Hitachi F7000 Instruction Manual
Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
A - 28 (6) As shown in Fig. F-2, set the triangular cell filled with Rhodamine B into the cell holder in the sample compartment, the filter holder to the sample compartment and the red filter (single-side frosted) in the filter holder. In so doing, pay attention to the direction of the triangular cell and filter. (Refer to Fig. F-3) Fig. F-2 Setting of Triangular Cell Containing Rhodamine B Fig. F-3 Setting Direction of Triangular Cell (top view) NOTE: It may take time to obtain stable data due to characteristics of the detector immediately after the photomultiplier voltage has been remarkably changed (from 700 V to 250 V for example). When more accurately calibrating the zero point, wait until stable data is obtained and then proceed to step (5). Triangular cell with fixture (containing Rhodamine B) Filter holde rRed filter Metal part Excitation light Triangular cellFluorescenceRed filter (single-side frosted) Frosted
A - 29 (7) Execute the Instrumental Response command from the Utility menu. The Instrumental Response dialog box will open as shown in Fig. F-4. Select “EX [200 to 600 nm].” (Refer to ① in Fig. F-4.) Fig. F-4 (8) Click the Run button. (Refer to ② in Fig. F-4.) A message box “Set the triangular cell with Rhodamine B to cell holder and set the red filter.” will appear. After setting the Rhodamine B and red filter, click the OK button. (Refer to ③ in Fig. F-5.) Fig. F-5 1 2 3
A - 30 (9) The EX wavelength is set at 640 nm and the instrument starts EX wavelength scan (from 198.6 to 601.4 nm). At this stage, the following measuring conditions are automatically set. • Scan mode : Excitation • Data mode : Fluorescence • EX slit : 5.0 nm • EM slit : 20.0 nm • Scan speed : 60 nm/min • PMT voltage : 400 V • Response : Auto (10) Upon completion of measurement, the conditions set before measurement are retrieved. Now, the instrument response on the excitation side has been measured and stored. Take out the rectangular cell.
A - 31 F.2 Measurement of Instrumental Response on Emission Side This is the function to obtain the instrumental response on the emission side such as wavelength characteristics of the emission monochromator or detector (photomultiplier). The instrumental response on the emission side is determined by measuring a combination of instrumental response on both the excitation and emission sides by synchronous wavelength scan, and dividing it by the instrumental response on the excitation side preliminarily measured. An emission spectrum is correctable within a range of 200 to 600 nm. (1) Click the (analysis method) button on the Measurement toolbar. A box for setting your analysis method will appear. (2) Select the General tab. On the General tab page, specify “Wavelength scan” for the measurement mode. (3) Select the Instrument tab. (4) Set “Fluorescence” for the data mode and “250 V” for the photomultiplier voltage. (5) Execute the Zero Adjust command from the Spectrophotometer menu to calibrate the zero point. (6) As shown in Fig. F-6, set the diffuser into the cell holder in the sample compartment. Fig. F-6 Setting of Diffuser F.2.1 Overview F.2.2 Operating Procedures Diffuser
A - 32 NOTES: 1. Measurement of the instrumental response on the excitation side must be completed before executing measurement of the instrumental response on the emission side. 2. It may take time to obtain stable data due to characteristics of the detector immediately after the photomultiplier voltage has been remarkably changed (from 700 V to 250 V for example). When more accurately calibrating the zero point, wait until stable data is obtained and then proceed to step (5). (7) Execute the Instrumental Response command from the Utility menu. The Instrumental Response dialog box will open as shown in Fig. F-7. Select “EM [200 to 600 nm].” (Refer to ① in Fig. F-7.) Fig. F-7 (8) Click the Run button. (Refer to ② in Fig. F-7.) A message box “Set diffuser to cell holder.” will appear. Click the OK button. (Refer to ③ in Fig. F-8.) Fig. F-8 1 2 3
A - 33 (9) The EX and EM wavelengths are set at 198.6 nm and the instrument starts synchronous wavelength scan (from 198.6 to 601.4 nm). At this stage, the following measuring conditions are automatically set. • Scan mode : Synchronous • Data mode : Fluorescence • EX slit : 5.0 nm • EM slit : 20.0 nm • Scan speed : 60 nm/min • PMT voltage : 250 V • Response : Auto (10) Upon completion of measurement, the conditions set before measurement are retrieved. Now, the instrument response on the excitation side has been measured and stored. Take out the diffuser.
