Home > Hitachi > Fluorescence Spectrophotometer > Hitachi F7000 Instruction Manual

Hitachi F7000 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F7000 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							6.1 
    6 - 1 
    6.  3-D TIME SCAN  
     
    6.1  Flow of Operation  
     
    This section explains the parameter setting and operating method for 3-
    D time scan.    Measurement proceeds along the following flow.   
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
    Fig. 6-1    Operational Flow (3-D time scan) Click 
       button and set your analytical conditions. 
     
     • Scan range 
    • Data mode 
    •  Scan speed, etc. 
    Click 
     button. 
     • Sample name 
    • Auto File 
    •  File name, etc. 
    Set a sample. 
    Click 
      button to execute wavelength scan. 
    • Spectrum tracing 
    • Peak detection 
    Click 
      button.    Measurement result will be displayed 
    in a report format. 
     
    Start 
    Method setting 
    (See 6.2.) 
    Sample name/ 
    comment input 
    (See 6.3.) 
    Measurement 
    (See 6.4.) 
    Data processing 
    (See Section 7.) 
    Result output in report 
    (See 6.9.) 
    End  
    						
    							6.2    Creating an Analysis Method 
    6 - 2 
    6.2    Creating an Analysis Method   
     
    Set the analytical conditions (method) prior to measurement.     
     
    (1) Click the 
       button on the Measurement toolbar.    The window 
    in Fig. 6-2 appears.   
    (Or select the Method command from the Edit menu.)     
     
     
     
    Fig. 6-2  Analysis Method  
     
     
    (2)  On the Analysis Method window, analytical conditions for the 3-D 
    time scan mode can be set.    The following tabs are provided on 
    this window.    These tabs are detailed on the subsequent pages.   
    • General  
    • Instrument  
    • Monitor  
    • Processing  
    • Report  
     
      
    						
    							6.2.1 
    6 - 3 
    The parameters shown in Fig. 6-3 are provided at the General tab.     
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 6-3  General Tab 
     
     
    1 Measurement (mode)   
    A measurement mode is selectable.    Here, select “3-D Time 
    scan.”  
    For Wavelength scan, Time scan, Photometry and 3-D Scan, refer 
    to Sections 2, 3, 4 and 5, respectively.   
    2 Operator  
    Enter an operator name.   
    3 Instrument  
    The name of the connected instrument is indicated.   
    4 Sampling 
    - Standard   
    Indicates that samples are to be exchanged manually.   
    5 Comments  
    Enter a description or notes on measuring conditions.   
    6.2.1 General Tab 
     
    1 
    2 
    3 
    4 
    5 
    6 
    7 
    8  
    						
    							6.2    Creating an Analysis Method 
    6 - 4 
    6 Load  
    Click this button to read out saved analytical conditions.     
    Upon clicking, the file opening window appears.    So, select an 
    analytical condition file.   
    7 Save  
    When you click this button after setting the parameters, the 
    analytical conditions can be saved by overwriting.    In case of a 
    new analysis method, clicking this button displays the “Save As” 
    window.    You must enter a file name for saving in this window.   
    8 Save As   
    Click this button for renaming the set conditions for saving.     
    The “Save As” window appears, so enter a file name.   
     
      
    						
    							6.2.2 
    6 - 5 
    The window shown below is based on that 3-D Time scan is specified 
    for Measurement at the General tab.    When you click the Instrument 
    tab, the window in Fig. 6-4 appears.   
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 6-4  Instrument Tab 
     
     
    1 Scan mode   
    Select a scan mode here.    In case of Synchronous, the abscissa 
    of the graph stands for excitation wavelength.    The unit indication 
    of the abscissa becomes ‘EX/EM (nm).’   
    2 Data mode   
    Select a data mode.    The Phosphorescence mode is for acquiring 
    only phosphorescence after extinction of excitation beam while 
    turning a chopper on the excitation side.   
    3 Chopping speed   
    Select a chopping speed.    Setting is allowed when 
    Phosphorescence is selected for Data mode.     
    The Phosphorescence mode is for acquiring only phosphorescence 
    after extinction of excitation beam while turning a chopper on the 
    excitation side.    An increase in chopping speed makes the 
    chopper turns faster.   
    6.2.2 Instrument 
    Tab 
    3 
    4 
    5 
    6 
    7 
    2 
    1 
    8 
    9 
    21 10 
    13
    14
    15
    16
    17
    12
    11
    18
    19
    20 
    						
