    							A - 48 
    G.3.1A Photometry:  Quantification Tab (Quantification type:  Wavelength) 
     
    Parameter Name  Setting Range 
    Quantification type  Wavelength, Peak area, Peak height, Derivative, Ratio 
    Calibration type  None, 1st order, 2nd order, 3rd order, Segmented 
    Number of wavelengths  
    Calibration Type  Setting Range 
    None  1 to 6 
    1st order 
    2nd order 
    3rd order 
    Segmented 1 to 3     
    Concentration unit  Up to 8 characters 
    Manual calibration  ON, OFF 
    ON: Factors are specifiable. 
    Factor Setting Range 
    A0 to A3  -100000 to 100000     
    Force curve through zero  ON, OFF 
    Digit after decimal point  0 to 3 
    Lower concentration limit  0 to 1e+009 
    Upper concentration limit  0 to 1e+009 
     
     
    G.3.1B  Photometry:  Quantification Tab (Quantification type:    Peak area, 
    Peak height) 
     
    Parameter Name  Setting Range 
    Quantification type  Wavelength, Peak area, Peak height, Derivative, Ratio 
    Calibration type  None, 1st order, 2nd order, 3rd order, Segmented 
    Peak apex  Wavelength range on Instrument tab page ± 0.1 to 700.0 nm
    Concentration unit  Up to 8 characters 
    Manual calibration  ON, OFF 
    ON: Factors are specifiable. 
    Factor Setting Range 
    A0 to A3  -100000 to 100000     
    Force curve through zero  ON, OFF 
    Digit after decimal point  0 to 3 
    Lower concentration limit  0 to 1e+009 
    Upper concentration limit  0 to 1e+009 
    Integration method  Rectangular, Trapezoid, Romberg 
    Threshold  0.001 to 1000 
    Sensitivity  1, 2, 4, 8  
    						
    							A - 49 
    G.3.1C Photometry:  Quantification Tab (Quantification type:  Derivative) 
     
    Parameter Name  Setting Range 
    Quantification type  Wavelength, Peak area, Peak height, Derivative, Ratio 
    Calibration type  None, 1st order, 2nd order, 3rd order, Segmented 
    Peak apex  Wavelength range on Instrument tab page 
    Concentration unit  Up to 8 characters 
    Manual calibration  ON, OFF 
    ON: Factors are specifiable. 
    Factor Setting Range 
    A0 to A3  -100000 to 100000     
    Force curve through zero  ON, OFF 
    Digit after decimal point  0 to 3 
    Lower concentration limit  0 to 1e+009 (1000000000) 
    Upper concentration limit  0 to 1e+009 (1000000000) 
    Derivative order  2, 4 
    Smoothing order  2 to 4 
    Number of points  5 to 101 
     
     
    G.3.1D Photometry:  Quantification Tab (Quantification type:  Ratio) 
     
    Parameter Name  Setting Range 
    Quantification type  Wavelength, Peak area, Peak height, Derivative, 
    Ratio 
    Subtract background [Wavelength 3] ON, OFF 
    K-Factor 0 
    K-Factor 1 
    K-Factor 2 -9999 to -0.001, 0.001 to 9999 
    Operand  + (plus), - (minus), * (times) 
      
    						
    							A - 50 
    G.3.2A  Quantification:  Instrument Tab (Quantification type:    Wavelength, 
    Ratio) 
     
    Parameter Name  Setting Range 
    Data mode  Fluorescence, Luminescence, Phosphorescence 
    Chopping speed  10 Hz, 20 Hz, 40 Hz 
    Wavelength mode  EX WL Fixed, EM WL Fixed, Both WL Fixed 
    Fixed WL  Input Range:    0, 200 to 900 nm 
    WL 1 to WL 6  Input Range:    0, 200 to 900 nm 
    EX Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    EM Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    PMT Voltage  Setting range:    250, 400, 700, 950 V 
    0 to 1000 V when “PMT Voltage 0-1000V” is turned ON 
    (checked). 
    PMT Voltage 0 - 1000 V  ON, OFF 
    Auto statistic calc. number  2 to 4000 
    Replicates  1 to 20 
    Integration time  0.1 to 10.0 s 
    Delay  0 to 9999 s 
     
