    							Presets and System setup
    Vivid7/Vivid7 PRO Users Manual427
    FC092326-03
    RemoteArch-RemoteHD Remote archive remote harddrive
    A remote database (either on 
    EchoPAC workstation or on 
    EchoServer) is used for patient 
    archiving. Images are stored to a 
    network image volume (either 
    internal HD on EchoPAC workstation 
    or EchoServer volume).
    Remote Archive - Remote 
    HD/DICOM ServerA remote database is used for 
    patient archiving. Images are stored 
    to a network image volume and to a 
    DICOM server.
    This dataflow 
    cannot be used 
    with an 
    ImageVault 3.0 
    server.RemoteArchive-MODRemote archive Magneto Optical 
    Disk
    A remote database (either on 
    EchoPAC workstation or on 
    EchoServer) is used for patient 
    archiving. Images are stored to a 
    MOD.
    WL-LA-DServ: 
    the local 
    database is not 
    searched, only 
    the DICOM 
    Modality Worklist.Worklist/LocalArchive-
    DICOMServer/Int.HDModality Worklist local archive 
    DICOM server and local harddrive
    Search in the DICOM Modality 
    Worklist, the patient found is copied 
    into local database. The patient 
    information and the examination 
    results are stored to the local the 
    database. Images are stored to a 
    DICOM Server and to an image 
    volume on the local harddrive.
    Worklist/RemoteArchive-
    DICOMServer/RemoteHDModality Worklist remote archive 
    DICOM server and remote harddrive
    Search in the DICOM Modality 
    Worklist, the patient found is copied 
    into a remote database. The patient 
    information and examination results 
    are stored to a remote database. 
    Images are stored to a DICOM 
    Server and to an image network 
    volume as pure DICOM in both 
    locations. Dataflow Description 
    						
    							Presets and System setup
    428Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Worklist/Remote Archive 
    - Remote StorageThis dataflow is used in a network 
    environment that includes Vivid HL7 
    Gateway. The patient list in the 
    Search/Create Patient window is 
    coming from Vivid HL7 Gateway 
    through DICOM Modality Worklist. 
    All patient data and images are 
    stored to EchoServer.
    Raw DICOM MODRaw image format to/from a DICOM 
    Magneto Optical Disk
    Read/Write images in raw format 
    from/to a DICOM formatted 5.25 
    MO-disk.
    DICOM MODPure DICOM image format to/from a 
    DICOM Magneto Optical Disk
    Read/Write images in “pure” DICOM 
    format from/to a DICOM formatted 
    5.25 MO-disk.
    DICOM CD readDICOM CD read
    Read DICOM Media from the CD-
    drive.
    EP MAC MOD: 
    the MOD must be 
    PC-formatted.
    Only images can 
    be read.EchoPAC Mac MOD readEchoPAC Macintosh Magneto 
    Optical Disk read
    Read MOD from EchoPAC 
    (Macintosh).
    DICOM ServerDICOM server
    Store pure DICOM images to a 
    DICOM device.
    DICOM PrintDICOM Print
    Send images to a DICOM printer.
    Query RetrieveQuery Retrieve
    Retrieve images from a DICOM 
    server Dataflow Description 
    						
    							Presets and System setup
    Vivid7/Vivid7 PRO Users Manual429
    FC092326-03
    To select the default dataflow
    1. Select the dataflow in the Name drop-down menu. (see 
    Figure 12-14).
    2.Check the Default box.
    The dataflow will be selected by default when restarting 
    the unit.
    3.Check the Direct Store box to have data stored 
    automatically to the archive (no buffer storage).
    To create a new dataflow
    1. Press Add.
    The Add dataflow window is displayed.
     Figure 12-15: The Add dataflow window
    2. Enter a name for the new dataflow.
    3. Press OK.
    4. Select an Input device in the Available inputs/outputs field 
    and press the Right arrow button to assign the service to 
    the dataflow.
    5. Repeat step 4 to assign input/output devices as desired.
    Most dataflows consist of one input device and one or 
    several output devices. However in two cases it is possible 
    to have two input devices, when defining a dataflow with 
    DICOM Worklist or DICOM Query/Retrieve in combination 
    with a database. 
    6. Adjust the device’s parameters as described below.
    7. To remove a device from the dataflow, select the device in 
    the Selected devices field and press the Left arrow button. 
    						
