Home > Hitachi > Fluorescence Spectrophotometer > Hitachi F7000 Instruction Manual

Hitachi F7000 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							9.2    Use of AS-3000 Intelligent Autosampler 
    9 - 10 
    9.2    Use of AS-3000 Intelligent Autosampler 
     
    Set the rack parameters, injection volume and other items with the   
    AS-3000 Intelligent Autosampler.    For details, refer to the instruction 
    manual attached to the AS-3000. 
     
     
     
     
     
     
     
     
     
     
     
    Fig. 9-14  Cord Connection 
     
     
    Parameter setting:    Set all parameters in the AS-3000. 
     
    NOTE:  When using a sample table, match the number of samples set 
    in the AS-3000 with that set in the sample table.    If the latter 
    is less than the former, some samples cannot be measured. 
     
     
    PCAS-3000 USBEXTIN
    COM1OUT1 IN1 Communication cable 
    (furnished with F-7000)Connecting cord   
    (furnished with AS-3000) 
    F-7000 
    						
    							9.2 
    9 - 11 
     
    (1)  Select the Measure command from the Spectrophotometer menu 
    of the FL Solutions program or click the 
     (measurement) 
    button on the toolbar.    A window as in Fig. 9-15 will open. 
     
     
     
    Fig. 9-15    For Start from Creation of Calibration curve 
     
     
    (2) Press the START key on the AS-3000. 
     
    (3)  Measurement will start. 
     
      
    						
    							9.3  Analog Output 
    9 - 12 
    9.3  Analog Output 
     
    For connection, use the analog output terminal.    Select the Analog 
    Output command from the Utility menu.    (The analog output I/F (option) 
    is required.) 
     
     
     
    Fig. 9-16 
     
     
    This is a setting of data value corresponding to analog output of 1 V. 
    Analog output value (V) becomes as follows. 
     
    Low :  (data value)/1000 
    High :  (data  value)/10000 
     
     
      
    						
    							A - 1 
    APPENDIX A    DETAILS OF QUANTIFICATION   
     
    A.1  Foreword 
     
    In the Photometry mode of the F-7000 spectrophotometer, the following 
    four calibration curve types are available. 
     
      Linear working curve 
      Quadratic working curve 
      Cubic working curve 
      Multiple segment working curve 
     
    These are explained in detail below. 
     
     
    A.2    Linear Working Curve (1st order) 
     
    A regression line is determined via the least squares method from a 
    maximum of 20 data.    The calculation formula is as follows: 
     
    x = A1  y + A0 
    ()
    Ayx
    nyx
    y
    nyii i i
    ii11
    122=−⋅
    −∑ ∑ ∑
    ∑ ∑, AxnAy
    nii01=−⋅∑∑ 
     
    where,  x :  Sample concentration (input value) 
      y :  Sample data (measured value) 
      n :  Number of samples 
     
      
    						
    							A - 2 
    A.3    Quadratic Working Curve (2nd order) 
     
    A quadratic curve is determined via the least squares method from a 
    maximum of 20 data.    The calculation formula is as follows: 
     
    x = A2  y
    2 + A1  y + A0 
    ()()()()
    ()()(){}
    ASy xSyy SyxSyy
    SyySy y Syy 2
    22
    22 22
    =
    
     
    ()()Syy yy
    nii=∑
    ∑22
     
    ()()()()
    ()()(){}
    ASyxSy y Sy xSyy
    SyySy y Syy 1
    22 2 2
    22 22
    =
    
     
    ()Syx y xyx
    niiii=−⋅∑ ∑
    ∑
     
    AAAx
    nx
    ny
    nii i012
    2
    = ∑∑ ∑
     ()Syy yyy
    niii
    232=−⋅∑ ∑
    ∑
     
    ()Sy x y xyx
    ni
    iii
    222=−⋅∑ ∑
    ∑
     
    ()
    ()Sy y yy
    nii
    22 422=∑
    ∑ 
     
    where,  x :  Sample concentration (input value) 
      y :  Sample data (measured value) 
      n :  Number of samples 
     
     
    A.4    Cubic Working Curve (3rd order) 
     
    A cubic curve is determined via the least squares method from a 
    maximum of 20 data.    The calculation formula is as follows: 
     
    x = A3  y
    3 + A2  y2 + A1  y + A0 
     
    where,  x :  Sample concentration (input value) 
      y :  Sample data (measured value) 
      n :  Number of samples 
     
      
    						
    							A - 3 
    A.5    Multiple Segment Working Curve (Segmented) 
     
    The calibration curve is apt to bend when measuring turbid samples or 
    the like.    Use of a quantification program allows correcting the 
    calibration curve by using up to 20 standards.    Figure A-1 shows an 
    example of curve correction.    For the part which exceeds the 
    measuring range of the standards, simply extend the line as it is.   
     
