Hitachi F7000 Instruction Manual
Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
1.8 Window Setting 1 - 40 1.8.7 Analysis Condition Tab For the settable range of each parameter, refer to Appendix I. Fig. 1-56 Analysis Condition Tab 1 Integration time for current data An integration time for reading the current data by the Read Data command of Tools menu can be set. 2 Change Instrumental Response Condition When this check box is turned on, the excitation slit, emission slit and photomultiplier voltage can be set on the instrument function measurement window. This window appears when clicking the Instrument Response command of Utility menu. 1 2
1.8.8 1 - 41 1.8.8 Report Tab Fig. 1-57 Report Tab Data List Printing Direction The direction of data list printing (with constant line spacing) can be set. In case of vertical printing: Wavelength or time and photometric values will be printed vertically in 3 rows. In case of horizontal printing: Wavelength or time and photometric values will be printed horizontally in 3 rows. That is, 3 data will be printed on each line.
1.9 Saving a Data File 1 - 42 1.9 Saving a Data File Data can or cannot be saved depending on setting. So perform setting, referring to the table below. Setting in any of patterns ① to ⑤ is recommended. “Use sample table in analysis method” Setting ON OFF Sample name/comment input dialog “Save file automatically after data acquisition” ⎯ ON OFF Analysis Method - Monitor tab “Open data processing window after data acquisition” ON OFF ON OFF ON OFF Display × × × Result Saving Automatic Automatic Automatic Automatic Manual Impossible Pattern ① ② ③ ④ ⑤ ⑥ Read and × symbols in the Result - Display row as follows. : Icons or data processing window is open immediately after measurement. × : The data processing window is not open immediately after measurement.
1.10 1 - 43 1.10 Photometer Menu Commands The Spectrophotometer menu contains the following commands. • Set Wavelength • Auto Pre-scan • Zero Adjust • Shutter Open/Close Set Wavelength This command sets the excitation/emission wavelength to a specified wavelength. Click the Set Wavelength command or button, and the dialog box shown below will come up. Fig. 1-58 Zero Adjust This command automatically closes the shutter and measures a zero point in detection. When the zero point is changed, e.g., after variation of the photomultiplier voltage, automatic zero-point adjustment will be carried out. NOTE: If the lid of sample compartment is opened or closed during zero- point adjustment, zero point may not be set exactly. In this case, perform zero-point adjustment again. Auto Pre-scan This command automatically searches for the optimum excitation/emission wavelength for the sample under measurement. Other peaks than the fluorescence peak (such as peaks of excitation scattered light and Raman spectrum) are ignored. On completion of automatic pre-scanning, the optimum excitation/emission wavelength is indicated on the monitor window (in the current wavelength indication field detailed in 1.6). Shutter Open/Close This command is available for opening/closing the excitation beam shutter. The current shutter state is indicated on the right side of the monitor window (⑨ in Fig. 1-43).
1.11 Utility Menu Commands 1 - 44 1.11 Utility Menu Commands The Utility menu contains the following command. • Instrument Response • Analog Output Gain • EX Power Monitor • Sensitivity Check • Lamp Usage • Lamp Off Instrument Response This command is used for spectral correction. The spectral correction accessory kit (P/N 251-0126) needs to be prepared. Analog Output Gain Use this command for setting a data value corresponding to 1 V of analog output. For details, refer to 9.3. EX Power Monitor Use this command for checking output of the monitor detector. When the EX Power Monitor is turned on (clicked), monitor value data is displayed regardless of measurement condition. Sensitivity Check Use this command for checking sensitivity. You can check sensitivity/drift with this command. For details, refer to the maintenance edition. Lamp Usage Use this command for checking/resetting the usage period of the Xenon lamp. A cumulative turn-on time of the Xenon lamp is indicated. Fig. 1-59
1.11 1 - 45 When the limit use time is exceeded, the following message appears. Fig. 1-60 After lamp replacement, reset this indication. For details, refer to the maintenance edition. Note: If the Model F-7000 power supply is turned on without starting up the PC, indication of the Xenon lamp turn-on time may not be updated correctly. Lamp Off Use this command for turning off the Xenon lamp. For turning on the Xenon lamp again, restarting the spectrophotometer is necessary.
1.12 Termination of FL Solutions Program 1 - 46 1.12 Termination of FL Solutions Program After completion of measurement, shut down the instrument in the following procedure. 1 From the File menu, select the Exit command. A window appears as shown below. Fig. 1-61 2 For termination without turning off the light source lamp, select “Close the monitor window, but keep the lamp operating?” For termination after turning off the light source lamp, select “Close (Turn off) the lamp, then close the monitor window?” Click Yes, and the FL Solutions program will then be terminated. 3 Turn off the power switch of the spectrophotometer main unit. 4 Click the start button of Windows XP, select “Turn Off Computer” and click OK. 5 Turn off power to the personal computer and the CRT display monitor. (At step 4, the personal computer may be turned off depending on system settings.) 6 Turn off power to the printer.
2.1 2 - 1 2. WAVELENGTH SCAN 2.1 Flow of Operation This section explains the condition setting and operating method for wavelength scan. Measurement proceeds along the following flow. Fig. 2-1 Operational Flow (wavelength scan) Start Method setting (See 2.2.) Sample name/ comment input (See 2.3.) Pre-scan (See 2.4) Measurement (See 2.5.) Data processing (See Section 7.) Result output in report (See 2.6.) End Click button and set your analytical conditions. • Scan range • Data mode • Scan speed, etc. Click button. • Sample name • Auto File • File name, etc. Set a sample. Click button to execute auto scale for peaks. Click button to execute wavelength scan. • Spectrum tracing • Peak detection, etc. Click button. Measurement result will be displayed in a report format.
2.2 Creating an Analysis Method 2 - 2 2.2 Creating an Analysis Method Set the analytical conditions (method) prior to measurement. Select the Method command from the Edit menu or click the button. A window shown in Fig. 2-2 will appear. Fig. 2-2 Analysis Method The following tabs are provided on the Analysis Method window for the wavelength scan mode. These tabs are detailed on the subsequent pages. • General • Instrument • Monitor • Processing • Report
2.2.1 2 - 3 The parameters shown in Fig. 2-3 are provided at the General tab. For the settable range of each parameter, refer to Appendix G. Fig. 2-3 General Tab 1 Measurement (mode) A measurement mode is selectable. Here, select “Wavelength scan.” For Time scan, Photometry, 3-D Scan and 3-D Time scan, refer to Sections 3, 4, 5 and 6, respectively. 2 Operator Enter an operator name. 3 Instrument The name of the connected instrument is indicated. 4 Sampling - None Indicates that samples are to be exchanged manually. 5 Comments Enter a description or notes on measuring conditions. 6 Use sample table (Set measurement sample) Apply a check mark here when a sample table is used. 2.2.1 General Tab 1 2 3 4 5 6 7 8 9