      
    						
    							A - 51 
    G.3.2B  Photometry:  Instrument Tab (Quantification type:    Peak area, Peak 
    height, Derivative) 
     
    Parameter Name  Setting Range 
    Scan mode  Excitation, Emission, Synchronous 
    Data mode  Fluorescence, Luminescence, Phosphorescence 
    Chopping speed  10 Hz, 20 Hz, 40 Hz 
    EX WL  
    Scan Mode  Input Range 
    Excitation  0, 200 to 900 nm 
    Synchronous  200 to 890 nm     
    EX Start WL  Input Range:    200 to 890 nm 
    EX End WL  Input Range:    210 to 900 nm 
    EM WL  Input Range:    0, 200 to 900 nm 
    EM Start WL  Input Range:    200 to 890 nm 
    EM End WL  Input Range:    210 to 900 nm 
    Scan speed (unit: nm/min)  60000 
    30000 
    12000 
    2400 
    1200 
    240 
    60 
    30 
    When Synchronous is specified for scan mode,   
    1200 nm/min or less is indicated. 
    When Phosphorescence is specified for data mode, select 
    from the following.    Chopping speed is in parentheses. 
    (40 Hz)  (20 Hz/10 Hz)
    240 
    60 
    30 60 
    30 
        
    Delay  0 to 9999 s 
    EX Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    EM Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    PMT Voltage  Setting range:    250, 400, 700, 950 V 
    0 to 1000 V when “PMT Voltage 0-1000V” is turned ON 
    (checked). 
    PMT Voltage 0-1000V  ON, OFF  
    						
    							A - 52 
    (cont’d) 
    Parameter Name  Setting Range 
    Response (unit: s)  0.002 
    0.004 
    0.01 
    0.05 
    0.1 
    0.5 
    2.0 
    4.0 
    8.0 
    Auto 
    Corrected spectra  ON, OFF 
    Shutter control  ON, OFF 
    Replicates  1 to 20 
    Auto statistic calc. number  2 to 4000 
     
      
    						
    							A - 53 
    G.3.3  Photometry:  Standard Tab 
     
    Parameter Name  Setting Range 
    Number of Samples  1 to 20 
    Sample  Up to 40 characters 
    Comment  Up to 255 characters 
    Concentration  0 to 1000000000 
     
     
    G.3.4  Photometry:  Monitor Tab 
     
    Parameter Name  Setting Range 
    Y Axis: Max.  -119999999.9 to 120000000.0 
    Y Axis:    Min.  -120000000.0 to 119999999.9 
    Open data processing window after data acquisition  ON, OFF 
    Print report after data acquisition  ON, OFF 
     
     
    G.3.5  Photometry:  Report Tab 
     
    Parameter Name  Setting Range 
    Output  Print Report, Use Microsoft(R) Excel, Use print 
    generator sheet 
    Print items: Include date  ON, OFF 
    Print items: Include method  ON, OFF 
    Print items: Include calibration curve  ON, OFF 
    Print items: Include standard data  ON, OFF 
    Print items: Include standard comment  ON, OFF 
    Print items: Include calibration data  ON, OFF 
    Print items: Include sample data  ON, OFF 
    Print items: Include sample comment  ON, OFF 
      
    						
    							A - 54 
    G.4.1  3-D Scan:  Instrument Tab 
     
    Parameter Name  Setting Range 
    Data mode  Fluorescence, Phosphorescence 
    Chopping speed  10 Hz, 20 Hz, 40 Hz 
    EX Start WL  Input Range:    200 to 850 nm 
    EX End WL  Input Range:    250 to 900 nm 
    EX Sampling interval  Input Range:    1 to 50 nm 
    EM Start WL  Input Range:    200 to 850 nm 
    EM End WL  Input Range:    250 to 900 nm 
    EM Sampling interval  Input Range:    1 to 10 nm 
    Scan speed (unit: nm/min)  60000 
    30000 
    12000 
    2400 
    1200 
    240 
    60 
    30 
    when Synchronous is specified for scan mode,   
    1200 nm/min or less is indicated. 
    When Phosphorescence is specified for data mode, select 
    from the following.    Chopping speed is in parentheses. 
    (40 Hz)  (20 Hz/10 Hz) 
    240 
    60 
    30 60 
    30 
        
