    							A - 38 
    (7)  Set the diffuser into the cell holder in the sample compartment.   
    (Refer to Fig. F-12.) 
     
     
     
    Fig. F-12 
     
     
    (8)  Execute the Instrumental Response command from the Utility 
    menu.    The Instrumental Response dialog box will open as 
    shown in Fig. F-13. 
    Select “EX [500 to 900 nm].”    (Refer to ①  in Fig. F-13.) 
     
     
     
    Fig. F-13 
     
     Diffuse
    r 
    1 
    2 
    						
    							A - 39 
    (9) Click the Run button.  (Refer to ② in Fig. F-13.)  A message 
    box “Set diffuser to cell holder.” will appear.    Click the OK button.   
    (Refer to ③  in Fig. F-14.) 
     
     
     
    Fig. F-14 
     
     
    (10)  The EX and EM wavelengths are set at 498.6 nm and the 
    instrument starts synchronous wavelength scan (from 498.6 to 
    901.4 nm). 
    At this stage, the following measuring conditions are automatically 
    set. 
    • Scan mode :  Emission 
    • Data mode :  Luminescence 
    •  EX slit  :  5.0 nm 
    •  EM slit  :  5.0 nm 
    •  Scan speed  :  60 nm/min 
    •  PMT voltage :  350 V 
    • Response  :  Auto 
     
    (11)  Upon completion of measurement, the conditions set before 
    measurement are retrieved. 
    Now, the instrument response on the emission side has been 
    measured and stored.    Take out the diffuser. 
     
    3 
    						
    							A - 40 
    F.5    Resetting of Instrumental Response 
     
    (1) Click the Reset button in the Instrumental Response dialog box.   
    (Refer to ①  in Fig. F-15.) 
     
     
     
    Fig. F-15 
     
     
    (2)  A message box “Erase instrumental response on 
    excitation/emission side. OK?” will appear.    Click the OK button.   
    (Refer to ②  in Fig. F-16.) 
     
     
     
    Fig. F-16 
     
     
    (3)  Now, the instrumental response has been erased. 
     
    1
    2 
    						
    							A - 41 
    APPENDIX G  PARAMETER SETTING RANGE 
    FOR ANALYSIS METHOD 
     
    G.1  General Tab:  Common to all measurement 
    modes 
     
    Parameter Name  Setting Range 
    Measurement  Wavelength scan, Time scan, Photometry, 3-D scan,   
    3-D Time scan 
    Operator  Up to 20 characters 
    Comments  Up to 255 characters 
    Use sample table  ON, OFF 
     
    ON :  Checked 
    OFF :  Unchecked 
     
     
    G.1.1  Wavelength Scan:  Instrument Tab 
     
    Parameter Name  Setting Range 
    Scan mode  Excitation, Emission, Synchronous 
    Data mode  Fluorescence, Luminescence, Phosphorescence 
    Chopping speed  10 Hz, 20 Hz, 40 Hz 
    EX WL  
    Scan Mode  Input Range 
    Excitation  0, 200 to 900 nm 
    Synchronous  200 to 890 nm   
    EX Start WL  Input Range:    200 to 890 nm 
    EX End WL  Input Range:    210 to 900 nm 
    EM WL  Input Range:    0, 200 to 900 nm 
    EM Start WL  Input Range:    200 to 890 nm 
    EM End WL  Input Range:    210 to 900 nm  
    						
    							A - 42 
    (cont’d) 
    Parameter Name  Setting Range 
    Scan speed   
    (unit: nm/min) 60000 
    30000 
    12000 
    2400 
    1200 
    240 
    60 
    30 
    When Synchronous is specified for scan mode, 1200 nm/min or less 
    is indicated. 
    When Phosphorescence is specified for data mode, select from the 
    following.   
    Chopping speed is in parentheses. 
    F-7000 (40 Hz)  F-7000 (20 Hz/10 Hz) 
    240 
    60 
    30 60 
    30 
      
    Delay  0 to 9999 s 
    EX Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    EM Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    PMT Voltage  Setting range:    250, 400, 700, 950 V 
    0 to 1000 V when “PMT Voltage 0-1000V” is turned ON (checked). 
    PMT Voltage   
    0 - 1000 V ON, OFF 
    Response (unit: s)  0.002 
    0.004 
    0.01 
    0.05 
    0.1 
    0.5 
    2.0 
    4.0 
    8.0 
    Auto 
    Corrected spectra  ON, OFF 
    Shutter control  ON, OFF 
    Replicates  1 to 99 
    Cycle time  0.0 to 180.0 min 
     
    ON :  Checked 
    OFF :  Unchecked  
    						
    							A - 43 
    G.1.2  Wavelength Scan:  Monitor Tab 
     
    Parameter Name  Setting Range 
    Y Axis: Max  -119999999.9 to 120000000.0 
    Y Axis: Min  -120000000.0 to 119999999.9 
    Open data processing window after data acquisition  ON, OFF 
    Open data processing window after acquisition 
    (Overlay) ON, OFF 
    Print report after data acquisition  ON, OFF 
    Overlay ON, OFF 
     
     
    G.1.3  Wavelength Scan:  Processing Tab 
     
    Parameter Name  Setting Range 
    CAT ON, OFF 
    Processing choices  Savitsky-Golay, Smooth, Mean Smooth,   
    Median Smooth, Derivative 
    Peak Finding: Integration method  Rectangular, Trapezoid, Romberg 
    Peak Finding: Threshold  0.001 to 1000 
    Peak Finding: Sensitivity  1, 2, 4, 8 
     
