Hitachi F7000 Instruction Manual
Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
4.2.2 4 - 5 Upon selection of the Quantitation tab, the window shown in Fig. 4-4 appears. For the settable range of each parameter, refer to Appendix G. Fig. 4-4 Quantitation Tab 1 Quantitation type Quantitation type is selectable among Wavelength, Peak area, Peak height, Derivative and Ratio. Spectra measured by the Peak area, Peak height and Derivative methods can be checked, referring to 7.19. (a) Wavelength Upon selecting ‘None’ for Calibration type in this quantitation method, photometric values can be obtained at the specified number of wavelengths for Number of wavelengths. A calibration curve will not be prepared. Upon selecting 1st order, 2nd order, 3rd order or Segmented in this quantitation method, calibration curve data and sample data will be obtained by one, two or three-wavelength calculation depending on your entry for Number of wavelengths. 4.2.2 Quantitation Tab 1 2 5 3 4 7 6 8
4.2 Creating an Analysis Method 4 - 6 (i) One wavelength calculation This is the most generally used method. When you specify one wavelength and enter it at the Instrument tab, a calibration curve and concentration are determined from the data I 1 at the specified wavelength as shown in Fig. 4-5 (a). Fig. 4-5 (a) (ii) Two wavelength calculation When you specify two wavelengths and enter them at the Instrument tab, a calibration curve and concentration are calculated through determination of I by the equation below assuming I 1 and I2 are the photometric values at two different wavelengths as shown in Fig. 4-5 (b). Fig. 4-5 (b) I1 Data I WL1 Data I1 I 2 I WL1 WL2
4.2.2 4 - 7 (iii) Three wavelength calculation As shown in Fig. 4-5 (c), assuming I 1, I2 and I3 are the data values at three different wavelengths, a calibration curve and concentration are calculated through determination of I by the equation below. Photometry is made in the order of WL1, WL2 and WL3. I = I 2 - But the following relation must hold: WL1 < WL2 < WL3 or WL1 > WL2 > WL3 Fig. 4-5 (c) (b) Peak area Enter the peak wavelength and its accuracy for Peak apex. Integration method (area detection method) is selectable among Rectangular, Trapezoid and Romberg. (Refer to “integration method” in Appendix D.2.) Then, set a threshold value and sensitivity for defining the peak, referring to 7.6. I 1 (WL3 - WL2) + I3 (WL2 - WL1) WL3 - WL1 Data I1 I 2 WL1 WL2 I 3 WL3I
4.2 Creating an Analysis Method 4 - 8 Measurement will be carried out within the wavelength range set at the Instrument tab. According to the specified threshold level and sensitivity, peak wavelength is sought. The area of the peak found within a range of the entered Peak apex ± wavelength accuracy is calculated, thereby preparing a calibration curve. Fig. 4-6 Quantitation Tab (peak area)
4.2.2 4 - 9 (c) Peak height Enter the peak wavelength and its accuracy for Peak apex. Measurement will be carried out within the wavelength range set at the Instrument tab. According to the threshold level and sensitivity specified for the “Peak area” quantitation method, peak wavelength is sought. A calibration curve is prepared using the height of the peak found within a range of the entered Peak apex ± wavelength accuracy. Fig. 4-7 Quantitation Tab (peak height)
4.2 Creating an Analysis Method 4 - 10 (d) Derivative Enter peak wavelength (or wavelength to be used for calibration curve preparation) for Peak apex. Measurement will be carried out within the wavelength range set at the Instrument tab. A calibration curve is created using the derivative value at the peak wavelength. Assign your values for derivative parameters; Derivative order, Smoothing order and Number of points. Fig. 4-8 Quantitation Tab (derivative)
4.2.2 4 - 11 (e) Ratio Upon selecting Ratio, an input column for three factors (K- factor 0, K-factor 1, K-factor 2) and a setting column for calculation method (+, -, ∗) are displayed. NOTE: By this quantitation method, a calibration curve cannot be created. Fig. 4-9 Quantitation Tab (peak ratio) Calculation is made by the following equation. RATIO = #K0 # (calculation method): +, -, ∗ WL1, 2, 3 : Specified wavelengths WL3 : Background correcting wavelength A ( ) : Measured value at specified wavelength 2 Calibration type Set a calibration type. For selecting factor input (when the formula of standard calibration curve and its factors are already known), put a mark in the check box for Manual calibration and enter values for A0, A1, A2 and A3. K1 (A (WL1) - A (WL3)) K2 (A (WL2) - A (WL3))
4.2 Creating an Analysis Method 4 - 12 (a) None Mode in which calibration curve is not used. When this mode is selected with Wavelength set for Quantitation type, multi-wavelength measurement at up to six wavelengths is possible. However, no spectrum can be displayed. (b) 1st order This specifies the most generally used calibration curve. A linear calibration curve is to be created. The formula of calibration curve is as follows. x = A 1y + A0 Here, x is sample concentration and y is sample data. Fig. 4-10 1st Order (c) 2nd order A calibration curve is created using a quadratic curve. The formula of calibration curve is as follows. x = A 2y2 + A1y + A0 Here, x is sample concentration and y is sample data. Fig. 4-11 2nd Order Data K0Conc Data K0Conc
4.2.2 4 - 13 (d) 3rd order A calibration curve is created using a cubic curve. The formula of calibration curve is as follows. x = A 3y3 + A2y2 + A1y + A0 Here, x is sample concentration and y is sample data. Fig. 4-12 3rd Order (e) Segmented A calibration curve created by connecting standard samples linearly according to the measured values of respective samples or input data. Fig. 4-13 Segmented Data K0Conc Data Conc For the part which exceeds the measuring range, extend the line just before it. Max. No. of standards: 20
4.2 Creating an Analysis Method 4 - 14 NOTE: A proper calibration curve can be created when the photometric value increases or decreases monotonously versus the concentration value. With a monotonic increase, the calibration curve is created with standard 1 (STD1) as origin. And with a monotonic decrease, the calibration curve is created with STD1 plotted on the photometric value axis and STD5 on the concentration axis. In these examples, the number of standards is set at 5. 3 Number of wavelengths Enter the desired number of wavelengths to be used for measurement. 4 Concentration unit Enter a concentration (in mg/mL for example) from the keyboard. Calibration curve with monotone increase Calibration curve with monotone decrease Data Data Conc Conc