Home > Hitachi > Fluorescence Spectrophotometer > Hitachi F7000 Instruction Manual

Hitachi F7000 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							A - 28 
    (6)  As shown in Fig. F-2, set the triangular cell filled with Rhodamine 
    B into the cell holder in the sample compartment, the filter holder 
    to the sample compartment and the red filter (single-side frosted) 
    in the filter holder.    In so doing, pay attention to the direction of 
    the triangular cell and filter.    (Refer to Fig. F-3) 
     
     
     
     
    Fig. F-2    Setting of Triangular Cell Containing Rhodamine B 
     
     
     
     
    Fig. F-3    Setting Direction of Triangular Cell (top view) 
     
     
    NOTE:  It may take time to obtain stable data due to characteristics of 
    the detector immediately after the photomultiplier voltage has 
    been remarkably changed (from 700 V to 250 V for example).   
    When more accurately calibrating the zero point, wait until 
    stable data is obtained and then proceed to step (5). 
      Triangular cell with 
    fixture (containing 
    Rhodamine B) 
    Filter holde
    rRed filter 
    Metal part
    Excitation light
    Triangular cellFluorescenceRed filter   
    (single-side frosted)
    Frosted  
    						
    							A - 29 
    (7)  Execute the Instrumental Response command from the Utility 
    menu.    The Instrumental Response dialog box will open as 
    shown in Fig. F-4. 
    Select “EX [200 to 600 nm].”    (Refer to ①  in Fig. F-4.) 
     
     
     
    Fig. F-4 
     
     
    (8) Click the Run button.  (Refer to ②  in Fig. F-4.) 
    A message box “Set the triangular cell with Rhodamine B to cell 
    holder and set the red filter.” will appear.    After setting the 
    Rhodamine B and red filter, click the OK button.  (Refer to ③ in 
    Fig. F-5.) 
     
     
     
    Fig. F-5 
     
     
    1 
    2
    3 
    						
    							A - 30 
    (9)  The EX wavelength is set at 640 nm and the instrument starts EX 
    wavelength scan (from 198.6 to 601.4 nm). 
    At this stage, the following measuring conditions are automatically 
    set. 
    • Scan mode :  Excitation 
    • Data mode :  Fluorescence 
    •  EX slit  :  5.0 nm 
    •  EM slit  :  20.0 nm 
    •  Scan speed  :  60 nm/min 
    •  PMT voltage :  400 V 
    • Response  :  Auto 
     
    (10)  Upon completion of measurement, the conditions set before 
    measurement are retrieved. 
    Now, the instrument response on the excitation side has been 
    measured and stored.    Take out the rectangular cell. 
      
    						
    							A - 31 
    F.2  Measurement of Instrumental Response on   
    Emission Side 
     
    This is the function to obtain the instrumental response on the emission 
    side such as wavelength characteristics of the emission monochromator 
    or detector (photomultiplier). 
    The instrumental response on the emission side is determined by 
    measuring a combination of instrumental response on both the 
    excitation and emission sides by synchronous wavelength scan, and 
    dividing it by the instrumental response on the excitation side 
    preliminarily measured.    An emission spectrum is correctable within a 
    range of 200 to 600 nm. 
     
     
    (1) Click the 
      (analysis method) button on the Measurement 
    toolbar.    A box for setting your analysis method will appear. 
     
    (2)  Select the General tab.    On the General tab page, specify 
    “Wavelength scan” for the measurement mode. 
     
    (3)  Select the Instrument tab. 
     
    (4)  Set “Fluorescence” for the data mode and “250 V” for the 
    photomultiplier voltage. 
     
    (5)  Execute the Zero Adjust command from the Spectrophotometer 
    menu to calibrate the zero point. 
     
    (6)  As shown in Fig. F-6, set the diffuser into the cell holder in the 
    sample compartment.   
     
     
     
    Fig. F-6    Setting of Diffuser 
    F.2.1 Overview 
    F.2.2 Operating 
    Procedures 
     
    Diffuser  
    						
    							A - 32 
    NOTES:  1.  Measurement of the instrumental response on the 
    excitation side must be completed before executing 
    measurement of the instrumental response on the 
    emission side. 
      2.  It may take time to obtain stable data due to 
    characteristics of the detector immediately after the 
    photomultiplier voltage has been remarkably changed 
    (from 700 V to 250 V for example).    When more 
    accurately calibrating the zero point, wait until stable 
    data is obtained and then proceed to step (5). 
     
    (7)  Execute the Instrumental Response command from the Utility 
    menu.    The Instrumental Response dialog box will open as 
    shown in Fig. F-7. 
    Select “EM [200 to 600 nm].”    (Refer to ①  in Fig. F-7.) 
     
     
     
    Fig. F-7 
     
     
    (8) Click the Run button.  (Refer to ② in Fig. F-7.)  A message 
    box “Set diffuser to cell holder.” will appear.    Click the OK button.   
    (Refer to ③  in Fig. F-8.) 
     
