A - 38 (7) Set the diffuser into the cell holder in the sample compartment. (Refer to Fig. F-12.) Fig. F-12 (8) Execute the Instrumental Response command from the Utility menu. The Instrumental Response dialog box will open as shown in Fig. F-13. Select “EX [500 to 900 nm].” (Refer to ① in Fig. F-13.) Fig. F-13 Diffuse r 1 2
A - 39 (9) Click the Run button. (Refer to ② in Fig. F-13.) A message box “Set diffuser to cell holder.” will appear. Click the OK button. (Refer to ③ in Fig. F-14.) Fig. F-14 (10) The EX and EM wavelengths are set at 498.6 nm and the instrument starts synchronous wavelength scan (from 498.6 to 901.4 nm). At this stage, the following measuring conditions are automatically set. • Scan mode : Emission • Data mode : Luminescence • EX slit : 5.0 nm • EM slit : 5.0 nm • Scan speed : 60 nm/min • PMT voltage : 350 V • Response : Auto (11) Upon completion of measurement, the conditions set before measurement are retrieved. Now, the instrument response on the emission side has been measured and stored. Take out the diffuser. 3
A - 40 F.5 Resetting of Instrumental Response (1) Click the Reset button in the Instrumental Response dialog box. (Refer to ① in Fig. F-15.) Fig. F-15 (2) A message box “Erase instrumental response on excitation/emission side. OK?” will appear. Click the OK button. (Refer to ② in Fig. F-16.) Fig. F-16 (3) Now, the instrumental response has been erased. 1 2
A - 41 APPENDIX G PARAMETER SETTING RANGE FOR ANALYSIS METHOD G.1 General Tab: Common to all measurement modes Parameter Name Setting Range Measurement Wavelength scan, Time scan, Photometry, 3-D scan, 3-D Time scan Operator Up to 20 characters Comments Up to 255 characters Use sample table ON, OFF ON : Checked OFF : Unchecked G.1.1 Wavelength Scan: Instrument Tab Parameter Name Setting Range Scan mode Excitation, Emission, Synchronous Data mode Fluorescence, Luminescence, Phosphorescence Chopping speed 10 Hz, 20 Hz, 40 Hz EX WL Scan Mode Input Range Excitation 0, 200 to 900 nm Synchronous 200 to 890 nm EX Start WL Input Range: 200 to 890 nm EX End WL Input Range: 210 to 900 nm EM WL Input Range: 0, 200 to 900 nm EM Start WL Input Range: 200 to 890 nm EM End WL Input Range: 210 to 900 nm
A - 42 (cont’d) Parameter Name Setting Range Scan speed (unit: nm/min) 60000 30000 12000 2400 1200 240 60 30 When Synchronous is specified for scan mode, 1200 nm/min or less is indicated. When Phosphorescence is specified for data mode, select from the following. Chopping speed is in parentheses. F-7000 (40 Hz) F-7000 (20 Hz/10 Hz) 240 60 30 60 30 Delay 0 to 9999 s EX Slit (unit: nm) 1.0 2.5 5.0 10.0 20.0 EM Slit (unit: nm) 1.0 2.5 5.0 10.0 20.0 PMT Voltage Setting range: 250, 400, 700, 950 V 0 to 1000 V when “PMT Voltage 0-1000V” is turned ON (checked). PMT Voltage 0 - 1000 V ON, OFF Response (unit: s) 0.002 0.004 0.01 0.05 0.1 0.5 2.0 4.0 8.0 Auto Corrected spectra ON, OFF Shutter control ON, OFF Replicates 1 to 99 Cycle time 0.0 to 180.0 min ON : Checked OFF : Unchecked
A - 43 G.1.2 Wavelength Scan: Monitor Tab Parameter Name Setting Range Y Axis: Max -119999999.9 to 120000000.0 Y Axis: Min -120000000.0 to 119999999.9 Open data processing window after data acquisition ON, OFF Open data processing window after acquisition (Overlay) ON, OFF Print report after data acquisition ON, OFF Overlay ON, OFF G.1.3 Wavelength Scan: Processing Tab Parameter Name Setting Range CAT ON, OFF Processing choices Savitsky-Golay, Smooth, Mean Smooth, Median Smooth, Derivative Peak Finding: Integration method Rectangular, Trapezoid, Romberg Peak Finding: Threshold 0.001 to 1000 Peak Finding: Sensitivity 1, 2, 4, 8