    							Presets and System setup
    430Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Adjusting the assigned devices
    1. Select the device in the Selected devices field.
    2. Press Properties.
    The Properties window is displayed.
    3. Adjust the device specific parameters as desired (see table 
    below). Not all the settings listed below apply to all devices.
    General settings Definition
    Name Free text: give a descriptive name for the device.
    IP address Select from drop-down menu
    Database Name Automatically selected according to the IP address
    File destination Automatically selected according to the IP address
    Removable Check the entry is the media is removable.
    Image settings Definition
    Allow raw data
    : save data in both raw and DICOM format.
    : save data in DICOM format only.
    Max Framerate Select 25, 30 or Full from the pop-up menu. Full (original 
    acquisition) is default.
    Compression Select compression type or no compression.
    Quality Set picture quality from 1 to 100%. A low picture quality 
    level allows high data compression, while a high picture 
    quality restrains the compression. 
    Allow Multiframe
    : allow cineloop storage.
    Connection settings Definition
    Retry Set maximum number of connection tentatives, time 
    interval between tentatives and time-out.
    DICOM settings Definition
    AE Title The Application Entity Title is set during DICOM 
    configuration. Refer to the network specifications. 
    						
    							Presets and System setup
    Vivid7/Vivid7 PRO Users Manual431
    FC092326-03
    Port The Port no. is allocated during DICOM configuration. 
    Refer to your network specifications.
    Verification Verify the connection to another DICOM application
    Storage commitment Send a request to a PACS, asking it to permanently 
    archive image(s)
    MPPS Modality Perform Procedure Step: send information 
    (typically to a HIS) that a scheduled exam has been 
    started, performed or interrupted. DICOM settings Definition 
    						
    							Presets and System setup
    432Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Additional outputs
    The Additional outputs sheet deals with configuration of the 
    PRINT and ALT keys on the control panel. Several outputs (e.g. 
    Video Print, Laser print, DICOM storage...etc.) can be 
    associated to the keys (i.e. hitting 
    PRINT can result in printing a 
    Color video print and storage to a DICOM media).
     Figure 12-16: The sublevel Additional outputs (example)
    1. Select between PRINT and ALT keys.
    2. Available output devices that can be assigned to 
    the current button.
    3. Output devices assigned to the current button.
    4. Add or remove selected device to/from the 
    current button.5. Adjust the device settings for the selected 
    assigned device
    6. Select the type of images to produce and adjust 
    image settings. These settings will overrule the 
    adjustment made in the Properties window. 
    						
    							Presets and System setup
    Vivid7/Vivid7 PRO Users Manual433
    FC092326-03
    The image configuration parameters
    The table below gives a list of the configuration parameters. 
    To configure the Print or Alt. Print key
    1. In Button field select Print or Alt. Print.
    2. Select an output device in the available outputs field and 
    press the Right arrow button to assign the service to the 
    dataflow.
    3. Press Properties.
    The Properties window is displayed.
    4. Adjust the device specific parameters as desired.
    Some of the settings can be changed directly in the Image 
    to Produce field in the sublevel Additional outputs.
    Configuration parameter
    FormatSelect between:
    • Raw DICOM
    • DICOM
    • AVI
    • JPEG
    Image framesSelect between:
    • Single: stores single frame only
    • Multiple: stores cineloop
    • Secondary Capture: screen shot
    Image compressionSelect compression mode from the pop-up menu.
    QualitySet picture quality from 1 to 100%. A low picture quality 
    level allows high data compression, while a high picture 
    quality restrains the compression. 
    Capture AreaSelect between:
    •  Video Area (1)
    •  Whole Screen (2) 
    						
    							Presets and System setup
    434Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    To remove a device, select the device in the Selected devices 
    field and press the Left arrow button. 
    						
    							Presets and System setup
    Vivid7/Vivid7 PRO Users Manual435
    FC092326-03
    Tools
    The Tools sublevel sheet deals with: 
    •  formatting of removable media (MO disk, CD-R or ZIP disk).
    •  verification of DICOM directory on removable media.
     Figure 12-17: The sublevel Tools
    1. Select removable media to verify/format
    2. Label a new removable MO disk (free text)
    3. Verify DICOM directory on removable DICOM 
    disk
    4. Start formatting
    5. Install a DICOM viewer6. Select the device for patient information 
    exported as Excel
    7. Set the path for the repository for HL7 export 
    (format: \\server-name\share-name)
    8. Set a user-defined remote path for Save as 
    function and export to Excel. 
    						
    							Presets and System setup
    436Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Formatting a removable media
    1. Select the removable media from the Media pop-up menu.
    2. Enter a name for the removable media in the Label field.
    3. Check DICOM viewer is desired.
    DICOM viewer enables the display of DICOM images from 
    the removable media without the need of any other 
    software.
    4. Press Format.
    5. Wait for the display of the Information window indicating 
    that the formatting process is completed.
    Do not eject the CD 
    using the button on 
    the CD drive6. Press ALT+E to eject the disk.
    Verifying a removable media
    1. Insert the removable media to verify.
    2. Select the pre-configured removable media from the Media 
    pop-up menu.
    3. Press Ve r if y.
    The unit is verifying the DICOM directory on the selected 
    media. 
    						