     
     
    Fig. F-8 
    1 
    2
    3 
    						
    							A - 33 
    (9)  The EX and EM wavelengths are set at 198.6 nm and the 
    instrument starts synchronous wavelength scan (from 198.6 to 
    601.4 nm). 
    At this stage, the following measuring conditions are automatically 
    set. 
    • Scan mode :  Synchronous 
    • Data mode :  Fluorescence 
    •  EX slit  :  5.0 nm 
    •  EM slit  :  20.0 nm 
    •  Scan speed  :  60 nm/min 
    •  PMT voltage :  250 V 
    • Response  :  Auto 
     
    (10)  Upon completion of measurement, the conditions set before 
    measurement are retrieved. 
    Now, the instrument response on the excitation side has been 
    measured and stored.    Take out the diffuser. 
     
      
    						
    							A - 34 
    F.3  Measurement of Instrumental Response in 
    Long Wavelength Range on Emission Side 
     
    When a corrected spectrum in the long wavelength area of 600 nm or 
    more is required on the emission side, the beam of a lamp whose 
    wavelength characteristics are known is allowed to be irradiated to the 
    emission monochromator, and the instrumental response in the long 
    wavelength range on the emission side is read for correction. 
    As a light source, a sub standard light source (P/N 250-0123) is used.   
    The correction range by it is 500 to 900 nm and, when combined with an 
    instrumental response in the short wavelength range, spectrum 
    correction is available over the entire wavelength range (200 to 900 nm) 
    on the emission side. 
     
     
     
    Refer to the instruction manual for the sub standard light source 
    (optional accessory). 
     
    (1)  Close the cover of the sample compartment. 
     
    (2) Click the 
      (analysis method) button on the Measurement 
    toolbar.    A box for setting your analysis method will appear. 
     
    (3)  Select the General tab.    On the General tab page, specify 
    “Wavelength scan” for the measurement mode. 
     
    (4)  Select the Instrument tab. 
     
    (5)  Set “Fluorescence” for the data mode and “400 V” for the 
    photomultiplier voltage. 
     
    (6)  Execute the Zero Adjust command from the Spectrophotometer 
    menu to calibrate the zero point. 
     
    (7)  Set the sub standard light source in the sample compartment with 
    reference to its instruction manual.    (Refer to Fig. F-9.) 
    Turn on the power switch of the sub standard light source and 
    wait for approx. 5 minutes until the lamp stabilizes. 
    F.3.1 Overview 
     
    F.3.2 Operating 
    Procedures  
    						
    							A - 35 
     
     
     
    Fig. F-9 
     
     
    (8)  Execute the Instrumental Response command from the Utility 
    menu.    The Instrumental Response dialog box will open as 
    shown in Fig. F-10. 
    Select “EM [500 to 900 nm].”    (Refer to ①  in Fig. F-10.) 
     
     
     
    Fig. F-10 
     
     
    1 
    2 
    						
    							A - 36 
    (9) Click the Run button.  (Refer to ② in Fig. F-10.)  A message 
    box “Set the sub standard light source.” will appear.    Click the 
    OK button.  (Refer to ③  in Fig. F-11.) 
     
     
     
    Fig. F-11 
     
     
    (10)  The instrument starts EM wavelength scan (from 498.6 to   
    901.4 nm). 
    At this stage, the following measuring conditions are automatically 
    set. 
    • Scan mode :  Emission 
    • Data mode :  Luminescence 
    •  EM slit  :  5.0 nm 
    •  Scan speed  :  60 nm/min 
    •  PMT voltage :  400 V 
    • Response  :  Auto 
     
    (11)  Upon completion of measurement, the conditions set before 
    measurement are retrieved. 
    Now, the instrument response on the emission side has been 
    measured and stored.    Take out the sub standard light source 
    and diffuser. 
     
     
    3 
    						
    							A - 37 
    F.4  Measurement of Instrumental Response in 
    Long Wavelength Range on Excitation Side 
     
    When a corrected spectrum in the long wavelength area of 600 nm or 
    more is required on the excitation side, the instrumental response in the 
    long wavelength range on the excitation side is read for the correction 
    with reference to the instrumental response obtained at the long 
    wavelength on the emission side. 
    This measurement is enabled only when the instrumental response at 
    the long wavelength on the emission side is available.    The correction 
    range by it is 500 to 900 nm and, when combined with an instrumental 
    response in the short wavelength range, spectrum correction is 
    available over the entire wavelength range (200 to 900 nm) on the 
    excitation side. 
     
     
    (1)  Close the cover of the sample compartment. 
     
    (2) Click the 
      (analysis method) button on the Measurement 
    toolbar.    A box for setting your analysis method will appear. 
     
    (3)  Select the General tab.    On the General tab page, specify 
    “Wavelength scan” for the measurement mode. 
     
    (4)  Select the Instrument tab. 
     
    (5)  Set “Fluorescence” for the data mode and “350 V” for the 
    photomultiplier voltage. 
     
    (6)  Execute the Zero Adjust command from the Spectrophotometer 
    menu to calibrate the zero point. 
     
    F.4.1 Overview 
     
    F.4.2 Operating 
    Procedures 
      
    						
    All Hitachi manuals Comments (0)