      
    						
    							A - 44 
    G.1.4  Wavelength Scan:  Report Tab 
     
    Parameter Name  Setting Range 
    Output  Print Report, Use Microsoft(R) Excel,   
    Use print generator sheet 
    Orientation Portrait, Landscape 
    Print items: Include date  ON, OFF 
    Print items: Include method  ON, OFF 
    Print items: Include graph  ON, OFF 
    Print items: Include data listing  ON, OFF 
    ON:    “Constant” or “Select data” is selectable. 
    When “Constant” is specified: 
    Parameter Setting Range 
    Data interval  Minimum sampling interval to 
    measuring wavelength interval 
    Data start  Wavelength range on 
    Instrument tab page 
    Data end  Wavelength range on 
    Instrument tab page 
    When “Select data” is specified: 
    Selectable from 12 wavelengths. 
    Print items: Include peak table  ON, OFF 
    ON:    Items to be output peak table are selectable. 
    Item Setting Range 
    Peak WL/Peak data  ON, OFF 
    Start WL/End WL  ON, OFF 
    Valley WL/Valley data  ON, OFF 
    Peak Area  ON, OFF   
     
      
    						
    							A - 45 
    G.2.1  Time Scan:  Instrument Tab 
     
    Parameter Name  Setting Range 
    Data mode  Fluorescence, Luminescence, Phosphorescence,   
    Phosphorescence Life Time,   
    Phosphorescence Life Time (short)   
    Chopping speed  10 Hz, 20 Hz, 40 Hz 
    EX WL  Input Range:    0, 200 to 900 nm 
    EM WL  Input Range:    0, 200 to 900 nm 
    Time unit  s, ms 
    Scan time  
    Time Unit  Input Range 
    s  10 to 9000 s 
    ms  100 to 9999 ms 
     
    When Phosphorescence Life Time (short) is specified for data 
    mode: 
    Chopping Speed  Scan Time 
    40 Hz  Fixed at 20 ms 
    20 Hz  Fixed at 40 ms 
    10 Hz  Fixed at 80 ms   
    Delay  0 to 9999 s 
    EX Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    EM Slit (unit: nm)  1.0 
    2.5 
    5.0 
    10.0 
    20.0 
    PMT Voltage  Setting range:    250, 400, 700, 950 V 
    0 to 1000 V when “PMT Voltage 0-1000V” is turned ON (checked). 
    PMT Voltage   
    0 - 1000 V ON, OFF 
    Response (unit: s)  0.002 
    0.004 
    0.01 
    0.05 
    0.1 
    0.5 
    2.0 
    4.0 
    8.0 
    Shutter control  ON, OFF 
    Replicates  1 to 99 
    Cycle time  0.0 to 180.0 min 
      
    						
    							A - 46 
    G.2.2  Time Scan:  Monitor Tab 
     
    Parameter Name  Setting Range 
    Y Axis: Max  -119999999.9 to 120000000.0 
    Y Axis: Min  -120000000.0 to 119999999.9 
    X Axis: Max (unit: s)  0.1 to 9999.0 
    X Axis: Min (unit: s)  0.0 to 9998.9 
    Open data processing window after 
    data acquisition ON, OFF 
    Open data processing window after 
    acquisition (Overlay) ON, OFF 
    Print report after data acquisition  ON, OFF 
    Overlay ON, OFF 
     
     
    G.2.3  Time Scan:  Processing Tab 
     
    Parameter Name  Setting Range 
    Processing choices  Savitsky-Golay Smooth, Mean Smooth,   
    Median Smooth, Derivative 
    Peak Finding: Integration method  Rectangular, Trapezoid, Romberg 
    Peak Finding: Threshold  0.001 to 1000 
    Peak Finding: Sensitivity  1, 2, 4, 8 
    Kinetics: Start time  0 to 3599 s, 0 to 9998 ms 
    Kinetics: End time  0 to 3600 s, 1 to 9999 ms 
    Kinetics: K-Factor  -9999 to -0.001, 0.001 to 9999 
      
    						
    							A - 47 
    G.2.4  Time Scan:  Report Tab 
     
    Parameter Name  Setting Range 
    Output  Print Report, Use Microsoft(R) Excel,   
    Use print generator sheet 
    Orientation Portrait, Landscape 
    Print items: Include date  ON, OFF 
    Print items: Include method  ON, OFF 
    Print items: Include graph  ON, OFF 
    Print items: Include data listing  ON, OFF 
    ON: “Constant” or “Select data” is selectable. 
    When “Constant” is specified: 
    Parameter Setting Range 
    Data interval  Minimum sampling interval to 
    measuring scan time 
    Data start  Scan time range on 
    Instrument tab page 
    Data end  Scan time range on 
    Instrument tab page 
    When “Select data” is specified: 
    Selectable from 12 time points (in measurement 
    range). 
    Print items: Include peak table  ON, OFF 
    ON:    Items to be output onto peak table are selectable.
    Item Setting Range 
    Peak time/Peak data  ON, OFF 
    Start time/End time  ON, OFF 
    Valley time/Valley data  ON, OFF 
    Peak Area  ON, OFF   
    Print items: Include kinetics  ON, OFF 
     
      
    						