    							6.2    Creating an Analysis Method 
    6 - 6 
    4 EM WL   
    For the Excitation scan mode, enter the fixed wavelength on the 
    emission side .    For the Synchronous scan mode, enter the start 
    wavelength on the emission side.   
    5 EX Start WL   
    This is the start wavelength on the excitation side in wavelength 
    scan.    Enter the shorter wavelength side of measuring wavelength 
    range.  
    6 EX End WL   
    This is the end wavelength on the excitation side in wavelength 
    scan.    Enter the longer wavelength side of measuring wavelength 
    range.  
    7 EX WL   
    For the Emission scan mode, enter the fixed wavelength on the 
    excitation side.   
    8 EM Start WL   
    This is the start wavelength on the emission side in wavelength 
    scan.    Enter the shorter wavelength side of measuring wavelength 
    range.  
    9 EM End WL   
    This is the end wavelength on the emission side in wavelength 
    scan.    Enter the longer wavelength side of measuring wavelength 
    range.  
    10 Scan speed   
    Set a scan speed.   
    11 EX Slit   
    Select a slit width on the excitation side.   
    12 EM Slit   
    Select a slit width on the emission side.   
    13 PMT Voltage   
    A function for controlling the voltage of the photomultiplier detector 
    (photomultiplier tube).    As a reference, changing from 400 V to 
    700 V increases the data value about two digits and changing from 
    700 V to 950 V increases it about 1 digit.   
    When a check mark is put for PMT Voltage 0-1000V, the 
    photomultiplier voltage is settable within a range from 0 to 1000 V.   
    14 PMT Voltage 0-1000V   
    When a check mark is put in this box, the photomultiplier voltage is 
    settable within a range from 0 to 1000 V.    
    						
    							6.2.2 
    6 - 7 
    15 Response  
    When “Auto” is selected, response will be set automatically 
    depending on scan speed.    Select “Auto” usually.   
    16 Corrected spectra   
    A function for determining the spectrum inherent to a sample by 
    correcting the photometer wavelength characteristic using the 
    saved instrument parameters, following measurement with the 
    instrument parameters for photometer control.    If there are no 
    instrument parameters for the wavelength range to be measured, 
    only the wavelength range that allows correction is scanned.     
    And if there is no wavelength area measurable, an error will occur.   
    The Instrument Response command is provided in the Utility menu 
    for operating this function.   
    When this setting is at ON, “Corrected spectra” appears on the 
    monitor window.    (See Fig. 1-43 in 1.6.)   
    17 Shutter control   
    The shutter can automatically be closed in other than measurement 
    for slowing down sample deterioration due to the energy of 
    excitation beam and opened when measurement starts.     
    18 Replicates  
    Set the number of repeat measurements.     
    19 Cycle time   
    Set a repetition interval.   
    20 Range of corrected spectra   
    Spectrum correction range is indicated.   
    21 Prediction measurement time   
    The predicted measurement time is indicated.    After change of a 
    parameter such as scan speed, click the Update button.   
    The predicted value will then be updated.   
     
      
    						
    							6.2    Creating an Analysis Method 
    6 - 8 
    When selecting the Monitor tab, the window in Fig. 6-5 appears.     
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 6-5  Monitor Tab  
     
     
    Assign the following parameters.   
     
    1 Y-Axis Max (maximum)   
    Enter a maximum value for Y-axis on the monitor window.   
    2 Y-Axis Min (minimum)   
    Enter a minimum value for Y-axis on the monitor window.   
    3 Contour interval   
    Function for setting intervals of drawing contour lines.     
    4 Open data processing window after data acquisition   
    Select whether or not to conduct data processing after sample 
    measurement.    When selected (check mark is applied), an icon 
    (as shown in 
    ) is displayed for the data 
    processing window at the end of measurement.    And by opening 
    this icon, data processing such as peak detection is available.   
     
    6.2.3 Monitor Tab 
     
    2 
    1 
    4 
    5 
    3  
    						
    							6.2.3 
    6 - 9 
    NOTE:  When the remaining memory capacity of PC in operation is 
    very small (system resource is less than 20%), the data 
    processing window will not open even if this item is turned ON.   
    Either close the presently opened data processing window or 
    restart the FL Solutions program.   
     
    5 Print report after data acquisition   
    Select whether or not to print a report after sample measurement.   
    When selected (check mark is applied), printing will be made 
    automatically after measurement.    The items set at the Report tab 
    will be printed.    When this item is not selected, data processing is 
    allowed by reading out a file via [File] - [Open].   
     
      
    						
    							6.2    Creating an Analysis Method 
    6 - 10 
    On selection of the Processing tab, the window in Fig. 6-6 will be 
    displayed.  
    For the settable range of each parameter, refer to Appendix G.   
     
     
     
    Fig. 6-6  Processing Tab  
     
     
    1 Peak Finding:  Integration method  
    Select an integration method for the peaks to be displayed in the 
    data processing window.   
    2 Threshold  
    Set a detection limit for the photometric value axis of peak and 
    valley.    Peaks or valleys below the setting will not be detected.   
    3 Sensitivity  
    Select the number of data points in the horizontal axis direction 
    (abscissa).    For detecting sharp peaks, select a sensitivity value of 
    “1.”    Or select “8” for detecting broad peaks.   
     
     
    6.2.4 Processing 
    Tab 
     
    1 
    3
    2 
    						
    All Hitachi manuals Comments (0)