instrument tab in the 
    Analysis Method window.    At the General tab in this window, 
    remove a check mark from “Use sample table.”   
     
    (2) Click the 
      button to make sure a sample has already been 
    defined.  
     
    (3) Click the 
      (measurement) button to start measurement.   
     
    (4)  When measurement starts, the window shown in 3-21 appears to 
    await an external start signal.     
     
     
     
    Fig. 3-21  Stopped Flow Measurement Window   
    (external signal wait status) 
     
     
    3.4.3 Measurement 
    in Stopped 
    Flow Mode  
    						
    							3.4  Conducting Measurement 
    3 - 24 
    (5)  On arrival of the external signal, measurement starts and a 
    spectrum is drawn in the window.   
     
     
     
    Fig. 3-22    Stopped Flow Measurement Window   
     
     
    For stopping the measurement halfway, click the 
     (stop) 
    button.  
     
    (6)  Upon completion of the measurement, the measured data will be 
    saved automatically at the specified saving location using the 
    specified file name when auto saving is selected.     
    If auto saving is not selected, select the “Save As” command from 
    the File menu and save the data.   
      
    						
    							3.4.3 
    3 - 25 
    NOTICE:  1.  The data measured under the following conditions will 
    be saved automatically in Program Files\FL 
    Solutions\Data.  (Example:  C:Program Files\FL 
    Solutions\Data)  
    (1)  Measurement was carried out using a sample table 
    with no file name specified in the File Name 
    column.  
    (2)  Measurement was carried out with “Auto File” 
    specified and a file name unspecified, avoiding use 
    of a sample table.   
    2.  The data measured under the following conditions will 
    be saved automatically in Program Files\FL Solutions.     
    (Example:  C:Program Files\FL Solutions)  
    (3)  There is not an adequate free area in the specified 
    storage medium (such as floppy disk or MO).     
    (4)  Writing is inhibited.   
    A file automatically saved under such a condition as above is 
    named according to the following rule.   
    Rule 1:  When a file name is given to analysis method: 
    File name of analysis method + serial No. (001, 
    002, ...)   
    Rule 2:  When a file name is not given to analysis method:   
    DEFLT + serial No. (001, 002, ...)   
     
      
    						
    All Hitachi manuals Comments (0)