Home > Hitachi > Fluorescence Spectrophotometer > Hitachi F7000 Instruction Manual

Hitachi F7000 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							4.2.2 
    4 - 5 
    Upon selection of the Quantitation tab, the window shown in Fig. 4-4 
    appears.  
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 4-4  Quantitation Tab 
     
     
    1 Quantitation type   
    Quantitation type is selectable among Wavelength, Peak area, 
    Peak height, Derivative and Ratio.    Spectra measured by the Peak 
    area, Peak height and Derivative methods can be checked, 
    referring to 7.19.   
    (a) Wavelength  
    Upon selecting ‘None’ for Calibration type in this quantitation 
    method, photometric values can be obtained at the specified 
    number of wavelengths for Number of wavelengths.     
    A calibration curve will not be prepared.   
     
    Upon selecting 1st order, 2nd order, 3rd order or Segmented 
    in this quantitation method, calibration curve data and sample 
    data will be obtained by one, two or three-wavelength 
    calculation depending on your entry for Number of 
    wavelengths.  
     
    4.2.2 Quantitation 
    Tab 
     
    1 
    2 
    5 
    3 
    4 
    7
    6
    8  
    						
    							4.2    Creating an Analysis Method 
    4 - 6 
    (i)  One wavelength calculation   
    This is the most generally used method.    When you 
    specify one wavelength and enter it at the Instrument 
    tab, a calibration curve and concentration are 
    determined from the data I
    1 at the specified wavelength 
    as shown in Fig. 4-5 (a).   
     
     
     
    Fig. 4-5 (a) 
     
     
    (ii)  Two wavelength calculation   
    When you specify two wavelengths and enter them at 
    the Instrument tab, a calibration curve and 
    concentration are calculated through determination of I 
    by the equation below assuming I
    1 and I2 are the 
    photometric values at two different wavelengths as 
    shown in Fig. 4-5 (b).     
     
     
     
    Fig. 4-5 (b)   
     
     
    I1
    Data 
    I 
    WL1 
    Data 
    I1  I
    2
    I 
    WL1  WL2 
    						
    							4.2.2 
    4 - 7 
    (iii)  Three wavelength calculation   
    As shown in Fig. 4-5 (c), assuming I
    1, I2 and I3 are the 
    data values at three different wavelengths, a calibration 
    curve and concentration are calculated through 
    determination of I by the equation below.     
    Photometry is made in the order of WL1, WL2 and WL3.     
     
    I = I
    2 -   
     
    But the following relation must hold:   
    WL1 < WL2 < WL3 or WL1 > WL2 > WL3   
     
     
     
    Fig. 4-5 (c) 
     
     
    (b) Peak area  
    Enter the peak wavelength and its accuracy for Peak apex.   
    Integration method (area detection method) is selectable 
    among Rectangular, Trapezoid and Romberg.    (Refer to 
    “integration method” in Appendix D.2.)    Then, set a 
    threshold value and sensitivity for defining the peak, referring 
    to 7.6.   I
    1 (WL3 - WL2) + I3 (WL2 - WL1) 
    WL3 - WL1 
    Data 
    I1  I
    2
    WL1  WL2 I
    3
    WL3I 
    						
    							4.2    Creating an Analysis Method 
    4 - 8 
    Measurement will be carried out within the wavelength range 
    set at the Instrument tab.    According to the specified 
    threshold level and sensitivity, peak wavelength is sought.   
    The area of the peak found within a range of the entered 
    Peak apex ± wavelength accuracy is calculated, thereby 
    preparing a calibration curve.   
     
     
     
    Fig. 4-6    Quantitation Tab (peak area) 
     
      
    						
    							4.2.2 
    4 - 9 
    (c) Peak height  
    Enter the peak wavelength and its accuracy for Peak apex.   
    Measurement will be carried out within the wavelength range 
    set at the Instrument tab.    According to the threshold level 
    and sensitivity specified for the “Peak area” quantitation 
    method, peak wavelength is sought.    A calibration curve is 
    prepared using the height of the peak found within a range of 
    the entered Peak apex ± wavelength accuracy.     
     