A - 44 G.1.4 Wavelength Scan: Report Tab Parameter Name Setting Range Output Print Report, Use Microsoft(R) Excel, Use print generator sheet Orientation Portrait, Landscape Print items: Include date ON, OFF Print items: Include method ON, OFF Print items: Include graph ON, OFF Print items: Include data listing ON, OFF ON: “Constant” or “Select data” is selectable. When “Constant” is specified: Parameter Setting Range Data interval Minimum sampling interval to measuring wavelength interval Data start Wavelength range on Instrument tab page Data end Wavelength range on Instrument tab page When “Select data” is specified: Selectable from 12 wavelengths. Print items: Include peak table ON, OFF ON: Items to be output peak table are selectable. Item Setting Range Peak WL/Peak data ON, OFF Start WL/End WL ON, OFF Valley WL/Valley data ON, OFF Peak Area ON, OFF
A - 45 G.2.1 Time Scan: Instrument Tab Parameter Name Setting Range Data mode Fluorescence, Luminescence, Phosphorescence, Phosphorescence Life Time, Phosphorescence Life Time (short) Chopping speed 10 Hz, 20 Hz, 40 Hz EX WL Input Range: 0, 200 to 900 nm EM WL Input Range: 0, 200 to 900 nm Time unit s, ms Scan time Time Unit Input Range s 10 to 9000 s ms 100 to 9999 ms When Phosphorescence Life Time (short) is specified for data mode: Chopping Speed Scan Time 40 Hz Fixed at 20 ms 20 Hz Fixed at 40 ms 10 Hz Fixed at 80 ms Delay 0 to 9999 s EX Slit (unit: nm) 1.0 2.5 5.0 10.0 20.0 EM Slit (unit: nm) 1.0 2.5 5.0 10.0 20.0 PMT Voltage Setting range: 250, 400, 700, 950 V 0 to 1000 V when “PMT Voltage 0-1000V” is turned ON (checked). PMT Voltage 0 - 1000 V ON, OFF Response (unit: s) 0.002 0.004 0.01 0.05 0.1 0.5 2.0 4.0 8.0 Shutter control ON, OFF Replicates 1 to 99 Cycle time 0.0 to 180.0 min
A - 46 G.2.2 Time Scan: Monitor Tab Parameter Name Setting Range Y Axis: Max -119999999.9 to 120000000.0 Y Axis: Min -120000000.0 to 119999999.9 X Axis: Max (unit: s) 0.1 to 9999.0 X Axis: Min (unit: s) 0.0 to 9998.9 Open data processing window after data acquisition ON, OFF Open data processing window after acquisition (Overlay) ON, OFF Print report after data acquisition ON, OFF Overlay ON, OFF G.2.3 Time Scan: Processing Tab Parameter Name Setting Range Processing choices Savitsky-Golay Smooth, Mean Smooth, Median Smooth, Derivative Peak Finding: Integration method Rectangular, Trapezoid, Romberg Peak Finding: Threshold 0.001 to 1000 Peak Finding: Sensitivity 1, 2, 4, 8 Kinetics: Start time 0 to 3599 s, 0 to 9998 ms Kinetics: End time 0 to 3600 s, 1 to 9999 ms Kinetics: K-Factor -9999 to -0.001, 0.001 to 9999
A - 47 G.2.4 Time Scan: Report Tab Parameter Name Setting Range Output Print Report, Use Microsoft(R) Excel, Use print generator sheet Orientation Portrait, Landscape Print items: Include date ON, OFF Print items: Include method ON, OFF Print items: Include graph ON, OFF Print items: Include data listing ON, OFF ON: “Constant” or “Select data” is selectable. When “Constant” is specified: Parameter Setting Range Data interval Minimum sampling interval to measuring scan time Data start Scan time range on Instrument tab page Data end Scan time range on Instrument tab page When “Select data” is specified: Selectable from 12 time points (in measurement range). Print items: Include peak table ON, OFF ON: Items to be output onto peak table are selectable. Item Setting Range Peak time/Peak data ON, OFF Start time/End time ON, OFF Valley time/Valley data ON, OFF Peak Area ON, OFF Print items: Include kinetics ON, OFF