A - 34 F.3 Measurement of Instrumental Response in Long Wavelength Range on Emission Side When a corrected spectrum in the long wavelength area of 600 nm or more is required on the emission side, the beam of a lamp whose wavelength characteristics are known is allowed to be irradiated to the emission monochromator, and the instrumental response in the long wavelength range on the emission side is read for correction. As a light source, a sub standard light source (P/N 250-0123) is used. The correction range by it is 500 to 900 nm and, when combined with an instrumental response in the short wavelength range, spectrum correction is available over the entire wavelength range (200 to 900 nm) on the emission side. Refer to the instruction manual for the sub standard light source (optional accessory). (1) Close the cover of the sample compartment. (2) Click the (analysis method) button on the Measurement toolbar. A box for setting your analysis method will appear. (3) Select the General tab. On the General tab page, specify “Wavelength scan” for the measurement mode. (4) Select the Instrument tab. (5) Set “Fluorescence” for the data mode and “400 V” for the photomultiplier voltage. (6) Execute the Zero Adjust command from the Spectrophotometer menu to calibrate the zero point. (7) Set the sub standard light source in the sample compartment with reference to its instruction manual. (Refer to Fig. F-9.) Turn on the power switch of the sub standard light source and wait for approx. 5 minutes until the lamp stabilizes. F.3.1 Overview F.3.2 Operating Procedures
A - 35 Fig. F-9 (8) Execute the Instrumental Response command from the Utility menu. The Instrumental Response dialog box will open as shown in Fig. F-10. Select “EM [500 to 900 nm].” (Refer to ① in Fig. F-10.) Fig. F-10 1 2
A - 36 (9) Click the Run button. (Refer to ② in Fig. F-10.) A message box “Set the sub standard light source.” will appear. Click the OK button. (Refer to ③ in Fig. F-11.) Fig. F-11 (10) The instrument starts EM wavelength scan (from 498.6 to 901.4 nm). At this stage, the following measuring conditions are automatically set. • Scan mode : Emission • Data mode : Luminescence • EM slit : 5.0 nm • Scan speed : 60 nm/min • PMT voltage : 400 V • Response : Auto (11) Upon completion of measurement, the conditions set before measurement are retrieved. Now, the instrument response on the emission side has been measured and stored. Take out the sub standard light source and diffuser. 3
A - 37 F.4 Measurement of Instrumental Response in Long Wavelength Range on Excitation Side When a corrected spectrum in the long wavelength area of 600 nm or more is required on the excitation side, the instrumental response in the long wavelength range on the excitation side is read for the correction with reference to the instrumental response obtained at the long wavelength on the emission side. This measurement is enabled only when the instrumental response at the long wavelength on the emission side is available. The correction range by it is 500 to 900 nm and, when combined with an instrumental response in the short wavelength range, spectrum correction is available over the entire wavelength range (200 to 900 nm) on the excitation side. (1) Close the cover of the sample compartment. (2) Click the (analysis method) button on the Measurement toolbar. A box for setting your analysis method will appear. (3) Select the General tab. On the General tab page, specify “Wavelength scan” for the measurement mode. (4) Select the Instrument tab. (5) Set “Fluorescence” for the data mode and “350 V” for the photomultiplier voltage. (6) Execute the Zero Adjust command from the Spectrophotometer menu to calibrate the zero point. F.4.1 Overview F.4.2 Operating Procedures