     
     
    Fig. 4-7    Quantitation Tab (peak height) 
     
      
    						
    							4.2    Creating an Analysis Method 
    4 - 10 
    (d) Derivative  
    Enter peak wavelength (or wavelength to be used for 
    calibration curve preparation) for Peak apex.     
    Measurement will be carried out within the wavelength range 
    set at the Instrument tab.    A calibration curve is created 
    using the derivative value at the peak wavelength.     
    Assign your values for derivative parameters; Derivative 
    order, Smoothing order and Number of points.   
     
     
     
    Fig. 4-8    Quantitation Tab (derivative) 
     
      
    						
    							4.2.2 
    4 - 11 
    (e) Ratio  
    Upon selecting Ratio, an input column for three factors (K-
    factor 0, K-factor 1, K-factor 2) and a setting column for 
    calculation method (+, -, ∗) are displayed.     
     
    NOTE:  By this quantitation method, a calibration curve cannot 
    be created.   
     
     
     
    Fig. 4-9    Quantitation Tab (peak ratio) 
     
     
    Calculation is made by the following equation.   
     
    RATIO =    #K0 
     
    # (calculation method):    +, -, ∗  
    WL1, 2, 3  :  Specified wavelengths   
    WL3  :  Background correcting wavelength   
    A ( )  :  Measured value at specified wavelength   
     
    2 Calibration type   
    Set a calibration type.    For selecting factor input (when the formula 
    of standard calibration curve and its factors are already known), put 
    a mark in the check box for Manual calibration and enter values for 
    A0, A1, A2 and A3.     
     K1 (A (WL1) - A (WL3))
    K2 (A (WL2) - A (WL3)) 
    						
    							4.2    Creating an Analysis Method 
    4 - 12 
    (a) None  
    Mode in which calibration curve is not used.    When this 
    mode is selected with Wavelength set for Quantitation type, 
    multi-wavelength measurement at up to six wavelengths is 
    possible.    However, no spectrum can be displayed.   
     
    (b) 1st order  
    This specifies the most generally used calibration curve.     
    A linear calibration curve is to be created.    The formula of 
    calibration curve is as follows.   
    x = A
    1y + A0  
    Here, x is sample concentration and y is sample data.   
     
     
     
    Fig. 4-10  1st Order 
     
     
    (c) 2nd order  
    A calibration curve is created using a quadratic curve.     
    The formula of calibration curve is as follows.   
    x = A
    2y2 + A1y + A0 
    Here, x is sample concentration and y is sample data.   
     
     
     
    Fig. 4-11  2nd Order 
    Data 
    K0Conc 
    Data 
    K0Conc  
    						
    							4.2.2 
    4 - 13 
    (d) 3rd order  
    A calibration curve is created using a cubic curve.     
    The formula of calibration curve is as follows.   
    x = A
    3y3 + A2y2 + A1y + A0 
    Here, x is sample concentration and y is sample data.   
     
     
     
    Fig. 4-12  3rd Order  
     
     
    (e) Segmented  
    A calibration curve created by connecting standard samples 
    linearly according to the measured values of respective 
    samples or input data.   
     
     
     
    Fig. 4-13  Segmented 
     
     
    Data 
    K0Conc 
    Data 
    Conc
    For the part which 
    exceeds the measuring 
    range, extend the line 
    just before it. 
    Max. No. of standards:    20  
    						
    							4.2    Creating an Analysis Method 
    4 - 14 
    NOTE:  A proper calibration curve can be created when the 
    photometric value increases or decreases 
    monotonously versus the concentration value.    With a 
    monotonic increase, the calibration curve is created with 
    standard 1 (STD1) as origin.    And with a monotonic 
    decrease, the calibration curve is created with STD1 
    plotted on the photometric value axis and STD5 on the 
    concentration axis.    In these examples, the number of 
    standards is set at 5.   
     
     
     
     
    3 Number of wavelengths   
    Enter the desired number of wavelengths to be used for 
    measurement.  
    4 Concentration unit   
    Enter a concentration (in mg/mL for example) from the keyboard.   
    Calibration curve with monotone increase
    Calibration curve with monotone decrease
    Data  Data 
    Conc
    Conc 
    						
    All Hitachi manuals Comments (0)