Home > Hitachi > Fluorescence Spectrophotometer > Hitachi F7000 Instruction Manual

Hitachi F7000 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							2.5  Conducting Measurement 
    2 - 24 
    (5)  Upon completion of the measurement, the measured data will be 
    saved automatically at the specified saving location using the 
    specified file name.    Opening the saved data file will display the 
    data processing window as shown below. 
     
     
     
    Fig. 2-20    Data Processing Window (wavelength scan) 
     
     
    ① Spectrum  
    The spectrum of a measured sample is displayed.   
     
    ② Peak table  
    Peak wavelength, peak height and valley wavelength are 
    listed.  
     
    ① Spectrum
    ②Peak table  
    						
    							2.5.2 
    2 - 25 
    NOTICE:  1.  The data measured under the following conditions will 
    be saved automatically in Program Files\FL 
    Solutions\Data.  (Example:  C:Program Files\FL 
    Solutions\Data)  
    (1)  Measurement was carried out using a sample table 
    with no file name specified in the File Name 
    column.  
    (2)  Measurement was carried out with “Auto File” 
    specified and a file name unspecified, avoiding use 
    of a sample table.   
    2.  The data measured under the following conditions will 
    be saved automatically in Program Files\FL Solutions.   
    (Example:  C:Program Files\FL Solutions)  
    (3)  There is not an adequate free area in the specified 
    storage medium (such as floppy disk or MO).     
    (4)  Writing is inhibited.   
    A file automatically saved under such a condition as above is 
    named according to the following rule.   
    Rule 1:  When a file name is given to analysis method: 
    File name of analysis method + serial No. (001, 
    002, ...)   
    Rule 2:  When a file name is not given to analysis method:   
    DEFLT + serial No. (001, 002, ...)   
     
      
    						
    							2.6  Data Printout 
    2 - 26 
    2.6  Data Printout  
     
    Printout can be provided even when the function for automatic printout 
    after measurement is not used.    The following two procedures are 
    available.  
     
    (1)  Click on the spectrum window, and the circumference turns blue.   
    In this status, select the Print command from the File menu.     
    Or click the 
      (print) button on the toolbar.    The field active 
    on the data processing window (e.g., ①  in Fig. 2-20) will be 
    printed.    What is displayed by clicking the Print Preview 
    command in the File menu will be printed.   
     
     
     
    Fig. 2-21   
     
     
    (2)  Select the Report command in the Data menu.    Or click the 
     
    button on the toolbar.    The Print Preview window will appear as 
    shown below, so click Print.   
    						
    							2.6 
    2 - 27 
     
     
     
    Fig. 2-22 
     
     
    NOTE:  If characters overflow from the print frame in repeat 
    measurement or in overlaying, then change the printing 
    orientation from vertical (Portrait) to horizontal (Landscape).   
      
    						
    							2.6  Data Printout 
    2 - 28 
    Select the items to be printed in the following procedure.   
    Click the 
      button, and the Spectrum Properties window will appear.   
    ↓ 
    Select the Report tab, and the window in Fig. 2-23 will open.   
    ↓ 
    In this window, select desired print items.   
     
     
     
    Fig. 2-23 
     
      
    						
    							3.1 
    3 - 1 
    3.  TIME SCAN  
     
    3.1  Flow of Operation  
     
    This section explains the parameter setting and operating method for 
    time scan.    Measurement proceeds along the flow shown below.   
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
    Fig. 3-1    Operational Flow (time scan) Click 
      button. 
     • Measuring wavelengths 
    •  Scan time, etc. 
    Click 
     button. 
     • Sample name 
    • Auto File 
    •  File name at destination of 
    saving, etc. 
    Set a sample.   
    Click 
      button to execute time scan. 
    • Rate calculation 
    • Peak detection 
    • Spectrum tracing 
    • Spectral calculation 
    Click 
      button.    Measurement result will be displayed 
    in a report format. 
     
    Start 
    Method setting 
    (See 3.2.) 
    Sample name/ 
    comment input 
    (See 3.3.) 
    Measurement 
    (See 3.4.) 
    Data processing 
    (See Section 7.) 
    Result output in report 
    (See 3.5.) 
    End  
    						
    							3.2    Creating an Analysis Method 
    3 - 2 
    3.2    Creating an Analysis Method   
     
    Set the analytical conditions (method) prior to measurement.     
    From the Edit menu, select the Method command or click the 
     
    button.    A window shown in Fig. 3-2 will appear.   
     
     
     
    Fig. 3-2  Analysis Method 
     
     
    The following tabs are provided on the Analysis Method window for the 
    time scan mode.    These tabs are detailed on the subsequent pages.   
    • General  
    • Instrument  
    • Monitor  
    • Processing  
    • Report  
     
      
    						
    							3.2.1 
    3 - 3 
    The parameters shown in Fig. 3-3 are provided at the General tab.     
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 3-3  General Tab 
     
     
    1 Measurement (mode)   
    A measurement mode is selectable.    Here, select “Time scan.”   
    For Wavelength scan, Photometry, 3-D Scan and 3-D Time scan, 
    refer to Sections 2, 4, 5 and 6, respectively.   
    2 Operator  
    Enter an operator name.   
    3 Instrument  
    The name of the connected instrument is indicated.   
    4 Sampling  
    - Standard   
    Indicates that samples are to be exchanged manually.     
    5 Comments  
    Enter a description or notes on measuring conditions.   
    6 Use sample table (Set measurement sample)   
    Apply a check mark here when a sample table is used.   
    3.2.1 General Tab 
     
    1 
    2 
    3 
    4 
    5 
    6 
    7 
    8 
    9  
    						
    							3.2    Creating an Analysis Method 
    3 - 4 
    7 Load  
    Click this button to read out saved analytical conditions.     
    Upon clicking, the file opening window appears.    So, select an 
    analytical condition file.   
    8 Save  
    When you click this button after setting the parameters, the 
    analytical conditions can be saved by overwriting.    In case of a 
    new analysis method, clicking this button displays the “Save As” 
    window.    You must enter a file name for saving in this window.   
    9 Save As   
    Click this button for renaming the set conditions for saving.     
    The “Save As” window appears, so enter a file name.   
      
    						
    							3.2.2 
    3 - 5 
    This window appears when Time scan is specified for Measurement at 
    the General tab on the Analysis Method window.    When you click the 
    Instrument tab, the window in Fig. 3-4 appears.   
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 3-4  Instrument Tab  
     
     
    1 Data mode   
    Select a data mode.    The Luminescence mode is for measuring 
    luminescence in the biological/chemical field.    In this mode, only 
    the sample side signal is acquired without performing ratio 
    photometry.    If phosphorescence life is shorter than 20 ms, 
    “Phosphorescence short life” should be selected.    In this case, unit 
    time and measuring time are automatically set to ms and 20 ms. 
    2 Chopping speed   
    Select a chopping speed.    Setting is allowed when 
    Phosphorescence is selected for Data mode.     
    The Phosphorescence mode is for acquiring only phosphorescence 
    after extinction of excitation beam while turning a chopper on the 
    excitation side.    An increase in chopping speed makes the 
    chopper turns faster.   
    3.2.2 Instrument Tab 
     
    2 
    3 
    4 
    5 
    6 
    1 
    7 
    11
    12
    13
    14
    15
    10
    9
    8
    16 
    						
    All Hitachi manuals Comments (0)