    EX Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    EM Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    PMT Voltage  Setting range:    250, 400, 700, 950 V 
    0 to 1000 V when “PMT Voltage 0-1000V” is turned ON 
    (checked). 
    PMT Voltage 0 - 1000 V  ON, OFF 
    Response (unit: s)  0.002 
    0.004 
    0.01 
    0.05 
    0.1 
    0.5 
    2.0 
    4.0 
    8.0 
    Auto 
      
    						
    							A - 55 
    (cont’d 
    Parameter Name  Setting Range 
    Corrected spectra  ON, OFF 
    Shutter control  ON, OFF 
     
    ON :  Checked 
    OFF :  Unchecked 
     
     
    G.4.2  3-D Scan:  Monitor Tab 
     
    Parameter Name  Setting Range 
    Y Axis: Max  -119999999.9 to 120000000.0 
    Y Axis: Min  -120000000.0 to 119999999.9 
    Contour interval  0.001 to 1000.000 
    Open data processing window after data acquisition  ON, OFF 
    Print report after data acquisition  ON, OFF 
     
     
    G.4.3  3-D Scan:  Processing Tab 
     
    Parameter Name  Setting Range 
    Peak Finding: Integration method  Rectangular 
    Peak Finding: Threshold  0.001 to 1000 
    Peak Finding: Sensitivity  1, 2, 4, 8 
     
      
    						
    							A - 56 
    G.4.4  3-D Scan:  Report Tab 
     
    Parameter Name  Setting Range 
    Output  Print Report, Use Microsoft(R) Excel, Use print 
    generator sheet 
    Orientation Portrait, Landscape 
    Print items: Include date  ON, OFF 
    Print items: Include instrument  ON, OFF 
    Print items: Include graph  ON, OFF 
    ON: Any of the following items is selectable. 
    • Contour 
    • Bird’s-eye view 
    •  Contour and Bird’s-eye view 
    •  Contour and Spectrum 
    Print items: Include peak table  ON, OFF 
    Print items: Data listing*1 ON, OFF 
    ON:  Setting can be made for excitation and emission 
    spectra. 
    Parameter Setting Range 
    Interval  Minimum data interval to 
    measuring wavelength interval 
    Start  Wavelength range on Instrument 
    tab page 
    End  Wavelength range on Instrument 
    tab page 
        
     
    *1:  Valid when “Use Microsoft(R) Excel” or “Use print generator sheet” is specified for 
    Output.    Always turned ON when “Use print generator sheet” is specified. 
      
    						
    							A - 57 
    G.5.1  3-D Time Scan:    Instrument Tab 
     
    Parameter Name  Setting Range 
    Scan mode  Excitation, Emission, Synchronous 
    Data mode  Fluorescence, Luminescence, Phosphorescence 
    Chopping speed  10 Hz, 20 Hz, 40 Hz 
    EX WL   
    Scan Mode  Input Range 
    Excitation  0, 200 to 900 nm 
    Synchronous  200 to 890 nm    
    EX Start WL  Input Range:    200 to 890 nm 
    EX End WL  Input Range:    210 to 900 nm 
    EM WL  Input Range:    0, 200 to 900 nm 
    EM Start WL  Input Range:    200 to 890 nm 
    EM End WL  Input Range:    210 to 900 nm 
    Scan speed (unit: nm/min)  60000 
    30000 
    12000 
    2400 
    1200 
    240 
    60 
    30 
    When Synchronous is specified for scan mode,   
    1200 nm/min or less is indicated. 
    When Phosphorescence is specified for data mode, select 
    from the following.    Chopping speed is in parentheses. 
    (40 Hz)  (20 Hz/10 Hz)
    240 
    60 
    30 60 
    30 
        
    EX Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    EM Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    PMT Voltage  250, 400, 700, 950 V 
    0 to 1000 V when “PMT Voltage 0-1000V” is turned ON 
    (checked). 
    PMT Voltage 0 - 1000 V  ON, OFF  
    						