Home > Hitachi > Fluorescence Spectrophotometer > Hitachi F7000 Instruction Manual

Hitachi F7000 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							4.6 
    4 - 45 
    4.6    Manual Input of Standard Data   
     
    The standard data can be entered manually.    This function is usable 
    after measurement of standards.    Either select Standard Data from the 
    Edit menu on the monitor window or click the 
     button.  A window 
    as shown in Fig. 4-47 will appear.    So enter values in the Data column.   
    After entry, click the OK or Update button.    Then, the calibration curve 
    is recalculated and displayed.   
     
     
     
    Fig. 4-47 
     
      
    						
    							4.7    Measurement Using Saved Calibration Curve 
    4 - 46 
    4.7    Measurement Using Saved Calibration Curve   
     
    Call a calibration curve in the following way.   
     
    (1)  Set measurement mode in Photometry.   
     
    (2)  From the File menu, select the calibration curve load command.   
    The following message is reported in a dialog box.     
    This message does not appear immediately after turning on 
    power supply or when measurement mode is other than 
    Photometry.  
    Click the OK button.   
     
     
     
    Fig. 4-48 
     
     
    (3)  A window as shown below will appear.    Select the file of the 
    calibration curve to be used.   
    After file selection, click the Open button.   
     
     
     
    Fig. 4-49 
      
    						
    							4.7 
    4 - 47 
    (4)  The window in Fig. 4-50 is now displayed.    The sample data 
    section is blank.     
     
     
     
    Fig. 4-50 
     
     
     (5)  Click the 
      button, and the window in Fig. 4-51 will appear.   
    Enter a sample name and comment.   
    Click the OK button.   
     
     
     
    Fig. 4-51  Sample data 
    section  
    						
    							4.7    Measurement Using Saved Calibration Curve 
    4 - 48 
    (6)  Set a sample to be measured.    Click the  button.   
    (Or select the Scan command from the Spectrophotometer 
    menu.)  
     
    (7)  The following message box appears.   
    Click the OK button.   
     
     
     
    Fig. 4-52 
     
     
    (8) The Sample button is now active.    Click this button.   
    Measurement starts.   
     
     
     
    Fig. 4-53 
     
     
    (9)  After measurement, click the End(F9) button for termination.    
    						
    							5.1 
    5 - 1 
    5.  3-D SCAN  
     
    5.1  Flow of Operation  
     
    This section explains the parameter setting and operating method for 3-
    D scan.    Measurement proceeds along the following flow.   
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
    Fig. 5-1    Operational Flow (3-D scan)   Click 
       button and set your analytical conditions. 
     
     • Scan range 
    • Data mode 
    •  Scan speed, etc. 
    Click 
     button. 
     • Sample name 
    • Auto File 
    •  File name, etc. 
    Set a sample. 
    Click 
      button to execute wavelength scan. 
    • Spectrum tracing 
    • Peak detection 
    • Spectral calculation 
    Click 
      button.    Measurement result will be displayed 
    in a report format. 
     
    Start 
    Method setting 
    (See 5.2.) 
    Sample name/ 
    comment input 
    (See 5.3.) 
    Measurement 
    (See 5.4.) 
    Data processing 
    (See Section 7.) 
    Result output in report 
    (See 5.9.) 
    End  
    						
    							5.2    Creating an Analysis Method 
    5 - 2 
    5.2    Creating an Analysis Method   
     
    Set the analytical conditions (method) prior to measurement.   
     
    (1) Click the 
      button on the Measurement toolbar.    The window 
    in Fig. 5-2 appears.   
    (Or select the Method command from the Edit menu.)     
     