     
     
    Fig. A-1    Correction of Curve Bending 
     
     
    Data
    Conc  
    						
    							A - 4 
    <Notes on Preparation of Multiple Segment Working Curve> 
    (1)  A correct calibration curve can be created only when the 
    measured value increases or decreases monotonically versus the 
    concentration value.    Especially when the inclination is negative-
    going, be sure to measure a blank having a data value larger than 
    the other standards.    A curve not showing a monotonic increase 
    will appear as in Fig. A-2.    And when the data value of the blank 
    is small, regardless of a negative-going inclination, the curve will 
    appear as in Fig. A-3. 
     
     
     
    Fig. A-2  Example of Curve without Monotonic Increase   
    (unsuitable curve) 
     
     
     
     
    Fig. A-3  Example of Negative-going Curve where Measured   
    Value of Blank is Small (unsuitable curve) 
     
     
    Data
    Conc 
    Data
    Conc  
    						
    							A - 5 
    (2)  Remeasurement of Standards 
     
    The standards can be remeasured after once preparing a 
    calibration curve.    The curve prepared in such a case will appear 
    as in Fig. A-4. 
     
     
     
    Fig. A-4    Calibration Curve when Standards are Remeasured 
     
     
     
    Data
    Conc
    Redrawn 
    calibration curve
    Remeasured STD1
    Initially measured STD1 
    						
    							A - 6 
    APPENDIX B  DETAILS OF RATE ANALYSIS 
    FUNCTION 
     
    B.1  Foreword 
     
    Rate analysis is used in the analysis of enzyme reactions.    It is utilized 
    for clinical and biochemical tests by reagent manufacturers, hospitals 
    and so on.    A computer is used to calculate the concentration from the 
    variation in data per unit time, and the result is displayed and printed out. 
     
     
    B.2  Calculation Method 
     
    A timing chart for rate analysis is shown in Fig. B-1.    Data is acquired 
    when the initial delay time has elapsed after pressing the Measure 
    button.    A regression line is determined from this data via the least 
    squares method, and the gradient and activity value are calculated.   
    The calculation formula is as follows. 
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
    Fig. B-1 
     
     
    Data 
    Td Tc
    Tt
    Tm
    A0
    A1
    A5
    A4
    A3
    A2
    A6
    A7
    Time
    Start of 
    measurement Data: 
    A0, A1, A2, A3, 
    A4... 
    Td  :  Initial delay time
    Tm :  Measurement time 
    Tc :  Sampling interval 
    Tt :  Calculation time  
    						
    							A - 7 
    Prepare a regression line via the least squares method from the 
    measured data, and obtain a determination coefficient. 
     
    y = ax + b 
    where, 
     
    ()
    axyxy
    n
    xx
    niiii
    ii=
    
    ∑ ∑
    ∑
    ∑
    ∑
    22
      
    ()byax
    nii=−∗∑
    ∑ 
     
    x
    i :  Time (s) of each data 
    y
    i :  Value of each data 
    n :  Number of samples 
     
    The determination coefficient CD becomes as follows: 
     
    ()
    () ()
    CDnxy x y
    nx x ny yii i i
    ii ii=−
    − ⎛
    ⎝ ⎜⎞
    ⎠ ⎟− ⎛
    ⎝ ⎜⎞
    ⎠ ⎟∑ ∑ ∑
    ∑ ∑∑∑
    2
    22
    22
     
     
      Gradient (variation per minute) 
    min) (/ a 60
    Tka
    Di= = 
     
     Activity 
    C
    i = k  Di 
     
      R (relative coefficient) 
    () ()⎟ ⎟
    ⎠ ⎞
    ⎜ ⎜
    ⎝ ⎛
    ∑∑−
    ⎟ ⎟
    ⎠ ⎞
    ⎜ ⎜
    ⎝ ⎛
    ∑∑−⎟ ⎟
    ⎠ ⎞
    ⎜ ⎜
    ⎝ ⎛
    ∑∑∑2
    i i2
    2
    i i22
    i i
    i iy y n x x ny x y
    R
      x n
    CD-
    = =
     
     
      R2 (determination coefficient) 
    ()
    ⎟ ⎟
    ⎠ ⎞
    ⎜ ⎜
    ⎝ ⎛
    ∑
    ⎟ ⎟
    ⎠ ⎞
    ⎜ ⎜
    ⎝ ⎛
    ∑ −
    ⎟ ⎟
    ⎠ ⎞
    ⎜ ⎜
    ⎝ ⎛
    ∑
    ⎟ ⎟
    ⎠ ⎞
    ⎜ ⎜
    ⎝ ⎛
    ∑ −⎟ ⎟
    ⎠ ⎞
    ⎜ ⎜
    ⎝ ⎛
    ∑∑∑−
    = =
    2
    i 2
    i 2
    i 2
    i2
    i i i i 2
    y y n x x ny x y x n
    CD R
     
     
     
     
    NOTE:  If the range for rate calculation does not coincide with the 
    actual measured data range, then use only the measured data 
    within the range for the calculation. 
      
    						
    All Hitachi manuals Comments (0)