     
     
    Fig. 5-2  Analysis Method  
     
     
    (2)  The following tabs are provided on the Analysis Method window 
    for the 3-D scan mode.    These tabs are detailed on the 
    subsequent pages.   
    • General  
    • Instrument  
    • Monitor  
    • Processing  
    • Report  
     
      
    						
    							5.2.1 
    5 - 3 
    The parameters shown in Fig. 5-3 are provided at the General tab.     
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 5-3  General Tab 
     
     
    1 Measurement (mode)   
    A measurement mode is selectable.    Here, select “3-D Scan.”   
    For Wavelength scan, Time scan, Photometry and 3-D Time scan, 
    refer to Sections 2, 3, 4 and 6, respectively.   
    2 Operator  
    Enter an operator name.   
    3 Instrument  
    The name of the connected instrument is indicated.   
    4 Sampling 
    - Standard   
    Indicates that samples are to be exchanged manually.     
    5 Comments  
    Enter a description or notes on measuring conditions.   
    5.2.1 General Tab 
     
    1 
    2 
    3 
    4 
    5 
    6 
    7 
    8  
    						
    							5.2    Creating an Analysis Method 
    5 - 4 
    6 Load  
    Click this button to read out saved analytical conditions.     
    Upon clicking, the file opening window appears.    So, select an 
    analytical condition file.   
    7 Save  
    When you click this button after setting the parameters, the 
    analytical conditions can be saved by overwriting.    In case of a 
    new analysis method, clicking this button displays the “Save As” 
    window.    You must enter a file name for saving in this window.   
    8 Save As   
    Click this button for renaming the set conditions for saving.     
    The “Save As” window appears, so enter a file name.   
     
      
    						
    							5.2.2 
    5 - 5 
    The window shown below is based on that 3-D Scan is specified for 
    Measurement at the General tab.    When you click the Instrument tab, 
    the window in Fig. 5-4 appears.   
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 5-4  Instrument Tab 
     
     
    1 Data mode   
    Select a data mode.    The Phosphorescence mode is for acquiring 
    only phosphorescence after extinction of excitation beam while 
    turning a chopper on the excitation side.   
    2 Chopping speed   
    Select a chopping speed.    Setting is allowed when 
    Phosphorescence is selected for Data mode.     
    The Phosphorescence mode is for acquiring only phosphorescence 
    after extinction of excitation beam while turning a chopper on the 
    excitation side.    An increase in chopping speed makes the 
    chopper turns faster.   
    3 EX Start WL   
    This is the start wavelength on the excitation side in wavelength 
    scan.    Enter the shorter wavelength side of measuring wavelength 
    range.  
    5.2.2 Instrument 
    Tab 
     
    3 
    4 
    5 
    6 
    7 
    17 
    2 
    1 
    8 
    9 
    12
    13
    14
    15
    16
    11
    10
    18  
    						
    							5.2    Creating an Analysis Method 
    5 - 6 
    4 EX End WL   
    This is the end wavelength on the excitation side in wavelength 
    scan.    Enter the longer wavelength side of measuring wavelength 
    range.  
    5 EX Sampling interval   
    In the 3-dimensional scan mode, emission spectrum measurement 
    is repeated while shifting the excitation wavelength.    Therefore, a 
    shorter sampling interval on the excitation side will result in a longer 
    measurement time.   
    6 EM Start WL   
    This is the start wavelength on the emission side in wavelength 
    scan.    Enter the shorter wavelength side of measuring wavelength 
    range.  
    7 EM End WL   
    This is the end wavelength on the emission side in wavelength 
    scan.    Enter the longer wavelength side of measuring wavelength 
    range.  
    8 EM Sampling interval   
    In the 3-dimensional scan mode, emission spectrum measurement 
    is repeated while shifting the excitation wavelength.    Therefore, a 
    shorter sampling interval on the emission side will result in a longer 
    measurement time.   
    9 Scan speed   
    Set a wavelength scan speed.   
    10 EX Slit   
    Select a slit width on the excitation side.   
    11 EM Slit   
    Select a slit width on the emission side.   
    12 PMT Voltage   
    A function for controlling the voltage of the photomultiplier detector 
    (photomultiplier tube).    As a reference, changing from 400 V to 
    700 V increases the data value about two digits and changing from 
    700 V to 950 V increases it about 1 digit.   
    When a check mark is put for PMT Voltage 0-1000V, the 
    photomultiplier voltage is settable within a range from 0 to 1000 V.   
    13 PMT Voltage 0-1000V   
    When a check mark is put in this box, the photomultiplier voltage is 
    settable within a range from 0 to 1000 V.    
    						
    All Hitachi manuals Comments (0)